Displaying report 1-1 of 1.
Reports until 13:45, Wednesday 05 February 2014
H1 ISC
alexan.staley@LIGO.ORG - posted 13:45, Wednesday 05 February 2014 - last comment - 19:10, Wednesday 05 February 2014(9846)
EX Measurements

(Sheila, Alexa)

1. We wanted to re-visit the EX PLL shot noise measurement we had previously done.

2. WIth the EX PLL locked, the arm cavity well aligned, and the green beam flashing, we looked at the PDH error signal out of the demod IMON and measured the peak-to-peak of the signal to be 230mVpp

3. We went on to investigate the fringe wrapping we saw in the PDH amplitude spectrum at the error point. With the arm locked, we only saw a slight difference in the spectrum with the HEPA fans on or off (Sheila will attach pictures). There did not seem to be fringe wrapping, and the acoustic noise seemed minimal. I will look at the PDH shot noise again when I can misalign ITMX and see if the fringe wrapping is still there.. 

Comments related to this report
sheila.dwyer@LIGO.ORG - 19:10, Wednesday 05 February 2014 (9867)
The reference (fainter trace) is with hepa filter on, current trace with hepa filter on.
Images attached to this comment