Displaying reports 241-260 of 86768.Go to page Start 9 10 11 12 13 14 15 16 17 End
Reports until 16:45, Friday 06 February 2026
LHO General
corey.gray@LIGO.ORG - posted 16:45, Friday 06 February 2026 (89063)
Fri Ops DAY Shift Summary

TITLE: 02/06 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
Super Bowl Weekend SHIFT SUMMARY:

JAC team continued with JAC alignment work in HAM1.  HAM3 AIP was swapped & started pumping over the weekend. 

(Go Seahawks!)

LOG:

 

H1 PSL
ryan.short@LIGO.ORG - posted 16:44, Friday 06 February 2026 (89072)
PSL Cooling Water pH Test - Monthly

FAMIS 28229

pH of PSL chiller water was measured to be between 10.0 and 10.5 according to the color of the test strip.

H1 SQZ (SQZ)
karmeng.kwan@LIGO.ORG - posted 14:11, Friday 06 February 2026 - last comment - 14:41, Friday 06 February 2026(89069)
HAM7 BL2 clipping model

From the previous alog post, Sheila measured a power drop across BL2 lens.

Here I modeled the loss of the 700um beam waist across the 2.5mm BL2 lens aperture.

The expression for the transmitted power obtained by modifying the Gaussian beam equation to include a displacement d= dx2 + dyis shown here.

And here's the contour plot for the loss (1-T) across the lens aperture for a beam transversing across various spots on the lens.

Images attached to this report
Comments related to this report
karmeng.kwan@LIGO.ORG - 14:41, Friday 06 February 2026 (89070)

Me and Sheila are checking on the geometry of the setup.

The BM3 mirror mount IXM100.2 has a 100 pitch screw adjustment ( correspond to 100 turn per inch translation), and the beam appeared to be clipped symmetrically for 1/8 rev. For the mirror mount of 1.87" width, this correspond to a mirror rotation of 1.3mrad.

The distance between BM3 and BL2 is 148mm, from the mirror rotation, the IR beam angular deflection is 2.6mrad, this implies an aperture size of 0.4mm. Something is not right, and it's a mystery that requires more investigation.

Images attached to this comment
H1 ISC (IOO)
jennifer.wright@LIGO.ORG - posted 14:11, Friday 06 February 2026 - last comment - 15:59, Tuesday 10 February 2026(89068)
IMC cavity axis now vs pre-vent

Following on from this log (alog #89046), we have already checked that the MC mirrors are restored to their previous positions from before the power cut in December. We know the seismic state now should be as close as we can manage to this time 2025-12-03 11:28:44 UTC when the mode cleaner was locked at 2W input power and before any HEPIs were locked for the vent.:

Summary of seismic state now:

HAM ISI locked. HAM 2, 3, 4, 5, 6 ISI isolated.

HAM1,2 and  BSC2 HEPI locked. HAM 3,4,5,6 HEPI isolated - ie. as per nominal operation.

I trended the IM osems and the MC2 and IM4 QPDs in HAM2 while the mode cleaner was locked for mode scans yesterday (reference time is 2026-02-05 18:44:36 UTC).

It seem as though the MC2 and IM4 trans powers we have are lower now, but also the IM mirrors have changed.

As Sheila did yesterday I will try and summarise these changes in a table and compare whether this makes sense with the greater HAM1 loss we have because of the JAC (~40 %).

Just posting an updated version of Sheila's table to compare this new time (3rd December).

  2025-12-03 2026-02-05  
seismic state nominal HAM1 ISI, HEPI locked, HAM2 HEPI locked. MC mirrors should be at or close to nominal assuming table alignment didn't change much.
power into HAM1 H1:IMC-PWR_IN_OUT16 2W 1W  
MC2 trans 310 counts 90 counts consistent with 60% HAM1 throughput measured with a power meter (extra loss due to temporary mirror for JM3)

IM4 trans nsum

1.8W 0.1W should be 0.54 W so not consistent with JAC loss.
IM1 position P, Y 3130, -679 changed by -11, -0.44 Off in pitch
IM2 P, Y 915, -226 -50, -3 Off in both, mainly pitch.
IM3 P, Y 44, -1634 +4, -13 Off in both, mainly yaw
IM4 P, Y -2815, -22 +64, -8 Off in both, mainly pitch.

So after this Sheila and I are going to try turning on the IMC ASC and working out if we can tune the alignment onto IM4-TRANS using the IMs.

