Displaying reports 49261-49280 of 87710.Go to page Start 2460 2461 2462 2463 2464 2465 2466 2467 2468 End
Reports until 19:10, Wednesday 29 November 2017
H1 PEM (SEI)
brian.lantz@LIGO.ORG - posted 19:10, Wednesday 29 November 2017 - last comment - 22:40, Friday 01 December 2017(39567)
direction of the anemometers?

Noted that at about 10 am - 10:04 am PST on top of End Y, the anemometer was pointing to about 20 deg to the -Y of +x and that sensor is not moving very smoothly.
The sensor is on the upper roof, about 1/2 way along Y, and about 8 feet from the +x edge, and about 6 feet up.

10 am = 1196013618 for 600 sec - epics data says that EY is about 360. so this is = north
Perhaps the EY sensor is aligned with north, but need to recheck. 

tconvert 11:00:00 pst -> 1196017218

at 10:59 am, the EX sensor was pointed in the +x direction. It then moved to point in the -Y direction. 

at GPS time 1196017200, the direction reading for EX is 183 deg. 
at ...17254, it moves up to 271 degrees. I think this means that EX wind direction is aligned with the observatory coordiatates. That wind coming from the +x direction reads +180, and wind coming from the -Y direction readys 270. This is not crazy, since wind blowing from -x towards +x would read 0. 

The EndX direction monitor is moving smoothly. I also note that the black hat thing has come off. looks like it used to cover a wire wound resistor and a toggle thing. I do not know what it is supposed to do, but it will probably be happier with the cover replaced. 

Comments related to this report
daniel.sigg@LIGO.ORG - 08:37, Thursday 30 November 2017 (39572)

I assume the black black hat thing is the cone that collects the rain. The toggle thing is the rain gauge. The heating resistor is there to melt the snow.

brian.lantz@LIGO.ORG - 22:40, Friday 01 December 2017 (39606)CDS, SEI

Went back and checked the EY anemometer this afternoon. Looks like the direction indicator for that one is not working correctly. At 3:39 pm I held the direction indicator so that it should read wind from the +X direction (180 according to  Robert's site standard, see alog 28456 ). I held it there for about a minute, then pointed it so that it appeared to be a wind from the -Y direction, and should read 270. Held again for about a minute, then pointed back to +x (which is about where I found it, and where the wind is actually coming from). In the attached plot below, you can see that it isn't reading correctly, and also seems pretty noisy. (at about t=0, it should read 180, not 240, and around t=60, it should move up and read about 270, not 340)

Anyway, the EY anemometer direction is not to be trusted until Richard has a chance to take a look.

Images attached to this comment
H1 ISC (ISC, SUS, SYS)
jeffrey.kissel@LIGO.ORG - posted 19:07, Wednesday 29 November 2017 (39566)
More By-Hand IFO Alignment Success: PRM, SRM, and ZM2 aligned
S. Dwyer, K. Kawabe, J. Kissel, B. Weaver, G. Moreno, C. Vorvick

We've completed the by-hand alignment work needed with the PSL beam, picking up where we left off yesterday (LHO aLOG 39553). That means, we've reconstructed an alignment (using steps 1-11 in LHO aLOG 39547) to get a well-aligned PSL beam from the IMs, centered through SRM and coming out of the OFI well, and then repeated steps 12-14 from LHO aLOG 39553 which where unsuccessful yesterday: 
(12) make sure ZM2 was well aligned by
    (a) inserting the temporary half-wave-plate into the output apeture of the OFI faraday rotator, then
    (b) mechanically adjusting ZM2 to hit the last (most -Y, closest to the future VOPO) iris on HAM6 as shown in Figure 4 of E1500469, then
    (c) sealing in the alignment reference by placing the other iris (most +Y, closest to HAM56 septum) in the beam path and aligning it to the beam.
(13) make sure that SRM is well-aligned by 
    (a) first mechanically adjusting the alignment *some* this morning with the aux laser -- see LHO aLOG 39556, then 
    (b) further mechanically adjusting the alignment the *rest* of the way, using the well-aligned PSL beam, then 
    (c) fiddling with the sliders, until we see fringing patterns at SR2 with the SRY cavity aligned again. Because Betsy is awesome, we see fringing with zero digital alignment bias on SRM. BOSS!!
(14) make sure that PRM is well-aligned by 
    (a) driving it digitially with slider bias,
    (b) looking for fringe fringe patterns at one of the ITMs (we chose to misalign ITMX, and look for fringing on ITMY from the PRY cavity)

