Displaying reports 4841-4860 of 89152.Go to page Start 239 240 241 242 243 244 245 246 247 End
Reports until 15:09, Monday 08 September 2025
H1 ISC
elenna.capote@LIGO.ORG - posted 15:09, Monday 08 September 2025 (86788)
No-squeezing noise comparison with OM2 hot and cold

In the past, when heating up the OM2 heater, we have noticed that the low frequency sensitivity improves, for example alog 78652.

I wanted to make this comparison again, but we have seen that the squeezing performance can be variable, so I first chose to compare a no squeezing time with OM2 hot and cold.

It's also important to note that the calibration has likely changed with the hot OM2. Some of that can be corrected with the TDCFs. However, we saw that the required SRCL offset to minimize SRC detuning with OM2 hot was significantly different, alog 86737. This means that the sensing function is probably significantly different, causing some change in the calibration that is not compensated. We didn't take any calibration measurements with hot OM2, so we have no idea how much or how to compensate the calibration change.

With OM2 hot, the coherence of DARM with SRCL and MICH increased, although SRCL can explain most of the excess noise between the two times. Interestingly, the coherence of DARM with jitter decreased with OM2 hot. There is no significant change in the coherence with PRCL.

Here is a comparison of the cold and hot OM2 times, using GDS CALIB STRAIN CLEAN.

I performed a simple coherence subtraction of SRCL alone, and SRCL and MICH together. I checked that there is no significant coherence between SRCL and MICH.

Given the possible issue with the calibration, I am not certain how comparable the low frequency noise is. There is a small increase in the noise up to 30 Hz with OM2 hot, even with the above subtractions applied.

Non-image files attached to this report
H1 General
anthony.sanchez@LIGO.ORG - posted 13:20, Monday 08 September 2025 (86787)
H1 ISI CPS Sensor Noise Spectra Check - Weekly FAMIS

H1 ISI CPS Sensor Noise Spectra Check - Weekly Famis 26655

BSC high freq noise is elevated for these sensor(s)!!!
    
ITMY_ST1_CPSINF_H3
 

Non-image files attached to this report
H1 SUS (ISC, SEI)
oli.patane@LIGO.ORG - posted 12:17, Monday 08 September 2025 (86784)
Estimators not causing extra ASC SRC2 Y CTRL noise

In 86589, Jeff points out mysterious noise seen in SRC2 Y CTRL between 1 and 3 Hz that is only seen when the SR3 estimator is fully on with the L to P contribution included.

To verify this, I plotted some other times with the estimators ON and OFF, and looking at those different times, I was able to confirm that the unknown increase in noise Jeff points out between 1 and 3 Hz is NOT caused by the estimator, but by some other noise source in passing.

Images attached to this report
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 12:10, Monday 08 September 2025 - last comment - 16:49, Friday 03 October 2025(86778)
filter cavity backscatter measurements

Camilla, Sheila

We did a series of quick measurements related to filter cavity backscatter, a follow up with plots will come soon.

Injection for linear coupling (varies from day to day, we are repeating to see how much it changes, 1st screenshot):

Fringe wrapping (scattered amplitude seems slightly lower than in 2024 when moving ZM2, slightly higher when moving ZM5):

Open loop gain (also varies from day to day):

FC2 M1/M3 cross over:

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 16:49, Friday 03 October 2025 (87274)

Here's the plot of the fringe wrapping measurements in displacement units.  It can be compared to a similar measurement made on ZM4 + 5 at LLO: 60856. Over email Peter asked some questions about the power levels needed to explain this. 

Power level heading towards HAM7 from OFI:

The power on the DCPDs is 47mW, and there is 12pW retro-reflected off the filter cavity, so the total isolation provided by OFI + SFI2 + SFI1 is 2.5e-10 in power ratio, or 96dB.  The OFI isolation ratio was measured to be 43dB in 79379.  If this is true it would imply that one of the SFIs is providing less than the 30dB isolation assumed in T1800447, and we should have 2uW of carrier light headed towards SFI2.

Our readback of the 1% pick off of light from the interferometer heading towards SFI2, B:PD1 (OFI PD A) says that we have 0.03mW on it, meaning 3mW from the IFO going towards SFI2, about 1mW of this would be carrier based on (87114),which seems too high.  

The responsivity of this PD was checked in 60284, and later double checked because it seemed low (the settings are still the same).  The similar PD OFI PDB has a measured responsivity of 0.25A/W and the excelitas website lists a peak responsivity of 0.6A/W at 850nm for these PDs.  (ffd-200h-si-pin) If we think that this calibration was mistaken and the real responsivity is more like OFI PD B, 0.25A/W, there is 0.72 mW of light from the OFI heading towards SFI2,  ~240 uW of carrier, the OFI isolation would only be 23dB, and the SFIs must be providing something like 36 dB each.  

