The HV at EndX was switched back on yesterday afternoon, I ran the OPLEV charge measurement for ETMX first thing this morning. The GRD state of ETMX was changed from ALIGNED to MISALIGNED in the last ~2 minutes of my measurement for some reason, I had unmanaged it from ALIGN_IFO. It was during the last LR measurement, LR Yaw has the largest error bars so that might have played into that. There was also a ground motion increase from a 5.7 earthquake near Tonga from the 2nd measurement till the 4th but my coherences were all >0.9 and the errors looked ok.
15:16 - 15:50 UTC EQ mode
The charge has decreased towards zero on 3/4 quadrants in both P and Y, UL being the odd one out mostly in Yaw. All quadrants/dofs are now under +/- 50 V.