REFL_A_RF9 phase adjusted

(Alexa, Daniel, Stefan)

 

The RD rotation was adjusted to 120deg. This was determined such that the sideband signal was seen in only I, with nothing appearing in Q.

 

Note: the RF9 frequency is set to 9099471Hz  (only 300Hz from the FSR). We still see RF AM.