Jeff B. & Kate G.

A horizontal 4" witness plate (wafer) and horizontal 1" optic (S/N 1247) were placed near the center of the SEI table on Feb 5th. A vertical 1" optic (S/N 1245) was mounted on the HR side of SR2. A vertical wafer still needs to be set near SR2. 

Important notes: Witness samples were not placed immediately after the chamber was opened. Jeff did some cleaning before SR2 was installed and the wafers and optics were placed. Also, the 1" optics were removed from their PET G containers, put in PEEK holders, double bagged, and stuck in a bag with other tools before going in chamber. I didn't see obvious scratches in the First Contact, so hopefully the samples weren't damaged. 
Images attached to this report
Comments related to this report
kaitlin.gushwa@LIGO.ORG - 10:29, Tuesday 11 February 2014 (9974)

On Feb 7th, I used a handheld particle counter in the cleanroom around HAM4 and in chamber before starting work. The counts seemed really high. The particle counter (used for trending data) on the +X side of the chamber was not working.

Size (um) Cleanroom Count Chamber Count
0.3 10725 17177
0.5 2518 3895
0.7 1075 1562
1.0 586 798
2.0 169 213
5.0 1 0

I placed a vertical wafer near the HR side of SR2, and inspected the cleanliness of the chamber with a flashlight array. There were a few particles on the witness samples (which had not seen work since they were placed two days earlier), and a lot of contamination on the TCS HWS optics and SEI table. The dark iPhone photos don't do it justice.There was only time for a quick cleaning using the TigerVac and IPA soaked wipes. It would be helpful to get a PET swipe sample the next time work is done in HAM4. 