Displaying reports 44181-44200 of 88572.Go to page Start 2206 2207 2208 2209 2210 2211 2212 2213 2214 End
Reports until 17:33, Wednesday 05 December 2018
H1 CAL (CAL, DetChar, ISC)
jeffrey.kissel@LIGO.ORG - posted 17:33, Wednesday 05 December 2018 (45726)
DARM OLG Model vs. Measurement -- A Reminder -- H1SUSETMX L3 / TST / ESD Actuation Strength droped by 2 Because We're Using 0.5 the ESD Bias Voltage
J. Kissel

While updating the DARM loop model, in order to compare with the measured loop (from LHO aLOG 45714), I found two things:
    (1) That the ETMX_L3_DRIVEALIGN_L2L gain had changed from -40 to -80, and
    (2) That the loop model was over estimating the measured DARM OLG TF  by a factor of two.

This lead me down a small rabbit hole, in which I found Peter Fritschel's brief entry from early in November (LHO aLOG 45193), which states that "we've reduced the ETMX bias voltage from -430 V_ESD (/40 = 10.75 V_DAC) to -218 V_ESD (/40 = 5.45 V_DAC)". 
However, this was permanent change implemented in guardian / the ETMX bias path in a sneaky way:
    (a) the H1:SUS-ETMX_L3_LOCK_INBIAS is now zero (the bias record was moved from INBIAS to the BIAS_OFFSET so we can use the ramping of the OFFSET)
    (b) the H1:SUS-ETMX_L3_LOCK_BIAS_OFFSET is -9.3 (V_DAC * 40 V/V = 372 V_ESD), and
    (c) the H1:SUS-ETMX_L3_LOCK_BIAS_GAIN is 0.5
    (d) the ETMX_L3_DRIVEALIGN_L2L gain is -80 (where it used to be 40)
such that the output is (including the cnts2V filter gain of 2^18 / 20 = 13107.2) = -60948.5 18-bit DAC counts = -4.65 V_DAC = 186 V_ESD. 
Remember also -- 
    (e) we compensate the fact that we have a 20 bit DAC by dividing the requested DAC counts by 4 in the ESDOUTF banks. 

#TangledWebsWoven.

Before I figured out the mystery, however, I was allowed some time to measure the actuation strength of the L3 stage with the PCAL system (we need this for ER13 anyways) to help me. See the 1st and 2nd attached results. They report that the actuation strength is now 2.149e-11 N/V^2, or 2.294e-12 N/ct_18bitDAC, which is half the value measured in October (LHO aLOG 44690), before we switched to a 20 bit DAC (which doesn't affect the estimate because of (e)), and before the above sneaky bias voltage reduction (a-d).

The problem for the calibration group to fix: yet again we've been out-smarted by the ISC team in its infinite ways of setting a  single number. 
The DARM loop model does not yet include (c), and because of (e) we haven't yet updated the model DAC gain from an 18 to 20 bit DAC gain thinking we could get away with ignoring it after the ESDOUTF compensation.
So -- I can make the DARM loop model work by 
     (i) accurately changing the ETMX_L3_DRIVEALIGN_L2L gain at -80
     (ii) leaving the model DAC gain at 2^18/20 V
     (iii) incorrectly leaving the bias voltage at -9.3 V_DAC (= 372 V_ESD)
     (iv) reducing the physical actuation strength of the ESD by a factor of 2 from 4.1303e-11 to 2.149e-11 N/V^2

3rd attached result is this DARM Open Loop Gain model vs. measurement.

