Displaying reports 53481-53500 of 83258.Go to page Start 2671 2672 2673 2674 2675 2676 2677 2678 2679 End
Reports until 01:31, Tuesday 18 October 2016
H1 ISC
sheila.dwyer@LIGO.ORG - posted 01:31, Tuesday 18 October 2016 (30581)
Jitter noise again

Sheila, Jenne, Daniel

  1. The beam jitter seen by the IMC WFS became worse below 100 Hz as soon as the HPO was turned on, and has stayed similar since. The jitter at 100 Hz isn't much worse than it used to be. 
  2. Today Jenne and I redid a few white noise injections for MICH, SRCL, and jitter from the IMC PZT. The aux loop noise seems to be low enough that it is not our main problem.
  3. We don't think that the main coupling of HPO jitter noise to DARM is through lock point errors on the refl diode.

More about the jitter measurements:

Jenne, Daniel and I made two types of jitter measurements this afternoon, first exciting the IMC PZT and making a projection using IMC WFS DC, then exciting IM4 and making projections using REFL WFS.  The jitter measurements in the attached noise budget are based on the IMC PZT measurements, the IM4 ones are not included, because we don't believe that the REFL WFS noise floor is jitter, so these measurements were more like upper limits. The HPO jitter is in a gouy phase more similar to the one we excited using IM4 we believe, while the PZT excitations are in a gouy phase more simliar to the peaks from structures on the PSL table. 

For both types of measurements, we wanted to see if it is plausible that the coupling mechanism to DARM is through the REFL diode.  To do this we measured transfer functions to both the REFL control signal and to DARM.  When we are exciting the IMC PZT, we cause sensor noise in the IMC loop which gets supressed by the CARM loop, so measuring the coupling to DARM CONTROL and multiplying by the previously measured REFL control to DARM coupling causes us to overestimate the coupling to DARM.  For the IM4 excitation, we injected a line at 150 Hz, because we had to make a rather large injection in DARM to see it in REFL control.  Using the REFL control signal to estimate the coupling to DARM we underestimate the coupling by about a factor of 5.  This means that at least in the gouy phase that we excite with IM4 (which we think is similiar to the gouy phase where the broadband HPO jitter shows up), the main coupling of jitter noise to DARM is not through lock point errors on the REFL diode. 

Images attached to this report
H1 AOS
nutsinee.kijbunchoo@LIGO.ORG - posted 00:02, Tuesday 18 October 2016 (30613)
Ops EVE shift summary

TITLE: 10/17 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Lock Aquisition
INCOMING OPERATOR: Patrick
SHIFT SUMMARY: We had trouble locking DRMI. ISI BS ST2 CPs looked noisy but they weren't the cause of the problem. Noise eater acted up once or twice. Jenne and Sheila working on the alignment to reduce some excess noise.

Images attached to this report
H1 TCS
nutsinee.kijbunchoo@LIGO.ORG - posted 23:37, Monday 17 October 2016 (30612)
TCSY chiller top off

The water was at 5.6 cm when I went out. I added ~500 ml of water and the level indicator went all the way to the top. I noticed the water had a hard time making it's way down the filter and indeed the filter was puffed up from the bottom -- just like what Jason has seen before (alog30351). I took out the filter to let the indicator fall to it's true reading and added ~1250mL of water (total -- without filter in place). This brought the indicator from 5.6cm to 8.9cm. I put the filter back in and closed the cover very carefully so it doesn't push the metal filter on the side.

H1 ISC
kiwamu.izumi@LIGO.ORG - posted 19:09, Monday 17 October 2016 (30610)
Various out-of-loop sensors for frequency noise

Daniel, Evan, Keita, Kiwamu,

We checked the level of frequency noise at various out-of-loop equivalent sensors and then projected them to DARM to see if they make sense at all.

If we accept the projected noises, it is unlikely that frequency noise is the limiting noise source in the 200 - 1 kHz band.


[Various out-of-loop sensors for freq. noise]

We wanted to know the out-of-loop stability of frequency noise. We measured frequency noise transfer functions for the following sensors relative to REFL9I.

