We had to place ten optics/opto-mechanics assemblies. The D1201098 installation kit (cookie cuters) parts were placed to facilitate placement and alignment of optic assemblies. The assemblies were then bolted into position, summarized below:
HWSX STEER M1 - 4" optic assembly installed without issue. This optic may need to be temporarily removed for the SR2 install - the dog clamps that are in-situ can be used to mark the position.
HWSX STEER M2 - 4" optic assembly installed without issue
HWSX STEER M3 - 2" optic assembly installed without issue
HWSX DCBS1 - we were missing 8/32 bolts for attaching the 2" lens mount to the top of this pedestal (D1101840). We opted to installed the pedestal anyway and leave the 2" lens mount off the table (perhaps 100g - 150g with lens) as the the mount can be attached in-situ.
HWSX VACLENS - we were missing 8/32 bolts for attaching the 2" lens mount to the top of this pedestal (D1101840). See HWSX DCBS1
Here are the full details ...
We had to place ten optics/opto-mechanics assemblies. The D1201098 installation kit (cookie cuters) parts were placed to facilitate placement and alignment of optic assemblies. The assemblies were then bolted into position, summarized below:
X-Arm Optics (see D1101083)
Y-Arm Optics (see D1101084)
Photos are attached below. Complete set is on ResourceSpace
We plan on installing the 4x 2" lens mounts and optics during the alignment of these optics or earlier.