(note, chronologically, today, we did steps 14, then 13, then 12).

Attached is a screen shot of OSEMs and sliders once we got SRM aligned, and SRY was flashing. PRM and ITMX are mis-aligned during this screen shot: their slider values should be PRM -995.6 in pitch (yaw should be as shown), and ITMX should be +189.3 in Yaw (pitch should be as shown).

Notes: 
- We're still using only one set of cables to control the OFI, OMs 1-3, or ZMs 1-2, and they all are fed into the CDS system as OM1. We had last used that set of cables to test the OFI (see LHO aLOG 39033). Which means when Betsy and Travis plugged in that set of cables into ZM2 to align and center ZM2's OSEMs this morning, the control system was still configured for an OFI. I've fixed all that, and it's now working as a standard HTTS, and we've confirmed that we can push ZM2 to steer the beam as needed into HAM6, AND, I've confirmed that damping works (though just with a quick look at the MEDM screen). I attach a screen shot of ZM2 functional in the OM1 infrastructure, functionally running as an HTTS.

- As of writing this log, Keita, Sheila, and Gerardo are still out there making the changes to the OM1, OM2, and OM3 alignment, as described in LHO aLOG 39504. I attach some notes about their re-assessment of the plan this evening, which had to be re-measured because LHO aLOG 39504 was thought out before the septum window was rotated.

Great work again today team, and good luck to those still out there!
Images attached to this report
Non-image files attached to this report
H1 AOS
betsy.weaver@LIGO.ORG - posted 18:46, Wednesday 29 November 2017 - last comment - 10:02, Thursday 30 November 2017(39565)
ZM2 HAM5 Tip Tilt BOSEMs centered up

While alignment work was happening in other areas, Travis and I worked on getting the ZM2 OSEMs useful.  Fil and Richard helped me hook ZM2 satellite cable up to the OM1 since the ZM2 medm is all white.  We took the OLVs and then set all 4 BOSEMs to the 50% OLV mark. 

Note, since the cables between the rack and the chambers have not yet been run, we will have to move the 1 temp cable between boxes, one at a time for OM1, OM2, OM3, ZM2, and the OFI during the imminent closeout testing.

 

Comments related to this report
betsy.weaver@LIGO.ORG - 10:02, Thursday 30 November 2017 (39573)

ZM2 BOSEM Details:

Location S/N OLV 50% OLV
UR 280 31000 15500
UL 252 30200 15100
LR 259 32767 16383
LL 027 28000 14000

 

Cables used are:

D1000234 S/N  S1000694

D1000225 S/N  S1106781

H1 SEI (PEM)
brian.lantz@LIGO.ORG - posted 18:44, Wednesday 29 November 2017 - last comment - 10:06, Monday 11 December 2017(39564)
Smoke candle test at EndX test fence

smoke test at EndX - Brian Lantz, Jim Warner, Siddhesh Pai, Brijesh Pant
Nov 29 2017, about 2:08 pm local
used 1 2B candle from superior signal

burn is ~60 sec.
https://youtu.be/XbouhfAK1Mg

smoke candle is ~10ft in front of the fence,

The smoke candle is about 11 ft. up off the ground, close to the vertical center of the fence material. The picture below shows the pole we used w/ Jim as scale standing next to the fence - The smoke emitter will be placed at the height of the blue tape on the pole.

We took several videos, the one from Brian's phone is linked above.

