Displaying reports 21-40 of 86532.Go to page 1 2 3 4 5 6 7 8 9 10 End
Reports until 15:30, Thursday 05 February 2026
H1 AOS
matthewrichard.todd@LIGO.ORG - posted 15:30, Thursday 05 February 2026 (89056)
OMC scans -- 2026-02-05

M. Todd, S. Muusse, C. Compton, S. Dwyer


Today we ran some more OMC scans with the ITM ring heaters on. At first we ran the OMC scan with the 9/45 sidebands on, single bounce off ITMX. Then we turned sidebands off and did both ITMX and ITMY.

Measurement Time Test Masses CO2 [W] Ring Heater (per segment) [W] SR3 [W] OM2 [W] FOM Notes
OMC Scan - Single Bounce off of ITMX 1454352294 Cold 0 2.45  0 0 Mismatch = 20.2% Sidebands on
OMC Scan - Single Bounce off of ITMX 1454359702 Cold 0 2.45  0 0 Mismatch = 23.9% Sidebands off
OMC Scan - Single Bounce off of ITMY 1454360545 Cold 0 2.00 0 0 Mismatch = 22.7% Sidebands off
Images attached to this report
H1 SEI
jeffrey.kissel@LIGO.ORG - posted 14:32, Thursday 05 February 2026 (89047)
Setting Lens Position for SPI In-vac SuK Fiber Collimator Beam Profile: Systematic Error in Model Identified
J. Kissel, J. Warner (with sage wisdom provided remotely by S. Koehlenbeck)

After mounting fiber collimator S0272503 within its predestined adapter ring into the actuator-less IXM100 mount (one of two from LHO:86299) -- creating a fiber collimator assembly (D2400146) -- we connected the integrated fiber patch cord of the fiber feedthru S3228003 (see LHO:89029) to the collimator.

With this final, "as it shall now be forever" integrated fiber-optic-to-free-space converter system for the SPI MEAS path, we want to adjust the collimator's lens position to achieve "optimal collimation;" having the waist of the beam at the fiber collimator. To avoid obnoxious factors of two mistakes, the variable w shall be used to refer to beam radii as a function of position, z, along the beam projection, and thus w0 shall refer to the beam radius at beam projection distance, z = 0. Whenever diameters = 2*radii are envoked, D, shall be used, e.g. the beam waist diameter shall be D0, and the beam diameter as a function of beam projection distance shall be D(z).

Previous attempts to adjust the lens position of the S0272502 collimator are described in LHO:84825 (dirty, using the AxcelPhotonics Butterfly Dioe Laser) and again in LHO:86350 (post-Class-B bake, also with AxcelPhotonics laser). 

The method used in those entries was repeated: 
    (0) Set up a 1064 [nm] fiber-coupled seed laser, and markers of beam projection distances at z = [0.508 0.991 1.499 2.007 3.251 4.496 5.41] [m] (Roughly even half-meter points that one gets from "finding 1 inch hole position that gets you close to 0.5 [m] increments", i.e. z = [1'8", 3'3", 4'11", 6'7", 10'8", 14'9", and 17'9"])

    (1) Calculate the expected beam diameter at furthest projection distance, 
            D(z = 5.41 [m]) = (2 * w0) * \sqrt(1 + (z / z_{R})^{2})
        assuming 
            z_{R} = \pi * w0^{2} / \lambda,
            \lambda = 1064-e9 [m], and
            w0 = 1.05e-3 [m]
        That works out to be D(z = 5.41 [m]) = 4073 [um].
    
     (2) Set up a beam profiler at the z = 5.41 [m] position (we again used the NanoScan, with the NanoScan-v2 software on a windows laptop), loosen the lens-position set screws with the tiny 1.2 [mm] flat-head screw driver, and use the proprietary SuK "eccentric key" cam adjuster to painstakingly iterate the position of the lens and gather a beam diameter measurement (using the "D4sigma" mean, waiting until the D4sigma standard deviation dropped below 1 [um]).

     (3) Once happy that the X and Y beam diameters are within ~50 [um] of the target D( z=5.41[m] ) = 4073 [um], fix the lens position by re-securing its tiny set screws, and confirm the beam X/Y diameter remain acceptable.

