Displaying reports 55361-55380 of 84655.Go to page Start 2765 2766 2767 2768 2769 2770 2771 2772 2773 End
Reports until 17:04, Friday 30 September 2016
H1 ISC
daniel.sigg@LIGO.ORG - posted 17:04, Friday 30 September 2016 - last comment - 11:26, Tuesday 04 October 2016(30119)
Jitter into the IMC

This is a plot of the jitter measured by the IMC WFS DC PIT/YAW sensors during last nights lock. The 280 Hz periscope peak reaches about 1x10-4/√Hz in relative pointing noise, or about 3x10-4 rms. The relative pointing noise out of the HPO is about 2x10-5/√Hz at 300 Hz. After the attenuation through the PMC this would correspond to a level below 10-6/√Hz. The jitter peaks show up in DARM, if they are high enough. This is clearly visible in the coherence spectra.

The ISS second loop control signal is an indication of the intensity noise after the mode cleaner with only the first loop on. The flat noise level above 200 Hz is around 3x10-6/√Hz in RIN, with peaks around 240 Hz, 430 Hz, 580 Hz and 700 Hz. Comparing this to the free-running noise in alog 29778 shows this RIN level at 10^-5/√Hz. We can also compare this with the DBB measurements, such as in alog 29754: the intensity noise after the HPO shows a 1/f behaviour and no peaks. Looking at the numbers it explains the noise below 300 Hz. It looks like a flat noise at the 10^-5 level including the above peaks gets added to the free-running intensity noise after the PMC. The peaks in the controls signal of the second loop ISS line up with peaks visible in the pointing noise. But, neither the numbers nor the spectral shape matches. These peaks have coherence with DARM.

Non-image files attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 10:30, Monday 03 October 2016 (30169)

Checking the calibration of the WFS DC readouts I noticed a calibration error of a factor of 0.065. So, all angles measured by the WFSs should be scaled by this number. This still makes the jitter after the PMC dominant, but one might expect to see some of the HPO jitter peak show through in places where the downstream jitter has a valley. In any case, we should repeat the PSL jitter measurement with the IMC unlocked.

Non-image files attached to this comment
daniel.sigg@LIGO.ORG - 11:26, Tuesday 04 October 2016 (30206)

A report of the measured beam jitter at LLO is available in T1300368.

An earlier measurement at LHO is reported in alog 21212. Using an IMC divergence angle of 1.6x10–4 rad, the periscope peak at 280 Hz is around 10^-4/√Hz. This is closer to the first posted spectrum with the "wrong" calibration. Here I post this spectrum again and add the dbb measurement of the jitter out of the HPO propagated through the PMC (1.6%), but scaled by a fudge factor of 2. The Sep 11, 2015, spectrum shows a more or less flat noise level below 80 Hz, whereas the recent spectrum shows 1/f noise. The HPO spectrum also shows as 1/f dependency and is within a factor of 2 of the first posted spectrum. If jitter into the IMC is the main coupling mechanism into DARM, the HPO jitter peaks above 400 Hz are well below the PSL table jitter after the PMC and the would not show up in the DARM spectrum.

Non-image files attached to this comment
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 16:55, Friday 30 September 2016 - last comment - 11:15, Monday 03 October 2016(30115)
Spectral shape of the noise we are tyring to hunt

I have done a cross correlation analysis of the two OMC DCPDs in order to study the spectral shape of the broad band noise that has been limiting the sensitivity in 200-1000 Hz band. Here are some results.

The red curve is cross-correlated noise of the two DCPDs in terms of the DCPD current in mA (with DARM loop effect removed).The data I used is a lock stretch from Sep. 26th which lasted for 6 hours in undisturbed. The frequency resolution or bandwidth is set to 0.1 Hz. The data length is 3 hours starting at 9:00:00 UTC of the day. The DARM suppression is taken out from OMC DCPD SUM. I have used the latest PreER10 DARM model (29931) to obtain a reasonable DARM suppression function. As we all know, there is high noise in 200-1000 Hz whose level is as high as shot noise (or null stream shown in green) in this frequency band.

