stefan.ballmer@LIGO.ORG - posted 19:25, Tuesday 21 January 2014 - last comment - 01:05, Wednesday 22 January 2014(9412)
First PRMI locks
Kiwamu, Stefan

We locked the PRMI on the sidebands today. The locks lasted for about 1min at a time. Lock acquisition is still slow, so more work is needed.
The first lock started at UTC 2014 Jan 22, 01:59:42, and lasted until 02:00:25. There were several additional locks after that.
Attached is a large screen shot with all relevant screens.

We managed to get a preliminary OLG measurement for PRCL, the UGF seems to be around 55Hz (plot attached.)
We were not successful with a MICH OLG measurement yet.

Finally, attached is also a plot of error and control signals for the two loops.
Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 01:05, Wednesday 22 January 2014 (9414)

It seems that the pitch motion in PRM and PR2 is excited a lot during the acquisition and this prevents us from a smooth and quick lock.

In fact, when I enabled the LSC output after the pitch motion of PRM and PR2 settled down, it often locked within 5 sec. However, when their pitch rang up to more than 1 urad in RMS, locking became extremely difficult -- it catches a fringe approximately once per 5-10 minutes. Also, it seems to me that the unlock events were more or less initiated by the pitch fluctuation. See an example in the attached.

Images attached to this comment