     (4) Move the beam profiler to each pre-designated beam projection position, z, and record the beam diameter.

We did a few extra things in our 2026-02-03 characterization as well:
     (0') Due to initial confusion of the wavelength of the AxcelPhotonics butterfly diode laser, we also set up an independent OzOptics DTS0081 fiber coupled laser as well, both per LHO Optics Lab SOP (M1800318). See attached picture of setup that allowed for easy interchange between the two seed lasers. We confirmed that they're both \lambda = 1064 [nm].

     (3') After the lens was secure and we confirmed that the beam diameter remained at 4073 [um], we wanted to check that disconnecting and reconnecting the fiber from/to the collimator wouldn't impact the lens position. As such, after the initial measurement [Trial 1], we disconnected / reconnected / remeasured three more times with the AxcelPhotonics laser, and once with the OzOptics laser, confirming that the beam diameter stayed at 
    DX (z = 5.41[m]) = 4079.6 +/- (17.3, 7.2)
    DY (z = 5.41[m]) = 4084.2 +/- (8.3, 3.7)
with the parentheticals being the standard deviation of the 5 values (matlab's std(vals)), and the standard error on the mean (std(vals) / sqrt(N=5)).

Confident that the lens position is where we wanted it to be, assuming w0 = 1.05e-3 and \lambda = 1064 [nm], we scanned the beam along the z position with both lasers. The raw scan data can be found in the attached OzOpticsLaser and AxcelPhotonics.txt files, in the JaMMT-friendly format of 
    Scan Position    X beam *radius*,     Y beam *radius*, 
    z, [cm]          wX(z) [um]           wX(z) [um]

Photos of raw data can be found in this album for the AxcelPhotonics measurement set, and that album for the OzOptics measurement set.

I then pushed that beam profile thru an updated version of my 2025 matlab code (which uses a la mode at its core; attached here for posterity) to fit the profile for waist radius, w0 and waist position, z0.
     The fit predicts a waist w0 of 
        (w0x, w0y)_OzOptics       = (0.7040, 0.6766) [mm] 
        (w0x, w0y)_AxcelPhotonics = (0.6807, 0.7049) [mm]
     at position 
        (z0x, z0y)_OzOptics       = (1.4940, 1.5837) [m]  
        (z0x, z0y)_AxcelPhotonics = (1.5803, 1.4869) [m]
See attached plot.

The good news is that the beam is of excellent Gaussian property (quite symmetric in X and Y), and the fit results of the profile measurements with two independent seed lasers agree. 

However -- even after fixing the method mistake of reading off the D4sigma as radii vs. diameter as was the case in the 2025-06-03 and 2025-08-07 measurement sets/fits -- there's a systematic error in this method :: setting the beam diameter to D(z=5.41 [m]) = 4073 [um] consistently puts the z position of the waist, z0 continuously ends up at ~1.5 [m], rather than at 0.0 [m] where we want it.

Having brought up this issue with our wise laser mage (Sina), she isn't surprised and helped guide us toward the source of systematic error: the statement in Figure 3 of Section 1 of the ISIK Assembly Procedure T2400413 about the waist of the beam being w0 = 1.05e-3 [m]. This is *an* ideal case, in which the parameters of the PM980 fiber as spec'd:
    - a core radius of 2.75 [um], 
    - a "nominal" numerical aperture of 0.12 [ ]
    - a the seed \lambda = 1064 [nm], 
that (using this MFD calculator) yields a "mode field diameter" MFD = 7.14 [um]. 
Setting the mode field diameter as the beam diameter, at lens D(1/e^2) = 7.14 [um] in the core, butt up against the f=11 [mm] focal length lens in the collimator, and that (using this Laser Spot Size Calculator, and an ideal M^2 = 1.0 [ ] beam quality) yields a "beam diameter at focus" -- the waist diameter, D0 = 2.09 [mm], or w0 = 1.045e-3 [m] ~ 1.05e-3 [m].