I also made another plot, which is an overlay of 11 cross correlated spectra each of which is from different time but from the same lock stretch. See below.

This time, the frequency resolution is set to 1 Hz, so that we can get several FFTs out of the same lock stretch. Each spectrum is produced by an 18 minutes integration. As shown in the plot, there was one spectrum which is polluted by some kind of glitch. Otherwise, the shape of noise stayed almost unchanged over 3 hours. Also fig files are available.

Images attached to this report
Non-image files attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 11:15, Monday 03 October 2016 (30176)

There was a question of whether the cross-spectrum noise was from readout electronics or not. I have checked it by looking at dark noise from this past weekend in which laser happened to be off.

I conclude that the excess noise we see at around 400 Hz is not from dark noise or similar readout noise. In fact the dark noise level above 100 Hz was found to be consistent with a measurement a year ago (20526). Things look good to me. The attached is a plot of the dark noise spectra.

DCPD settings:

  • First stage of the whitenings is engaged.
  • High-Z (400 Ohms)
Images attached to this comment
H1 PSL
edmond.merilh@LIGO.ORG - posted 16:46, Friday 30 September 2016 - last comment - 17:37, Friday 30 September 2016(30118)
PSL tripped off

@ ≈23:45 UTC

Comments related to this report
edmond.merilh@LIGO.ORG - 17:37, Friday 30 September 2016 (30120)

The Diode Chiller was in error - Flow Sensor 1. With Jason's phone guidance I was able to restart the LASER with no issues. The thought is that this is a pre-cursor to a control panel failure on the Diode chiller. I was informed by Jason that there is one on it's way. It should be here some time late next week.

H1 General
edmond.merilh@LIGO.ORG - posted 16:10, Friday 30 September 2016 (30114)
Shift Summary - Eve Transition
 
 
TITLE: 09/30 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Commissioning
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    Wind: 7mph Gusts, 3mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.11 μm/s 
QUICK SUMMARY:
LHO General (OpsInfo)
corey.gray@LIGO.ORG - posted 16:07, Friday 30 September 2016 (30113)
Shift Summary

Summary:  Morning was spent recovering from Power Glitch.  Afternoon H1 was handed over for Commissioning.  Here are some notable activities from shift:

Notes:

Locking Notes:

H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 15:08, Friday 30 September 2016 (30112)
CDS model and DAQ restart report, Thursday 29th September 2016

model restarts logged for Thu 29/Sep/2016
2016_09_29 03:35 h1nds0
2016_09_29 14:10 h1nds0
2016_09_29 14:27 h1nds0
2016_09_29 14:32 h1nds0

installed more memory in h1nds0 to make it the same as h1nds1, no further unexpected restarts of h1nds0 for the rest of the day (and none on Friday at time of writing).

H1 ISC (DetChar, ISC)
gabriele.vajente@LIGO.ORG - posted 14:04, Friday 30 September 2016 (30110)
Jitter coupling stable over time

BruCo scans (see here for the most recent one) showed a large coherence with IMC alignment signals, in the region between 100 and 1000 Hz. The most significant coherence is with IMC-WFS_B_DC_PIT.

I computed the transfer function between this IMC beam spot position signal and the OMC DCPD SUM signal, for a few different time periods in the past couple of days. The transfer functions are basically non changing over time in the different locks. This suggests that, if the increased coupling is due to a IFO misalignment, this is quite repeatable and constant over time (fig. 2)

Not much of a smoking gun, but some hints.

Images attached to this report
LHO VE
chandra.romel@LIGO.ORG - posted 13:55, Friday 30 September 2016 - last comment - 16:45, Friday 30 September 2016(30108)
CP4 experiment
Overfilled CP4 again:

1. Opened exhaust bypass valve
2. Doubled LLCV (66% open)
3. Took ~17 min. to fill reservoir from 92% to 100%.
4. Took another hour to see LN2 pour out the exhaust. 
5. TC 252B did not respond to temp. change (and I broke the other one yesterday)
6. Set LLCV to nominal at 32% open in manual mode after the overfill
7. Would like to see how long it takes CP to recover to 92% at nominal LLCV (but may not happen this time since it's the weekend)

I used the borescope to look at the TC junction on the exhaust to see if it's measuring the pipe or the fluid flow. I couldn't see much but I suspect it's measuring just the pipe since the time delay on CP3 is long and temps never drop below -12degC on CP3 (response is faster when heating pipe with heat gun). 

