FAMIS 11527 I think I accidentally put too much water in the TCS Y chiller. I didn't see it do anything abnormal though. TCS X Before: 28.9 After: 30.5 Added: 150 mL TCS Y Before: 9.3 After: 10.8 Added: 100 mL
End-Y - Trip level was lowered to 8 10/16. The level is 8 14/16 - No change from last check. End-X - The level is 7 14/16 - No change from last check. CS - The level is 6 6/16 - No change from last check. No new leaks were observed on any platforms. Closing FAMIS #13005
TITLE: 01/21 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Preventive Maintenance
OUTGOING OPERATOR: Cheryl
CURRENT ENVIRONMENT:
SEI_CONF state: SC_OFF_NOBRSXY
Wind: 5mph Gusts, 3mph 5min avg
Primary useism: 0.04 μm/s
Secondary useism: 0.51 μm/s
QUICK SUMMARY:
Maintenance Begins!
Looks like X-arm access is clear all the way to EX except for a pile right at EX (tyler/scott are on it).
TITLE: 01/21 Owl Shift: 08:00-16:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Preventive Maintenance
INCOMING OPERATOR: Corey
SHIFT SUMMARY: locked most of the shift, relocking took a while, H1 reached NLN and then I ran the PEM injections (cold H1)
LOG:
attached are the SDF Diffs that I had when H1 reached NLN, CALCS? about 20 diffs, and ETMY SUS offsets, the same diffs that Camilla aloged, alog 54597.
MY AH REHEAT 20 days
locked in Observe, 2 EQs, EQ mode, damping ETMY violin mode 20
TITLE: 01/21 Owl Shift: 08:00-16:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Observing at 119Mpc
OUTGOING OPERATOR: Ed
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 4mph Gusts, 3mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.45 μm/s
QUICK SUMMARY: H1 locked in Observe
TITLE: 01/21 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
INCOMING OPERATOR: Cheryl
SHIFT SUMMARY:
LOG:
LOG:
Wasn't going to jinx myself by logging that everything was ok.....until:
04:13 Lockloss - no apparent reason - initiating recovery now
DCHILFLOW seems to be "slinkying" it's way down the "stairs". I suppose eerything else looks ok. Camila had the FSS go into oscillations on her twice during an initil alignment (during PRC align, i believe).
TITLE: 01/21 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 118Mpc
OUTGOING OPERATOR: Camilla
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 0mph Gusts, 0mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.44 μm/s
QUICK SUMMARY:
H1 locked (4:52) and Observing 118Mpc
D. Gustafson on site
TITLE: 01/21 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 117Mpc
INCOMING OPERATOR: Ed
SHIFT SUMMARY:Spent some time struggling to get back to NLN after lockloss (not sure reason for lockloss yet), and calebration measuremnets in the afternoon.
LOG:
CP6 Dewar level started alarming at 15% full. We have an LN2 delivery scheduled for Wednesday now that X1 is clear of tumbleweeds, and want to avoid Dewar level falling below 10%, at which point contaminate (ice crystals) buoyancy level meets CP supply line and thus potential for ice plugs is high.
Ignore alarms for now. LN2 consumption averages 1.2% per day.
Glad to hear operators are not being pestered with these alarms at console. I've asked Dave Barker to silence text alarms.
As requested by Chandra, I've bypassed the CP6 LN2 Dewar cell phone alarm for 24 hours, we can extend this as needs be.
Bypass will expire:
Tue Jan 21 15:49:45 PST 2020
For channel(s):
H0:VAC-MX_CP6_LT355_DEWAR_LEVEL_PCT
Thank you Chandra. We have not yet had any CR alarms.
Actually since the delivery is not scheduled until Wednesday this week, I've extended the bypass till then.
Bypass will expire:
Wed Jan 22 16:03:24 PST 2020
For channel(s):
H0:VAC-MX_CP6_LT355_DEWAR_LEVEL_PCT
Attached is a dtt file with measured transfer functions from master outs to the newly installed noise monitors for ITMX. I started but didn't finish making the same set of measurements for ITMY.
Today I copied the AntiLPL2 filters from the coilout filters to the noise mon filters, and engaged them. When we are locked the pum coil drivers are in state 3, Acq filter on, low pass filter off. According to T1100378-v12 the noise monitor is after the low pass filter, but before the acquisition filter. In low noise the coil drivers are in state 3, LP on Acq off, so I set them into that state and measured the transfer functions from master out to noise monitor for each coil on both ITMX and ITMY.
Attached are the new measurements.
The levels of the water after your filled look okay. I wouldn't worry about it being overfilled here.
She'll be 'right.