This answer to the calculation for w0 was originally pulled from the SPI conceptual design Slide 60 of G2301177, before we even settled on the particular parameters of the 60FC line of SuK collimators, hence the "nominal" value for numerical aperature.

In fact, the technical data for the 60FC-0-A11-03-Ti SuK collimator we're using says the numerical aperture is NA = 0.25 [ ]. That puts 
    - MFD = 4.67 [um], 
    - the waist diameter at D0 = 3.19 [mm], 
    - waist radius at w0 = 1.595 [mm],
    - Rayleigh Range at zR = 7.5 [m],
    - Diameter D(z=5.41 [m]) = 3931.3 [um]
... quite different.

In fact, if that is the case, and the Rayleigh Range is 7.5 [m] instead of 3.25 [m], then I need to extend our range of measurement positions well beyond 5.41 [m]...

But in short -- the core radius and *actual* numerical aperture may be different than even this specific FC model, and the focal length of the lens might not be exactly 11.00 [mm]. 

As such, we need to 
    - use this measured beam profile to backward propagate through the lens and *fit* for the MFD (convert it beam radius MFw) within the fiber (constant as a function of z along the fiber), rendering he need for a spec'd numerical aperture and core radius value moot.
    - use the *fit* MFw to then forward propagate out (again) through fiber collimator lens, and determine the waist radius, w0, of the out-going free-space beam.
    - from there, recalculate the Rayleigh Range and thus target Beam Diameter at each measurement position.

Back to the drawing board!
Images attached to this report
Non-image files attached to this report
H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 14:07, Thursday 05 February 2026 (89051)
HAM1: PM1 up and working/damping (health checks ongoing)

Turns out that my PM1 investigation (drop in bosem counts by 20%, as reported in 89017) was totally unnecessary and this was due to the coil driver swap in Dec 2025 (LHO alog 88998)  and I had totally forgotten about it.  

Long story short - this afternoon I went into HAM1 chamber and re-attached the BOSEMs, connected the cables, centered the flags and re-attached the beam dump behind it. I re-took the open light counts, applied the offsets and the gain for PM1 (see screenshot attached). I have accepted the new settings in the SDF (safe).

Latest OLC for PM1 are as follows,

UL 20800
LL 23100
UR 19100
LR 23200

Healths checks on PM1 is still ongoing (chamber side measurements shows peaks are in the right place, magnitude is off so we are looking into the new Coil Driver and if everything is consistent or not digitally), will post plots as soon I am able to process them.

 

Images attached to this report
H1 ISC (IOO, ISC, SYS)
keita.kawabe@LIGO.ORG - posted 13:11, Thursday 05 February 2026 (89049)
JAC heater problem (one pin in an in-vac tri-cable D2500336 shorts to the shielding) (Masayuki, Jennie, Daniel, Keita)

Masayuki found that the JAC heater driver shuts itself down as soon as it is turned on. 

Daniel traced this back to that the heater elements are grounded somewhere. Turns out that the DB15-DB9-DCPD to DB25 in-vac tri-cable (D2500336) has a defect. One pin for Heater B Terminal D is shorted to the shield. Since the shield is ultimately connected to the chassis ground of the heater driver, and since the driver is differential, this effectively short-circuits the output of one power amplifier to the ground. Masayuki has a spare, we'll check it in the lab and will replace it.

From outside of the chamber, using DB25 breakout board on D4-F10, when everything is connected in chamber, in-air pin 13 (shield of the cable inside the chamber) is shorted to pin 8, 9 and 22 with only 2 or 3 Ohm. These are all connected to Heater B Terminal D, which is one leg of the Heater B.  In addition, pin13 and pin 10 and 23 show 54 Ohm, these are Heater B Terminal C, so this is expected (nominal resistance of the heater is 50 Ohm). Nothing is shorted to the chamber gound, which is a good news.

When the DB15 connector (that goes to the two heater elements and a thermistor) was disconnected in chamber, the only short-circuit that was observed from outside the chamber was pin 13 to pin 22 (still 3 Ohm).

When the DB25 connector was disconnected at the cable bracket, nothing shorts to anything.

