ETMX still looking gooooood

Took transfer functions again on Friday (sept 25) as another verification of ETMX L1 LL still no longer rubbing (and entire suspension still being healthy). Measurements are still looking good.

Settings:
- SEI nominal
- HEALTH_CHECK
- DAMP OFF
- alignment offsets ON

M0
Data:
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMX/SAGM0/Data/2026-09-25_1730_H1SUSETMX_M0_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMX/SAGM0/Results/2026-09-25_1730_H1SUSETMX_M0_ALL_TFs.pdf
r13181

R0
Data:
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMX/SAGR0/Data/2026-09-25_1800_H1SUSETMX_R0_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMX/SAGR0/Results/2026-09-25_1800_H1SUSETMX_R0_ALL_TFs.pdf
r13180

Images attached to this report