L3 / TST Actuation Function Measurements:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/MeasurementsFullIFOActuationTFs/
    2018-12-05_H1SUSETMX_L3_PCAL2DARM_5min.xml
    2018-12-05_H1SUSETMX_L3_iEXC2DARM_5min.xml

Data Processing scripts that makes the attached plots:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Scripts$
    DARMOLGTFs/process_darmolg_20181205.py
    FullIFOActuationTFs/process_actuationmeas_20181205.py
    FullIFOSensingTFs/process_sensingmeas_20181205.py

New DARM loop Model that informs the attached final 3rd plot:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/params
    modelparams_H1_20181205.py
Non-image files attached to this report
H1 TCS (TCS)
daniel.brown@LIGO.ORG - posted 17:26, Wednesday 05 December 2018 (45727)
ETMX weak point absorber

Danny, Dan

There is a weak point absorber on ETMX we saw in the 20W power up today. This seems significantly weaker than the current ITMY absorbers. The first image shows the point absorber and ifo beam heating, the cross marks the center of the beam heating. By choosing a reference 350s later over the same timeframe we can extract the ifo beam heating only, as shown in the second image.

Images attached to this report
H1 ISC
stefan.ballmer@LIGO.ORG - posted 16:13, Wednesday 05 December 2018 - last comment - 16:43, Wednesday 05 December 2018(45724)
OMC dither alignment does not maximize optical gain

Recently (at least since we started using the new SOFT offsets for the new TCS pre-loading) the OMC alignment dither loops ran away during power-up to 20W, leading to an OMC-SUS saturation. So we dug a little deeper:

- The loop that runs away during power-up to 20W is H1:OMC-ASC_ANG_Y_OUTPUT, driven by dither lines.
- So we set the QPD offsets to match the dither alignment at low power (2W), and increased the power running on the QPD for OMC alignment. That worked fine and did not saturate the OMC SUS.
- Indeed the QPD OMC alignment produced 5% higher optical gain at 20W (!, that does translate into range - 80Mpc at 20W into the interferometer).
- We switch back and forth between QPD alignment and dither alignment (stopping short of saturating the OMC SUS), and repeatably saw more optical gain on the QPDs
- Just to verify that it is not a dither SNR problem, we increased the dither line strengh. This just confirment that the dither alignment has an offset (i.e. pulls us away place with less optical gain).

So, we are forced to conclude that the simple dither alignment somehow is affected by non-arm carrier light. It might be time to revive the idea of a beacon alignment.

 

For now I commented out the dine in the OMC_LOCK guardian that switches to dither alignment.

Comments related to this report
sheila.dwyer@LIGO.ORG - 16:43, Wednesday 05 December 2018 (45725)

Two other interesting observations from this test:

When Stefan increased the dither line amplitudes we saw a broad 50 Hz peak downconverted into DARM (the OMC dithers are 4 lines spaced by 25Hz)

When the OMC alignment is bad we get increased noise in the darm blms from 10-20 Hz. 

H1 CDS
jenne.driggers@LIGO.ORG - posted 16:03, Wednesday 05 December 2018 - last comment - 08:32, Thursday 06 December 2018(45723)
Camera copy script down for Yarm

The Yarm camera copy script is frozen.  Thankfully DiagMain warned us, so DaveB can fix it before he heads home for the day.  We don't need it now since we have a nice ~80Mpc 20W lock, but we'll need it for our next initial alignment.

Comments related to this report
david.barker@LIGO.ORG - 08:32, Thursday 06 December 2018 (45741)

DIAG_MAIN was issuing an error because the ITMY camera data which is copied to the ALS had frozen following our digital video work early Wed afternoon (we are getting both red and green ITMY cameras operational). I reverted the digital video settings at 5pm to restore the camera-copy data for overnight locking. We will take another look at this today.

H1 General
jim.warner@LIGO.ORG - posted 16:01, Wednesday 05 December 2018 (45722)
Shift Summary

17:00 Chandra to HAM6/HAM1

19:00 Nutsinee to ISCT6

20:00 Danny TJ to EX

23:00 Danny TJ to EY

 

 

H1 CAL (ISC)
jeffrey.kissel@LIGO.ORG - posted 15:33, Wednesday 05 December 2018 - last comment - 15:33, Wednesday 05 December 2018(45714)
Calibration Sensing Function Update: Optical Gain Consistent; First Quality Measure of O3 Spring Detuning
J. Kissel

