Tried to do some FSS related measurements to see why it was the FSS kept losing lock.
Unfortunately it seems the FSS fieldbox signals of interest - which is just about all
of them as far as transfer function measurements are concerned - are not of any use
in their current state.
In doing a measurement this morning, the injection locking was broken which in
turn tripped the high power oscillator's power watch dog. The laser restarted without
any issue.
The missing equipment hunt concluded with limited success. There are still a few items missing. Fred, Vern, and Richard thanked all for the efforts. PSL – Looking into an instability with the FSS. CDS – Working on frame writer problem VAC – Will need to bake out the End-Y RGA before O2. All other subsystems report no problem. There will be no maintenance window on Thursday. There will be a safety meeting this afternoon at 15:00.
The ITM-X ISI Stage 1 and Stage 2 WDs tripped at 12:35:25 and 12:35:30 (UTC) respectively. No problem resetting. Was not highlighted on Ops Overview MEDM.
painful - we need to make the relocking robust
To do next: TCS tuning (diff?) see whether the new noise depends on it.
Updated the guardian script to reach low noise in both 25W and 40W mode.
Attached is a DCPD cross-power spectrum at 40W (references) and at 25W (live data). Red is the PD_A-PD_B cross power. Blue is the DCPD_SUM power spectrum cast in the same units to give a reference of where shot noise lies.
Note: the excess noise we currently see is also present at 25W. It is almost identical to the 40W noise up to 200Hz. However, the notch structure between 200Hz and 700Hz is different at different power levels - no idea why.
Plot 2 shows the MICH coherence with DARM at 25W - the MICH-correction is not fine-tuned for that power level, but it is not the dominant contribution.
Guardian changes:
ISS_ON: based on IMC-PWR_IN_OUT16, set 25W or 40W offset.
# crude way to make it work at 25W and 40W (no intermediate!)
if ezca['IMC-PWR_IN_OUT16'] > 30:
ezca['PSL-ISS_SECONDLOOP_REF_SIGNAL_ANA']= -0.9326934814453125 # 40W
else:
ezca['PSL-ISS_SECONDLOOP_REF_SIGNAL_ANA']= -0.588 # 25W
LOWNOISE_ASC: The state used to only run >35W. Changed it to always run, except for the 40W compensation filters on CHARD_Y and DHARD_Y (FM6).
Couple of things that would be good to check:
Executive summary: * Good news - as expected, the 16-Hz comb due to the OMC length dither is gone (at least at this sensitivity level) * Bad news - low-frequency 1-Hz combs remain, and some new low-frequency combs & lines have appeared Some details:
I analyzed the 56.8406Hz comb with coherence tool and here are the results. The same structure is found to be significant in 35 channels in ER9, distributed in ISI, SUS, PEM and LSC subsystems. Among all the 35 channels, 22 of them does not have a range up to its 11th harmonic, 625.25 Hz.
Keith indicated in his slog entry that a DAQ malfunction is suspected to be the ultimate source of this, and these findings suggest it's in an EX electronics crate.
Here are a few interesting observations:
The 9th harmonic at 511.56Hz is the weakest in most channels, sometimes buried in noises.
In some PEM channels, there are missing lines at low frequency (< 200 Hz) and high frequency (> 500 Hz).
In PEM and ISI channels, there seems to be another comb structure with a frequency slightly larger than 56.8406Hz coexists. That one is usually most significant at its third harmonics.
Generally, the structure is more clearly seen in LSC, SUS and ISI channels
Sample plots from each subsystem:
Figure 1: We can see the 56.8406Hz comb structure exists with its 9th harmonic weakest in ISI.
Figure 2: PEM channels have more noises and, as in ISI channels, the other comb structure coexists.
Figure 3: SUS channels do not have enough range up its 11th harmonic but we can see its first and second harmonic here.
Figure 4: There is only one channel from LSC but the structure is very clear.
All plots and a list of channels are attached in the zip file.
