Displaying report 1-1 of 1.
Reports until 10:35, Thursday 16 January 2014
H1 SEI
hugh.radkins@LIGO.ORG - posted 10:35, Thursday 16 January 2014 (9326)
ETMX/ITMX ISI X motion compared to Optic Oplev Pitch

Attached is a coherence plot.  Basically all X motions on the ISI on the two platforms have similar coherence with their Optical Lever Pitch.  Interesting though, the ITM with a 0.56Hz notch has much broader coherence below this frequency whereas the ETM has no notches and much less coherence elements below 1Hz.  Maybe there is area here to improve.

Images attached to this report
Displaying report 1-1 of 1.