Reports until 21:51, Monday 14 September 2026
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 21:51, Monday 14 September 2026 - last comment - 16:31, Tuesday 15 September 2026(91903)
quick look at recovering squeezer
Comments related to this report
sheila.dwyer@LIGO.ORG - 16:31, Tuesday 15 September 2026 (91928)

Camilla and I had another look at the filter cavity alignment this afternoon.  We see green flashes of 7.1 counts, where we would expect about 30 counts based on the launch power and O4 FC transmission.  These slider values are in the attached screenshot.  Went to SQZT7 to align the beam onto FCGS refl diode, didn't see any signal in the DC readback of that diode.  However, now there are mini locks of the filter cavity on the 00 mode, so this looks like the diode is working.  