The conclusion is that something is wrong with the DB15-DB25 part of the tri-cable.

Oh, D2500119-v2 says that the JAC control is on D4-F9 of HAM1, but it's on D4-F10 at LHO. This is already captured in D1002872-v10.

Images attached to this report
H1 SQZ (CDS)
camilla.compton@LIGO.ORG - posted 12:52, Thursday 05 February 2026 (89048)
HAM7 Ground Loop Checks

Sheila and I followed T2200048 to check for ground loops in the SQZ rack yesterday while the HAM7 door was being put back on. Did not check picos. Results attached.

Strange ones noted below: 

Non-image files attached to this report
H1 IOO
sheila.dwyer@LIGO.ORG - posted 11:37, Thursday 05 February 2026 (89046)
IMC loop gain and power checks this morning

Sheila, Matt, Masayuki

This morning Matt and I brought the SR785 back to the PSL racks to check the IMC OLG.  We think that the increase in IMC gain that's needed to restore the ugf is consistent with what's expected (so the modulation depth for 24MHz is similar to before the JAC work), but there is only about 1/4 of the expected power on IM4 trans sum.  

  before JAC install  today  
power into HAM1 H1:IMC-PWR_IN_OUT16 2W 1W  
H1:IMC-REFL_SERVO_IN1GAIN 5dB 11dB we would need about 15dB to restore 35kHz ugf, which is consistent with what's expected for the power drop.  
IMC ugf 35kHz 23.3kHz (see attachment)  
MC2 trans 310 counts 90 counts consistent with 60% HAM1 throughput measured with a power meter (extra loss due to temporary mirror for JM3)
IM4 trans nsum 1.8W 0.1W We should expect 0.54W here if alignment was the same as Nov 2025
AS_C with single bounce power into HAM6   0.6mW expect 1W *0.6 HAM1 throughput * 0.03 PRM * 0.25 BS * 0.3 SRM = 1.3mW, so we are seeing half the light we expect in HAM6

The misalignment onto IM4 trans might be explained by HEPI being locked, when this shift happened in December the alignment onto IM4 trans QPD shifted so that the beam was all on the upper two quadrants. 

 

Non-image files attached to this report
H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 08:37, Thursday 05 February 2026 - last comment - 15:15, Thursday 05 February 2026(89039)
HAM1 alignment/mode-match finalization and align the TRANS PD path

[Keita, Jennie, Masayuki]

Summary

After completing the EOM alignment, we realigned the beam to the IMC using JM2 and JM3. During this process, a large shift of JM3 was observed and corrected. By iteratively moving JM3 toward the PSL side, the mode mismatch was improved to below 1%. We also confirmed the presence of a diffracted s-polarization beam from the EOM with an angle consistent with expectations. Finally, we measured the power throughput and completed the alignment of the TRANS PD path.

Details

After finishing the EOM alignment, we attempted to align the beam to the IMC using JM2 and JM3. At this point, we noticed that JM3 had moved significantly. We re-tightened the dog clamp and the mirror mount, after which the alignment recovered. The mode mismatch (ratio of 2nd-order mode height to TEM00) after this realignment was 0.93/48 = 1.94%. This is actually worse than we measured yesterday, would be because the EOM crystal clipping was solved and it changed the beam shape.

To further improve the mode matching, we decided to move JM3. First, JM3 was shifted by 0.5 inch toward the PSL side, which improved the mode mismatch to 0.57/41.8 = 1.36%. We then moved JM3 by an additional 0.5 inch, resulting in 0.32/39 = 0.821 %. A further 0.5 inch shift improved the mismatch to 0.25/38.8 = 0.644%. Since this level was sufficient, we stopped the adjustment at this point.

We observed an additional beam separated by approximately 1 cm from the main beam at a distance of about 1 m from the EOM. This is likely the s-polarization beam diffracted by the EOM. The relative angle between the two beams was approximately 0.5 degrees, which is consistent with expectations.

We then measured the power throughput using a power meter. The measured powers were 94 ± 2 mW at the EOM output, 95 ± 3 mW at the EOM input, 7 ± 1 mW at JAC REFL, and 100 ± 1 mW at JAC input which indicates no significant loss.