To do this, we excited one of the inputs of the summing node board with broad band noise from 100 Hz to 4 kHz. The relative transfer functions between REFL 9I and these sensors were found to be almost flat as expected. See the upper left panel of the third attachment for the flatness. In the reset of this report, the following scaler values will be used and we do not use the actual measured transfer functions.

We then converted the sensor spectra (taken when the interferometer was in low-noise) by using the above values to (uncalibrated) frequency noise as seen in REFL9I. The below shows (uncalibrated) frequency noise estimated by these sensors.

Good agreement is seen above 3 kHz where all the sensors see the loop-gain-limited frequency noise. For POPs, they are found to be limited by shot noise as expected at frequencies below 2 kHz or so, except that POP 9 sees some extra frequency noise in the jitter band (200 Hz- 1kHz). Since the coherence is not low below 50 Hz, we should not believe the spectra below 50 Hz or so.

[Projection to DARM]

The below plot shows projections of frequency noises to DARM using the different sensors. The coupling transfer function of REFL9I -> DARM was once measured by exciting the same excitation point as the above measurement. We confirmed that the coupling grows proportionally to frequency (29893). Also, see the upper middle panel in the third attachment for the coupling transfer function. For noise projection, I have used a simplified transfer function of 1e-5 x ( f /700Hz) [DARM meters / REFL9I counts]. Since all the out-of-loop sensors are already calibrated to equivalent of REFL 9I digital counts, all of them are just propagated to DARM by multiplying the same coupling transfer functions.

Again, the projection below 50 Hz is not trustable because of low coherence in the measurements. High frequency excess in DARM above 4 kHz seems to explainable by the frequency noise -- all the projections more or less agree with DARM. In the jitter band, 200 Hz - 1 kHz, the closest projection was given by POP 9I which showed some acoustic type structure. Although POP9I is still not high enough to entirely explain what we see in DARM. Moreover, the peak structure seen in POP 9I do not quite match the ones in DARM. Probably those structures in POP 9I are likely some kind of sensing noise.

In any case, if we believe this plot, frequency noise does not seem to be limiting the current DARM in the 200 Hz - 1 kHz band.

Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 17:46, Monday 17 October 2016 (30609)
All models compiled against RCG2.3

WP6251 RCG3.2 upgrade.

In preparation for the RCG3.2 upgrade I have compiled all 106 models against it. I did not install the new code, this was just a check that all models compile against the new system. I took the opportunity to rebuild the H1.ipc file (current has 5544 lines, new has 5096 lines).

Procedure was:

create new build area /opt/rtcds/lho/h1/rtbuild-3.2

save H1.ipc to archive/H1.ipc.17oct2016 and empty H1.ipc

do first round of compiles (make -i World) to fully populate H1.ipc (many compiles abort on ipc error)

do second round of compiles to actually compile each model (took 49 minutes)

check that all error.log files are empty (indeed they are)

save the new H1.ipc into archive/H1.ipc.rcg3_2

restore original H1.ipc file (in case any models are restarted tonight)

H1 CDS (CDS, ISC, PSL, SUS)
jeffrey.kissel@LIGO.ORG - posted 17:38, Monday 17 October 2016 - last comment - 18:49, Monday 17 October 2016(30608)
Cleared PEM, ITM, PSL ISS, and CS ECAT1PLC1 of NOT INIT and NOT FOUND Channels in SDF System
J. Kissel

After taking a look at the SDF overview screen, I noted a plethora of yellow lights indicative of many channels not found or not initialized because of front-end model changes. As such, I've initialized and monitored all channels not initialized, and I've updated all (currently in use; see attachment) reference files to rid them of channels that no longer exist. The affected sub-systems are:
PEM EX, PEM EY
PSL ISS, and
CS ECAT1PLC1
Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 18:49, Monday 17 October 2016 (30611)

[Sheila, Jenne, JeffK]

With the IFO in Down after a nice lock, we have accepted all of the SDF diffs (in monitored and not monitored channels), so that we'll reboot tomorrow in a much-closer-to-good state. 

