Reports until 09:01, Friday 31 January 2014
H1 General
andres.ramirez@LIGO.ORG - posted 09:01, Friday 31 January 2014 (9702)
PSL Check
PSL Check:
Laser Status: 
SysStat is good
Output power is 29.3 W (should be around 30 W)
FRONTEND WATCH is Active

PMC:
It has been locked 2 day, 16 hr 11 minutes (should be days/weeks)
Reflected power is 1.1 Watts  and PowerSum = 12.16 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 11 h and 22 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 0.94V (should be 0.9V)

ISS:
The diffracted power is around 5.4% (should be 5-15%)
Last saturation event was 11 h and 28 minutes ago (should be days/weeks)