Displaying report 1-1 of 1.
Reports until 02:06, Friday 31 January 2014
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 02:06, Friday 31 January 2014 (9695)
RF modulation frequency changed to 9099170 Hz

I saw a big RAM offset in REFLAIR_RF45 and didn't like it. So I changed the modulation frequency to 9099170 Hz where the RAM offset became zero. This frequency is consisent to what Alexa got (see alog 9679) in her IMC length measurement.

Before this change, the mod frequency was at 9099471 Hz and the RAM offset was 3000 counts in REFLAIR_RF45_Q with its whitening gain at 21 dB and with PRM aligned.

Displaying report 1-1 of 1.