Displaying reports 45281-45300 of 83756.Go to page Start 2261 2262 2263 2264 2265 2266 2267 2268 2269 End
Reports until 10:10, Friday 01 December 2017
H1 CDS
patrick.thomas@LIGO.ORG - posted 10:10, Friday 01 December 2017 - last comment - 13:18, Friday 01 December 2017(39592)
h1ecaty1 IOC crashed, restarted
Screenshot attached.
Images attached to this report
Comments related to this report
patrick.thomas@LIGO.ORG - 10:15, Friday 01 December 2017 (39593)
FRS 8220
keith.thorne@LIGO.ORG - 10:32, Friday 01 December 2017 (39594)
Looking at the screen, I see "Error code 13 ~~~MYSTERY ERROR~~~ Go Google "ADS Return Codes"

So I did. ADS return codes are from Beckhoff - error 13 is "Port Not Connected"
david.barker@LIGO.ORG - 12:53, Friday 01 December 2017 (39596)

Once the IOCs were started the corresponding SDF systems crashed. This is unusual behaviour, but we have seen it before. I restarted the SDF processes as user controls on h1build

starth1sysecaty1plc1sdf

starth1sysecaty1plc2sdf

starth1sysecaty1plc3sdf

keith.thorne@LIGO.ORG - 13:18, Friday 01 December 2017 (39598)CDS
I think we now have a fix for that in RCG 3.4 and beyond.  Due to EPICS sequencer changes, I had to re-write the EZCA parts in our front-end sequencer code so they are now robust against the (remote) IOC going offline/online.  This will fix the problem when the PSL Beckhoff OPI<->EPICS IOC is restarted, causing the non-real-time part of the PSLISS model to stop.  Thinking about this, it may also apply in this case, where we communicate with a remote IOC on the IFO Beckhoff TwinCAT<->EPICS IOCs
H1 PSL (IOO, PSL)
peter.king@LIGO.ORG - posted 08:42, Friday 01 December 2017 (39591)
Laser power at base of IO periscope
I measured the laser power at the base of the periscope this morning.  The power meter was located
in front of the mirror directing the beam upwards.

    With an Ophir PD300-3W-SH power meter (S/N 171141, P/N 1Z02411), the highest value measured
over a ~2 minute interval was 160.4 mW.

    With an Ophir 20C-SH power meter (S/N 1711175, P/N 7Z02602), the highest value measured over
a ~ 2 minute interval was 169 mW.

    Both readings were taken using a 10 second average, with the zero offset removed using the
Vega display unit.
H1 SUS (ISC, SEI, TCS)
jeffrey.kissel@LIGO.ORG - posted 16:59, Thursday 30 November 2017 - last comment - 09:19, Friday 20 June 2025(39590)
B&K Hammer Done in HAM3 Baffles, SR2 Cage, and of Primary/Final/Large in-vac TCS Steering Mirrors in BSC2
J. Kissel, S. Pai

Siddhesh and I B&K Hammer'ed the new MC2 and PR2 Scrapper Baffles, SR2 Cage, and of Primary/Final/Large? in-vac TCS Steering Mirrors in BSC2. More details, pictures, and results to follow.

Notes for myself later: 
- SR2 cage accelerometer was oriented with X Y Z aligned with the IFO global coordinates. 
- All other measurements had acc Y aligned with IFO Y, and acc Z aligned with IFO X.
Comments related to this report
jeffrey.kissel@LIGO.ORG - 09:19, Friday 20 June 2025 (85195)
Adding an oldy-but-goldy picture of MC2 (background) and PR2 (foreground) taken by Conor Mow-Lowry the day of this BnK exercise. The picture is taken from the -Y HAM 3 door, so IFO +X and the beam splitter is to the right, and +L for both suspensions is to the left as their HR surfaces point back toward HAM2 (to the left).

Critical to today's yak shaving: the OSEM measuring Transverse at the top mass (M1) stage is toward the camera for both suspensions, i.e. in their +T direction, which means it's the "SIDE," or SD OSEM, per E1100109 (rather than the OPPOSITE SIDE, or OS).

(yak-shaving: while putting together the derivation of the OSEM2EUL matrices in G2402388), I'm reminded that some HSTS were assembled with the Transverse sensor on in the OS position, and I'm on the hunt as to which ones -- and making sure the (SD to T) & (T to SD) element OSEM2EUL & EUL2OSEM matrices, respectively, are correctly -1.0 if SD and +1.0 if OS.)

