Because Stefan jinxed me, the opsws1 crashed while I was on the last few super-long data points of the H1 SUS BS and H1 SUS PRM M2 L to M3 PY TFs that I'd been running all evening. Curses. I've restarted the transfer functions, after rebooting the computer and clearing all relevant test points. I've also started them with a more limited range (from 0.2 to 5 [Hz] instead of 0.1 to 10 [Hz]) and reduced the number of data points (from 128 to 111). These were started around 12:30 am local time. In case these new templates fail again, the amplitude envelop I'm using is Freq [Hz] PRM [ct] PR2 [ct] BS [ct] 0.1 3e5 3e6 2e4 0.2 2e4 0.3 3e5 3e6 5e3 0.5 1e5 1e6 1e3 0.7 2e5 1e5 6e3 1.0 5e4 1e5 3e4 2.0 1e5 1e6 1e5 5.0 1e5 1e6 1e6 10. 5e5 5e6 1e6 15 5e6 20. 5e5 5e6 5e6 Drive Channels H1:SUS-PRM_M2_LOCK_L_EXC H1:SUS-PR2_M2_LOCK_L_EXC H1:SUS-BS_M2_LOCK_L_EXC Witness Sensors: PRM: REFL WFS A DC, Pitch and Yaw H1:ASC-REFL_A_DC_PIT_OUT H1:ASC-REFL_A_DC_YAW_OUT PR2: M3 OSEMs, Pitch and Yaw H1:SUS-PR2_M3_WIT_P H1:SUS-PR2_M3_WIT_Y BS: Optical Lever, Pitch and Yaw H1:SUS-BS_M3_OPLEV_PIT_OUT H1:SUS-BS_M3_OPLEV_YAW_OUT