Displaying report 1-1 of 1.
Reports until 06:15, Tuesday 11 February 2014
H1 ISC
sheila.dwyer@LIGO.ORG - posted 06:15, Tuesday 11 February 2014 - last comment - 11:03, Tuesday 11 February 2014(9970)
green team work yesterday

Yesterday we made some measurements for the noise model (end sation PDH spectra and open loop), and ALS COMM slow path, fast path and open loop). 

Then we moved on to trying to make some noise measurements.  We tuned the demod phase of RF A 9 to put the signal in the Q phase (we don't have enough range with the phase shifter to put it in the I phase).  The demod is now on local control (remote controls not working) with al switches down except for 8, 2, 1, and 1/2 (11.5 ns delay). 

We also tried to check the IR alingment into the arm cavity by locking at the transmitted PD.  We found the 00 mode with the COMM PLL offset at 1.23, and saw up to 90 counts in transmission.  We saw a misalingment mode (between pitch and yaw) at 6.48, with 17 counts in transmission.  We strugged to improve this, mostly because the transmitted power was fluctuating so much from the frequency noise. 

We engaged the boost on the PLL readback path, (adding more DC gain to ALS CARM).  This reduced the noise significantly, although we were still moving by mre than a linewidth, so we can't really make a linear measurement when locked with ALS COMM. 

Comments related to this report
daniel.sigg@LIGO.ORG - 11:03, Tuesday 11 February 2014 (9982)

Evan, Daniel

We found that the controls cable is missing from the second delay line, The cable we need is ISC-332 (DB25). We did not find any cable with this number, but we found two ISC-323. One is a DB25, the other a DB15. ISC-323 is for a shutter on ISCT6 (DB15). We think that ISC-332 (DB25) has been mislabeled. Unfortunately, it was pulled to the AS port rather than REFL.

Displaying report 1-1 of 1.