H1 CDS (DAQ, DCS, GRD, SYS)
jeffrey.kissel@LIGO.ORG - posted 17:09, Monday 17 October 2016 (30607)
We Need Improvements to the SDF System
J. Kissel, J. Driggers, S. Dwyer

I've already mentioned this a few months ago in LHO aLOG 27517, but I'll repeat these requests again, because we've thought a little bit more on how we'd like them implemented.

Thing we need to make SDF more maintainable:
1) Remove the distinction between SAFE and DOWN for SUS and SEI. Because SEI and SUS come up with the watchdog tripped, and the start up state for both systems guardian has the masterswitch off, there's no difference between these states. 
2) A rapid way to assess and change a group of front-end model's SDF reference files. I.e. we should be able to switch between OBSERVE and DOWN quickly instead of having the 4 screen, 15 clicks per front-end process it is now.
3) A easy, built in way to find out what channels of a given front-end are controlled by all guardians, such that we can un-monitor them.
4) A much easier way to reconcile non-initialized and not-found channels than the still-confusing SDF Save and SDF Restore screens.

Regarding 
(1) -- This has been completed. The SUS that *have* down.snaps all now have their safe.snaps soft-linked to their down.snaps.
(2) -- we now have a rapid way to assess at which reference file a front end is looking, because I've modified the overview. However, we still need a rapid way to change them all. Sheila and Jenne have programmed the enforcement of those front end models that *have* down.snaps to reference them in the DOWN state of the ISC_LOCK guardian, using shell scripts like
/opt/rtcds/userapps/release/isc/h1/guardian/
All_SDF_down.sh
All_SDF_observe.sh

These are rapid ways to change them all, once created, but because the SDF system identifies models by DCUID number instead of model name, it's arduous to create and maintain such a script.

(3) -- We still don't have this. Jamie and Jenne (and several others, I'm sure) have already discussed that parsing guardian code to find what channels it touches is intractable. 
As such, the new idea is that we have some script that 
     (a) takes two burt snapshots of an entire front-end model EPICs settings database
     (b) compares them -- any channel that has a change between those two snap files gets flagged for not-monitoring, all others get flagged for monitoring.
     (c) Those flags are absorbed into a *single* SDF file (with none on this unmaintainable several-file nonsense), and then
     (d) all monitored settings values are accepted.
We imagine this script to be run once a week, or so.

(4) -- We can now initialize a value by just flipping over to the "CHANS NOT INIT" menu, and accepting them. However, by default, they go into the list of channels *not* monitored. In order to initialize them as monitored, one has to remember to select *MON* (all) and then confirm. Until we get (3) addressed, the functionality should be that hitting *accept* for a non-initialized channel should automatically monitor it. For CHANS NOT FOUND We still must do the multi-screen, confusing save and reload.

This entry is posted as a request to the CDS software team to help us out. We don't think we can accomplish the solutions to (3) or (4), and we're mildly unhappy with the work-arounds we've done for (1) and (2). Thanks ahead of time.
H1 ISC (SUS)
jeffrey.kissel@LIGO.ORG - posted 16:25, Monday 17 October 2016 (30605)
SUS Alignement Offsets during This Afternoon's Lock Stretch, in prep for tomorrow's RCG Upgrade Recovery
The title says it all!
Images attached to this report
H1 OpsInfo
terra.hardwick@LIGO.ORG - posted 16:18, Monday 17 October 2016 (30602)
'Bouncing' PI modes

Operators: If PI modes bounce during your shift - such as Mode26 and Mode18 - tune phase based on overall slope of the top of the bounces. The bouncing is not real (it's a filter issue I'm working on), but the overall height of the bounces is, so damp that as usual. This should only happen when two modes close in frequency are rung up at the same time, so you might need to be changing both phases during this time. Note that this will likely occur within the first half hour of powering up from a cold(er) state. 