I've gather a new measurement suite for the DARM loop / Sensing Function -- Results are attach, and sensing function parameters are quoted below.
The interferometer conditions during the measurement:
   - 20 W input power
   - DARM actuator configuration: EY L1, EY L2, EX L3. EX L3 ESD Bias set to -4.65 volts at the (20 bit) DAC, or -4.65 * 40 V_ESD/V_DAC = -186 V at the ESD Bias Electrode
   - all modern (red) ASC loops were engaged including SRC (36 MHz BS control, 72 MHz SRC control). 
   - let the IFO thermalize for ~30 minutes or so before I began taking the measurement. 
   - all calibration lines were OFF (PCAL and SUS).

The optical gain is consistent with what has recently been done to match PCAL calibration lines to DELTAL EXTERNAL (i.e. adjusting the inverse sensing function gain from 1.0 to 1.1).
This is the first quality measurement of the H1 IFO's SRC detuning since we replaced ITMX and the ETMs between O2 and O3 -- and results look good -- the optical spring frequency appears to have been reduced from between ~6-8 Hz (depending on the alignment and thermal state of the IFO) to f_s = 3.6 +/- 0.1 Hz (informed by only one measurement, 68% C.I.). The Q remains unresolved (though we expect the Q to be very close to critically coupled at ~1/2).

More details below.

The processed results based on today's measurements for the sensing function parameters are as follows:

Parameter                              | Quantiles (0.15, 0.50, 0.84)
---------------------------------------------------------------------
Cavity gain, H_c (ct/m)                | 3.245e+06, 3.249e+06, 3.253e+06
Cavity pole, f_cc (Hz)                 | 423.7, 425.2, 426.8
Detuned SRC spring frequency, f_s (Hz) | 3.441, 3.603, 3.758
Detuned SRC spring quality factor, Q_s | 30.71, 23.81, 19.41
Residual time delay, tau_c (usec)      | -2.499, -1.806, -1.112

See previous values (that informed what is currently installed and calibrating DELTAL_EXTERNAL) in LHO aLOG 44364.
Specifically, I draw your attention to the optical gain ratio of (previous / current) = 3.552e6 ./ 3.249e+06 = 1.093 ~~ 1.1, which is what's installed currently in the CAL-CS GAIN field.

If we so chose to update the sensing function filters in the CAL-CS model (probably not yet), those filters should be:
Inverse Sensing FOTON values:
H1:CAL-CS_DARM_ERR Bank
SRCD2N: zpk([425.2376;3.5282;-3.6796],[0.1;0.1;7000],1,"n")gain(1298.18)
Gain: gain(3.078e-07)

Inverse Sensing without cavity pole FOTON values for CFTD path:
H1:CAL-CS_DARM_CFTD_ERR Bank
SRCD2N: zpk([3.5282;-3.6796],[0.1;0.1],1,"n")gain(1298.18)
Gain: gain(3.078e-07)

Data was taken with the templates:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs
2018-12-05_H1DARM_OLGTF_23to800Hz_10min.xml
2018-12-05_H1_PCAL2DARM_TF_23to800Hz_2min.xml

Results were processed and produced with
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Scripts/FullIFOSensingTFs$ python3.5 process_sensingmeas_20181205.py


And updated DARM loop model comparison to come -- but we know we at least need to make the updates described in LHO aLOG 44667.
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 13:27, Wednesday 05 December 2018 (45717)
Forgot to mention -- it looks like the systematic error between PCAL and DELTAL EXTERNAL is at ~5% / 5 deg at maximum, indicating our frequency dependent systematic error in the response function is quite small -- almost at the level of O2. Nice!
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 15:29, Wednesday 05 December 2018 (45721)
Attached is a screenshot of relevant channels that state the DARM loop parameters during this measurement suite.
Images attached to this comment
H1 SUS (GRD, ISC, OpsInfo)
jeffrey.kissel@LIGO.ORG - posted 15:04, Wednesday 05 December 2018 (45719)
H1 SUS ETMY Bias Found ON -- Turned it OFF, Accepted in SDF (Also Inputs to Violin Mode Damping Filters now Ignored in SDF)
J. Kissel (in concurrence with others in the control room)

