Displaying reports 35241-35260 of 89220.Go to page Start 1759 1760 1761 1762 1763 1764 1765 1766 1767 End
Reports until 11:22, Sunday 15 March 2020
H1 General
camilla.compton@LIGO.ORG - posted 11:22, Sunday 15 March 2020 - last comment - 15:09, Sunday 22 March 2020(55601)
H1 Lockloss 17:53:18 UTC
Lost lock, after 29h30m lock. No obvious reason why. Environmental conditions are all fine.
 
No green light on the y-arm yet so it will be interesting to see how increase flashes copes with that, I may have to intervene.
We saw this before after a long 58 hour lock with dramatic temperature change (alog 54580). Though I'm not sure we can blame temperature here as the outside temurature has only changed +3°C since the last lock. -16°C  change from a few days go it if takes the equipment a long time to change temperature.
Comments related to this report
camilla.compton@LIGO.ORG - 12:23, Sunday 15 March 2020 (55602)

Had to move both ETMY and TMSY a decent way to get light onto the camera (20 counts on the slider for ETMX Pit and 15 for TMSY Pit).

camilla.compton@LIGO.ORG - 14:54, Sunday 15 March 2020 (55604)

Had dragged TMSY too far away from it's normal position so that we were not getting enough light on the QPD's and had ALS_Y in fault with message "PZT trig servo not on". I think this is a different problem to the one disrupted by TJ in alog alog 54526 but it may be related.

Keita helped me to get light back on the QPDs:

From sitemap>ALS> End Y overview > QPD_A (FE) and QPD_B (FE), the NSUM should be ~1 for green arms locking, was closer to 0.05. after deciding that the suspensions had bee moved too much, we the QPD NSUMs and the alignment sliders and reverted them to their values at the time of last green arms locking, then did the same for the BIAS on the PZT1 and PZT2 PIT and YAW filters. NSUMS then came up to 1. Rs-adjusted the PZT filter bias numbers to be closer to their outputs and then continued locking.

We have just lost PRMI lock (not total lock) 3 times after I noticed that the FIND_IR TR_Y trace was way higher than it should be attached photo (20 rather than 1). Something strange must have happened so I will try and initial alignment.

Images attached to this comment
camilla.compton@LIGO.ORG - 15:16, Sunday 15 March 2020 (55606)

DRMI locked straight away after a smooth initial alignment. Unsure what went on before!

camilla.compton@LIGO.ORG - 15:51, Sunday 15 March 2020 (55607)ISC
22:44 Observing 
A very long lock acquisition, will try to understand this week what are the best steps to take in the future when there is no light on the Green camera.
Accepted the SDFs related the the PZT BIAS changes I made with Keita, attached. tagging ISC ?
Images attached to this comment
camilla.compton@LIGO.ORG - 15:09, Sunday 22 March 2020 (55727)ISC

At the start of this locking acquisition, there was no light on ALS-Y_QPD_B (NSUM ~  0.05). ALSY PZT1 and PZT2 were then not adjusted. It can be seen from the top 2 plots of the attached image  that they were being significantly adjusted at the start of each green arms locking to maximize QPD light. Keita helped me move the PZT1 and PZT2 values closer to the values which maximized NSUM in the past so in future they should not be so far away. 

It's still not clear why there was no light on QPD_B to start with. The values of PZT1 and PZT2 do drift over time to give the maximum light on the QPDs so it's possible that to was just out of range. Maybe these biases should be checked and adjusted more regularly. Tagging ISC

Images attached to this comment
H1 General
camilla.compton@LIGO.ORG - posted 08:11, Sunday 15 March 2020 (55600)
Shift transition to Day
TITLE: 03/15 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 114Mpc
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_CONF state: WINDY
    Wind: 24mph Gusts, 15mph 5min avg
    Primary useism: 0.08 μm/s
    Secondary useism: 0.31 μm/s
QUICK SUMMARY: Locked 26h45. Steady winds of 20+mph.
Our range is a little lower than expected but I can't see a reason why so putting it down to the wind. 
H1 General
cheryl.vorvick@LIGO.ORG - posted 00:18, Sunday 15 March 2020 - last comment - 08:46, Monday 16 March 2020(55599)
OPS Eve Summary

TITLE: 03/15 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 114Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: Locked in Observe, range trending downward, no obvious reason
LOG:

Comments related to this report
sheila.dwyer@LIGO.ORG - 13:50, Sunday 15 March 2020 (55603)

After Rhaul loaded his change to the guardian code on Friday 55575, I expected that the guardian would no longer be setting gains to 0.  It looks like we have lost lock and relocked since the guardian change was loaded so it should have gone into place.  Did you find the gain set to 0?  

rahul.kumar@LIGO.ORG - 08:46, Monday 16 March 2020 (55614)

Here Guardian is setting it to max gain which is 0 for ITMY 12 (nominal is 1200).