We also checked the beam position at JM3 and confirmed that it was shifted by approximately 1/4 inch in the +y direction. The iris after JM3 and the iris after the periscope were then centered.

Finally, we aligned the TRANS PD path. Using JACT_BS1, we temporarily installed an HR mirror in place of the laser window to obtain sufficient light to the TRANS PD. With this configuration, we aligned the photodiode such that the reflected beam from the PD was properly dumped into the beam dump. Also, we make sure the transmission beam from the laser window will be cought by the same beam dump by removing the mirror and make sure that the beam coming from the JAC is directly hitting the beam dump.

Pictures:
JM3 beam position, JM3 position, and L2 position

Images attached to this report
Comments related to this report
masayuki.nakano@LIGO.ORG - 17:49, Wednesday 04 February 2026 (89040)

The JAC scan gives 0.347(TEM00) to 0.00425+0.0013+0.009= (other small peaks). The total fraction of small peaks is 4%. So, 4mW of the 7mW at the reflection includes all of these fractions.

Images attached to this comment
keita.kawabe@LIGO.ORG - 22:56, Wednesday 04 February 2026 (89041)

EOM alignment

This was a 2-day's worth of job. It was briefly reported in the alog from the first day (89018) but I'll repeat what was already reported so people can see what was done concerning EOM alignment in a single log.

Day 1:

After we thought we completed the mode matching yesterday, we found that the beam has a halo that looked like a weird horizontal streak (horizontalstreak.jpg). It seemed as if it came from the EOM itself.

Eventually we found that the beam coming out of JAC looked as if it's higher than the EOM crystal center by more than 1mm (sorry no picture). We raised the entire EOM assembly by about 1mm by inserting shims under the EOM base plate at 3 locations. shims.jpg is the top view of the EOM, see shim_front.jpg and shim_back.jpg for the close up of the shims. (Each shim is actually two 91080A026 flat slotted washers, each washer is 0.02 to 0.026" thick, so the EOM got higher by anywhere from 0.04 to 0.052" or 1 to 1.3mm.)

After this, the horizontal streak was gone but there was still a vertical streak that was hard to photograph. We checked the horizontal beam position on the EOM input aperture and it looked awfully close to what is supposed to be the edge of the crystal (EOM_IN_horizontally_off.jpg). 

We pushed the EOM in -Y direction by 1mm or so, the input beam position looked good, we realigned the beam downstream of the EOM, measured the mode matching, that was great and we were happy. But I thought that the beam still looked a bit weird vertically (though better than before), it was better than before but weird. We checked the beam position on the EOM output and it was off (EOM_OUT_horizontally_off.jpg).

At this point we wanted to make an YAW adjustment for the EOM pivot plate. It turns out that we had to undo the hard-to-access screw I reported before (caution.jpg) and it was impossible to access when the SMA elbow was connected, a regular Allen key (or even the ball end one) interfere with the connector. It's not a huge interference but I worried that I'll damage the SMA, so we stopped it and called it the day.

(Note for the future design: Why don't we relocate the bolt to the opposite corner (relocate_bolt.png)? )

Day 2:

Ibrahim found us a cut Allen key that fits under the SMA (short_allen.jpg, short_allen2.jpg). We loosened three bolts circled in green in three_bolts.jpg and rotated the pivot plate. It was tedious and we needed three iterations, but we managed to reasonably center the beam position at both the input and the output of the EOM (good_in.jpgok_out.jpg).

(Note for the future design: As of now there is no visual guide for the beam center position for the output side plate (the guide is on the opposite face that is not visible). This is because the input and the output side plate are the same thing (https://dcc.ligo.org/D2500128) and it only has visual guide on one face and not on the opposite face. Give me the visual guide like in visual_guide.png.)

Note: We had to loosen the strain relief such that the cables can slide inside the viton pads, otherwise the tension and stiffness of the cable act as tough springs and the pivot plate will spring back after rotation, so everything will be tedious. For each strain relief, I left one of the two tiny screws somewhat loose, and made the jam nut finger tight. The cable won't go anywhere and it still acts perfectly as a strain relief.

