Summary about missing power in REFL beam
Evan, Stefan, Kiwamu, Lisa

There has been some confusion about factors of two in previous entries, so here are our "final" numbers:

- Pin = 8.8 W entering the IMC, measured on the PSL table
- with PRM aligned, R = 98.6%
-  from Paul's calibration  of the MC2Trans diode (we measured 645 uW), it looks like there is T = 88% from the PSL to the transmission of the input mode cleaner;
- given the HAM1 path, we expect 1.25% of the REFL power to reach the LSC RF diode P_exp = 8.8 * 0.986 * 0.88 * 1.25e-2 = 95 mW;


 LSC in vac REFL A (LSC RF PD)  (1.25% of REFL power) = 15900 counts (maximum found by moving RM2)

==> 15900 [counts] * (1/1638) [V/counts] * 2 [diff to single input] * (1/200) [1/Ohm] * 0.8 [W/A] =  78 mW 

 LSC REFL AIR A (LSC RF PD)  (0.625% of REFL power) = 7955 counts 

==> 32.5 mW measured directly in front of the diode on table

==> 7955 [counts] * (1/1638) [V/counts] * 2 [diff to single input] * (1/200) [1/Ohm] * 0.8 [W/A] = 39 mW

 LSC REFL AIR B (BBPD)  (0.625% of REFL power) = 21770 counts 

==> 34 mW measured directly in front of the diode on table


==>  Conclusion : The discrepancy between what we see and what we expect is not a factor of 2, but more like 20%-30%, and in vac and in air diodes gives us consistent numbers. 



Comments related to this report
lisa.barsotti@LIGO.ORG - 18:02, Saturday 08 February 2014 (9933)ISC
Nope..we didn't get the right factors this time either:

W/A  = 1/0.8, and the factor of 2 diff to single is not there (and, even if it doesn't matter, R for PRM is 97.3%, not 98.6%).
I will post another entry with the right numbers..somehow I can't edit this one..

P.S: Still need to calibrate these numbers, measurements were taken all at the same time. It might be also worth to check if the T = 88% transmission from PSL to IMC throughput is correct.

 LSC REFL AIR B (BBPD)  (0.625% of REFL power) = 21770 counts 

==> 34 mW measured directly in front of the diode on table

WFS in vac REFL A (0.625% of REFL power) = 34200 counts
WFS in vac REFL B (0.625% of REFL power) = 36400 counts