- - - - - 

Since editing filters to help with the crossing modes - Mode18 and Mode26 - we've seen PI signals bounce on the PI StripTool several times now. I think this occurs during crossing when both modes are rung up; in this case, there are two peaks of similar amplitude within a few tenths of a Hz apart, the amplitudes are enough to engage the PLL, and the PLL(s) get confused about which peak is 'theirs' and end up jumping back and forth between the two. At the same time, bandpass filters are being turned on/off by the guardian, based on the frequency readback of the PLL's; the bounciness is caused by alternatively right and wrong BP filters being turned on, so the amplitude of peak appears to change dramatically as its within or outside of the bandpass. 

Attached are spectra during bouncy time when both peaks are elevated and during normal damping time (ten minutes later) when only one peak is high. Also attached are trends of relevant channels during this time. Bounciness seen in first bump in RMSMON channels, no bounciness ten minutes later in second bump. The FREQ_MON channels show the first bump correlating with strong switchbacks in frequency (especially in Mode18), versus single frequency tracking ten minutes later.

Still working on a good solution to handle this. For now, Mode's 18 and 26 have only been staying so close and rung up for periods of ~10 min and only during colder lock starts (after that Mode18 rises in frequency away from Mode26); we only saw this once over the full weekend of locking. And if we damp based on overall slope of bouncing, it's been damping well. 

Images attached to this report
LHO General
thomas.shaffer@LIGO.ORG - posted 16:01, Monday 17 October 2016 (30588)
Ops Day Shift Summary

TITLE: 10/17 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Lock Aquisition
INCOMING OPERATOR: Nutsinee
SHIFT SUMMARY: Lost lock a few times just after getting to NLN or just after powerup, not really sure why (once was me trying to adjust the ISS diffracted power too late in the game). Been locked for 4hours, commissioners hard at work.
LOG:

H1 CAL (DetChar)
jeffrey.kissel@LIGO.ORG - posted 15:47, Monday 17 October 2016 - last comment - 16:17, Monday 17 October 2016(30597)
PCALX Roaming Calibration Line Frequency Changed to 2001.3 Hz
J. Kissel for S. Karki

I've changed the > 1 kHz PCALX calibration line frequency, a.k.a. the "long duration sweep" line, from 2501.3 to 2001.3 Hz. Recall this started moving again on Oct 6th (see LHO aLOG 30269) I report the progress towards completing the sweep below. 

Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00      
1501.3       35k                      02:00     
2001.3       35k                      02:00                   39322.0           Oct 17 2016 21:22:03 UTC
2501.3       35k                      05:00                   39322.0           Oct 12 2016 03:20:41 UTC    Oct 17 2016 21:22:03 UTC      days
3001.3       35k                      05:00                   39322.0           Oct 06 2016 18:39:26 UTC    Oct 12 2016 03:20:41 UTC      days
3501.3       35k                      05:00                   39322.0           Jul 06 2016 18:56:13 UTC    Oct 06 2016 18:39:26 UTC      months
4001.3       40k                      10:00
4301.3       40k                      10:00       
4501.3       40k                      10:00
4801.3       40k                      10:00  
5001.3       40k                      10:00
Images attached to this report
Comments related to this report
evan.hall@LIGO.ORG - 15:22, Monday 17 October 2016 (30601)

I had changed the pcal y line heights yesterday around 10 am local for a contrast defect test and have only now reverted them back to their old values.

jeffrey.kissel@LIGO.ORG - 16:17, Monday 17 October 2016 (30604)DetChar
Saying more words for Evan:

He changed the 7.9 Hz and 1083.7 Hz line heights, by adjusting the H1:CAL-PCALY_PCALOSC4_OSC_[SIN,COS]GAIN and H1:CAL-PCALY_PCALOSC3_OSC_[SIN,COS]GAIN (respectively) oscillator gains. They we changed starting at Oct 16 2016 17:27 UTC and restored by  Oct 17 2016 22:23 UTC. So any undisturbed time processed from this period should be excised from the collection of data for the 2501.3 Hz analysis for fear of confusion on the optical gain.

