Displaying reports 51261-51280 of 85417.Go to page Start 2560 2561 2562 2563 2564 2565 2566 2567 2568 End
Reports until 13:57, Tuesday 28 March 2017
H1 SUS
thomas.shaffer@LIGO.ORG - posted 13:57, Tuesday 28 March 2017 (35148)
Charge Measurements For EX and EY

EY charge is still growing and the angular actuation strength continues to slowly change, but the longitudinal relative actuation strength is still around 3-4%. This is basically the same as last week (look at Jeff's alog, not mine).

Images attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 13:56, Tuesday 28 March 2017 (35149)
Serviced Compressors #1 and #2 at Y-Mid Vent/Purge-Air Skid

Unit #1 and #2 are done, grease was applied to scroll pump for unit #1 and #2, and to all electrical motors.

Compression test results:  
- Unit #1 pressure 125 psi
- Unit #2 pressure 120 psi
- Unit #3 pressure 120 psi
- Unit #4 pressure 125 psi
- Unit #5 pressure 125 psi

Scroll pumps #3, #4 and #5 need to be greased.

Unit #2 needs a new one-way valve.

Unit #5 needs a new relief valve, but all other relief valves are OK.

Done under WP 6541.

 

H1 SEI
hugh.radkins@LIGO.ORG - posted 13:43, Tuesday 28 March 2017 - last comment - 14:10, Tuesday 28 March 2017(35147)
More Insulation added to BRSY Seismo Table

Following improvements seen with insulating the Legs of the table, 34805, & 34869, 1 to 2 inches of insulation has been added to the table held by the legs which is holding the T240.  See the photos in 34805 and compare with the photo attached here.

Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 14:10, Tuesday 28 March 2017 (35150)

Sad thing though, I should have realized, the decline of the driftmon now puts the camera image in jeopardy of being too close to the edge.  I had the enclosure open for a time and the cool down has pulled the drift signal down to -15660.  It could be 24 hours before things warm up enough to bring it back to before incursion level of -14000.

Ed has the BRSY engaged to nominal now but we are watching to make sure it is doing some good for the IFO.

We'll need to make another centering effort within a couple weeks.

H1 SEI
hugh.radkins@LIGO.ORG - posted 13:16, Tuesday 28 March 2017 (35145)
WBSC1 ITMY Stage2 Corner2 CPS Bumps from T240, now gone

As reported in 34677, the wandering bump in the ST2 Corner2 CPS Spectra showed again.  First seen at this location in June 2016, it was gone after a site power outage. A couple weeks ago, test suggested its origins were inside the vacuum system, 34820.  I speculated other in-vacuum sensors might be the perpetrators.

Today, with the ISI offline, I sequentially powered down and powered back up the Corner2 GS13s, the L4Cs, and then the T240.  The bump while steadily wandering around, remained proudly above background until the T240 was off.  So far it has not returned.

See the attached plots for the Sensor signals for time line and the CPS Spectra (note T0 time.)

I will update and close FRS 7576.

Images attached to this report
LHO General
thomas.shaffer@LIGO.ORG - posted 12:43, Tuesday 28 March 2017 (35144)
LVEA Swept

I unplugged a few unused extension chords and turned off the CS wifi, but that was all that I touched.

There are much more PEM cables strewn across the floors of the LVEA, at least compared to last time I was in there. Please be cautious of this tripping hazard.

H1 ISC
keita.kawabe@LIGO.ORG - posted 12:10, Tuesday 28 March 2017 (35142)
Commissioning window for Wed. Mar. 29

We're planning to spend up to 8 hours for high power PRMI/DRMI test to narrow down the hot spot location in the HWSX chain.

Wed. Mar. 29 2017, from 8AM Pacific, 10AM Central, 11AM Eastern.

We're coordinating with LLO so they can do their commissioning tasks during this window to minimize the down time for double coincidence.

H1 CDS (SEI)
david.barker@LIGO.ORG - posted 12:03, Tuesday 28 March 2017 (35141)
removed most cron jobs on the HEPI pump controllers to remove glitching

Most crontab jobs were removed on h1hpipumpctrl last Wednesday, 3/22  ( https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=35001 ) and this resolved the daily and weekly 4am glitches on the fluid pressure.