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 15:59, Tuesday 10 February 2026 (89104)

Since the IMC is nicely locked, and we're only seeing 0.2 W on IM4 trans at a time when we expect ~1W, I moved the IM1, IM2, and IM3 to these positions.  Still not very much on IM4_Trans.

H1 CDS
david.barker@LIGO.ORG - posted 13:05, Friday 06 February 2026 - last comment - 15:11, Friday 06 February 2026(89067)
NDS1 restart, offload TW1 raw minute trends

WP13007

The rawminute trend copy to permanent archive completed 04feb2026. Today I reconfigured NDS1 to serve these data from their permanent archival location and started the deletion of the files from TW1.

NDS1 was restarted at 12:51 for the new configuration.

Comments related to this report
david.barker@LIGO.ORG - 15:11, Friday 06 February 2026 (89071)

Deletion of last 6 months of raw minute trend files took 1hr56min. Disk usage reduced from 93% to 2%.

H1 TCS
ryan.short@LIGO.ORG - posted 11:22, Friday 06 February 2026 (89066)
TCS Monthly Trends

FAMIS 38837

CO2 lasers have been off for the past month for vent activities and HWS SLEDs were just turned on a few days ago, so this month's trends don't show much activity.

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:48, Friday 06 February 2026 (89065)
Fri CP1 Fill

Fri Feb 06 10:08:33 2026 INFO: Fill completed in 8min 30secs

 

Images attached to this report
H1 PSL (PSL)
corey.gray@LIGO.ORG - posted 07:40, Friday 06 February 2026 (89064)
PSL Status Report (FAMIS #39750)

This is for FAMIS #39750.

Laser Status:
    NPRO output power is 1.841W
    AMP1 output power is 70.43W
    AMP2 output power is 139.4W
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 0 days, 21 hr 23 minutes
    Reflected power = 27.19W
    Transmitted power = 103.9W
    PowerSum = 131.1W

FSS:
    It has been locked for 0 days 18 hr and 54 min
    TPD[V] = 0.4791V

ISS:
    The diffracted power is around 4.0%
    Last saturation event was 0 days 0 hours and 0 minutes ago

Possible Issues:
    PMC reflected power is high

LHO General
corey.gray@LIGO.ORG - posted 07:36, Friday 06 February 2026 (89062)
Fri Ops Day Transition

TITLE: 02/06 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 0mph Gusts, 0mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.31 μm/s 
QUICK SUMMARY:

HAM1 JAC work will continue.  HAM7 is closed but pumpdown has not started yet.  LVEA continues in the Bifurcated Laser Safe state.

H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 18:50, Thursday 05 February 2026 - last comment - 21:30, Thursday 05 February 2026(89060)
JAC Heater functioning

While Keita will report the detail about the heater wiring trouble shooting, here is the quick report for the JAC heater functionality.
The attatched plot shows the two thermistor signals (top/bottom) and the heater input (middle). The temperature is reasonably changing with the heater driver input.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 21:30, Thursday 05 February 2026 (89061)

JAC in-vac tri-cable was swapped. (Masayuki, Jennie, Jason, Keita)

The cable that failed (D2500336-V2, S2501241) was pulled out of the chamber. (In the process, one of the body mode damper crossbar had to be temporarily removed to release the cable. ) The cable was wrapped and put in a bag without further testing (yet).

Jennie and I tested the new tri-cable (D2500336-V2, S2501242) in the optics lab and it was good (i.e. every pin was connected to the pin it is supposed to be connected, no cross-wiring, no short circuits).

The new cable was installed in chamber. 

I checked the in-chamber connection from the in-air side of the D4F10 feedthrough. Pin13 wasn't connected to anything, ditto for the chamber ground, thermistors showed about 11kOhm each, heaters were about 50 Ohm each, PD anode and cathode were good, no cross-wiring and no unintended short circuits.

After connecting the in-air cable to the feedthrough, PZT and Trans PD worked right away.

After connecting the heater cable to the driver chassis, we confirmed (using the breakout board and a DVM) that the voltage across the heater elements was ~1.4V when H1:JAC-HEATER_DRV_VSET~3.6[V?],  H1:JAC-HEATER_DRV_VMON~1.6[V?] and H1:JAC-HEATER_DRV_IMON~-0.098[A?]. Maybe 1.4V is close enough to 1.6V (0.2V might be the voltage drop of the in-air cable?) but I don't understand the logic about VSET being 3.6V nor why IMON is negative (in my mind it seems logical if IMON~1.6V/25Ohm = 0.064A).