Just to be clear. Here are the channels that the coherence tool is finding the comb. This is what is supporting Keith's assumption that the problems could be in an EX electronics crate. Channels List: H1:ISI-ETMX_ST2_BLND_RX_GS13_CUR_IN1_DQ_data H1:ISI-ETMX_ST2_BLND_RY_GS13_CUR_IN1_DQ_data H1:ISI-ETMX_ST2_BLND_RZ_GS13_CUR_IN1_DQ_data H1:ISI-ETMX_ST2_BLND_X_GS13_CUR_IN1_DQ_data H1:ISI-ETMX_ST2_BLND_Y_GS13_CUR_IN1_DQ_data H1:ISI-ETMX_ST2_BLND_Z_GS13_CUR_IN1_DQ_data H1:LSC-X_TR_A_LF_OUT_DQ_data H1:PEM-EX_ACC_BSC9_ETMX_Y_DQ_data H1:PEM-EX_ACC_BSC9_ETMX_Z_DQ_data H1:PEM-EX_ACC_ISCTEX_TRANS_X_DQ_data H1:PEM-EX_ACC_VEA_FLOOR_Z_DQ_data H1:PEM-EX_MIC_VEA_MINUSX_DQ_data H1:PEM-EX_MIC_VEA_PLUSX_DQ_data H1:ISI-ETMX_ST1_BLND_Y_T240_CUR_IN1_DQ_data H1:ISI-ETMX_ST1_BLND_Z_T240_CUR_IN1_DQ_data H1:ISI-GND_STS_ETMX_X_DQ_data H1:ISI-GND_STS_ETMX_Y_DQ_data H1:PEM-EX_MAINSMON_EBAY_1_DQ_data H1:PEM-EX_MAINSMON_EBAY_2_DQ_data H1:PEM-EX_MAINSMON_EBAY_3_DQ_data H1:PEM-EX_SEIS_VEA_FLOOR_X_DQ_data H1:PEM-EX_SEIS_VEA_FLOOR_Y_DQ_data H1:SUS-ETMX_L1_WIT_Y_DQ_data H1:SUS-ETMX_L2_WIT_L_DQ_data H1:SUS-ETMX_L2_WIT_P_DQ_data H1:SUS-ETMX_L2_WIT_Y_DQ_data H1:SUS-ETMX_M0_DAMP_L_IN1_DQ_data H1:SUS-ETMX_M0_DAMP_P_IN1_DQ_data H1:SUS-ETMX_M0_DAMP_T_IN1_DQ_data H1:SUS-ETMX_M0_DAMP_V_IN1_DQ_data H1:SUS-ETMX_M0_DAMP_Y_IN1_DQ_data
I chased Comb 23 (type K) in Keith’s post, shown in Keith's original post as
This comb has an offset of 153.3545 Hz and a fundamental frequency of 0.0884Hz. It starts at 153.3545 Hz and goes up to its 11th harmonic, 154.3272 Hz. As is listed in Keith's txt file:
Comb 23 (type K, offset=153.354500): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) K 153.3545 ( 0 X 0.0884) 1.844961e-19 **** K 153.4429 ( 1 X 0.0884) 1.949756e-19 **** K 153.5314 ( 2 X 0.0884) 2.165192e-19 ***** K 153.6198 ( 3 X 0.0884) 2.181833e-19 ***** K 153.7082 ( 4 X 0.0884) 2.457840e-19 ***** K 153.7966 ( 5 X 0.0884) 2.243089e-19 ***** K 153.8851 ( 6 X 0.0884) 2.709562e-19 ***** K 153.9735 ( 7 X 0.0884) 2.499596e-19 ***** K 154.0619 ( 8 X 0.0884) 2.562208e-19 ***** K 154.1503 ( 9 X 0.0884) 1.945817e-19 **** K 154.2388 ( 10 X 0.0884) 1.951777e-19 **** K 154.3272 ( 11 X 0.0884) 1.703353e-19 ****
I found the comb structure in two channels of ISI subsystem.
Figure 1 shows the plot of channel H1:ISI-HAM6_BLND_GS13RZ_IN1_DQ. Descriptions of this channel can be found here:
https://cis.ligo.org/channel/314371
Figure 2 shows the plot of channel H1:ISI-HAM6_BLND_GS13Z_IN1_DQ. Descriptions of this channel can be found here:
https://cis.ligo.org/channel/314374
In the plots of both channels, we can see a comb structure stands out at the positions of harmonics. We are wondering about the reason for this:
Why these seismic isolation channels?