Upon investigating parameters for the DARM Loop Model, I found that the ETMY ESD Bias was ON. While some previous noise investigations have indicated that the bias doesn't impact the noise (e.g. LHO aLOG 45113), it seems bad practice (due to general charging concerns) to leave this on. Thus, I've turned OFF the ETMY ESD bias, and captured that in the SDF system (in the "safe" snap) so it sticks. Note that the HV driver is still *on*, we're just not requesting it to drive any voltage. The binary switches are in the configuration where the the low-voltage (LV) output is selected, the high-voltage input to the LV driver is disconnected, and the LV driver's low-pass filter is engaged.

Also -- while I was in there, I unmonitored the distracting "inputs to the violin mode damping filters" for all 4 quads which are controlled by guardian. There are 40 of them, so they pollute the check against the safe.snap and OBSERVE.snap SDF file when looking for other problems.
H1 SEI
hugh.radkins@LIGO.ORG - posted 12:37, Wednesday 05 December 2018 (45713)
More HAM1 TT L4C Sensor Comparisons

Following on from the Monday Log, thought it prudent to look closer at the V1 sensor.  Plot one of that alog suggested the V1 and possibly the H1 sensors were outliers.

The first attachment here shows high and low gain responses during quiet ground motion times.  You'd expect the signals to be the same unless the GM is different.  All sensors show differences at lower frequencies with V1 starting to diverge at 150 mHz.  Below about 70mHz, the other two vertical traces diverge while the horizontal stays similar to as low as 40mHz.

The second attachment shows all High Gain responses between quiet and large EQ times.  On this plot, the before noted problem of the V1 sensor is not presenting itself, that is, all the vertical signals look similar well below 10mHz.  Does this suggest the V1 low gain signal path has issues?  Otherwise, why would not all the vertical sensors reflect the same GM condition...?

Images attached to this report
H1 ISC
jenne.driggers@LIGO.ORG - posted 11:32, Wednesday 05 December 2018 (45706)
Wed AM locking
Images attached to this report
H1 SUS (SUS)
corey.gray@LIGO.ORG - posted 10:26, Wednesday 05 December 2018 (45674)
Charge Measurement on Dec 4 (before Maintenance)

Came in a little early to run ETM Charge Measurements before Maintenance & followed the usual procedure

Log:  (Times in PDT)

Notes on the procedure/measurements:

Here are the original Slider Values (which I returned to after the measurement):

ETMx:

ETMy:

Images attached to this report
H1 SQZ
nutsinee.kijbunchoo@LIGO.ORG - posted 10:21, Wednesday 05 December 2018 - last comment - 17:48, Wednesday 05 December 2018(45708)
OPO loop TF bad and then good again

Attached a transfer function from yesterday and today along with the pump laser output.

Basically when we had >840mW, something  was noisy and EOM couldn't handle it.

Now that we're down to 770-780mW the loop behaves okay again.

This is the laser output and green production out of the SHG (1day)

SHG scan looked fine. There was no obvious multimode or misalignment yesterday.

We had >840mW yesterday with less green production out of the SHG compared to today. Funny.

The transmission of green from the coupler to SQZT6 also has dropped to 20%. I was able to recover some yesterday with the two pick-off mirrors before the coupler but once the power started to fluctuate it's difficult to do any optimization. I'm still hopeful that this is an alignment drift issue. Would be nice to install additional picomotors there.