But it's good to know that it's doing something.

H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 18:42, Thursday 05 February 2026 (89059)
HAM1 beam damp campaign part 1.

[Jason, Masayuki]

We still need to find the back reflection beam from the septum viewport, and discuss what we will do for the vertically separated beams.

 

H1 General (Lockloss)
oli.patane@LIGO.ORG - posted 17:06, Thursday 05 February 2026 (89058)
Lock duration and ETM glitch lockloss stats through all of O4c

I never made updated plots for the lock duration and prevelance of the DARM glitch locklosses for the last couple months of O4c (up to Sept 1st, 2025, up to Oct 1, 2025), so here they are. The performance during October and November mostly line up with what we had been seeing in August and September following the bias change (86027).

Through October and November, the frequency of DARM glitch locklosses, and locklosses in general, was still lower than they had been before the ESD bias change. October went realy well. November wasn't the best month for ETM glitch locklosses - we had 8 locklosses due to ETM glitch during the month. That's only slightly worse than the amount we saw in August right after the bias was doubled, but during November we were locked for less time overall due to O4c ending on November 18th and PEM week not requiring us to keep the IFO locked overnight or on weekends, so that's not the best. However, we still had longer locks on average during November vs before the bias change.

O4 DARM Glitch vs All Locklosses
O4c Lock Length Distribution
O4c Lock Stats

Images attached to this report
LHO General
ryan.short@LIGO.ORG - posted 16:33, Thursday 05 February 2026 (89057)
Ops Day Shift Summary

TITLE: 02/06 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: Work continued in HAM1 on/around the JAC, including ghost beam hunting and fixing a shorting cable. OMC scans were also taken today at points when the IMC could be locked (manually, since the IMC_LOCK Guardian is struggling to lock the IMC itself). HAM7 sits with its doors on, but pumpdown timing is still being discussed.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
22:49 SAF LVEA is Laser SAFE LVEA NO* LVEA is Laser SAFE *BIFURCATED HAM1/2 Ongoing
16:03 FAC Kim LVEA N Technical cleaning 17:22
16:40 FAC Randy LVEA N Cleanup in East bay 17:09
17:11 ISC Sheila, Matt LVEA - Opening light pipe; plugging in SR785 to IMC board 17:38
17:53 FAC Randy LVEA N Cleanup in East bay 18:11
18:03 TCS Camilla, Sophie Prep Lab Local CHETA work 19:52
18:03 ISC Jason LVEA - Turning up input power to 1W 18:17
18:23 JAC Keita, Masayuki LVEA YES JAC EOM work 20:39
18:23 JAC Jennie LVEA YES JAC EOM work 18:53
18:32 FAC Kim LVEA N Technical cleaning 19:18
20:30 SUS Rahul LVEA - Working on PM1 21:21
21:13 ISC Matt, Sophie LVEA - Turning sidebands on 21:17
21:17 CAL Tony Pcal Lab Local Restarting voltmeters 21:28
21:21 VAC Travis LVEA - Taking picture of HAM3 21:25
21:36 JAC Keita, Jennie Opt Lab N JAC cable work 23:25
21:37 JAC Masayuki, Jason LVEA YES JAC EOM work Ongoing
21:53 CAL Tony PCal Lab Local Closing shutter 21:58
21:59 CDS Dave MER N Scouting rack install 23:18
22:18 VAC Travis LVEA N Grabbing parts 22:23
22:25 SEI Jim LVEA N Looking at feedthrus on HAM3 23:25
23:13 ISC Matt LVEA - Unplugging SR785 23:16
23:25 JAC Keita, Jennie LVEA YES JAC work Ongoing
23:51 CDS Dave CER N Scouting rack install 00:12
00:23 CDS Fil, Marc LVEA N Measuring cables Ongoing
H1 AOS
matthewrichard.todd@LIGO.ORG - posted 15:30, Thursday 05 February 2026 (89056)
OMC scans -- 2026-02-05

M. Todd, S. Muusse, C. Compton, S. Dwyer


Today we ran some more OMC scans with the ITM ring heaters on. At first we ran the OMC scan with the 9/45 sidebands on, single bounce off ITMX. Then we turned sidebands off and did both ITMX and ITMY.

