Laser Status: SysStat is good Output power is 29.3 W (should be around 30 W) FRONTEND WATCH is Active HPO WATCH is red PMC: It has been locked 6 day, 16 hr 34 minutes (should be days/weeks) Reflected power is 1.1 Watts and PowerSum = 12.1 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 1 h and 6 min (should be days/weeks) Threshold on transmitted photo-detector PD = 0.74V (should be 0.9V) ISS: The diffracted power is around 6.7% (should be 5-15%) Last saturation event was 22 h and 26 minutes ago (should be days/weeks)