TJ, Camilla
I put SR3 into it's O4 alignment (P 458, Y -135) changed from (P 524, Y -135), checked no saturating pixels (ITMX stayed at 1Hz frame rate, ITMY increased to 5Hz), took new HWS references and started the HWS code.
We are running a ring heater test with 2W/segment for 4 hours this evening. Using scripts below on tmux sessions:
Ring heaters should be back to nominal by ~1am. Purpose of this is to check HWS alignment before we align the CO2s (using HWS).
TJ, Camilla ITMY HWS signal looks great. ITMX HWS signal looks noisy, the power on the CCD camera is much smaller than expected, expect this needs realigning after the BS is different.