Today during maintenance I removed the last unneeded cronjob on EY, the one that runs monthly. The EY configuration is:

periodicity jobs which run change
hourly none no change
daily logrotate removed 8 jobs
weekly none removed 2 jobs
monthly none removed 1 job

The old cron directories have been renamed /etc/oldcron.[daily, weekly, monthly] (hourly was not changed)

I made this the standard configuration on h1hpipumpctrl[ex, l0] this morning during maintenance.

H1 CDS (SYS)
david.barker@LIGO.ORG - posted 11:36, Tuesday 28 March 2017 (35140)
Timing glitch on EX comparator unit, rack work most probably the cause

both channels on the EX comparator went out of nominal (by a very large margin) at 17:04:40 UTC (10:04:40 PDT) for just one second. It looks like the reference signal went bad at this time, which is coincidence with cable pulling activities at EX. Signals recovered after the one second glitch, with no re-occurence.

H1 CDS (DAQ, ISC)
david.barker@LIGO.ORG - posted 11:33, Tuesday 28 March 2017 - last comment - 14:16, Tuesday 28 March 2017(35139)
h1calcs, h1omc new code, DAQ restart

WP6544

Daniel, Kiwamu, Dave:

new model code was installed for h1calcs and h1omc. This added one Dolphin IPC channel (omc sender, calcs receiver) and added two 4kHz DAQ channels from calcs.

Models were restarted, followed by a DAQ restart.

Comments related to this report
david.barker@LIGO.ORG - 14:16, Tuesday 28 March 2017 (35151)

BTW: prior to the DAQ restart it had been running for 28 days 0 hours and 38 minutes.

H1 CAL
travis.sadecki@LIGO.ORG - posted 11:21, Tuesday 28 March 2017 (35137)
PCal ETMx beam position photos

As a complement to last week's ETMy photos that Sudarshan took (aLog 34980), today I took PCal beam position photos of ETMx.  I have sent them to Sudarshan for processing.  Stay tuned for results.

H1 DetChar (DetChar, PEM)
keita.kawabe@LIGO.ORG - posted 10:45, Tuesday 28 March 2017 - last comment - 08:50, Monday 03 April 2017(35136)
glitches that are seemingly n*12.1Hz

See red and blue VS green. Bumps are at around [4, 5, 6, 7, 8]*12.125 Hz or so.

When it was really bad (red), it was easily identifiable in detchar summary page, not so when it was mildly bad (blue).

Images attached to this report
Comments related to this report
duo.tao@LIGO.ORG - 23:49, Friday 31 March 2017 (35256)DetChar

I ran coherence tool on these lines. I first tried to find a comb over there but failed. Then I tried to find them as single lines. I looked for 6 single lines: 12.125Hz, 48.5Hz, 60.625Hz, 72.75Hz, 84.875Hz, 97Hz. Two of them are not found in the coherence tool. Here are the results:

12.5Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_12.125/index.html (Weak results and the structure does not look like a single line)

48.5Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_48.5/index.html (Found in four of the weeks, in some EX and EY magnetometers. There are some channels that do not look like a single line. But there seems to be something going on so I showed them.)

60.625Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_60.625/index.html (Not found)

72.75Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_72.75/index.html (Not found)

84.875Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_84.875/index.html (Weak. Only found in one channel and the structure does not look a single line)

97Hz: https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/O2_line_97/index.html (Found in many EX and EY magnetometers)

keita.kawabe@LIGO.ORG - 08:50, Monday 03 April 2017 (35286)

Robert's measurements seem to suggest that this is the resonance of one of the baffles.

https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=35166

LHO VE
chandra.romel@LIGO.ORG - posted 09:25, Tuesday 28 March 2017 (35133)
CP4 LLCV setting reduced

Lowered actuator value to 34% open for Dewar fill this morning.