MC2 and PR2's OSEM2EUL matrix value H1:SUS-[MC2,PR2]_M1_OSEM2EUL_2_6 is correctly -1.0.
Images attached to this comment
H1 General
jeffrey.bartlett@LIGO.ORG - posted 16:12, Thursday 30 November 2017 - last comment - 10:33, Friday 01 December 2017(39589)
Ops Day Shift Summary
Ops Shift Log: 11/30/2017, Day Shift 16:00 – 00:00 (08:00 - 16:00) Time - UTC (PT)
State of H1: Unlocked - Vent
Intent Bit: Engineering
Support: N/A
Incoming Operator: N/A
Shift Summary: Continuing vent work
 
Activity Log: Time - UTC (PT)
16:00 (08:00) Start of shift
16:15 (08:15) Jeff B. – Into LVEA to reset DM #2
16:20 (08:20) Jeff B. – Out of the LVEA
16:45 (08:45) Cheryl – Going to HAM2 for IO alignment
17:05 (09:05) Terry – Going into the Squeezer Bay
17:17 (09:17) Travis – Going into the Biergarten
17:42 (09:42) Vern – Escorting Indian visitor for LVEA tour
17:48 (09:48) Betsy – Going to LVEA to check on activities
17:52 (09:52) Betsy – Out of the LVEA
18:06 (10:06) Betsy – Going to the Biergarten
18:45 (10:45) Cheryl & Ed – Out of the LVEA
19:08 (11:08) Richard – Leading a contractor job walk through LVEA
19:56 (11:56) Jeff K. & Connor – Out of the LVEA
19:57 (11:57) Richard – Has contractor job walk at End-X
20:11 (12:11) Travis & Betsy – Out of the LVEA
20:39 (12:39) Jeff K. & Brijesh – B&K hammering on several suspensions on input arm
20:41 (12:41) Jeff K. – Transition LVEA to Laser Safe
20:48 (12:48) Bubba – Removing arm from HAM5
20:50 (12:50) Betsy – Going into the LVEA
21:14 (13:14) Travis – Going into the Biergarten
21:19 (13:19) Jason – Going to Biergarten
21:21 (13:21) Ed – Going into LVEA for IO and alignment work
21:39 (13:39) Rick – Going into LVEA
21:40 (13:40) Krishna & Mike – Going to End-X to work on BRS
21:42 (13:42) Bubba – Finished at HAM5
22:10 (14:10) Travis – Out of the LVEA
23:13 (15:13) Kyle – Back from End-X
00:00 (16:00) End of Shift
 

 

Comments related to this report
jeffrey.bartlett@LIGO.ORG - 10:33, Friday 01 December 2017 (39595)

Correction: To the 20:39 (12:39) entry -  Siddhesh was with Jeff K. not Brijesh. 

H1 SUS
betsy.weaver@LIGO.ORG - posted 15:35, Thursday 30 November 2017 - last comment - 15:44, Thursday 30 November 2017(39581)
BSC 2 Baffles aligned

Today, Travis and I used the red visible AUX laser to final align the ITMY elliptical baffle and the BS AR and HR baffles.  (Note, we had already done the BS baffles and the ITMX but wanted to check and document.  Note, the front back, and center targets of the elliptical baffle boxes do not all exactly line up, but the tolerance is 7mm so we are center to better than that on all of them.

Also, note, the BS baffles do not look very centered on the mounting to the BS structure, but the beam is centered thru the baffle, so can't do much about that.

Pictures are attached.

Comments related to this report
betsy.weaver@LIGO.ORG - 15:36, Thursday 30 November 2017 (39582)

ITMX Elliptical baffle front and back targets.

Images attached to this comment
betsy.weaver@LIGO.ORG - 15:37, Thursday 30 November 2017 (39583)

ITMY Elliptical Baffle front and back targets.

Images attached to this comment
betsy.weaver@LIGO.ORG - 15:38, Thursday 30 November 2017 (39584)

BS HR Baffle side position wrt to mounting screws and target pix.

Images attached to this comment
betsy.weaver@LIGO.ORG - 15:39, Thursday 30 November 2017 (39585)

BS AR Baffle position (sorry, no target pic).

Images attached to this comment
betsy.weaver@LIGO.ORG - 15:40, Thursday 30 November 2017 (39586)

View ~on beamline ITMX elliptical to BS AR and down to PR3.  Likely better pix coming...

Images attached to this comment
betsy.weaver@LIGO.ORG - 15:41, Thursday 30 November 2017 (39587)

View ~on beamline ITMX elliptical to BS AR and down to PR3.  Likely better pix coming...

betsy.weaver@LIGO.ORG - 15:44, Thursday 30 November 2017 (39588)

View ~on beamline ITMY elliptical to BS HR and down to SR3.  Likely better pix coming...

