Displaying reports 45481-45500 of 86626.Go to page Start 2271 2272 2273 2274 2275 2276 2277 2278 2279 End
Reports until 15:45, Wednesday 06 June 2018
LHO VE
kyle.ryan@LIGO.ORG - posted 15:45, Wednesday 06 June 2018 (42357)
GV12 open

After three more iterations today of accelerating then stopping GV12's VFD-driven motor following 0 - 300 rpm, stop, wait (hours) 0 - 300 rpm, stop, wait etc. pattern, the MEDM screen position indication finally transitioned from RED (CLOSED) to YELLOW (not CLOSED, not OPEN). 

I had noted initially that, when fully closed, the ball screw is recessed below the pulley nut(?) about 10 inches.  Today, when the CLOSED limit switch transitioned states, I noted that the ball screw was flush with the pulley nut.  Hmmmm...if this is by design, then this could be helpful in the future when assessing the degree of closure for these valves.  On related note, when fully open, the ball screw measured 37 7/16" above the pulley nut.  The hand marked measurement (written in ink pen in the vicinity) for this left by whoever inititally commissioned this valve back in 1997/8 was 37 9/16".  I wonder if this discrepancy is due to casual measuring, or, would I have achieved 37 9/16" today had the valve been driven to the OPEN limit switch at full speed? 

 

Non-image files attached to this report
H1 General
jeffrey.bartlett@LIGO.ORG - posted 15:24, Wednesday 06 June 2018 (42359)
Fisinshed Removing the In-House Vacuum system
   This morning the last of the in-house vacuum system was removed from Mid-y and End-X. This completes the removal of the in-house vacuum system from across the site. Closing WP #7440. 
H1 PEM
jeffrey.bartlett@LIGO.ORG - posted 15:21, Wednesday 06 June 2018 (42358)
Shutdown Dust Monitor End-X VEA-2
   Shutdown and removed the chamber side dust monitor (VEA-2) from End-X.  
H1 General
yannick.lecoeuche@LIGO.ORG - posted 14:21, Wednesday 06 June 2018 - last comment - 15:16, Wednesday 06 June 2018(42355)
LVEA Transitioning to Laser Hazard (WP #7628)

Jeff is transitioning the LVEA to laser hazard.

Comments related to this report
jeffrey.bartlett@LIGO.ORG - 15:16, Wednesday 06 June 2018 (42356)

LVEA is laser hazard - WP #7628.

H1 DAQ (CDS)
david.barker@LIGO.ORG - posted 13:22, Wednesday 06 June 2018 (42353)
h1tw1 died 01:45 PDT today

our old (gentoo 2.6.35) DAQ trend writer h1tw1 died at 01:45 this morning. There were messages on the console and the cpu was locked up. The console messages did not quite look like the 208.5 day bug, but on restart it forced an fsck because it had been running in excess of 246 days, so this may be the cause.

We have the replacement unit for h1tw1 on hand, FRS10834 covers its installation.

h1nds0 was in fact NFS mouting h1tw1, I changed this to h1tw0. If anyone needs raw minute trends for this morning, please switch your NFS to h1nds0.

LHO FMCS
bubba.gateley@LIGO.ORG - posted 08:07, Wednesday 06 June 2018 - last comment - 08:41, Wednesday 06 June 2018(42351)
E X Temperature Adjustment After Chamber Closeout
I adjusted the temperature at E X after the chamber was closed up and the purge air compressors were turned off.
The temperature at both end stations is set at 65 degrees F. 
Comments related to this report
chandra.romel@LIGO.ORG - 08:41, Wednesday 06 June 2018 (42352)

Brrr!

H1 IOO
sheila.dwyer@LIGO.ORG - posted 20:21, Tuesday 05 June 2018 - last comment - 13:58, Wednesday 06 June 2018(42350)
IMC spot positions measurement

Gabriele, Georgia, Sheila, Craig

After the model changes yesterday, and after relocking the mode cleaner after today's work inside the PSL, we worked on a script to measure the spot positions in the IMC.  Jenne had started a script for this, but since we needed to make the model changes we edited it today.  

