Chandra, Ken (with help from John & Richard) Removed the pneumatic actuators from LLCVs on CP 2,3,5,6 and replaced with the electronic version. Upgraded fuses to 2.5 A. Will finish up WP 5906 with CP3 tomorrow morning. CP 2,4,5,6 are in PID mode.
Kiwamu, Nutsinee, Jim, Dave
At noon we power cycled the h1oaf0 IO Chassis to hard reset the 16bit DAC card. We were suprised that both TCS chillers stopped running. During last week's IOP dac-disable events (which zeroed the DAC drives), the chillers kept on running for about an hour before tripping. As a further test, at 14:00 PDT we again powered down h1oaf0 CPU followed by its IO Chassis. At this point the chillers were operational. We then power cycled the AI chassis, which tripped the chillers. We then powered up the IO Chassis and computer, at which point the chillers tripped again. We were unsure if the trip was at the time the IOP model started, or h1tcscs model started driving the DAC channels.
Take home message: before restarting h1oaf0 IOP, CPU, IO Chassis or DAC-AI unit please contact Nutsinee.
Nutsinee, Kiwamu,
Today, we made a final touch on the alignment of the CO2Y table optics. As a result, we got a 5.7 W output power coming out of the table which is twice bigger than the value we measured back in April (26645). The beam profile now looks good -- almost axis symmetrical about the center of the beam area.
The next step: the alignment of the CO2 beam with respect to the interferometer beam.
[Details]
At the beginning, as a coarse alignment between the CO2 beam and the aperture mask, we moved the position of the mask by approximately a few mm in order to improve the horizontal beam profile. We then touched M5 in both pitch and yaw to further refine the alignment. Later we repositioned the beam dump which catches the reflection from the mask. This time, we used the FLIR camera as a reference which is much more sensitive than an image card with the UV light. The attached images are the ones after the fine adjustment.
Once we optimized the intensity profile, we re-aligned the beam through the two irises that have been serving as the fiducial points for projecting the beam to the test mass. After the alignment, we placed a power meter right behind the first iris which read 5.7 W when the rotation stage was at 18 deg (which should give us almost the maximum transmission). Before closing the table, we put the beam dump back in place to block the beam to the FLIR camera.
TITLE: 05/31 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: Jim
SHIFT SUMMARY: After PSL was brought back to life and maintenance activties subsided, we are having issues getting past DRMI locked.
LOG:
14:56 Ken to MY working on HVAC
15:06 Cristina and Karen to LVEA
15:35 Fil and Ed pulling cables in beer garden
15:43 Hugh to ends HEPI maint.
15:50 turned of BRS at both ends for maint. work
15:50 Gerardo and Manny to all VEAs, pulling extension cords in prep for power outage
16:12 Kiwamu working on TCS alignment
16:34 Hugh done
16:48 Nutsinee to TCS table
17:00 Karen to EY
17:05 Fil and Ed out
17:54 Fil working on ESD in beer garden
18:00 Gerardo and Manny back from ends, to LVEA
18:02 Jason to diode room
18:05 PSL back up
19:00 Kiwamu done
19:51 Fil to beer garden
20:21 Bubba to LVEA checking 3IFO N2 purge
21:07 Fil out
21:25 Kiwamu and Nutsinee to TCS table
22:27 turned BRS back on at both ends
22:39 Kiwamu and Nutsinee out
Required use of N. crane to stage equipment (~0900 - 0930 hrs. local). Temporary aux. pumps removed and HAM4 AIP system in nominal, as found, configuration.
In the morning we were wondering what the SRM composite mass looks like, in case we want to drive it to find out the frequency and Q of the first internal mode.
https://dcc.ligo.org/LIGO-D1200886
There is a ring shaped holder which has two nubs and a set screw. This holds the 2" optic. This holder is then bolted into the larger aluminum mass. Two views from the solidworks drawings are shows with some parts made transparent. If this drawing is right, I expect that the resonant frequency is in the 500-1000 Hz range.