Reflectivity:

If my interpretation of the fringe wrapping measurements into power are correct (12 nW of power is retroreflected from the path that includes ZM5), we are reflecting 50ppm of the carrier scattered toward HAM7 using the (recalibrated) 240uW value from OFI PDB, or 0.6% if we believe the isolation ratio measurement for the OFI and use the 2uW value.  B:BS1 is a 1%, so the maximum reflectivitiy we could get from scatter in the B:PD1 path would be 0.01%.  This means that the B:PD1 path can't explain the reflectivity needed if there is 2uW headed towards HAM7, and even if there is 240uW heading towards HAM7 this PD seems unlikely to explain the scatter, since it would need to reflect half the light that's incident on the PD.  Camilla did alog the check of the alignment (and the beam dump catching the retro-reflection off this diode: 65006

Images attached to this comment
H1 SQZ
camilla.compton@LIGO.ORG - posted 11:07, Monday 08 September 2025 - last comment - 11:24, Monday 08 September 2025(86777)
SQZ FIS Data Set, repeat of last week with cold OM2
Sheila, Camilla.
Repeating 86739, with OM2 in it's nominal cold state. Plot saved in camilla.compton/Documents/sqz/templates/dtt/20250908_SQZdata.xml and attached.
First I touched up the SHG temperature as the SHG power lower than nominal after it was increased yesterday 86767, increased from 93mW to 106mW, sdf attached. Than checked OPO temperature and while measuring NLG.
 
Type Time (UTC) Angle DTT Ref
No SQZ 16:09:30 - 16:19:00 N/A ref 0
FIS SQZ (tuned for 1kHz) 16:22:00 - 16:25:00 (-)137.5 ref 1
FIS Mid + SQZ (tuned to no sqz) 16:26:30 - 16:29:30 (-)165.0 ref 2
FIS Mid - SQZ (tuned to no sqz) 16:32:00 - 16:35:00 (-)120.4 ref 3
FIS ASQZ 16:39:00 - 16:42:00 (+)252.2 ref 8
Mean SQZ (ADF off) 16:42:30 - 16:45:30 N/A ref 9
FIS tuned for 100Hz (to look at thermal noise limits) 16:52:00 - 16:55:00 (-)119.5 ref 10
FIS tuned for 100Hz +5deg 16:55:30 - 16:58:30 (-)124.2 ref 11
FIS tuned for 100Hz -5deg 16:59:00 - 16:02:00 (-)113.8 ref 12

We tuned FIS around 100HZ and then changed the SQZ angle +/-5deg, plot attached. Shiela thinks this could help us constrain thermal noise in the future.

Plot attached of today's data (SRCL offset at nominal -382) compared to last weeks while OM2 was hot (with SRCL offset tuned at -235). Solid lines are cold OM2, dashed lines are hot OM2. FIS is better with cold OM2 as expected.

We noticed that mean SQZ was different <100Hz with the OM2 hot vs Cold. This persisted if I changed the averages so wasn't caused by a glitch and also seemed to be increased looking at Sheila's 86736 data too. Changed SRCL offset to see if this was the cause but it didn't appear to be. Plot attached, so the cause is OM2.

Type Time (UTC) SRCL Offset DTT Ref
Mean SQZ (ADF off) 16:42:30 - 16:45:30 -382 ref 9
Mean SQZ (ADF off) 17:17:00 - 17:20:00 -235 ref 15
Mean SQZ (ADF off) 17:08:30 - 17:10:00 -100 ref 13
Mean SQZ (ADF off) 17:11:00  - 17:14:00 0 ref 14

 

OPO Setpoint Amplified Max Amplified Min UnAmp Dark NLG
80 0.0395963 0.00050063 0.00170524 -2.611e-5 22.9
Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 11:24, Monday 08 September 2025 (86782)OpsInfo

Note for Operators, changing the SRCL offset put the THERMALIZATION guardian into error. This is fine.

Once the SRCL offset is back to nominal, just load the THERMALIZATION GRD and it's will go back to normal, I'm not sure what would happen if this was done early in the lock when the GRD was still ramping the offset though.

LHO VE
david.barker@LIGO.ORG - posted 10:17, Monday 08 September 2025 (86781)
Mon CP1 Fill

Mon Sep 08 10:07:09 2025 INFO: Fill completed in 7min 5secs

Gerardo confirmed a good fill curbside. TC-A became good again at 05:56 this morning and looks nominal in today's fill.

Gerardo inspected the TCs and suspects TC-A wiring had been moved by animal activity. Why it corrected itself this morning is a mystery.

