(Travis, Gerardo)
The ion pump for HAM1 was incorporated to the HAM volume yesterday, and today the SS-500 pump cart was isolated from the chamber (turbo pump and SS-500 remain on), and as expected the pressure went up and now we are waiting for the ion pump to take over and maintain the vacuum pressure for HAM1. We will be monitoring the progress of the ion pump.
(Travis, Gerardo)
We stopped pumping against the isolation valve for HAM1 turbo, now the turbo needs to spin down before we can disconect the cable from the magnetically levitated turbo pump. The flex hose was removed and the scroll pump was stopped, but needs to remain on until the turbo stops spinning.
Travis removed the aux-cart, SS-500, decoupled cables. Flex hose was removed the previous day. The SS-500 cart then got moved and now is installed on HAM6 turbo.