I would like to drill and tap TCs directly into the pipe and use the existing signal lines to measure a more direct temp. for CP3 automation.

Images attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 13:59, Friday 30 September 2016 (30109)
Attached is temp and pressure trend for a CP3 overfill
Images attached to this comment
chandra.romel@LIGO.ORG - 16:45, Friday 30 September 2016 (30117)
CP4 is now in PI mode with lower limit set to 32%. Need to set this back to 20% on Monday. Want to see how long it takes to fall to 92% full at this nominal value.
H1 IOO (IOO)
cheryl.vorvick@LIGO.ORG - posted 13:43, Friday 30 September 2016 (30107)
ISS Second Loop QPD orientation of pitch and yaw signals now corrected

The orientation of the ISS second loop qpd was rotated 90 degrees and reversed in sign when compared to the orientation of IM4 Trans QPD.

I changed the QPD matrix and tested the change by exercising IM1 in pitch and yaw.

I've updated SDF.

Details attached.

Images attached to this report
H1 IOO (IOO)
cheryl.vorvick@LIGO.ORG - posted 13:23, Friday 30 September 2016 (30106)
IM1-4 alignment slider gains corrected -

IM alignment slider gains were reversed, and set the urad / slider count to 0.045.

I've corrected the alignment slider gains for all IMs, and used measured values for each optic such that the urad / slider count is 0.05.

I've attached a snapshot of the new gains, and a snapshot of SDF before I acepted the new gains.

With the gain change, the slider values also change, so any burt restore prior to 30 September 2016 at 20:01UTC (13:01PT) will fail to return the IMs to their correct alignment slider values.

Images attached to this report
H1 CAL (CAL)
sudarshan.karki@LIGO.ORG - posted 12:49, Friday 30 September 2016 (30105)
Pcal beam optimization at Yend

RickS, TravisS, SudarshanK

Using the same procedure used at END-X (LHO alog 29873), we optimized the Pcal beam locations at END-Y today. 

In each of these figure, the optimal locations, 111.6 mm from the center of the face of the optic in both positive and negative y direction, are denoted by red cross and the actual pcal beam locations by cyan circle.

Attached figure (1) shows the beam before they were optimized.  The offsets ([xoffset, yoffset]) from the optimal locations for upper and lower beam were: U[1.36, 7.58] and L[3.18, -8.60] in mm and uses standard co-ordinate sign convention.

Attached figure (2) shows the beam after they were steered to their optimal positions using the Pcal transmitter module steering mirrors.  The new offsets from the optimal locations are U [-0.96, -0.67] and L[-0.59, 0.39].

 

mmoffsetx =
 
    0.9558
    0.5913
 
 
mmoffsety =
 
   -0.6705
    0.3884
mmoffsetx =
 
    0.9558
    0.5913
 
 
mmoffsety =
 
   -0.670
Images attached to this report
H1 SEI
hugh.radkins@LIGO.ORG - posted 12:39, Friday 30 September 2016 (30104)
LHO BRS Checks/Status after Power Glitch

Everything looks to have restarted on its own or taken care of by the likes of Patrick (EndX) but I haven't been able to pin him down on that.  Actually, just managed to break into those guys--the computers did not stop so there was no restart.  Manual intervention to restart at EndX would be required but this glitch did not drop the BRSs out.

The Ion Pump at EndY has been reprogrammed to restart on its own and that was verified to still be running by EE this morning as well as the drift trends.  The attached shows these trends and the medms of the BRSs.  Again, things look to be running fine.

Images attached to this report
H1 SEI (CDS, SEI)
krishna.venkateswara@LIGO.ORG - posted 12:32, Friday 30 September 2016 (30103)
All BRS' are recovered

Krishna

BRS-X, BRS-Y and c-BRS seem to be functioning normally at the moment. I don't know if they were restarted manually or on their own, but ligo-dv web shows normal looking drift channels on BRS-X and BRS-Y and CTRL channel on c-BRS (trends attached).