This was the end of the EOM alignment.

The beam shape looks better than at the end of Day 1, not sure if it's great though, it's hard to photography but something faint might be coming out of the EOM. 

The last picture shows the ghost beam which is likely in the wrong (S) polarization.

Images attached to this comment
keita.kawabe@LIGO.ORG - 22:58, Wednesday 04 February 2026 (89042)

EOM crystal serial number

Marking on the RTP container: #B1913109, 20000488M (the former is S/N)?

Images attached to this comment
corey.gray@LIGO.ORG - 15:15, Thursday 05 February 2026 (89055)EPO

Tagging for EPO.

LHO General
ryan.short@LIGO.ORG - posted 07:39, Thursday 05 February 2026 (89044)
Ops Day Shift Start

TITLE: 02/05 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: MAINTENANCE
    Wind: 11mph Gusts, 7mph 3min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.39 μm/s 
QUICK SUMMARY: JAC output alignment continues in HAM1 today, and the HAM7 door is on and will start pumping down soon.

The PSL FSS had been searching for resonance with the autolocker on all night with no success, so I turned it off for now. I imagine the PZT ramp frequency was manually changed yesterday for JAC work and just was not reverted, so the autolocker wouldn't work.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 03:51, Thursday 05 February 2026 - last comment - 15:15, Thursday 05 February 2026(89043)
HAM7-Y Door is on

(Randy, Jordan, Travis, Gerardo)

After lunch we installed HAM7 -Y door.  Note regarding this door to keep it on our memory, there are nicks and a few scratches on the door's flange surface, between 10 to 2 O'clock, but the scratches and nicks are away from the O-ring sealing area.

BTW, all the bolts are on the door, and they were torqued.

We also installed the two 12" OD blanks on the +Y door access ports.  Both blanks will need to be tested for leaks once the chamber is pumped down.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 15:15, Thursday 05 February 2026 (89054)EPO

Tagging for EPO.

H1 TCS
matthewrichard.todd@LIGO.ORG - posted 17:07, Wednesday 04 February 2026 (89038)
Turning on the Ring heaters again!

M. Todd, S. Muusse, C. Compton, S. Dwyer


I wanted to get another measurement of what the HWS think the coupling factor is for thermal lens from ring heater power.

The HWS were not on, so after filling out a work permit Camilla and I went out and turned on the HWS SLEDs. Then we restarted the HWS codes in the individual computers. We also asked Jim to take ITMY ISI to fully isolated.

After waiting about 25 minutes for the HWS to get a baseline reading we turned up both ITMY and ITMX ring heaters by 4W (2W/segment). The HWS will track the defocus and I will compare with my models tomorrow morning. I wanted to do both ITM ring heaters to get a self-consistent measurement.

We also plan on doing single bounce OMC scans tomorrow morning with the ITMs being sufficiently thermalized after the RH turn on. This should give us another lens (punny pun here) to look at the thermalization business.

H1 IOO (ISC)
ryan.short@LIGO.ORG - posted 16:56, Wednesday 04 February 2026 - last comment - 15:14, Thursday 05 February 2026(89037)
IOT1 (JAC Table) Ready for Roll-Up

J. Oberling, J. Wright, R. Short

The new IOT1 in-air optics table, a.k.a. the "JAC table," now has all of its optics, wavefront sensors, and photodiodes mounted, cables run, and is ready to be rolled up chamberside. This table is slated to be on the -Y side of HAM1 and contain PDs and WFS for the JAC reflected beam path.