Here is a picture of the actual SRM surrogte optic in it's inner holder (the LLO one shown). The 2" x 0.75" thick optic is held in the ring by what I believe are 2 black carbon peek set screws which are not shown in the assembly drawings.
Saturday 28th
08:47 PSL Diode Chiller monitor signals flat-line (compuer had been running for about 24 hours since last reboot)
10:24 TCS DAC output drops to zero volts (recovered soon after)
17:04 h1nds1 kernel panics (computer rebooted)
23:46 TCS DAC output drops to zero volts (recovered Monday)
Sunday 29th
Monday 30th
11:00 PSL laser power drops to zero (recovered Tuesday)
TCS Y-table FLIR camera
Kiwamu, Dave:
the TCS-Y infrared camera was not responding. Restarting the controller unit at the table appears to have resolved it.
h1nds1 restart
h1nds1 stopped running. Monit was not monitoring it, we manually instructed monit to restart daqd and it is running again.
h1oaf0 power cycle
Following three occurences of the TCS DAC being driven to zero volts, this morning we power cycled the IO Chassis for h1oaf0. Power was removed from the chassis using the front panel switch, it was down for about one minute. h1tcscs was manually burt restored to 09:10 PDT this morning.
roof camera
Richard, Jim, Dave:
We went onto the roof and verified the CDS AUX vlan was operational at the fiber media converter, it was.
J. Oberling, J. Bartlett, P. King (via phone)
Following yesterday's initial investigation into the PSl diode chiller issues (see alogs here), we swapped the control panel for the diode chiller with the one from the chiller we recently removed from service (this is a known working unit, just installed last year). After installation the chiller restarted without issue and ran for several minutes, also without issue. The serial number of the control panel we installed in the diode chiller is 44806P605; the faulty unit we removed from the diode chiller had no SN on it.
We then took the time to restart the PSL Beckhoff PC to unstick the frozen diode chiller channels. According to Dave Barker these channels froze sometime on Saturday morning. Fortunately the PSL interlocks are not tied to these channels, otherwise we would not have the PSL shutting down when the diode chiller shuts down (as we did yesterday). Once restarted the channels appeared to be reading OK. It is possible that these channels freezing for no reason can be used as an early warning sign of imminent chiller failure. The chiller communicates with the PSL Beckhoff PC via a serial RS-232 interface and we have seen channels freeze when the cables are unplugged and plugged back in but the PC not restarted (which is expected behavior as RS-232 is not hot swappable), but this is the first time I've seen the chiller spontaneously loose communication with the PSL Beckhoff PC. Jeff Bartlett is setting up a temporary Strip Tool on the PSL monitor in the control room that will monitor these channels. If anyone sees that these are flatlined (there should always be some variation), please let someone on the PSL team (Peter King, Jeff Bartlett, Ed Merilh, Rick Savage, or myself) know immediately. Thank you.
We then turned on the HPO, which came up with no problems. We let it sit and run for a little while, then restarted the rest of the PSL. As of now, everything is up an running. We are going to continue to monitor over the next couple days to ensure that everything is working correctly. The removed control panel will go back to Termotek with the chiller we are sending back and will be replaced as part of the service being done on that chiller.
Completely forgot to mention that after performing the above front panel swap and restarting the laser, we adjusted the calibration of the vortex flow sensors in both chillers. Using the chiller we just recently removed from service, we hooked it up to an external flow meter and compared the 2 readings (1 from the chiller's internal flow meter and 1 from our external flow meter) and calculated a new pulses/liter calibration for the vortex flow sensors. According to that measurement the vortex flow sensors should be set to 494 pulses/liter (they were originally set to ~970 pulses/liter, a number we got from LZH). The flow information from both chillers is now accurate.
1420 - 1440 hrs. local -> To and from Y-mid Opened exhaust check valve bypass, opened LLCV bypass 1/2 turn ccw -> LN2 at exhaust in 40 seconds -> Closed LLCV bypass, waited a few minutes and closed exhaust check valve bypass. Next overfill nominally Wednesday afternoon.