The smoke trail clear to the fence, but often difficult to see downstream of the fence. Why?
1. partly due to blue sky behind early trail and clouds behind late trail. note this for future film angles
2. trail is darker behind fence - might be some shade?
3. trail is getting bigger.
4. the place where the candle burned though was next to the post.

 - next tests we might try 2 candles together, and try to shoot against blue sky. i'm not sure I have enough candles to use 2 each for all my desired tests.

We see that the smoke moves through the fence. It seems like the fence might be "softening" the edges of the trails my generating lots of short scale mixing, but the structure at the ~1 foot scale seems unaffected by the fence.

Is the air moving more slowly behind the fence? Maybe? I sort of looks this way, but I can't understand how this would be true, if none of the air flow is getting diverted up-and-over or sideways-and-around the fence.
1. It is not going up over the fence -> if you assume the same density, then conservation of mass suggests that the speed on the two sides needs to match.
2. It might be slipping sideways instead of going over - this could allow slower speeds and would not be visible from this camera view -
3. It sort of looks slower - this could be the result of the smoke getting further away from the camera as it moves downwind. how to watch for this at EY?
 - maybe have a camera on the ground at the corner of the building looking up?

4. The structure of the plume moves right though this fence. The fence can't be adding large, immediate changes to the vertical velocity

More videos and pictures will be added, and we are planning to do some additional tests on Friday.

Jim set up an anemometer a bit (something like 50 feet) upwind of the fence to record the wind speed. Data download is pending.

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 10:06, Monday 11 December 2017 (39720)

Attached plot shows the wind speed recorded by the upstream ground anemometer during the smoke test. I'm also including the CSV data from the sensor. Apparently the alog won't allow CSVs, so I renamed it to a text file. w/e

Images attached to this comment
Non-image files attached to this comment
LHO General
patrick.thomas@LIGO.ORG - posted 16:24, Wednesday 29 November 2017 (39562)
Ops Shift Summary
TITLE: 11/29 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Planned Engineering
LOG:

16:48 UTC Terry to squeezer bay, laser hazard (all day)
16:49 UTC Peter to optics lab
17:22 UTC Nutsinee to squeezer bay
17:37 UTC Kyle to mid stations
17:38 UTC Rakesh, Brian and Jim to end X
17:39 UTC TJ to LVEA and then optics lab
17:40 UTC Betsy and Travis to LVEA, start of SRM alignment
Filiberto to CER, cabling for ZM2
17:57 UTC Kyle to end stations
18:09 UTC Hugh to HAM6, CPS shortening issue
18:33 UTC Karen and Vanessa to mid Y to clean
18:37 UTC Kyle back. Kyle and Gerardo to LVEA to check orientation of gate valves.
18:37 UTC TJ out of optics lab
19:00 UTC Jason to HAM5 to build AWC mounts
19:09 UTC Peter out of optics lab
19:16 UTC Rakesh, Brian and Jim back
19:28 UTC Kyle out of LVEA
19:29 UTC Karen and Vanessa leaving mid X
19:50 UTC Greg to TCSY table
20:24 UTC Keita and Jeff K. to LVEA
20:37 UTC Travis and Betsy back
20:38 UTC Jason back
20:47 UTC TJ to optics lab
20:58 UTC Corey taking guest on tour through LVEA
21:03 UTC Betsy to LVEA
21:15 UTC Jim and Brian to end X to take videos of smoke trails
21:48 UTC Betsy and Richard to CER
21:54 UTC Corey back, taking guest to overpass
22:09 UTC Gerardo to HAM5
22:51 UTC TJ back
22:54 UTC Jim and Brian back
23:00 - 23:38 UTC Site weekly meeting
23:52 UTC Hugh to HAM5, cabling for OFI
H1 SYS
betsy.weaver@LIGO.ORG - posted 14:02, Wednesday 29 November 2017 - last comment - 14:56, Wednesday 29 November 2017(39560)
VP Protection on doors during install

All doors that were removed (HAM2,3,4,5,6 N and S, and BSC 3 and BSC1 biergarten side) had proper viewport protectors on them either already, or when the tables were disconnected and removed.  See pictures of all doors that are hanging around LVEA, removed from chambers.