Measurement Time Test Masses CO2 [W] Ring Heater (per segment) [W] SR3 [W] OM2 [W] FOM Notes
OMC Scan - Single Bounce off of ITMX 1454352294 Cold 0 2.45  0 0 Mismatch = 20.2% Sidebands on
OMC Scan - Single Bounce off of ITMX 1454359702 Cold 0 2.45  0 0 Mismatch = 23.9% Sidebands off
OMC Scan - Single Bounce off of ITMY 1454360545 Cold 0 2.00 0 0 Mismatch = 22.7% Sidebands off
Images attached to this report
H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 08:37, Thursday 05 February 2026 - last comment - 15:15, Thursday 05 February 2026(89039)
HAM1 alignment/mode-match finalization and align the TRANS PD path

[Keita, Jennie, Masayuki]

Summary

After completing the EOM alignment, we realigned the beam to the IMC using JM2 and JM3. During this process, a large shift of JM3 was observed and corrected. By iteratively moving JM3 toward the PSL side, the mode mismatch was improved to below 1%. We also confirmed the presence of a diffracted s-polarization beam from the EOM with an angle consistent with expectations. Finally, we measured the power throughput and completed the alignment of the TRANS PD path.

Details

After finishing the EOM alignment, we attempted to align the beam to the IMC using JM2 and JM3. At this point, we noticed that JM3 had moved significantly. We re-tightened the dog clamp and the mirror mount, after which the alignment recovered. The mode mismatch (ratio of 2nd-order mode height to TEM00) after this realignment was 0.93/48 = 1.94%. This is actually worse than we measured yesterday, would be because the EOM crystal clipping was solved and it changed the beam shape.

To further improve the mode matching, we decided to move JM3. First, JM3 was shifted by 0.5 inch toward the PSL side, which improved the mode mismatch to 0.57/41.8 = 1.36%. We then moved JM3 by an additional 0.5 inch, resulting in 0.32/39 = 0.821 %. A further 0.5 inch shift improved the mismatch to 0.25/38.8 = 0.644%. Since this level was sufficient, we stopped the adjustment at this point.

We observed an additional beam separated by approximately 1 cm from the main beam at a distance of about 1 m from the EOM. This is likely the s-polarization beam diffracted by the EOM. The relative angle between the two beams was approximately 0.5 degrees, which is consistent with expectations.

We then measured the power throughput using a power meter. The measured powers were 94 ± 2 mW at the EOM output, 95 ± 3 mW at the EOM input, 7 ± 1 mW at JAC REFL, and 100 ± 1 mW at JAC input which indicates no significant loss.

We also checked the beam position at JM3 and confirmed that it was shifted by approximately 1/4 inch in the +y direction. The iris after JM3 and the iris after the periscope were then centered.

Finally, we aligned the TRANS PD path. Using JACT_BS1, we temporarily installed an HR mirror in place of the laser window to obtain sufficient light to the TRANS PD. With this configuration, we aligned the photodiode such that the reflected beam from the PD was properly dumped into the beam dump. Also, we make sure the transmission beam from the laser window will be cought by the same beam dump by removing the mirror and make sure that the beam coming from the JAC is directly hitting the beam dump.

Pictures:
JM3 beam position, JM3 position, and L2 position

Images attached to this report
Comments related to this report
masayuki.nakano@LIGO.ORG - 17:49, Wednesday 04 February 2026 (89040)

The JAC scan gives 0.347(TEM00) to 0.00425+0.0013+0.009= (other small peaks). The total fraction of small peaks is 4%. So, 4mW of the 7mW at the reflection includes all of these fractions.

Images attached to this comment
keita.kawabe@LIGO.ORG - 22:56, Wednesday 04 February 2026 (89041)

EOM alignment

This was a 2-day's worth of job. It was briefly reported in the alog from the first day (89018) but I'll repeat what was already reported so people can see what was done concerning EOM alignment in a single log.

Day 1:

After we thought we completed the mode matching yesterday, we found that the beam has a halo that looked like a weird horizontal streak (horizontalstreak.jpg). It seemed as if it came from the EOM itself.

Eventually we found that the beam coming out of JAC looked as if it's higher than the EOM crystal center by more than 1mm (sorry no picture). We raised the entire EOM assembly by about 1mm by inserting shims under the EOM base plate at 3 locations. shims.jpg is the top view of the EOM, see shim_front.jpg and shim_back.jpg for the close up of the shims. (Each shim is actually two 91080A026 flat slotted washers, each washer is 0.02 to 0.026" thick, so the EOM got higher by anywhere from 0.04 to 0.052" or 1 to 1.3mm.)