Images attached to this report
X1 DTS
james.batch@LIGO.ORG - posted 11:44, Friday 30 September 2016 (30100)
Restarted DTS after power glitch
Jonathan, Jim

The front end computers on the DAQ test stand have been restarted following this morning's power glitch.
H1 ISC (CDS, SYS)
jeffrey.kissel@LIGO.ORG - posted 11:44, Friday 30 September 2016 - last comment - 14:13, Friday 30 September 2016(30099)
AS Port Fast Shutter Functionality Checked
J. Kissel, K. Kawabe, T. Shaffer, C. Gray

After a message popped up on the ISC_LOCK guardian to review the last check of the AS port fast shutter, and we realize the last check should have been during the power glitch recovery, we manually checked the functionality of the fast shutter. The fast shutter appears functional, stopping light into the OMC breadboard and the GS13s show an impulse.

The manual check: 
Setup:
- The IMC is locked, and the IFO aligned enough to see flashes on the AS port camera
- Watching the OMC Breadboard QPDs MEDM screen to see a spot moving around (channels H1:ASC-OMC_A_[YAW,PIT]_OUTMON)
- Watching the HAM6 GS13s on a dataviewer session to watch for fast physical motion of the shutter (channels H1:ISI-HAM6_BLND_[X,Y,Z]_GS13_CUR_IN1_DQ)

The test:
- Use the fast shutter detail MEDM screen (sitemap > LSC > shutters > bottom middle "Detail" under the "HAM6 Fast Shutter" name)
- Hit "open" then "close" several times, watch for light to disappear from the OMC Breadboad QPDs, and watch for large impulse on GS13s.

We need to do a similar test after every power outage, no matter how small.
TJ has agreed to create a guardian "PRECHECK" state such that we can do manual check of the fast shutter functionality. (To date we've only done the automatic checks every 48 hours).
Note: There already exists a guardian check in the FAST_SHUTTER node, however, the check is programmed to fail without the IFO MUCH further into the lock acquisition sequence (i.e. DRMI locked), and we'd prefer to check the functionality far before this.
Comments related to this report
richard.mccarthy@LIGO.ORG - 14:13, Friday 30 September 2016 (30111)
Not sure of the significance of a power outage.  We often reboot computers turn on and off power supplies that control this device, do we have to do functional tests everytime?
H1 TCS
betsy.weaver@LIGO.ORG - posted 10:20, Friday 30 September 2016 - last comment - 12:00, Friday 30 September 2016(30092)
TCS chiller and lasers restarted

This morning around 8:30am, Nutsinee and I went out and restarted the TCS chillers and the TCS lasers.  We also took the opportunity to swap the greenish filters.  Recall, we had flushed both chillers last week, but when we installed a new set of filters we noticed that they were tinging green immediately even though we had just flushed the system.  These filters sat for ~a week and we swapped them today in hopes that whatever immediately turned them green last week was just still pushing throgh the system.  We also had to add another 750mL (to top off the half way full indicator to full) and again inspected for leaking water (none found).

We'll need to look up how to clean these recyclable filters.

Comments related to this report
alastair.heptonstall@LIGO.ORG - 11:20, Friday 30 September 2016 (30097)

Is this on top of the 500ml added yesterday?   https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=30074

vernon.sandberg@LIGO.ORG - 11:45, Friday 30 September 2016 (30101)

Added scales to the sides of the fill tubes. This will allow better monitoring of the chiller water levels. The TCS-X scale runs from 21 - 30, linearly, and the TCS-Y scale runs 0-10, linearly.

Images attached to this comment
betsy.weaver@LIGO.ORG - 12:00, Friday 30 September 2016 (30102)

I have just opened and inspected the TCSY table - All water lines look good and no puddles found.  The tubes also appear to be pretty solid with not many air bubbles.

Displaying reports 55361-55380 of 84655.Go to page Start 2765 2766 2767 2768 2769 2770 2771 2772 2773 End