Images attached to this report
H1 PSL
ryan.short@LIGO.ORG - posted 09:54, Monday 08 September 2025 (86780)
PSL 10-Day Trends

FAMIS 31102

The RefCav alignment has been drifting again this week and causing the signal on the TPD to drop, so this could use a touchup. I'll try to find some opportunistic time to do this after a lockloss either today or later this week.

No other major events of note.

Images attached to this report
H1 General
ryan.crouch@LIGO.ORG - posted 07:39, Monday 08 September 2025 - last comment - 09:39, Monday 08 September 2025(86775)
OPS Monday day shift start

TITLE: 09/08 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.09 μm/s
QUICK SUMMARY:

Comments related to this report
ryan.crouch@LIGO.ORG - 08:44, Monday 08 September 2025 (86776)

15:20 UTC Superevent S250908y came through, so we delayed the start of commissioning till 15:43 UTC even though there was no automated stand down generated from the event. It seems the event was actually at 15:02 according to GraceDB, so there was some kind of delay? We probably could have gone ahead at 15:30.

ryan.crouch@LIGO.ORG - 09:39, Monday 08 September 2025 (86779)PEM

The dust monitor in the diode room and the optics lab were reporting that the counts were stuck, I physically restarted the DR dust monitor then its' IOC as the power cycle didn't help. The optics lab was frozen so a power cycle fixed it, I restarted the IOC as well, I probably didn't need to. I reset the alarms levels after restarting the IOC. The reference wiki page on how to restart the dust monitor IOCs.

H1 AOS
ibrahim.abouelfettouh@LIGO.ORG - posted 22:00, Sunday 07 September 2025 (86774)
OPS Eve Shift Summary

TITLE: 09/08 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY:

IFO is in NLN and OBSERVING as of 14:13 UTC (37 hr lock!)

LOG:

 

H1 General
ryan.crouch@LIGO.ORG - posted 16:31, Sunday 07 September 2025 (86772)
OPS Sunday Day shift summary

TITLE: 09/07 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 159Mpc
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: We stayed locked the whole shift, over 31 hours.
LOG: No log.

 

Images attached to this report
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:00, Sunday 07 September 2025 (86773)
OPS Eve Shift Start

TITLE: 09/07 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 14mph Gusts, 9mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.12 μm/s
QUICK SUMMARY:

IFO is in NLN and OBSERVING as of 14:13 UTC (31 hr lock!)

 

LHO VE
david.barker@LIGO.ORG - posted 10:13, Sunday 07 September 2025 (86771)
Sun CP1 Fill

Sun Sep 07 10:07:43 2025 INFO: Fill completed in 7min 39secs

Good fill. TC-A continues to be all over the shop, occasionally giving believable readings. Similar to yesterday it was TC-A which had a larger Delta-temp and triggered the end-of-fill, closely followed by TC-B.

Images attached to this report
H1 General
ryan.crouch@LIGO.ORG - posted 07:31, Sunday 07 September 2025 - last comment - 08:02, Sunday 07 September 2025(86769)
OPS Sunday DAY shift start

TITLE: 09/07 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 4mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.12 μm/s
QUICK SUMMARY:

Comments related to this report
ryan.crouch@LIGO.ORG - 08:02, Sunday 07 September 2025 (86770)

I was restarting some FOMs and noticed that PIMON looked frozen, sure enough the last lockloss data it saved in /ligo/data/pimon/locklosses/ was from Aug 19 2025 06:54:13 UTC. I restarted it at 15:00 UTC.

H1 SQZ
camilla.compton@LIGO.ORG - posted 07:15, Sunday 07 September 2025 - last comment - 07:16, Sunday 07 September 2025(86767)
Oli's SQZ Troubleshooting, SHG power low: TEC Setpoint adjusted, plus other things...

Oli, Camilla (Oli did the work, I just wrote down what they did!)

Oli had been troubleshooting SQZ and found that when the OPO ISS tried to engage, it go to 10V and then unlocked. The OPO Trans output was only ~50uW on LOCKED_CLF_NO_ISS. This was meaning the ISS was maxing out it's 0-10V range to get to the nominal setpoint of 80uW. 

We could see on SQZT0 medm that the SHG output power was 81.2mW when nominal is ~ 100mW. This will reduce the available power to the OPO path. We followed the wiki to increase the SHG power (I should make the instructions of when to do this clearer) by adjusting the SHG SETTEMP. Doing this, Oli brought SHG power up from 81mW to 92mW. 

Then the OPO Trans output was ~65uW on LOCKED_CLF_NO_ISS (nominal 80uW). So Oli moved the waveplate between the +200MHz AOM and the SHG Rejected PD in the "to OPO" path to increase the available power sent into SHG launch. Then the AOM had enough range and the OPO could lock in LOCKED_CLF_DUAL_NO_ISS. 

