QPD Transimpedance Amplifier - EX
Looked at the QPD Transimpedance Amplifier units at EX, S1102832 and S1102821.  We wanted to confirm the board versions. All boards D1001974 are version 2 and have the following configuration:
- C15 is removed
- R15 is 0 ohms
- JP1 has a jumper installed

Filiberto Clara
Comments related to this report
keita.kawabe@LIGO.ORG - 12:28, Tuesday 04 February 2014 (9798)

Sorry, I was wrong, but things still look dubious.

From the attached dark spectra of QPD segments,  TR_A needs digital dewhite, it doesn't look like TR_B has too much dewhite.

But TR_B still looks like it has some kind of problem, maybe white/dewhite mismatch for seg3 and seg4 (or seg1 and seg2). Needs to be checked later.

Images attached to this comment