This assay is in response to something LLO brought up in FRS https://services.ligo-la.caltech.edu/FRS/show_bug.cgi?id=9479

 

Images attached to this report
Comments related to this report
kyle.ryan@LIGO.ORG - 14:56, Wednesday 29 November 2017 (39561)
LIGO-M1100039-v5 (LHO HAM Door Removal and Installation Procedure) and LIGO-M1100068-v5 (LHO BSC Door Removal and Installation Procedure) already require that the Lexan shutters be in the closed position prior to door removal/installation.
H1 TCS (TCS)
greg.grabeel@LIGO.ORG - posted 13:19, Wednesday 29 November 2017 (39558)
New Periscope Optic Mount Installed

The TCS CO2 tables have had the new periscope optic mount installed. Y-arm required a small modification to allow parts D1600106-V1 and D1201310-V2-01 to fit together. These new pieces have not been aligned to the table, or in-vacuum as yet.

Images attached to this report
H1 SEI
patrick.thomas@LIGO.ORG - posted 12:56, Wednesday 29 November 2017 - last comment - 16:29, Wednesday 29 November 2017(39557)
H1 ISI CPS Noise Spectra Check - Weekly
FAMIS 6926

ETMY_ST1_CPSINF_H2 and ETMY_ST2_CPSINF_V2 are elevated.

Plots posted for end X and end Y. All others ISI platforms are locked.
Images attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 13:52, Wednesday 29 November 2017 (39559)

The CPS noise check should still be informative with locked platforms (since it is electronic noise we are usually chasing).

patrick.thomas@LIGO.ORG - 16:29, Wednesday 29 November 2017 (39563)


		
		
Images attached to this comment
H1 SUS
betsy.weaver@LIGO.ORG - posted 12:47, Wednesday 29 November 2017 (39556)
SRM SUS mechanically realigned, bias relieved

Travis and I spent the morning realigning the SRM suspension in order to bring the pitch up onto IFO beam paths.  We used the red visible AUX1 laser to do this since it was much easier to hunt in the HAM4-5 Tube.  After walking the ~3" diffuse beam around everywhere we finally managed to set it on the right path.  We realigned all of the OSEMs which then saw this repointing and then looked at a few quick TFs to see that it still looked free.  We're handing back off to IR alignment team who are re-mapping the PSL beam to HAM5 for continued OFI alignment.

H1 PSL
patrick.thomas@LIGO.ORG - posted 11:56, Wednesday 29 November 2017 (39555)
Weekly PSL Chiller Reservoir Top-Off
FAMIS 6551

Added 200 mL H2O to the crystal chiller.

The fault light on the diode chiller is not lit and the water level indicator on the controller screen indicates the water level is OK.

The canister filters appear white and free of debris.
H1 ISC (INS, IOO, PSL, SUS)
jeffrey.kissel@LIGO.ORG - posted 18:27, Tuesday 28 November 2017 - last comment - 19:26, Friday 01 December 2017(39553)
By-hand IFO Alignment Success Finally
S. Dwyer, J. Kissel, K. Kawabe, E. Merilh, G. Moreno, C. Vorvick

With the help of all listed above, we were able to complete the entire by-hand, bypassed-input-mode-cleaner, IFO alignment plan, as described in LHO aLOG 39547. This means that the we are confident that the mechanical alignment of the following optics are good enough that we will be able to recover some alignment after we pump down:
    IM1, IM2, IM3, IM4,
    PR2, PR3,
    ITMX, ITMY, BS
    SR2, SR3.
I.e. all of these optics have enough range on their sliders that it's possible to find *an* alignment that sends a beam through the output faraday, can create a michelson fringe, and can retroreflect beams from the ITMs back to PRM.
In closing of the procedure -- after aligning SR2, SR3, and ITMY to steer a michelsom fringing beam through SRM into the OFI -- we went back to PRM, and checked whether the ITMs were still retroreflecting back to PRM. We found their beams in the PRM iris, but a little bit +Y in yaw and a little bit low in pitch. We suspect that this is merely a result of the drift of optics or the input beam over the course of a day's work. 