Images attached to this comment
H1 SYS
betsy.weaver@LIGO.ORG - posted 15:31, Thursday 30 November 2017 (39580)
HAM5 OFI Cable routing

This afternoon, while helping Jason install the new AWC optics, I worked on routing the AOSEM cables for the OFI.  Hugh had preassembled the D1000225 SRS cable to the D1000234 Quadrapus cable and mounted it in a cable table bracket.  So all I had to do was route it all on the table, plug it into the feedthru and start connecting AOSEMs.  I've plugged in 1 AOSEM and will start looking at medms next.

H1 SEI
hugh.radkins@LIGO.ORG - posted 12:31, Thursday 30 November 2017 (39579)
Working on noisy WHAM6 CPS cabling

In LHO aLog 39484, it was reported that the WHAM6 Corner3 CPSs was sitting on a rail and it would intermittently come off the rail when the cables were messed with.  I managed to get the channels off the rail but V3 was clearly noisier than the others.

Yesterday, I worked the cables over to eliminate this and maybe succeeded.  The CPS cables are in the right turn port of the 5-way cross and one of the cables was looped back in the hidden turn and although it could not be seen, I suspect the copper shield was close to the BNC Bayonet and this was the source of the intermittence.  Pulling and twisting the cable un-kinked the loop and it looks like it should remain.  Difficulty here is that now that the feedthru is on the five way cross, there is no connection to break to aid in untwisting the run.  At least for the past 24 hours however it has remained stable.

The attached 10 day trend of the CPS locals shows the elevated noise of V3 and how all sensors have a similar noise level.  Also notice how the H3 signal has jumped onto the page during this work yesterday indicating that channel was also affected although it was not on the rail (32k cts.)  But, it was being influenced on the absolute level readout, that is, actual reported position.  When I trend this back to before the vent, the reading was never at this level and I think that the sensor must have been exposed to some erroneous signals which was giving a false position.  I will say that both channels are much better now!

Will update ticket 9477 to monitor these channels for a time before closing.

Images attached to this report
H1 General (IOO)
edmond.merilh@LIGO.ORG - posted 11:06, Thursday 30 November 2017 (39577)
Current I/O Alignment Values/Positions

The figure below is a shot of the current positions after morning work. Cheryl has left for the dentist.

We can now hand off to Rick S. for ISS alignment.

IO work will continue this afternoon

Images attached to this report
H1 SUS (CDS)
jeffrey.kissel@LIGO.ORG - posted 10:24, Thursday 30 November 2017 (39575)
H1 SUS ZM2 Electronically Functional (using OM1 signal Chain), and Mechanically Free
J. Kissel, B. Weaver, S. Pant

After yesterday's connection of ZM2 to the single functional 4-OSEM electronics chain in HAM5/6 --> readout as "OM1", and re-arranging the control setup (i.e. changing the OSEM2EUL and EUL2OSEM matrices) from an OFIS to an HTTS, we were able to grab the first transfer functions functions of the brand-new HTTS in HAM5 -- ZM2. The results look excellent. The suspension is free, resonances look quite similar (read: close enough) to a previous reference of OM1, and the DC scale looks right-on target. Nice!!

I attached screen-shots of the results.

We'll wait for chamber close-out to really make these TFs official with the full matlab analysis suite, but for now, the new data (again, still using channels that claim "OM1" as the suspension) live here:
    /ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/ZM2/SAGM1/Data/
        2017-11-30_1724_H1SUSZM2_M1_WhiteNoise_L_0p01to50Hz.xml 
        2017-11-30_1724_H1SUSZM2_M1_WhiteNoise_P_0p01to50Hz.xml
        2017-11-30_1724_H1SUSZM2_M1_WhiteNoise_Y_0p01to50Hz.xml

(Note that, in committing these to the SVN, I've created two new corners of the SVN to make space for ZM1 and ZM2).
Images attached to this report
H1 ISC (INS, ISC)
keita.kawabe@LIGO.ORG - posted 22:23, Wednesday 29 November 2017 - last comment - 10:31, Thursday 30 November 2017(39570)
HAM5/HAM6 pre-vent work all done (Gerardo, Cheryl, JeffK, Sheila, Keita)

ZM path was redone.

After centering the beam on SR2 and on the OFI, HAM6 iris closer to HAM5 was temporarily removed to the side.

ZM2 was adjusted so that the beam is centered on it while the reflection goes to the center of the HAM6 iris closer to VOPO.

HAM6 iris closer to HAM5 was put in place, it went to slightly different location than it was originally at, but that's perfectly acceptable.

First round of OMC path realignment convinced us that the rest could be done in HAM6 after the corner is pumped down.