 

Images attached to this report
Comments related to this report
nutsinee.kijbunchoo@LIGO.ORG - 10:32, Wednesday 05 December 2018 (45710)

Note that the current temperature readout off the controller box is 29.71 C, current is 2.116A. This number has drifted since Haocun and Sheila adjusted the knob last week (alog45604, the 25 was probably 29 but I believe the decimal point to be accurate). 

nutsinee.kijbunchoo@LIGO.ORG - 17:48, Wednesday 05 December 2018 (45728)SQZ

Turned out the pump laser monitor PD sees the back reflected beam from SHG. So the plot above makes sense. When amount of green power and red trans goes up, back reflected red goes down. I found that the faraday holes were never opened to let rejected light out. I also installed extra beam dump to dump the beam that shoots upward. Both beam (see attachments) shows in the photos belong to back reflection. Opening the holes seems to help with the back reflected problem that leaks into the monitor PD (though it doesn't solve the problem, hopefully it makes diagnosing issue less confusing).

 

I installed another Thorlabs PD to look at the Mephisto (pump) reflected beam off the first QW. This is as close to the Mephisto as I could get before it hits the faraday. The output is plugged into the extra AI channel 3 on the SQZ rack. This is just in case the monitor PD is still contaminated with the SHG back reflection (I think it still is). And might help with the diagnostic if faraday is the problem. Hopefully not...

Images attached to this comment
H1 CDS (CDS, ISC)
keita.kawabe@LIGO.ORG - posted 20:25, Tuesday 04 December 2018 - last comment - 11:07, Wednesday 05 December 2018(45701)
Strange dark noise for LSC-REFL_A_RF9 demod

Probably this is a readback problem and thus not that important, but there's a bump at around 10kHz in dark noise of IOP channels of LSC-REFL_A_RF9_I_IN1 and Q_IN1, coherence is large between these two channels, and both the shape of the bump and the coherence are dependent on whitening gain.

In the attached, TP_CH28 and 29 correspond to RF9_Q_IN1 and I_IN1 respectively.

Red/blue is with 12dB of whitening gain, green/brown 45dB, pink/cyan 0dB, all with two stages of whitening filters. I don't understand this.

It seems to be a small effect and these digital signals are not used for control, but we need to check if this is present in channels that are actually used for control and on the same whitening board (H1:IOP-LSC0_MADC1_TP24=POP_A_RF9_Q, TP25=I, TP26=POP_A_RF45_Q, TP27=I, TP30=REFL_A_RF45_Q, TP31=I).

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 11:07, Wednesday 05 December 2018 (45711)

When MC unlocked for whatever reason, I misaligned MC2 and PRM and measured the IOP channels that are on the same whitening chassis as LSC-REFL_A_RF9 (1st attachment), and nothing looks too crazy except REFL_A_RF9 channels (28 and 29). POP_A_RF9_I and Q (24 and 25) have a bit of a bump at around 4.2kHz or so, which shows up in POP_A_RF9_Q in-lock (2nd attachment) but it looks harmless there.

For IOP-frontend channel name relation see top left of the 1st attached.

I'll look at the analog signal of REFL_A_RF9 again on the floor, but if I find nothing I won't investigate further.

Images attached to this comment
H1 ISC
stefan.ballmer@LIGO.ORG - posted 20:12, Tuesday 04 December 2018 - last comment - 11:36, Wednesday 05 December 2018(45699)
Green initial alignment references updated and ITMY green camera moved to recenter the spot

Keita, Stefan

We re-centered the green QPD offsets in full-lock (2W but with 20W soft offsets) and physically moved the Green camera back to the screen center.  (work permit 7985 with Chandra's verbal approval).

Comments related to this report
keita.kawabe@LIGO.ORG - 20:13, Tuesday 04 December 2018 (45700)
keita.kawabe@LIGO.ORG - 11:36, Wednesday 05 December 2018 (45712)

We need a sturdier camera enclosure than this.

Green camera enclosure is the higher and smaller of two boxes in the attached picture.

The problem is that the entire structure is too flimsy.

You'll notice in the third picture that the screws for the lid of the enclosure are loose. Tightening these was unproductive as it twisted the structure, changing the camera pointing, and it was difficult to compensate for this by the manual adjustment of the camera (which, apparently, we have to do with the lid removed).

If you push the enclosure by your finger the camera pointing changes (though it mostly goes back to the original position after removing the pressure). If the tension of the cable coming out of the box changes probably the pointing changes though we haven't tested this.