We are currently running the measurement, the script is at userapps/isc/common/scripts/decoup/a2l_min_IMC.py

It runs 6 dither lines on pit and yaw for MC1, MC2, and MC3.  

Comments related to this report
sheila.dwyer@LIGO.ORG - 13:58, Wednesday 06 June 2018 (42354)

The spot positions from this measurement are:

alpha =

    0.0573    0.1809    0.0549   -0.0706    0.0200    0.1241


measurements = 

    'MC1 P'    'MC2 P'    'MC3 P'    'MC1 Y'    'MC2 Y'    'MC3 Y'


mis-centering (mm) =

    2.4169    7.6333    2.3162   -2.9808    0.8459    5.2365

In this coordinate system positive distances are closer to the upper right osem than the center of the optic. Script is attached.  

Non-image files attached to this comment
LHO VE
kyle.ryan@LIGO.ORG - posted 18:48, Tuesday 05 June 2018 (42349)
Continued "bumping" GV12's motor

Operated GV12's VFD motor in the OPEN stroke three times throughout the day today following 0-300 rpm, stop, wait a few hours, repeat process.

H1 PSL
jason.oberling@LIGO.ORG - posted 18:00, Tuesday 05 June 2018 (42346)
PSL Work Today

In addition to the work logged by Keita here and here, and the cooling work logged by Jeff here, we performed the following:

-Physically installed the 70W amplifier external shutter.  The shutter has not been connected electrically yet, we will do that once we can turn the laser and PSL Control Box OFF.  Since the shutter is not connected electrically, the shutter flag was temporarily removed (as the shutter's default position is CLOSED).  Once the shutter has been connected electrically and proper operation confirmed, the shutter flag will be reinstalled.  See attached pictures.

-Removed the power supply cable for the existing IO_AB_PD3, as the connector is needed for the new power cable for the large-area PD that will be installed as the new IO_AB_PD3.

-As part of the bullseye PD installation, we had to swap the HR mirror that directs a pick-off beam into the PD for the 70W amp power watchdog with a beamsplitter.  As a result of this, we re-calibrated the watchdog PD for proper watchdog operation.

-Had Bubba look at the Laser Room door.  Over the last several weeks the door has become difficult to move; it is likely we need to have the manufacturer take a look at it.

Images attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 17:28, Tuesday 05 June 2018 (42345)
Corner Station NEG Pump PT-191 Valve Replacement

(Kyle R, Gerardo M)

Removed o-ring valve from housing for NEG pump PT-191, replaced it with a new metal seal valve.  Aux cart will be left running overnight pumping on the new volume only, will continue working on it tomorrow.

H1 PSL (ISC, PSL)
keita.kawabe@LIGO.ORG - posted 17:25, Tuesday 05 June 2018 (42344)
Bull's eye detector reinstalled (Jason, Rick, Keita)

We reinstalled the bull's eye detector on PSL.

Caveat: When the power on the detector was too much, three segments of the outer ring railed at ADC (i.e. -32k counts), but the center segment only reached -30k or so (attached), it's not like the power on the central segment was smaller than the others, it was quite contrary. Maybe there was some analog problem going on.

We installed an ND filter to reduce the power, plus RG850 to make sure that no pump light reaches bull's eye.

Right now the center segment is at -23k counts, ring segments are about 10k each. Calibration of the sensor will be worked on by Thomas and Georgia.

 

 

H1 PSL (ISC, PSL)
keita.kawabe@LIGO.ORG - posted 17:15, Tuesday 05 June 2018 - last comment - 18:09, Tuesday 05 June 2018(42333)
Main path and ALS path power increased (Jason, Rick, Keita)

As was reported in  alog 42311, last week ALS path power on the ISCT1 was found to be much lower than it used to. Suspicion was that the power was throttled on the PSL table intentionally, which turned out to be the case. We went to the PSL room today to increase the power.

Before we started working, ALS path power on the PSL table was about 180mW. Rick adjusted HWP right after the PMC to minimize the rejected beam on the polarizer, and the power in the ALS path increased to ~900mW.

