At 21:31:21utc, H1 was dropped out of OBSERVING. I did a quick scan and did not see anything obvious. There have been several Guardian Nodes which are NOT green: ISI (for a TRamp, Violins (from violin activities), Inj Trans (it's waiting to go back to Success state), etc, but these remained as they have been. A few notes:
Will keep the Guardian and SDF medm windows open in case there's another drop---so I can hopefully catch what flashes and sheds ligth as to the culprit here.
21:33:04utc Went back to OBSERVING
This morning (17:08utc) we had an EQ alert for an earthquake on the small side (according to the SEISMON EQ Response plot). On the other operator work station, I run with Seismon, SEI_CONF, and SEI channels up on ndscope to monitor EQs; I generally only transition to the EARTHQUAKE state when I see H1 begin to show effects from the EQ (via IMC, ASC, and seismometers).
The earthqake turned out to in fact be a small one with seismicity being elevated from 17:30-18:04utc as seen on the STSs. So I ignored this EQ.
H1 Transitioned to EARTHQUAKE, But By Whom?
At about 17:40utc received call from Eyal S. at LLO via TeamSpeak letting us know that we were still in the EARTHQUAKE state (Thanks for catching this, Eyal!). I checked the SEI_CONF log and it transitioned from WINDY to EARTHQUAKE at 17:12utc. I'm certain I did not transition it, and Cheryl was here, but she did not tell me she transitioned, so I'm going to guess an operator didn't transition. There are plans to automate how we handle earthquakes (see Jim's alog from last Thursday), but I don't think this is in effect yet.
18:09: Since this EQ was negligible & the small seismic waves were long gone, I transitioned H1 back to the WINDY state.
Timeline Summary:
Unless TJ has worked on this in the last couple of days, there is no automatic transition of SEI_CONF.
None of this is automated yet, this was done by a person.
Addressed TCS Chillers (16:48-16:56utc & with Cheryl as my Buddy since she was also here while I did this):
HAM & BSC CPS: All CPS' look good for high-freq range from attached plots.
TITLE: 12/22 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 117Mpc
OUTGOING OPERATOR: Cheryl
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 9mph Gusts, 6mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.37 μm/s
Microseism continues downward trend for last 24hrs (was squarely on 90th percentile @ ~0.5um/s, and is now in between 90th & 50th percentile @ ~0.3um/s); Yay. Other than that very quiet sitewide with only the one EQ Ed rode through during the EVE shift.
QUICK SUMMARY:
H1's been locked for over 41hrs with a nice range just under 120 Mpc.
TITLE: 12/22 Owl Shift: 08:00-16:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: H1 in Observe all shift
LOG:
Violin Damping, non-nominal values:
The change of guardian code to allow violin mode damping gain changes during Observe has paved the way to damp previously undamped modes, reducing the peaks in DARM. To capture what I've learned, changes I've made, planned changes, I'm putting in a summary.
Overall, I've found a number of filters that need to be adjusted, both damping and monitor filters, and typically the damping filters need to be more narrow, and monitor filters need to be biased to reduce the overlap for two close peaks.
STATE of H1: Observing at 114Mpc, Observe clock a 7.5 hours
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 4mph Gusts, 3mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.42 μm/s
CHECKUP ON CORNER STATION PUMP: when listening from the kitched door, all is well, normal background noise
TITLE: 12/22 Owl Shift: 08:00-16:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
OUTGOING OPERATOR: Ed
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 5mph Gusts, 4mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.37 μm/s
QUICK SUMMARY: H1 in Observe, and I noticed the chiller yard at the Corner Station making a loud clicking sound
When I arrived for my shift I noticed a loud clicking sound coming from the Corner Station chiller yard. Ed and I walked out there, and there is a "Switching Failure Alarm," on the pannel near the motor that's making the noise. There's also a periodic release of air. This is not normal for the chiller yard. There's no alarms in the CR, however, this is new since this morning. Calling Richard.
I called Kyle, and he's coming to fix the motor, since it's been running continuously for more than an hour. It was running when I arrived on site at 7:55 UTC, and is still running.
Kyle and I ran through all potential issues, results are below:
ACTIONS as result of the motor running continuously: Kyle is on his way to the site
Approx 9:30 UTC Kyle completed adjustments that put the motor back into a mode of running for a few minutes every few hours.
I'm reposting one picture for OPS, that shows the dial gauge on the tank. This gauge shows the pressure in the tank, and is related to instrument air, in that instrument air can be at this pressure, or above this pressure, but not below this pressure.
When I arrived, I found that the Corner Station instrument air compressor was running continously and the receiver tank's solenoid drain valve was venting every few seconds. I noted that user-adjustable settings (vent duration and frequency) on the new "fancy" adjustable auto-drain valve were incorrectly set. The drain valve was venting the receiver tank too long (6 seconds) and too frequently (every few seconds) such that the compressor was never ever able to increase the system pressure to the shut-off setpoint and the two were in equilibrium at 84 psi (see aLOG entry from Friday) -> THE FIX -> I minimized the duration setting and maximized the frequency setting to allow the compressor to "catch up" with the loss to atmoshpere and, soon, the system pressure reached the setpoint and the compressor shut-off.
TITLE: 12/22 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 117Mpc
INCOMING OPERATOR: Cheryl
SHIFT SUMMARY:
All of the fun stuff is in the previous logs from this shift
It has been drizzling rain off and on and Danny Sellers contaced me from LLO earlier to inform me of their lockloss following a "rung-up" BRS. They are back up and running
Cheryl has informed me of a mechanical sounding, clicking noise eminating from the corner station.
LOG:
04:56 The squezer fell out of lock again
04:57 H1 back to Observing
03:38 Verbal Alert and Telephone call
03:39 Conferrred with Joe Hanson at Livingston and confirmed the Superevent.
03:40 Conference in EM Follow-up channel on TS is concurring that it looks like a clear BBH and that both observatories were in nominally good/non-glitchy condition at the time of the event.
01:14UTC The new coefficients were loaded as advised by Cheryl's alog , pursuant to the squeezer lockloss and the opportunity afforded during temporary loss of the Oberving Intent Bit.
After watching the manager try and fail to re-lock a number of times, I took the very un-scientific approach while I looked through alogs.
01:23 H1 back to Observing
Perhaps the timing is a bit to quick for the system to stabilize for a solid locking sequence? anyone?
TITLE: 12/22 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 116Mpc
OUTGOING OPERATOR: Camilla
CURRENT ENVIRONMENT:
SEI_CONF state: WINDY
Wind: 17mph Gusts, 12mph 5min avg
Primary useism: 0.05 μm/s
Secondary useism: 0.43 μm/s
QUICK SUMMARY:
You can always look at the guardian log of any node using guardctrl command (see below).
In this case, TCS ETMX ring heater Beckhof was unresponsive for a few seconds though I don't know why.
Edit later: If you're in front of the workstation when something happens, you don't have to use command line, you can open MEDM screen for whatever node (in this case IFO node by pressing "GRD IFO" button at the top of the guardian overview screen) and then press the "log" button to see what's going on in real time.