We placed the components on the table according to the layout as close as possible, but some adjustments had to be made to account for things like cable routing or base sizes (for example, the singular GigE camera is not currently in its proposed position due to the network cable being hit by the table enclosure door; this will need to find a different spot). Routing of the RF cables for the WFS and picomotor cables may also need to be adjusted after final positions are decided. Many of the optics needed to be cleaned before being placed in their mounts. Two of the SMA connectors on the REFL PD were damaged to the point where cables could not be screwed in well enough, but Marc was able to fix these by (carefully) bending the connectors back with pliers. Once the door is back on HAM1 and the JAC reflected beam is exiting the viewport, we can move IOT1 into place and proceed with fine alignment of components on the table, after which the table layout should be updated with an "as-built" version.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 15:14, Thursday 05 February 2026 (89053)EPO

tagging for EPO

H1 CAL
anthony.sanchez@LIGO.ORG - posted 16:40, Wednesday 04 February 2026 - last comment - 15:26, Thursday 05 February 2026(89035)
PCAL EY End Station Measurement

Dripta and I went to EY yesterday (Feb 3rd) to do both an ES and a TX module maintanence. We followed T1500062-v21 with out much deviation until the end when we started the TX module maint.

Obligitory Before and After beam spots on the apature of RX sphere.

Data Analysis: 
python3 generate_measurement_data.py --WS PS4 --date 2025-11-03 
Reading in config file from python file in scripts
../../../Common/O4PSparams.yaml
PS4 rho, kappa, u_rel on 2025-11-03 corrected to ES temperature 299.4 K :
-4.701912257515925 -0.0002694340454223 2.686163396659873e-05
Copying the scripts into tD directory...
Connected to h1daqnds1
martel run
reading data at start_time:  1454177475
reading data at start_time:  1454177902
reading data at start_time:  1454178300
reading data at start_time:  1454179000
reading data at start_time:  1454179400
reading data at start_time:  1454179750
reading data at start_time:  1454179900
reading data at start_time:  1454180530
reading data at start_time:  1454180888
Ratios: -0.5341330662181019 -0.5436335114505099
writing nds2 data to files
finishing writing
Background Values:
bg1 =        18.796205; Background of TX when WS is at TX
bg2 =        5.033949; Background of WS when WS is at TX
bg3 =        18.801656; Background of TX when WS is at RX
bg4 =        5.198797; Background of WS when WS is at RX
bg5 =        18.803508; Background of TX
bg6 =        -0.514446; Background of RX

The uncertainty reported below are Relative Standard Deviation in percent 

Intermediate Ratios
RatioWS_TX_it      = -0.534133;
RatioWS_TX_ot      = -0.543634;
RatioWS_TX_ir      = -0.526715;
RatioWS_TX_or      = -0.535124;
RatioWS_TX_it_unc  = 0.054072;
RatioWS_TX_ot_unc  = 0.053357;
RatioWS_TX_ir_unc  = 0.053158;
RatioWS_TX_or_unc  = 0.054774;
Optical Efficiency
OE_Inner_beam                      = 0.986243;
OE_Outer_beam                      = 0.984385;
Weighted_Optical_Efficiency        = 0.985314;

OE_Inner_beam_unc                  = 0.041515;
OE_Outer_beam_unc                  = 0.041813;
Weighted_Optical_Efficiency_unc    = 0.058922;

Martel Voltage fit:
Gradient      = 1637.852893;
Intercept     = 0.265584;


 Power Imbalance = 0.982524;

Endstation Power sensors to WS ratios::
Ratio_WS_TX                        = -0.927845;
Ratio_WS_RX                        = -1.384820;

Ratio_WS_TX_unc                    = 0.044117;
Ratio_WS_RX_unc                    = 0.038945;

=============================================================
============= Values for Force Coefficients =================
=============================================================

Key Pcal Values :
GS           =      -5.135100; Gold Standard Value in (V/W)             
WS           =      -4.701912; Working Standard Value             

costheta     =      0.988362; Angle of incidence
c            =      299792458.000000; Speed of Light
             
End Station Values : 
TXWS         =        -0.927845; Tx to WS Rel responsivity (V/V)
sigma_TXWS   =        0.000409; Uncertainity of Tx to WS Rel responsivity (V/V)
RXWS         =        -1.384820; Rx to WS Rel responsivity (V/V)
sigma_RXWS   =        0.000539; Uncertainity of Rx to WS Rel responsivity (V/V)

e            =        0.985314; Optical Efficiency
sigma_e      =        0.000581; Uncertainity in Optical Efficiency

