2500 L/s ion pump controller settings
For the record, the following are the nominal values used for the large ion pump controllers (listed in the order queried by the MultiVac firmware Flow 7):  

Torr
Spare
7000V 
200mA 
200watt 
Protect
10mA 
STEP 
5000V 
1.0 x 10-3amps 
3000V 
4.0 x 10-5amps 
(StPt 1) 1.0 x 10-5torr (not used?) 
(StPt 2) 1.0 x 10-6torr (not used?) 
1.0 x 10-10 
1.7 x 10-9 
8.0 x 10-7 
2.0 x 10-4 
Comments related to this report
kyle.ryan@LIGO.ORG - 14:53, Wednesday 08 January 2014 (9159)
Correction:  

1 x 10-3 amps (should be) 3 x 10-3 amps