elenna.capote@LIGO.ORG - posted 20:19, Friday 09 October 2026 (92271)
Trouble getting back to resonance, is the recycling gain too low?

During the latter part of the day, Louis and I tried to get back to resonance, but have yet to get there. First, we had problems with TR carm (again), and then we struggled at the carm offset reduction.

For TR CARM, we thought the gain might be a bit low in start_tr_carm, so we upped the analog H1:LSC-REFL_SUM_A_IN1GAIN by 3 dB. This seems to help! Louis also increased the time in the servo to 10 seconds.

CARM_OFFSET_REDUCTION problems:

We had one lockloss at carm 5 pm after completing carm_offset_reduction by hand. Don't know why, but this was another state we sometimes had to carefully do by hand.

Want to reiterate that the IMC is still being fussy when relocking.