Sheila, Terra, Craig
The PSL is off, is has been since about 17:30 UTC. The chillers don't seem to be tripped. Confusingly the laser screen indicates that the chillers are fine, (two green boxes) while the PSL_STATUS screen has a red box for the crystal chiller (screenshot attached). Jason will come out to the site to investigate/restart the laser.
We also noted that the temperature trends for the PSL have been usual since thursday morning's incursion. (2nd screnshot) I went to the controller box and saw that the north AS unit was on, which was probably unintentional (the south unit was off, and they are normally both off in science mode). I turned it off at noon local time. Terra noted that the PSL microphone has seen an elevated level of noise in the last few days, which went back to normal as soon as the AC unit was off. (In the third attached screenshot, blue traces are from the time when the AC was on). The montors on the RF AM stabilization also changed when we turned off the AC, and some channels on the AM stabilization box seem to have been sensitive to some kind of switching of the PSL HVAC over the last few days.
It seems like we need some kind of a better way to monitor if the PSL environment settings are correct, maybe adding them to diag main if we can find a good set of tests to write. It also is suprising to me that our RF system seems to be so sensitive to accoustic pick up in the PSL. Has anyone in detchar looked at PSL PEM monitors to see if glitches there are correlated to the "RF45" glitches seen durring O1?
The TCS chillers are also tripped, with the same DAC problem we have been having. (27435) This happened about 36 hours ago.
Bottom chiller screen; flashing between 'temperature' and 'warning'
The laser was up and running this morning when I checked it around 6 am (local). The gibberish message on the diode chiller controller I've never seen before and is most likely a controller malfunction. To fix the problem, I would try (in order): - power cycling the chiller with the power switch located at the rear of the chiller - replacing the chiller controller (if Jeff Bartlett happens to have a spare handy) - install the spare chiller (which will take a bit of work because ... ) * the turbine flow sensors need to be replaced with the vortex ones * the 3-phase power plug needs to be installed * some filters need to be removed * the coolant lines will need to have any air pockets removed The problem with the first solution is that it is hard to gauge how long the "fix" might be valid for before the laser could trip out again.
We used Sheila's very instructive alog to kill and restart all the models on the OAF machine, reset the TCS chillers and restart the TCS laser.
J. Oberling, S. Dwyer
Attempted to bring the PSL back up but were ultimately unsuccessful. Came in and found the crystal chiller running and the diode chiller off, although the Laser MEDM screen indicated the diode chiller was up and running. EPICS channels frozen again?
The diode chiller turned on without an issue, although the weirdness on the main screen, seen in Terra's photos above, did not go away. We let the chiller run for several minutes and then attempted to power on the HPO. Approximately midway during the pump diode power up everything stopped and we found the diode chiller shut off. To see if it was a coincidence we reset the interlocks and attempted to turn the HPO on again, this time monitoring the chillers. The HPO got to its second stability range and the diode chiller immediately shut off. We power cycled the diode chiller (which by the way cleared the funky front panel issue seen in the photos above). This time the HPO acheived the second stability range for 10 whole seconds before the diode chiller shut off again; it almost seems as if the chiller is shutting off as soon as it sees a heat load. During all this the crystal chiller remained up and running without issue.
At this time I'm out of ideas, although the chiller behavior copuled with the front screen weirdness makes me think we may have a control panel problem with the diode chiller (as Peter mentions above); I seem to recall that when we had the chiller flow sensor issues last year (April/May 2015) we also had some weird issues with the chiller (the one we just recently removed from service) that were solved by replacing the control panels. I left the PSL off; the diode chiller is also off and the crystal chiller is running. Please do not attempt to turn it on, we will investigate more fully tomorrow morning.
Filed FRS #5605.
(see attached) Will investigate Tues.
AIP = "annulus ion pump"