We were even able to get a little bit further than planned: we checked 
(12) that the IFO beam, incorrectly polarized with a temporary half-wave plate on the OFI, went to ZM2. It did, but the check admittedly not very precise.
(13) SRM alignment by looking for the reflected beam off of SRM. Sadly SRM is out of range of its sliders in pitch. Not a terrible surprise, since this brand new monolithic optic has not yet had its mechanical alignment fine tuned. We'll work with the suspension team to fix this in the coming days.
(14) to see if we could align PRM, by looking for secondary retro-reflections from the ITMs back at PRM. We found secondary reflections, and were able to move them with PRM, but it was too late and we got confused as to which beam was which ITM, so we decided to call it a night. (Note: before playing we restored the alignment to the values that got the REFL beam back into HAM1, so that path is still good -- so far.)

All-in-all, a really successful, power-house of a day. Nice work team!!

What does this leave next?
- Mechanically aligning SRM to get a reflected beam back to the ITMs.
- Exploring PRM sliders to align secondary beams into PRM iris if not well-co-aligned with the forward input beam and the primary reflection.
- Move OMs to catch the new alignment through the OFI
Unfortunately, that means we'll have to reconstruct this alignment tomorrow, but hopefully now that we've gotten this far, it'll be easier to get back to this point faster tomorrow.

Please remember -- no computer / front-end restarts until further notice.


I attach a screenshot of all OSEM values and suspension (plus PSL PZT) slider values at the end of the day, with all optics mentioned above in their desired alignment state.
Also, I attach a more detailed log of the activity.
Images attached to this report
Non-image files attached to this report
Comments related to this report
cheryl.vorvick@LIGO.ORG - 19:26, Friday 01 December 2017 (39605)IOO

The IM alignment posted here is not final.  The final alignment before bias relief and adding o-rings is posted in alog 39577.

H1 TCS
filiberto.clara@LIGO.ORG - posted 16:35, Tuesday 28 November 2017 (39552)
HWS Camera Work Update

WP 7222
FRS 4559

Continued with WP 7222, isolating the HWS cameras. Both end stations were completed today. Same steps taken as work done in LVEA, see alog 39532. All temporary power cables removed from enclosures and cameras connected to Hartmann Corner/End Station Breakout Power Chassis.

F. Clara, T. Shaffer

H1 General
jeffrey.bartlett@LIGO.ORG - posted 15:58, Tuesday 28 November 2017 (39551)
Ops Day Shift Summary
Ops Shift Log: 11/28/2017, Day Shift 16:00 – 00:00 (08:00 - 16:00) Time - UTC (PT)
State of H1: Unlocked - Vent
Intent Bit: Engineering
Support: N/A
Incoming Operator: N/A
Shift Summary: Continuing vent work
 
Activity Log: Time - UTC (PT)
15:52 (07:52) Richard – Going into the CER
16:00 (08:00) Start of shift
16:05 (08:05) Richard – Out of CER
16:35 (08:35) Filiberto & Peter – Going to electronics rack by HAM2
16:40 (08:40) John Feicht from CIT on site
17:09 (09:09) Filiberto & Peter – Out of the LVEA
17:21 (09:21) Filiberto & TJ – Going to both end stations to isolate the HWS camera
17:58 (09:58) Cheryl, Keita, Ed, & Jeff K, - Going to HAM2/3 for IO alignment
18:02 (10:02) Brandon King – Dychem on site to service blue mat
18:08 (10:08) Jeff B – Give Dychem rep a quick tour of LVEA
18:40 (10:40) Terry & Nutsinee – Going to Squeezer Bay
18:45 (10:45) Brandon King – Leaving site
19:05 (11:05) End-Y Laser Hazard
19:36 (11:36) Gerardo – Going into LVEA and then into the Optics Lab
20:00 (12:00) TJ & Filiberto – Finished at End-Y – Both Ends are back to Laser Safe
20:32 (12:32) Ed – Out of the LVEA
20:33 (12:33) Jeff K. – Out of the LVEA
21:30 (13:30) Jeff K. & Ed - Back to the LVEA for IO alignment work
22:32 (14:32) Kyle – Going to Mid-Y
23:56 (15:56) Klye - Back from Mid-Y
00:00 (16:00) End of shift
 
