Toward optimizing the JAC injection mode-match at 62 W, the 2 W LDWFS data (LIGO-P2600491) taken before the EOM swap, after it, and after the IO_MB_L2 move were re-analyzed with a full Finesse model of the injection and REFL paths. Mode mismatch power at 2 W: 1.8 % → 5.8 % → 4.5 % for the three states, consistent with the REFL PD lock/unlock ratio (1.9 / 5.7 / 4.9 %), which didn't enters the fit. The current lens position did not improve as much as predicted earlier because that prediction used the symplified model, which misses the near-resonance of the RF sideband's first-order modes in this cavity. The Finesse model includes it and now reproduces all three states. The 62 W analysis and the lens solution derived from it come next.
The table gives the fitted injected beam as a deviation from the cavity eigenmode, both taken at the output of IO_MB_L3. The eigenmode there has a waist of 545 µm (YAW) / 551 µm (PIT) located 4.93 m downstream of L3. dw is how much larger the injected waist is than that; dz is the position of the injected waist relative to the cavity waist, positive toward the cavity (downstream), so the negative values below mean the injected beam focuses before the cavity waist. Values come from the joint fit over the three data sets; the errors are statistical.
| YAW dw [µm] | YAW dz [mm] | PIT dw [µm] | PIT dz [mm] | |c20| / |c02| | mismatch (Finesse) | REFL lock/unlock | |
|---|---|---|---|---|---|---|---|
| before EOM swap (09/14) | +87 ± 6 | -94 ± 18 | +48 ± 4 | -138 ± 22 | 0.108 / 0.076 | 1.8 % | 1.87 % |
| after EOM swap (09/15) | +136 ± 6 | -511 ± 36 | +65 ± 4 | -143 ± 25 | 0.212 / 0.090 | 5.8 % | 5.73 % |
| after L2 move (09/16) | +100 ± 5 | -449 ± 28 | +30 ± 3 | -234 ± 21 | 0.183 / 0.092 | 4.5 % | 4.93 % |
Three data sets were taken at 2 W input, one per state of the injection path: before the EOM swap, after the EOM swap, and after the IO_MB_L2 move. Each set is a series of 30 s records (24 records, 12 for the last one).
Two angular actuators are dithered, each in both planes, at separate frequencies: JM1 at 9.7 Hz (PIT) and 11.3 Hz (YAW), the injection PZT mirror at 13.1 Hz (PIT) and 14.9 Hz (YAW). A length dither is applied to the JAC PZT at 809 Hz. In addition, during each record one pico motor sweeps the beam spot on one WFS head along one axis with a slow triangle wave (fundamental about 1.5 Hz), cycling through the four head/axis combinations over the series.
The recorded channels are the WFS quadrant I/Q outputs of both heads (H1:JAC-WFS_{A,B}_{I,Q}{1-4}_ERR_DQ). Offline, the amplitude of each angular line is read in the WFS signal (RF) and in the WFS signal demodulated once more at 809 Hz (DD: double demodulation), each normalized by the head's DD sum: 2 actuators × 2 planes × 2 heads × (RF, DD) = 16 line observables per data set.
The pico sweep adds, for each head/axis combination, the ratio RF/DD of the head's response to the spot motion. Being a ratio it needs no pico calibration, and it is a direct mode-mismatch observable: it measures the c2 content (together with the detuning x0) as seen at that head (4 observables). Each data set thus contributes 20 observables.
| before EOM swap | after EOM swap | after L2 move | |
|---|---|---|---|
| head Gouy phase, WFS A / B, PIT (shared) | 124.8 ± 0.7° / 189.1 ± 0.9° | ||
| head Gouy phase, WFS A / B, YAW (shared) | 112.7 ± 0.7° / 173.9 ± 1.1° | ||
| mismatch power (Finesse) | 1.8 % | 5.8 % | 4.5 % |
All parameters, errors and the per-record fits are in the detailed analysis note (to be posted).
Only IO_MB_L2 (and IO_MB_L1 by 3 mm) was moved between the 09-15 and 09-16 sets. Back-propagating the fitted 09-15 beam through the injection telescope and asking which L2 displacement reproduces the fitted 09-16 beam gives 18 ± 3 mm toward L1. The survey gives the L1–L2 spacing as 60 mm before and 26.5 mm after, i.e. an L2 move of 33.5 mm, so the error is ~1.5cm. This error is bigger comparing to the quoted error, which is statistical only (from the fit χ2); lens focal-length tolerances, the optic positions (even between PSL and HAM1) and any thermal lensing are not included.
Figures
Figure 1: Beam radius along the injection path from the IO_MB_L3 output to the cavity, YAW (left) and PIT (right). Black: the cavity eigenmode (waist 545 / 551 µm, 4.93 m from L3). Colored: the injected beams from the joint fit for the three states. In every state the injected beam focuses before the cavity waist with a larger waist, and the two planes differ.
Figure 2: The 16 line observables per data set — RF (top) and DD (bottom) amplitudes of the four angular lines on WFS A (blue) and WFS B (orange), normalized by the DD sum — for the three states. Bars: data with 1 σ errors; black marks: the joint-fit model (one shared WFS layout, per-state injected beam and x0).