After this, the horizontal streak was gone but there was still a vertical streak that was hard to photograph. We checked the horizontal beam position on the EOM input aperture and it looked awfully close to what is supposed to be the edge of the crystal (EOM_IN_horizontally_off.jpg). 

We pushed the EOM in -Y direction by 1mm or so, the input beam position looked good, we realigned the beam downstream of the EOM, measured the mode matching, that was great and we were happy. But I thought that the beam still looked a bit weird vertically (though better than before), it was better than before but weird. We checked the beam position on the EOM output and it was off (EOM_OUT_horizontally_off.jpg).

At this point we wanted to make an YAW adjustment for the EOM pivot plate. It turns out that we had to undo the hard-to-access screw I reported before (caution.jpg) and it was impossible to access when the SMA elbow was connected, a regular Allen key (or even the ball end one) interfere with the connector. It's not a huge interference but I worried that I'll damage the SMA, so we stopped it and called it the day.

(Note for the future design: Why don't we relocate the bolt to the opposite corner (relocate_bolt.png)? )

Day 2:

Ibrahim found us a cut Allen key that fits under the SMA (short_allen.jpg, short_allen2.jpg). We loosened three bolts circled in green in three_bolts.jpg and rotated the pivot plate. It was tedious and we needed three iterations, but we managed to reasonably center the beam position at both the input and the output of the EOM (good_in.jpgok_out.jpg).

(Note for the future design: As of now there is no visual guide for the beam center position for the output side plate (the guide is on the opposite face that is not visible). This is because the input and the output side plate are the same thing (https://dcc.ligo.org/D2500128) and it only has visual guide on one face and not on the opposite face. Give me the visual guide like in visual_guide.png.)

Note: We had to loosen the strain relief such that the cables can slide inside the viton pads, otherwise the tension and stiffness of the cable act as tough springs and the pivot plate will spring back after rotation, so everything will be tedious. For each strain relief, I left one of the two tiny screws somewhat loose, and made the jam nut finger tight. The cable won't go anywhere and it still acts perfectly as a strain relief.

This was the end of the EOM alignment.

The beam shape looks better than at the end of Day 1, not sure if it's great though, it's hard to photography but something faint might be coming out of the EOM. 

The last picture shows the ghost beam which is likely in the wrong (S) polarization.

Images attached to this comment
keita.kawabe@LIGO.ORG - 22:58, Wednesday 04 February 2026 (89042)

EOM crystal serial number

Marking on the RTP container: #B1913109, 20000488M (the former is S/N)?

Images attached to this comment
corey.gray@LIGO.ORG - 15:15, Thursday 05 February 2026 (89055)EPO

Tagging for EPO.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 03:51, Thursday 05 February 2026 - last comment - 15:15, Thursday 05 February 2026(89043)
HAM7-Y Door is on

(Randy, Jordan, Travis, Gerardo)

After lunch we installed HAM7 -Y door.  Note regarding this door to keep it on our memory, there are nicks and a few scratches on the door's flange surface, between 10 to 2 O'clock, but the scratches and nicks are away from the O-ring sealing area.

BTW, all the bolts are on the door, and they were torqued.

We also installed the two 12" OD blanks on the +Y door access ports.  Both blanks will need to be tested for leaks once the chamber is pumped down.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 15:15, Thursday 05 February 2026 (89054)EPO

Tagging for EPO.

H1 CAL
anthony.sanchez@LIGO.ORG - posted 16:40, Wednesday 04 February 2026 - last comment - 15:26, Thursday 05 February 2026(89035)
PCAL EY End Station Measurement

Dripta and I went to EY yesterday (Feb 3rd) to do both an ES and a TX module maintanence. We followed T1500062-v21 with out much deviation until the end when we started the TX module maint.

Obligitory Before and After beam spots on the apature of RX sphere.

Data Analysis: 
python3 generate_measurement_data.py --WS PS4 --date 2025-11-03 
Reading in config file from python file in scripts
../../../Common/O4PSparams.yaml
PS4 rho, kappa, u_rel on 2025-11-03 corrected to ES temperature 299.4 K :
-4.701912257515925 -0.0002694340454223 2.686163396659873e-05
Copying the scripts into tD directory...
Connected to h1daqnds1
martel run
reading data at start_time:  1454177475
reading data at start_time:  1454177902
reading data at start_time:  1454178300
reading data at start_time:  1454179000
reading data at start_time:  1454179400
reading data at start_time:  1454179750
reading data at start_time:  1454179900
reading data at start_time:  1454180530
reading data at start_time:  1454180888
Ratios: -0.5341330662181019 -0.5436335114505099
writing nds2 data to files
finishing writing
Background Values:
bg1 =        18.796205; Background of TX when WS is at TX
bg2 =        5.033949; Background of WS when WS is at TX
bg3 =        18.801656; Background of TX when WS is at RX
bg4 =        5.198797; Background of WS when WS is at RX
bg5 =        18.803508; Background of TX
bg6 =        -0.514446; Background of RX