Thus the new table of times for valid analysis is


Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00      
1501.3       35k                      02:00     
2001.3       35k                      02:00                   39322.0           Oct 17 2016 22:24:00 UTC
2501.3       35k                      05:00                   39322.0           Oct 12 2016 03:20:41 UTC    Oct 16 2016 17:27:00 UTC      days
3001.3       35k                      05:00                   39322.0           Oct 06 2016 18:39:26 UTC    Oct 12 2016 03:20:41 UTC      days
3501.3       35k                      05:00                   39322.0           Jul 06 2016 18:56:13 UTC    Oct 06 2016 18:39:26 UTC      months
4001.3       40k                      10:00
4301.3       40k                      10:00       
4501.3       40k                      10:00
4801.3       40k                      10:00  
5001.3       40k                      10:00
Images attached to this comment
LHO VE
kyle.ryan@LIGO.ORG - posted 14:38, Monday 17 October 2016 (30599)
Manually over-filled CP3
LN2 @ exhaust in 60 seconds after opening LLCV bypass valve 1/2 turn -> closed LLCV bypass valve. 

Next over-fill to be Wednesday, Oct. 19th.  
LHO VE (PEM)
thomas.shaffer@LIGO.ORG - posted 14:26, Monday 17 October 2016 - last comment - 15:51, Monday 17 October 2016(30598)
EX Pump NLN Test

WP#6221

While locked in Low Noise, Kyle headed to EX to start and stop a pump in two 5 min intervals. The pump is connect to the North side of BSC5.

Times in UTC (GPS):

Start1 - 20:52:52 (1160772789)

Stop1 - 20:57:52 (1160773089)

Start2 - 21:02:52 (1160773389)

Stop2 - 21:07:52 (1160773689)

 

The purpose of this test is to determine if these pumps can be run without interferring with commissioners.

Comments related to this report
kyle.ryan@LIGO.ORG - 15:51, Monday 17 October 2016 (30603)
"Going once...going twice....SOLD!"  

Having heard no complaints, I will forge ahead and plan on running these pumps for 5-7 days (starting tomorrow morning) in support of the required bake-out of the X-end RGA.
LHO FMCS
john.worden@LIGO.ORG - posted 14:01, Monday 17 October 2016 (30596)
Mid station chillers

The Mid station chillers were turned off today and will likely remain off until March or so when the warmer weather returns.

H1 ISC (CAL, ISC)
evan.hall@LIGO.ORG - posted 14:05, Sunday 16 October 2016 - last comment - 17:53, Monday 17 October 2016(30573)
DARM contrast defect in full lock

I wanted to find out how much contrast defect light we have in DARM at the moment. It seems to be about 2.0(5) mA at the moment. Since we run with 20 mA of total photocurrent, this implies a homodyne angle that is mistuned by about 6° away from the nominal value of 90°. I did not check how stable it is over the course of several hours.

To measure the contrast defect, I watched the height of 332 Hz pcal line in DARM while varying the dc offset.

Non-image files attached to this report
Comments related to this report
evan.hall@LIGO.ORG - 19:55, Sunday 16 October 2016 (30575)

Also, I found that the DARM residual is microseism-dominated at 50 W of input power (the current blrms is about 0.5 µm/s). So I turned on a boost in FM6 of LSC-OMC_DC. We should incorporate this into the DARM filter modules.

jeffrey.kissel@LIGO.ORG - 17:53, Monday 17 October 2016 (30600)CAL
Expanding more on Evan's methods here:

Optical gain values in [mA/pm] were obtained by taking the magnitude of the transfer function at 331.9 [Hz] between H1:CAL-PCALY_RX_PD_OUT_DQ (pre-calibrated into [m] / zpk([],[1,1],1)) and H1:OMC-DCPD_SUM_OUT_DQ (pre-calibrated into [mA] to ~10% accuracy).

Total light on the OMC DCPD values in [mA] were pulled directly from H1:OMC-DCPD_SUM_OUT_DQ (again, pre-calibrated into [mA]).