H1 SEI
jim.warner@LIGO.ORG - posted 14:31, Tuesday 28 November 2017 (39549)
Sensor correction tests on ETMY

SamC and ConorML have designed some sensor correction filters using some matlab swarm optimization routines. Their designs are summarized in SEI log 1234 (yes, really) and a technical note in the DCC (https://dcc.ligo.org/DocDB/0147/T1700541/001/T1700541-V1.pdf ). I've installed these to test on ETMY and I'm leaving them running overnight. Microseism is pretty high right now so it's a good time to test. I've already collected some data, but the differences aren't dramatic, but I wasn't comparing very long stretches, I found a long window from last January with high microseism, that I'll try to use for comparison tomorrow.

First attached plot is the St1 Y motion measured by the T240 (thick lines), their RMS (dashed) and the ground motions (thin), red traces are for the swarm filters, blue are for the broadband filter we used during O2. Second plot shows the comparison of the CPS (SCSUM_CPS, thick lines), the cps RMS (dashed) and ground again (thin lines), again red traces are for the swarm filters, blue are for the O2 configuration. Last attached plot shows the two filter designs in displacement, and their complements, stolen from the DCC doc. Blends were not changed during these measurements. 

Sam says that the filters were design when .1-.3 hz gnd blrms was around 450 nm/s. I've put ETMY back to damped to give them some time with higher microseism to design with.

Images attached to this report
H1 AOS
jeffrey.kissel@LIGO.ORG - posted 13:39, Tuesday 28 November 2017 (39547)
By-hand IFO alignment Procedure Plan (Assuming IMC Bypass is IN, and IM1,2 alignment is Good)
K. Kawabe, J. Kissel, E. Merilh, C. Vorvick

Hoping not to have a repeat of yesterday's failed attempt at aligning the IFO by hand, we hit the whiteboard this morning to clearly outline our plan after Cheryl and Keita thought hard about what went wrong and what to do over night. Here's the game plan we're going forward with today:

Prep: misalign SRM, put face shield on ITMY

#      Step                            Drive Optic(s)               Alignment Metric
1      Input Align                     PSL PZT & (IM3 or IM4)       PRM and PR2 irises, IR card

2      Check PR3 spot                  (if bad, move PR2)           Target on PR3, camera viewer

3      Check ITMX Spot                 (if bad, move PR3)           Face shield on ITMX, camera viewer
  
4      Retro Reflect back to PRM       Walk PR2 and PR3             Return Beam overlap with forward beam @ PRM Iris, IR card
       For walk, 
       move PR2 opposite of PR3 in Pitch
       move PR2 same as PR3 in Yaw
       For walk, move with a ratio of 9.7 : 1 in PR2 : PR3 step sizes

5      Check ITMX Spot                 (if bad, walk PR2 and PR3)   Alternate between (face shield on ITMX, camera viewer) and (return beam Return Beam overlap with forward beam @ PRM Iris, IR card)

6      Check PR3 Spot                                               Target on PR3, camera viewer

--------- PRM to ITMX is set. Do not touch Input and PR optics ---------

Prep: put face shield on ITMX, keep SRM misaligned

7      Check ITMY Spot                 (if bad, roughly adjust BS)  Face Shield on ITMY

8      Retro Reflect IY back to PRM    Walk ITMY & BS               Return Beam overlap with forward beam @ PRM Iris

--------- PRM to ITMY is set. Do not touch Input, PR, ITMs, or BS optics until Step 11 ---------

Prep: remove ITMX face shield

9      Check SR3 spot                                               Target on SR3, camera viewer

10     Check SR2 spot(s)                                            IR card

11     Fine-tune ITMX and ITMX beam on SRM                              IR card
                                      Walk (ITMY & BS) for differential spot change
                                      Walk (SR2 & SR3) for common spot change / centering
       For walk, move with ratio 5.8 : 1 in SR2 : SR3 step sizes

As we break for lunch, we're mid-way through step 5.