This is really not a good system to be used as our initial alignment reference. It would be extremely useful if the camera enclosure is much sturdier (made out of metal plates instead of sheet metal, or sheet metal covering the rigid frame).

FRS 11931

Images attached to this comment
H1 PSL
jason.oberling@LIGO.ORG - posted 13:59, Tuesday 04 December 2018 - last comment - 13:24, Wednesday 05 December 2018(45683)
PSL Cooling Flow Rates Changed

J. Oberling, R. Schofield

At Robert's request I lowered the flow rates of the PSL cooling circuits by ever so slightly opening the bypass valve in the chiller room.  Our target was to get the flows close to 1.5 lpm.  In the past, the flow meters were happy with flows at this level (unlike when we were running with a flow closer to 1.0 lpm and the meters became very glitchy).  Robert will assess how this small change effects PSL table motion; I will keep an eye on the flow meter readings to see if they become glitchy again.

Flow rates before change:

Flow rates after change:

Comments related to this report
jason.oberling@LIGO.ORG - 13:24, Wednesday 05 December 2018 (45715)

Attached are the relevant flow rates for the last 2 days; the change in flow is clearly seen.  What looks like a single glitch shortly after the change is actually not.  While we were in the PSL enclosure Robert requested that we slightly open the bypass valve on the manifold (the valve on the old HPO crystal cooling circuit) for a few seconds to release any trapped air bubbles.  This is the cause of the apparent flow glitch, not an actual glitch in the reading from the flow meters.  Therefore, after >24 hours there have been no glitches in the flow rate.  I will continue to monitor this, but so far everything looks stable.

Images attached to this comment
H1 TCS (TCS)
daniel.brown@LIGO.ORG - posted 03:58, Sunday 02 December 2018 - last comment - 14:38, Wednesday 05 December 2018(45625)
ITMY HWS without point absorbers, HWS self absortion spot motion, CO2 tests

Danny, Craig, Dan Brown

We had a long lock run in low noise tonight. During this we were adjusting the PRC1 offset resulting in some slight alignment drifts of the BS and ITMs (45623). You can see the affect this has on the HWS spherical power here. The alignment offsets started around -3000 seconds on the time axis, first in pitch and then in yaw.

This drift was enough to see how the self heating spot moves relative to the HWS probe beam: ITMY before and after. The red cross is where the IFO beam was before the vent. The other thing to notice is that by choosing the reference time for when the point absorbers have heated up we can get the wavefront distortion without them included, so we see the broader absorption. We also ramped the CO2 heater to get a position of the CO2 relative to the interferometer beam, which is noticeably offset.

We did the same alignment shift images for ITMX, before and after. For ITMX the self heating spot wasn't in a bad position to begin with, however with the new offsets it's become misaligned to the HWS probe beam, so we need to do some picoing.

Lastly, we ramped CO2Y up and down to find the position on HWSY but also to see how the PRC gain and noise was affects. Turning the CO2Y power up 2X resulted in the ASC becoming unstable and noise increased at low frequencies, PRC gain didn't really change. Turning it down 1/2X did the same, DHARD yaw became unstable and we lost lock. We did notice a high frequency bump appearing at several kHz, similar to the frequency noise bump when changing the CO2 laser significantly. More investigation needed on that though. When HWSX is sorted out again we'll run some better comm and diff CO2 tuning.

Images attached to this report
Comments related to this report
daniel.brown@LIGO.ORG - 14:38, Wednesday 05 December 2018 (45718)

Added movie of ITMY point absorber heatup. First movie is with a reference before powerup. The second is using a reference when the point absorbers have reached a steady state. Zip files contains Numpy compressed npz files, each of these have the wavefront aberration data for the CO2 reduction and the ITMY heatup without point absorbers.

Non-image files attached to this comment
Displaying reports 44181-44200 of 88572.Go to page Start 2206 2207 2208 2209 2210 2211 2212 2213 2214 End