Main path power after EOM was 39.5W and H1:PSL-PWR_PMC_TRANS_OUTPUT was displaying 42.7 W (and H1:PSL-PWR_PMC_REFL_OUTPUT 18.9W, which is pretty bad BTW) , which we can use to calibrate things later. For now PSL rotator calibration as well as H1:PSL-EOM_A_DC_POWERMON calibration are both totally bogus.

After this we've done many things including but not limited to bull's eye reinstallation, and somehow the alignment into PMC got even worse (right now PMC transmission is only 20W). Jason says we need to realign into PMC.

 

Comments related to this report
jason.oberling@LIGO.ORG - 18:09, Tuesday 05 June 2018 (42347)

I'll tweak the beam alignment into the PMC at the first opportunity, likely in the morning before commissioning activities start in earnest.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 17:14, Tuesday 05 June 2018 (42343)
Corner Station NEG Pumps Valved In

Pumped out the gas out of the 3 NEG pump housing on top of the OMC tube, PT-191, PT-192 and PT-193.  There was no issues pumping down the volume on all 3 NEG pumps.  All 3 NEG pumps were valved into the main volume, PT-193 was first then PT-192, and PT-191 was last.

Unfortunately soon after the aux cart was decoupled from the PT-191 the pressure started raising (NEG was isolated from the aux cart via both of its valves), an indication of a leak somewhere.  We decided to isolate PT-191 from the rest of the vacuum system until we can resolve that issue, PT-192 and PT-193 remain valved in.

 

LHO VE
kyle.ryan@LIGO.ORG - posted 17:10, Tuesday 05 June 2018 (42342)
Replaced GV12's annulus ion pump

Replaced pump had been in operation since Oct. 2002.

LHO VE
chandra.romel@LIGO.ORG - posted 15:40, Tuesday 05 June 2018 - last comment - 18:13, Tuesday 05 June 2018(42340)
MTP ON @ EX

Spun turbo up this afternoon, backed by QDP80. Pressure currently at 2.6e-5 Torr.

Comments related to this report
chandra.romel@LIGO.ORG - 18:13, Tuesday 05 June 2018 (42348)

Installed fitting for EX MTP scroll so we can quickly transition tomorrow. Closed the 10" GV during this installation, so pressure rose for a few minutes.

H1 SEI (DetChar)
hugh.radkins@LIGO.ORG - posted 13:17, Tuesday 05 June 2018 - last comment - 16:20, Tuesday 05 June 2018(42334)
WHAMs 5 & 6 HEPIs Unlocked

No difficulty unlocking except for the access restrictions from ISC tables and laser assemblies.  Signs have been changed; please be mindful of the suspended parts.

Running Range-of-Motion tests; will report shortly.

Comments related to this report
hugh.radkins@LIGO.ORG - 14:15, Tuesday 05 June 2018 (42335)

For HAM6 X & Y motion, see the first attachment.  It looks like the X motion is clear to 0.8mm but the Y direction is showing some interference around +560um and -730um.  Yes this bothers me but it is more than enough range for operations.

For the Z stroke, see the second attachment.  You see some separation of the local sensors around 380 microns but that does not seem as bad as the separation at -460.  The platform tripped during the minus stoke and this was from the actuator drive railing.  Again, more than enough range for normal operation but I should investigate and mitigate at a low priority.  When Isolated, the HAMs should never move that much (except HAM1 which does have AC isolation.)

As far as the current running position relative to the long term locked position, all are within a few micro units except X which is about 9um from the locked position.  Putting the isolated position at the locked position increased the output drives which we like to keep low.  I doubt anyone will notice the difference between the long term locked position and the current isolated position; but, if anyone needs to shift the position, it can be done.  At worst, we'll need to put some turns on the big springs.

Images attached to this comment
hugh.radkins@LIGO.ORG - 16:20, Tuesday 05 June 2018 (42341)

For HAM5 see the attachments:  The horizontal local sensors show near perfect overlay supporting the assessment that there is no interference.  The second plot shows the verticals and it is evident that there is some interference around the -700 micron point.  More than good enough for nominal operations.

Images attached to this comment
Displaying reports 45481-45500 of 86626.Go to page Start 2271 2272 2273 2274 2275 2276 2277 2278 2279 End