Reports until 16:46, Tuesday 29 September 2026
H1 SQZ
ryan.short@LIGO.ORG - posted 16:46, Tuesday 29 September 2026 (92103)
Grid of SQZ Beam Profiles at Various ZM PSAMS Settings

Yesterday I took several beam profiles of the SQZ beam on SQZT7 using the ThorLabs M2 device at various settings for the ZM4 and ZM5 psams as we've done in the past to create a "grid" of values. The psam strain gauge values chosen for this set of measurements was:

ZM4 = [-3.5, -1.5, 0.5, 3.0], ZM5 = [-9.3, -7.0, -5.0, -1.3]

Between Sheila, Camilla, and myself, we were able to produce the attached figures comparing these profile results with the OMC scans taken last week (alog92065) to show the predicted mode mismatch between the SQZ beam and the OMC. The results indicate running on the very low end of the psams range with both ZM4 and ZM5 gives the best mode matching into the OMC with the current ZM2 setting.

Images attached to this report