We checked the beam position in HAM6 in the OMC path and moved OM1 and OM2. What we did is mostly consistent with what I wrote in alog 39504 right after the septum window was rotated 180 degrees:

After this,

As soon as we put the fast shutter in place, though, it was apparent that the beam from HAM5 was too close to the fast shutter structure (see attached). This could be mitigated by pushing OM2 back into -X direction by 10mm or so. This will push the OM1-OM2 line by about 4mm or so in -X direction at the fast shutter position, so the clearance will be that much larger, and the clearance between OM1-OM2 line and OM3 cage should still be OK.

Before we actually did this exercise we were afraid that there would be no good solution because of the fast shutter unless we push OMC in -X direction, but we convinced ourselves that moving OMC is unnecessary.

OM2 is in its non-final location, and it only has two dog clamps right now because the original dog clamp was too short for this new position. We might need a longer one.

What needs to be done after pumping down

Lock IMC and send a stable beam into HAM6.

Reposition everything so that the entire beam path has OK clearance from any edge (push OM2 in -X by 10mm or so, rotate OM1, rotate OM2, reposition fast shutter).

Rotate OM3.

Fine-tune to send the beam to both of the QPDs on OMC. For this, we probably have to remove OMC shroud.

Reposition OM2 transmission beam dump.

Move all mirrors/lens/ASC-AS_C in OM1 transmission path by 14mm in -X direction.

Do the OMC REFL path beam dump properly.

Comments related to this report
keita.kawabe@LIGO.ORG - 22:34, Wednesday 29 November 2017 (39571)INS, ISC

Pictures were taken after the FS was placed as close to OM3 as possible and aligned.

1: FS viewed from HAM5 side. The beam from HAM5 is very close to the edge of the aluminum structure of FS. It's hard to see from this but the clearance between the beam and the wire is good.

2: FS and OM2 viewed from OM1 side. The beam reflected by OM1 is actually hitting the center of the FS mirror though you cannot see it in this picture.

3: FS from OM1 side, with a card to show the beam position on the FS mirror. The wire loop on the right is more or less how I expect it to be, but the loop on the left is not.  The left loop is actually going under the left stopper arm. You'll understand this better if you look at the first picture again (note that left on this picture is right on the first picture). Should we fix this?

Images attached to this comment
keita.kawabe@LIGO.ORG - 10:31, Thursday 30 November 2017 (39574)

Position of ham6 zm path iris closer to vopo (1st) and to HAM5 (2nd).

Images attached to this comment
LHO FMCS (VE)
gerardo.moreno@LIGO.ORG - posted 20:01, Wednesday 29 November 2017 - last comment - 12:15, Thursday 30 November 2017(39568)
HAM5/6 Cleanroom HEPA Fans Tweak

(Kyle R, Gerardo M)

Last night while doing some work around the HAM5 I noticed the door cover for HAM5 being "sucked in" (see photo), that occurred every time someone removed the door cover to any other HAM chamber (4,3 or 2), it turns out that the new cleanroom creates great positive pressure causing the door cover for HAM5 to appear as being "sucked in".  So I ran a little test, while one of the other door covers was off and HAM5/HAM6 door covers on, I pushed some of the cleanroom curtains open, and that relieved some of the positive pressure created by the cleanroom HEPA fans, that allowed the purge air to "balloon" the door covers on HAM5, and as soon as I closed the curtains the door covers went back to getting pushed in towards the chamber.

Today Kyle and I used a long ladder to get access to some of the HEPA fans, there are 20 of them on that cleanroom, we dialed down 7 of them and there is a noticeable difference already, we plan to dial down some more tomorrow as work on that chamber allows it.

Images attached to this report
Comments related to this report
john.worden@LIGO.ORG - 21:43, Wednesday 29 November 2017 (39569)

You guys should check that the gap at the foot of the curtains is correct - refer to PSI drawings - this is intended to exhaust the air without building pressure in the cleanroom. It may be better to leave the flows at nominal and relieve pressure that way. By dialing back the hepa units you are decreasing the volume flow rate of clean air which may not be the right thing to do.

chandra.romel@LIGO.ORG - 10:37, Thursday 30 November 2017 (39576)

This cleanroom is the "gerbig" style with 20 HEPA filters and 1 ft clearance at bottom (non adjustable legs). We checked the particle count after dialing down flow; cleanroom is still running well below Class 100.

john.worden@LIGO.ORG - 12:15, Thursday 30 November 2017 (39578)

Thanks!

Displaying reports 45281-45300 of 83756.Go to page Start 2261 2262 2263 2264 2265 2266 2267 2268 2269 End