H1 SEI
edmond.merilh@LIGO.ORG - posted 08:16, Tuesday 28 March 2017 - last comment - 11:28, Tuesday 28 March 2017(35130)
BRSY Sensor Correction swithced OFF in prepeation for Maintenance

15:!5UTC

Comments related to this report
edmond.merilh@LIGO.ORG - 08:27, Tuesday 28 March 2017 (35131)

That's 15:15UTC

15:20 UTC I also moved SUS ITMY to "SAFE" and took ITMY ISI to "OFFLINE" for CPS Noise investigations.

edmond.merilh@LIGO.ORG - 08:38, Tuesday 28 March 2017 (35132)

15:30UTC Due to impending PCAL work. BRSX is also off. Current state: "VERY_WINDY_NOBRSXY"

edmond.merilh@LIGO.ORG - 11:28, Tuesday 28 March 2017 (35138)

18:24 Hugh is finished his work and ITMY has been returned to it's nominal state(s)

H1 DetChar (DetChar)
miriam.cabero@LIGO.ORG - posted 15:28, Monday 27 March 2017 - last comment - 12:26, Tuesday 28 March 2017(35116)
Blip-like injections

Evan, Miriam,

While L1 was having a small lock loss, we made a series of injections (with H1 in comissioning mode) in the H1:SUS-ETMY_L2_DRIVEALIGN_Y2L_EXC channel. The injections are single sine-gaussian pulses that simulate blip glitches (see https://ldas-jobs.ligo-wa.caltech.edu/~miriam.cabero/sine-gaussian.png ).

We started very quiet and slowly increased the amplitude, so that only the last 3 of the 9 injections appear in GDS-CALIB_STRAIN. The GPS times of the injections are:

1174684631

1174684680

1174684715

1174684745

1174684793

1174684854

1174684889 *

1174684945 *

1174685355 *

* Can be seen in CALIB_STRAIN

 

We will be repeating this kind of injections at opportunistic times (L1 not observing) in the next days, taking different blip morphologies, and different amplitudes.

Comments related to this report
andrew.lundgren@LIGO.ORG - 12:26, Tuesday 28 March 2017 (35143)DetChar, ISC, SUS
Only the loudest of these saturated the noisemons, and it did so by hitting an analog limit at plus/minus 22000 counts.

I projected the drive signal into noisemon counts and looked at the last three injections on the list. In the first two, the noisemon signal tracks the drive, and the subtraction of the two is just noise. In the last (loudest) injection, the noisemon hits an analog saturation at both plus and minus 22,000 counts leaving a huge glitch in the subtracted data.

This is good because it suggests that the only important analog limit in the noisemon is this threshold. I don't have time to document it now, but I've tried the same with a set of loud detchar injections, which go up to hundreds of Hz, and I get the same behavior. So when the drive signal does not push the noisemon beyond 22,000 counts, we can trust the subtraction, and anything we see has entered the signal between the DAC and noisemon; it's a glitch in the electronics and not a result of the DARM loop.

Attached are the three subtractions, noisemon minus projected drive signal.
Images attached to this comment
H1 CAL
jeffrey.kissel@LIGO.ORG - posted 16:12, Tuesday 21 March 2017 - last comment - 10:50, Tuesday 04 April 2017(34984)
2017-03-21 New Calibration Sensing Function Measurement Suite
J. Kissel

I've gathered our "bi-weekly" calibration suite of measurements to track the sensing function, and ensure all calibration is within reasonable uncertainty and to have corroborating evidence for a time-dependent detuning spring frequency & Q. Trends of previous data have now confirmed time dependence -- see LHO aLOG 34967.

Evan is processing the data and will add this day's suite to the data collection.

We will begin analyzing the 7.93 Hz PCAL line that's been in place since the beginning of ER10, using a method outlined in T1700106, and check the time dependence in a much more continuous fashion. My suspicion is that the SRC detuning parameters will change on the same sort of time scale as the optical gain and cavity pole frequency.

Note also, that I've grabbed a much longer data set for the broad-band injection, as requested by Shivaraj -- from 22:50:15 UTC to 22:54:20 UTC, roughly 4 minutes.

The data have been saved and committed to:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Measurements/SensingFunctionTFs
    2017-03-21_H1DARM_OLGTF_4to1200Hz_25min.xml
    2017-03-21_H1_PCAL2DARMTF_4to1200Hz_8min.xml

    2017-03-06_H1_PCAL2DARMTF_BB_5to1000Hz_0p25BW_250avgs_5min.xml
The data have been exported with similar names to the same location in the repo.