Martel Voltage fit : 
Martel_gradient         =        1637.852893; Martel to output channel (C/V)
Martel_intercept   =        0.265584; Intercept of fit of     Martel to output (C/V)

Power Loss Apportion : 
beta          =        0.998844; Ratio between input and output (Beta)  
E_T          =        0.992056; TX Optical efficiency 
sigma_E_T          =        0.000292; Uncertainity in TX Optical efficiency 
E_R          =        0.993204; RX Optical Efficiency 
sigma_E_R          =        0.000293; Uncertainity in RX Optical efficiency 

Force Coefficients : 
FC_TxPD          =        9.154540e-13; TxPD Force Coefficient 
FC_RxPD          =        6.225064e-13; RxPD Force Coefficient 
sigma_FC_TxPD          =        4.888564e-16; TxPD Force Coefficient 
sigma_FC_RxPD          =        3.063586e-16; RxPD Force Coefficient 
data written to ../../measurements/LHO_EndY/tD20260203/

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
anthony.sanchez@LIGO.ORG - 15:26, Thursday 05 February 2026 (89052)

TX module maintenance of End Y was done with reference to T1600436-v12

Date   Feb 3rd 2026
Laser Shutter Check   Pass
Max OFS Offset   8
95% OFS Offset   7.6
Operating OFS Offset   3.8
Laser Output Power   1.94W
After-Laser Rejected Power   3.96mW
AOM Input Power   1.88W
Max Diffracted Power   1.58 W
Un-Diffracted Power   155mW
AOM Diffraction Efficiency   84.04%
After-AOM Rejected Power   13.9mW
TxPD Power   13.7mW
OFSPD Power   6.59mW
Outer Beam Power   0.747W
Inner Beam Power   0.744W
Output Beam Power Ratio   0.995
OFS Gain   37.79
OFS Phase Margin   57.6
Non-image files attached to this comment
H1 AOS (EPO, SQZ, SUS, SYS)
corey.gray@LIGO.ORG - posted 15:55, Wednesday 04 February 2026 - last comment - 08:42, Thursday 05 February 2026(89034)
HAM7 Closeout Photos

Squeezed in a quick photoshoot of HAM7 (right after ISI-unlocking/measurements by Jim + purge air being turned up and before Door Team sealed it up).  HAM7 was open roughly from 11:16am-11:48amPDT for photos.  Since there was only one door off, lighting wasn't great and limited to shots from the -X side of HAM7.  Used Canon DSLR camera followed by iPhone shots; snapped a handful of macro lens shots for fun.  

Photo Album (109photos) uploaded to Google folder HERE.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 08:42, Thursday 05 February 2026 (89045)

Attaching photo of the photographer! 

Images attached to this comment
H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 15:15, Tuesday 03 February 2026 - last comment - 13:49, Thursday 05 February 2026(89017)
HAM1: SUSPM1 investigation - OLC of all four BOSEMs dropped by 20%

Fil, Rahul

Out of chamber investigation

This morning Fil and I unplugged Tip Tilt PM1 cable going to its Satamp and connected it to Tip Tilt JM3 and found no change in its Bosem counts (was still low for UL and UR, like before). We connected JM3 to PM1's electronics chain and it was looking good. We also checked the long grey cable and found it to be healthy. Thus we came to this conclusion that in-air electronics chain is healthy for PM1.

In-chamber investigation

During lunch time when the chamber was empty, I took the OLC of all four BOSEMs for PM1 and found their values to be lower by 20% (last taken in April 2025).

Bosem OLC (April 2025) OLC (Feb 2026)
UL 264040 20800
LL 28894 23100
UR 24529 19100
LR 29276 23200

This tells me that either the in-vacuum cable is not working properly or all four Bosems are malfunctioning. I need more in-chamber time to further investigate this issue.

Comments related to this report
rahul.kumar@LIGO.ORG - 13:49, Thursday 05 February 2026 (89050)SUS

PM1 issue was due to coil drive swap in Dec 2025 - https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=88998

Nothing wrong with the BOSEMs. cable or suspension. 

Displaying reports 21-40 of 86532.Go to page 1 2 3 4 5 6 7 8 9 10 End