The DARM DC offset was varied by adjusting the "fringe offset" or H1:OMC-READOUT_X0_OFFSET (pre-calibrated into [pm] to ~20-30% accuracy). This EPICs record can be found on the "IFO DC READOUT" sub-screen (called OMC_DC_READOUT.adl) of the OMC_CONTROL.adl overview screen. The nominal value is 10.65623 [pm], and to obtain the above data Evan varied the DARM DC offset from 6 to 13 in 1 [pm] increments.

     zeta = homodyne angle [rad] = arccot( contrast defect [mA] / total light on DCPDs [mA] )

where the contrast defect [mA] is the y intercept of the parabola.

The subsequent IFO optical gain vs. DC power on the DCPDs was then fit (by-eye) to blind/simple quadratic function with a DC offset. to arrive at the answer.

From Evan's presentation G1601599, which nicely distills the famous-yet-cryptic 2001 Buonnano&Chen paper, the response of a DRFPMI interoferometer using detuned resonant sideband extraction can be parametrized into a pair of complex poles (for the optical spring, at frequency |p| and quality factor Q), a pair of real poles (for the coupled cavity, or "RSE" pole, at frequency xi) and zero, at frequency z, which can potentially (and typically does for low detuning) cancel one of the RSE poles: 
     dP                    1 + i f / z 
     -- = g * ---------------------------------------           (5)
     dL       (1 + if/|p|Q + (f/|p|)^2) - (xi / f)^2
The zero, in his presentation, is composed of the following fundamental parameters,
                 cos(phi + zeta) - r_s cos(phi - zeta)
      z = f_a * ------------------------------------------      (8)
                 cos(phi + zeta) + r_s cos(phi - zeta)
where f_a is the arm cavity pole frequency (assumed to be the same for both arms), phi is the SRC detuning phase, and zeta is the homedyne angle.

One of the outputs of the above measurement, is that, if the homodyne angle, zeta, is consistently 90 +/- 6 [deg], then we can used Eq. (8) to simply fix the zero frequency in the overall IFO response (5), assuming the arm cavity pole frequency and SRC detuning phase also remain constant. This would reduce the parameter space over which the calibration group would have to MCMC in fits to measurements of the overall response (e.g. LHO aLOG 28302).

However, 
(1) This is, again, *one* measurement of the homodyne angle, zeta. We're going to have to measure it multiple times, and quantify the uncertainty in the estimate better, to make sure that we're confident it stays there.
(2) The SRC detuning phase, phi, and the arm cavity pole frequency, f_a, also need measuring with quantifiable uncertainty. These are also parameters believed to be fixed, but the question is always to what level. f_a has been measured before to be ~83 Hz, using several techniques (e.g. LHO aLOG 7054), but rarely with quantified uncertainty. Further, those measurements are typically taken of a single cavity, and there is worry that the pole frequency may change a bit in the full IFO due to different spot centering*. The detuning phase "can be determined by the spring frequency."
To me, this is quickly going down a rabbit hole of another independent MCMC parameter estimation fitting regime, but I'm still quite ignorant on the topic.

*word on the street is that LLO has a technique, once in full lock, of "kicking the SRM fast enough" such that full IFO remains locked but simple as an PRFPMI. I couldn't find an aLOG on it, and these discussions with Evan today were the first I'd heard of it. 

Worth exploring!
Images attached to this comment
gabriele.vajente@LIGO.ORG - 16:50, Monday 17 October 2016 (30606)

Doing a real least square fit gives different results, depending on what you assume

  1. If we require the minimum of the curve to be at z=0 (meaning that we have a good calibration of the optical gain zero), give a contrast defect of 1.6 mA. Sum of residuals is 1.3 mA^2
  2. If we allow the most generic second order polynomial, the fit is better (sum of residuals is 0.68 mA^2), but the minimum is at optical gain equal to 0.8 mA/pm and the contract defect is larger, 4.3 mA
Images attached to this comment
Displaying reports 53481-53500 of 83258.Go to page Start 2671 2672 2673 2674 2675 2676 2677 2678 2679 End