Then Oli had issues with getting the FC to lock in Green so followed the wiki instructions to in do that too, trending to last good alignments with ZM3 (had moved 20urad on FC lockloss from ASC turning off) and then manually locking FC and increasing power with FC2, (FC1, ZM3) alignments. Oli increased the power in the 0,0 mode form 35 to 65 with alignments. Nominally this is > 80 but it had only been at 50 in the last lock, probably from the SHG power being low. 

Back to Observing 7:13amPT. 

Comments related to this report
oli.patane@LIGO.ORG - 07:16, Sunday 07 September 2025 (86768)
Images attached to this comment
H1 SUS (ISC, SEI)
jeffrey.kissel@LIGO.ORG - posted 16:12, Tuesday 26 August 2025 - last comment - 12:20, Monday 08 September 2025(86589)
H1 SUS SR3 M1 Pitch and Yaw Estimator: According to ASC -- Missing L to P Sus. Point Contribution Restores Damping on 0.64 and 0.75 Hz Mode
J. Kissel

After today's H1 SUS SR3 Pitch Estimator's inclusion of the Sus. Point L to M1 P feed forward from the HAM5 ISI GS13s (where we had mistakenly only installed Sus. Point P to M1 P) -- see LHO:86567, I now compare the times of ASC signals (using 0.02 Hz binwidth, 64 sec FFT chunks, 30 averages, and a Hanning window with 50% overlap):
    2025-08-20 18:29 UTC - Both P and Y Estimators are OFF 
    2025-08-24 18:18 UTC - Both P and Y Estimators are ON, but for the P estimator, only the Sus. Point P to M1 P contribution is included in the GS13 FF
    2025-08-26 22:00 UTC - Both P and Y Estimators are ON, and for the P estimator, both P to P and L to P contributions are included in the GS13 FF.

I'm showing only the control signals, and only the ASC DOFs which are impacted by SR3: DHARD, MICH, SRC1, and SRC2; both pitch and yaw.

In All DOFs, the 0.64 and 0.75 Hz modes that had been made worse with only the P to P model of suspension point contribution have now been restored to no-estimator levels.
In some DOFs, the 1-10 Hz broadband motion is just barely improved, but enforces the conclusion that SR3 is only partially, if not 'not the dominate contribution' to the ASC signals. 
Here's a fun one for you -- look at SRC2 Y CTRL -- and look at how much the SRC2 *yaw* motion has changed from including the *longitudinal* suspension point contribution to M1 *pitch* top mass. #ThrowsHandsInAir

But -- overall -- with the improvements and all design intent included -- the SR3 P and Y estimators slightly improve the ASC noise from 1 to 10 Hz. Great!

As Oli notes in LHO:86551 comparing all of these different configurations from totally different days and times is dubious. We'll get "official" versions of all of these ON vs. OFF configurations on Thursday 8/28.

I also attach the local estimator metrics for pitch described in LHO:86553, comparing "only Sus Point P to M1 P modeled contribution" against "P to P and L to P sus point to M1 contribution." That similarly shows that L to P contribution forms a good fraction of the signal needed for damping.
Images attached to this report
Comments related to this report
edgard.bonilla@LIGO.ORG - 14:18, Wednesday 27 August 2025 (86600)

Great to see that the inclusion of the SUSpoint L to M1 P block makes a difference.

Unfortunately, the P2Y coupling is a wierd non-reciprocal coupling we were seeing on the measurements [See page 21 about the SR3 measurements from june after an OSEM calibration test]. The Pitch to Yaw transfer function is consistent with the conversion of pitch motion to observed yaw signals on the LF and RT OSEMs, this apparent motion gets turned into yaw feedback drive and creates the transfer function you see in the measurements linked.

Brian and I discussed it before and we had decided against adding a cross-term in the estimator to address it because it is both not well understood (at least we don't know why the LF and RT OSEMs see Pitch signals) , and because it is possible that we might only be able to address it if we commission the estimators in a fully serialized way, which would be even more time consuming for (probably) minimal benefit.

Depending on schedule, we can figure out if it is worth adding an M1 drive P to M1 Y estimator path, and probably commission a baby version of it with the measurements that Oli already took to see if it makes a difference.

oli.patane@LIGO.ORG - 12:20, Monday 08 September 2025 (86786)

I've confirmed in 86784 that the excess noise seen in the red trace between 1-3 Hz is not due to the SR3 Estimator having L2P compensation

Displaying reports 4841-4860 of 89152.Go to page Start 239 240 241 242 243 244 245 246 247 End