For time-tracking, this suite took ~38 minutes from 2017-03-21, 22:18 - 22:56 UTC.
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 17:37, Tuesday 21 March 2017 (34987)CAL, DetChar, GRD, ISC
J. Kissel

Because the calibration suite requires one to turn OFF all calibration lines before the measurements then back ON after, the time-dependent correction factor computation is spoiled temporarily. In the GDS pipeline, which uses FIR filters, it takes about 2 minutes for the calculation to return to normal functionality and produce sensible results (Good! this is what's used to correct h(t)). However, because the front-end's version of this calculation (NOT used in any corrections of any astrophysical or control room product) uses IIR filters, it remains polluted until one manually clears the history on all filter banks involved in the process. 

Normally, as the ISC_LOCK guardian runs through the lock acquisition sequence, it clears these filter banks history appropriately. However, the calibration suite configuration is still a manual action.

Moral of the story -- I'd forgotten to do this history clearing until about 1 hr into the current observation stretch. The history was cleared at approximately  2017-03-22 00:10 UTC.

Why am I aLOGging it? Because clearing this history does NOT take us out of observation mode. Rightfully so in the case, because again the front-end calculation is not yet used in any control system, or to correct any data stream, it is merely a monitor. I just aLOG it so that the oddball behavior shown at the tail end of today's UTC summary page has an explanation (both 20170321 and 20170322 show the effect).

To solve this problem in the future, I'm going to create a new state in the ISC_LOCK guardian that does the simple configuration switches necessary so no one forgets in the future.
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 18:48, Tuesday 21 March 2017 (34991)CAL, ISC
J. Kissel

On the discussion of "Why Can't LLO the same SNR / Coherence / Uncertainty below 10 Hz for These Sensing Function Measurements?"

It was re-affirmed by Joe on Monday's CAL check-in call that LLO cannot get SNR on 5- 10 Hz data points. There are two things that have been investigated that could be the reason for this:
   (1) The L1 DARM Loop Gain is too large ("much" larger than H1) at these frequencies, which suppresses the PCAL and SUS actuator drive signals.
   (2) Because of L1's choice in location of applying the optical plant's DC readout DARM offset and avoiding-DAC-zero-crossing-glitching SUS offset means there are single vs. double precision problems in using a very traditional DARM_IN1/DARM_IN2 location of the open loop gain transfer function.
both investigations are described in LHO aLOG 32061.

They've convinced me that (2) is a small effect, and the major reason for the loss in SNR is the loop gain. However, Evan G. has put together a critique of the DARM loop (see G1700316), which shows that the difference in suppression between 5-10 Hz is only about a factor of 4. I put a screen cap of page 4 which shows the suppression.

I attach a whole bunch of supporting material that shows relevant ASDs for both during the lowest frequency points of the DARM OLG TF and the PCAL 2 DARM TF:
     - DACRequest -- shows that a factor of 4 increase in drive strength would not saturate any stage of the ETMY suspension actuators
     - SNR_in_DARM_ERR -- shows the loop suppressed SNR of the excitation
     - SNR_in_DELTAL_EXT -- shows the calibrated displacement driven
     - SNR_in_OMC_DCPDs -- shows that a factor of 4 increase in drive strength would not saturate the OMC DCPDs 

 So ... is there something I'm missing?
Images attached to this comment
shivaraj.kandhasamy@LIGO.ORG - 09:39, Tuesday 28 March 2017 (35134)

Here with attached a plot showing comparison of PCal, CAL-DELTAL_EXTERNAL and GDS for the broad band injection. As expected, GDS agrees better with PCal injection signal. The code used to make the plot is added to svn,

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Scripts/PCAL/PcalBroadbandComparison20170321.m
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 10:39, Tuesday 28 March 2017 (35135)
Just to close out the question in Comment #2 above, LLO was indeed able to use LHO-like templates and drastically improve their SNR at low-frequency; check out LLO aLOG 32495. 

Hazaah!
jeffrey.kissel@LIGO.ORG - 10:50, Tuesday 04 April 2017 (35311)
J. Kissel, E. Goetz

The processed results for this data set are attached.

For context of how this measurement fits in with the rest of measurements taken during ER10 / O2, check out LHO aLOG 35163.
Non-image files attached to this comment
Displaying reports 51261-51280 of 85417.Go to page Start 2560 2561 2562 2563 2564 2565 2566 2567 2568 End