H1 General
cheryl.vorvick@LIGO.ORG - posted 20:38, Saturday 14 March 2020 (55598)
Mid-shift Update:

H1 locked in Observe, winds are below 20mph, and temperatures are below freezing, around 29F, -1.2C.

H1 General
cheryl.vorvick@LIGO.ORG - posted 16:17, Saturday 14 March 2020 (55597)
OPS Eve Transition

TITLE: 03/14 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 117Mpc
OUTGOING OPERATOR: Camilla
CURRENT ENVIRONMENT:
    SEI_CONF state: WINDY
    Wind: 22mph Gusts, 20mph 5min avg
    Primary useism: 0.06 μm/s
    Secondary useism: 0.32 μm/s
QUICK SUMMARY: locked in Observe

H1 General
camilla.compton@LIGO.ORG - posted 16:07, Saturday 14 March 2020 (55596)
Shift Summary - Day

TITLE: 03/14 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
INCOMING OPERATOR: Cheryl
SHIFT SUMMARY: Locked 10h45. Quiet day with steady 20+mph wind.
LOG: 20:27 UTC Trolley rolled down corridor (noting incase it caused noise in IFO)

H1 General
camilla.compton@LIGO.ORG - posted 12:03, Saturday 14 March 2020 (55595)
Mid Shift Summary

Observing for 6h45. Wind is still ~20mph and microseism has started to increase over the last few hours.

H1 General
camilla.compton@LIGO.ORG - posted 08:19, Saturday 14 March 2020 (55594)
Shift transition to Day
TITLE: 03/14 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 117Mpc
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_CONF state: WINDY
    Wind: 18mph Gusts, 14mph 5min avg
    Primary useism: 0.05 μm/s
    Secondary useism: 0.23 μm/s
    Less than 1/2 inch of snow outside, it has mainly blown off the 240 but has stuck to Rt 10. Currently 26degF
QUICK SUMMARY: Last night (10:48 UTC) a 6.7M EQ from Tonga knocked us out of lock 30 minutes after SEI_ENV had taken us to EQ mode. Back to observing at 12:15UTC.
H1 General
cheryl.vorvick@LIGO.ORG - posted 00:04, Saturday 14 March 2020 (55593)
OPS Eve Summary

TITLE: 03/14 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 113Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: H1 currently locked in Observe, out once to relock SQZ
LOG:

H1 SQZ
cheryl.vorvick@LIGO.ORG - posted 22:51, Friday 13 March 2020 (55592)
relocking SQZ, ring up preceded by a 2V change in PZT_1_VOLTS

Attached are 3 plots showing the change in voltage on the PZT, then SQZ rining up, and then relocking.  The first relock was unchanged, second was a tiny bit better, and third caught and SQZ rang down.  H1 out of Observe at 05:20:57UTC, and back in Observe at 05:33:45UTC.

Images attached to this report
H1 AOS (DetChar)
adrian.helmling-cornell@LIGO.ORG - posted 20:12, Friday 13 March 2020 (55591)
A (very late) DQ Shift Report for March 2 - 9

Shifter: Adrian Helmling-Cornell

Fellow: Phil Jones

DQ Shift Summary:

DQ Shift Page

H1 General
cheryl.vorvick@LIGO.ORG - posted 20:06, Friday 13 March 2020 (55590)
OPS Mid-Shift Update:

Locked in Observe, winds have come down to less than 10mph, range is 117Mpc.

H1 General
cheryl.vorvick@LIGO.ORG - posted 16:22, Friday 13 March 2020 (55589)
OPS Eve Transition

TITLE: 03/13 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 119Mpc
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_CONF state: WINDY
    Wind: 13mph Gusts, 10mph 5min avg
    Primary useism: 0.04 μm/s
    Secondary useism: 0.18 μm/s
QUICK SUMMARY: locked in Observe

LHO General
thomas.shaffer@LIGO.ORG - posted 16:13, Friday 13 March 2020 (55582)
Ops Day Shift Summary

TITLE: 03/13 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
INCOMING OPERATOR: Cheryl
SHIFT SUMMARY: Two lock losses this shift for unknown reasons. Both relocks were a non-issue.
LOG:

H1 General
thomas.shaffer@LIGO.ORG - posted 14:57, Friday 13 March 2020 - last comment - 15:37, Friday 13 March 2020(55587)
Lock Loss 2154

Unknown reason for lock loss at this time.

Comments related to this report
thomas.shaffer@LIGO.ORG - 15:37, Friday 13 March 2020 (55588)

Observing 2234 UTC.

40min autonomous relock from lock loss to observing. Maybe an O3 LHO record? Well done everyone!!

Displaying reports 35241-35260 of 89220.Go to page Start 1759 1760 1761 1762 1763 1764 1765 1766 1767 End