The uncertainty reported below are Relative Standard Deviation in percent 

Intermediate Ratios
RatioWS_TX_it      = -0.534133;
RatioWS_TX_ot      = -0.543634;
RatioWS_TX_ir      = -0.526715;
RatioWS_TX_or      = -0.535124;
RatioWS_TX_it_unc  = 0.054072;
RatioWS_TX_ot_unc  = 0.053357;
RatioWS_TX_ir_unc  = 0.053158;
RatioWS_TX_or_unc  = 0.054774;
Optical Efficiency
OE_Inner_beam                      = 0.986243;
OE_Outer_beam                      = 0.984385;
Weighted_Optical_Efficiency        = 0.985314;

OE_Inner_beam_unc                  = 0.041515;
OE_Outer_beam_unc                  = 0.041813;
Weighted_Optical_Efficiency_unc    = 0.058922;

Martel Voltage fit:
Gradient      = 1637.852893;
Intercept     = 0.265584;


 Power Imbalance = 0.982524;

Endstation Power sensors to WS ratios::
Ratio_WS_TX                        = -0.927845;
Ratio_WS_RX                        = -1.384820;

Ratio_WS_TX_unc                    = 0.044117;
Ratio_WS_RX_unc                    = 0.038945;

=============================================================
============= Values for Force Coefficients =================
=============================================================

Key Pcal Values :
GS           =      -5.135100; Gold Standard Value in (V/W)             
WS           =      -4.701912; Working Standard Value             

costheta     =      0.988362; Angle of incidence
c            =      299792458.000000; Speed of Light
             
End Station Values : 
TXWS         =        -0.927845; Tx to WS Rel responsivity (V/V)
sigma_TXWS   =        0.000409; Uncertainity of Tx to WS Rel responsivity (V/V)
RXWS         =        -1.384820; Rx to WS Rel responsivity (V/V)
sigma_RXWS   =        0.000539; Uncertainity of Rx to WS Rel responsivity (V/V)

e            =        0.985314; Optical Efficiency
sigma_e      =        0.000581; Uncertainity in Optical Efficiency

Martel Voltage fit : 
Martel_gradient         =        1637.852893; Martel to output channel (C/V)
Martel_intercept   =        0.265584; Intercept of fit of     Martel to output (C/V)

Power Loss Apportion : 
beta          =        0.998844; Ratio between input and output (Beta)  
E_T          =        0.992056; TX Optical efficiency 
sigma_E_T          =        0.000292; Uncertainity in TX Optical efficiency 
E_R          =        0.993204; RX Optical Efficiency 
sigma_E_R          =        0.000293; Uncertainity in RX Optical efficiency 

Force Coefficients : 
FC_TxPD          =        9.154540e-13; TxPD Force Coefficient 
FC_RxPD          =        6.225064e-13; RxPD Force Coefficient 
sigma_FC_TxPD          =        4.888564e-16; TxPD Force Coefficient 
sigma_FC_RxPD          =        3.063586e-16; RxPD Force Coefficient 
data written to ../../measurements/LHO_EndY/tD20260203/

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
anthony.sanchez@LIGO.ORG - 15:26, Thursday 05 February 2026 (89052)

TX module maintenance of End Y was done with reference to T1600436-v12

Date   Feb 3rd 2026
Laser Shutter Check   Pass
Max OFS Offset   8
95% OFS Offset   7.6
Operating OFS Offset   3.8
Laser Output Power   1.94W
After-Laser Rejected Power   3.96mW
AOM Input Power   1.88W
Max Diffracted Power   1.58 W
Un-Diffracted Power   155mW
AOM Diffraction Efficiency   84.04%
After-AOM Rejected Power   13.9mW
TxPD Power   13.7mW
OFSPD Power   6.59mW
Outer Beam Power   0.747W
Inner Beam Power   0.744W
Output Beam Power Ratio   0.995
OFS Gain   37.79
OFS Phase Margin   57.6
Non-image files attached to this comment
Displaying reports 241-260 of 86768.Go to page Start 9 10 11 12 13 14 15 16 17 End