WP 4405. I modified the IOP model for h1sieh45 (h1iopseih45) to fix a copy-paste error with the IOP watchdogs. For now, since SR2 is not sending an IPC status I have "wired the watchdog open" by setting the dackill input to unity. When the new IOP watchdog software using the new targeted parts go in I will remove this hard-wire.
14:00 all models on h1seih45 killed, IOP model restarted, all user models restarted (h1hpiham4,5 h1isiham4,5)
14:07 DAQ restarted due to IOP model INI file change
This closes out WP4405