Reports until 13:00, Friday 28 August 2026
H1 AOS
louis.dartez@LIGO.ORG - posted 13:00, Friday 28 August 2026 (91726)
PRM beam spot position
At Sheila's suggestion this morning I measured the beam spot position on PRM in PRMI. The drivealign gains that reduce the A2L coupling are P:0.38, Y:0.81. I used the same procedure as in 91701. This corresponds to an offset from center of 0.713 mm in PIT and 1.52 mm in YAW.

Images attached to this report