I attach a screenshot of our alignment offsets and OSEM values as we left it. #TouchWood
Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 12:49, Tuesday 28 November 2017 - last comment - 14:33, Tuesday 28 November 2017(39545)
long baseline minute trend dataviewer plots drawing issues

Chandra noticed some strange dataviewer plots when asking for long look-back minute trends (see attached). After some investigation, I found:

The problem is with h1nds1's configuration of archived raw minute trends. The most recent archive spans the dates 5/26 to 10/18. The older archives are not being accessed, which I think is a result of my reordering the archived directories in the daqdrc file. After Dan and I had moved all the minute trends to their own RAID, I noticed that raw minute trends are ordered oldest-to-most-recent which would not be efficient since most data requests are for recent (within the last year). I reordered the list to most-recent-to-oldest and now only the most-recent directory is being read. When we are able to restart the NDS, I'll reverse the order to verify this. In the mean time, minute trends before 5/26/2017 are unavailable.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 12:54, Tuesday 28 November 2017 (39546)

opened FRS9506 for this ticket.

keith.thorne@LIGO.ORG - 14:33, Tuesday 28 November 2017 (39550)CDS, DAQ
The daqd process assumes that the list is from oldest to newest.   This is how it has been since Alex Ivanov implemented it.  Yes, likely need to improve documentation but usage has been limited to LLO and LHO.
H1 ISC (INS, ISC)
keita.kawabe@LIGO.ORG - posted 18:12, Monday 27 November 2017 - last comment - 09:09, Wednesday 29 November 2017(39535)
HAM5 ZM path beam dump is done (Jeff, Sheila, Keita)

Since the reflection of IR beam on septum window AR is so small, we shot a red laser pointer beam from HAM5 through septum so the beam clears the HAM6 iris closer to HAM5, and falls on the second iris closer to VOPO within a mm from the center. Sheila calculated the deflection difference between IR and red, and it's so small it doesn't make much difference over the lever arm (a meter or two) in this case.

There were two beams coming through the septum, we took the first one (brighter, and towards +X) as the main beam.

There were three beams reflected off of the septum into HAM5. We took the first two towards +X as the main beams to dump.

We also confirmed that the scribe mark showing the thickest side of the septum is -X for the ZM path and +X for the OMC path.

Pictures will be added later.

Comments related to this report
keita.kawabe@LIGO.ORG - 18:22, Monday 27 November 2017 (39536)

1. Looking at the zm path septum window from HAM6. Laser pointer is in HAM5, already aligned. The beam spot on the septum is all inch or so away from the +X edge.

 2. How the two AR reflections are on the beam dump but the third beam which should be AR of AR is not caught. This should be ok.

3. ZM path in HAM5.

4. ZM path septum scribe mark, -X.

5. OMC path septum scribe mark, +X.

Images attached to this comment
keita.kawabe@LIGO.ORG - 09:09, Wednesday 29 November 2017 (39554)AOS, SYS

SYSTEMS alert:

Potentially, the location of the HAM5 septum baffle needs to be adjusted accordingly because the septum window for the main OMC path is rotated 180 degrees WRT the design.

The first reflection by the septum window is not affected but the second reflection (i.e. reflection by the surface on the HAM6 side) will move from -X side to +X side of the first reflection.

TJ is contacting systems to figure that one out.

Displaying reports 49261-49280 of 87710.Go to page Start 2460 2461 2462 2463 2464 2465 2466 2467 2468 End