H1 ETMX Optical Lever Calibration
Attached is a linear response graph of the ETMX OL to convert the QPD signals to uradians.

The calibration numbers are placed into the gain fields:
H1:SUS-ETMX_L3_OPLEV_PIT_GAIN == 76.72
H1:SUS-ETMX_L3_OPLEV_YAW_GAIN == 65.33