This post is supplementary to the first post about coherence analysis result for the 56.8406Hz Comb at
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=28619
The first post is addressing the 56.8406Hz comb found in Keith's original post (marked as D comb):
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=28364
Information about this comb from the txt file in Keith's post:
Comb 35 (type D, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) D 56.8406 ( 1 X 56.8406) 3.968800e-17 *********** D 113.6811 ( 2 X 56.8406) 1.773964e-17 ********** D 170.5217 ( 3 X 56.8406) 7.121580e-18 ********* D 227.3622 ( 4 X 56.8406) 3.232935e-18 ******** D 284.2028 ( 5 X 56.8406) 1.166094e-18 ******* D 341.0433 ( 6 X 56.8406) 1.007273e-18 ******* D 397.8839 ( 7 X 56.8406) 5.962059e-19 ****** D 454.7245 ( 8 X 56.8406) 3.752194e-19 ***** D 511.5650 ( 9 X 56.8406) 2.577108e-19 ***** D 568.4056 ( 10 X 56.8406) 1.964393e-19 **** D 625.2461 ( 11 X 56.8406) 1.891774e-19 **** --------------------------------------------------------------
Besides the 35 channels found in the original post, 7 more channels are found to be relevant to the 56.8406Hz Comb. Two new subsystems, ASC and HPI are involved.
These new channels are:
H1:ASC-X_TR_A_NSUM_OUT_DQ
H1:ASC-X_TR_B_NSUM_OUT_DQ
H1:HPI-ETMX_BLND_L4C_Y_IN1_DQ
H1:HPI-ETMX_BLND_L4C_Z_IN1_DQ
H1:PEM-EX_ACC_BSC9_ETMX_X_DQ
H1:SUS-ETMX_L1_WIT_L_DQ
H1:SUS-ETMX_L1_WIT_P_DQ
So updated channel list is (42 channels in total):
H1:ASC-X_TR_A_NSUM_OUT_DQ
H1:ASC-X_TR_B_NSUM_OUT_DQ
H1:HPI-ETMX_BLND_L4C_Y_IN1_DQ
H1:HPI-ETMX_BLND_L4C_Z_IN1_DQ
H1:ISI-ETMX_ST1_BLND_RX_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST1_BLND_RY_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST1_BLND_RZ_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST1_BLND_X_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST1_BLND_Y_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST1_BLND_Z_T240_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_RX_GS13_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_RY_GS13_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_RZ_GS13_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_X_GS13_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_Y_GS13_CUR_IN1_DQ
H1:ISI-ETMX_ST2_BLND_Z_GS13_CUR_IN1_DQ
H1:ISI-GND_STS_ETMX_X_DQ
H1:ISI-GND_STS_ETMX_Y_DQ
H1:LSC-X_TR_A_LF_OUT_DQ
H1:PEM-EX_ACC_BSC9_ETMX_X_DQ
H1:PEM-EX_ACC_BSC9_ETMX_Y_DQ
H1:PEM-EX_ACC_BSC9_ETMX_Z_DQ
H1:PEM-EX_ACC_ISCTEX_TRANS_X_DQ
H1:PEM-EX_ACC_VEA_FLOOR_Z_DQ
H1:PEM-EX_MAINSMON_EBAY_1_DQ
H1:PEM-EX_MAINSMON_EBAY_2_DQ
H1:PEM-EX_MAINSMON_EBAY_3_DQ
H1:PEM-EX_MIC_VEA_MINUSX_DQ
H1:PEM-EX_MIC_VEA_PLUSX_DQ
H1:PEM-EX_SEIS_VEA_FLOOR_X_DQ
H1:PEM-EX_SEIS_VEA_FLOOR_Y_DQ
H1:SUS-ETMX_L1_WIT_L_DQ
H1:SUS-ETMX_L1_WIT_P_DQ
H1:SUS-ETMX_L1_WIT_Y_DQ
H1:SUS-ETMX_L2_WIT_L_DQ
H1:SUS-ETMX_L2_WIT_P_DQ
H1:SUS-ETMX_L2_WIT_Y_DQ
H1:SUS-ETMX_M0_DAMP_L_IN1_DQ
H1:SUS-ETMX_M0_DAMP_P_IN1_DQ
H1:SUS-ETMX_M0_DAMP_T_IN1_DQ
H1:SUS-ETMX_M0_DAMP_V_IN1_DQ
H1:SUS-ETMX_M0_DAMP_Y_IN1_DQ
Attached images are sample plots from ASC and HPI subsystem.
Full results are also attached.
Here are the coherence search results of all the single lines in ER9 data, which are listed in Keith’s post. I found 29 of all the 198 lines on the list and posted the results on my homepage here:
https://ldas-jobs.ligo-wa.caltech.edu/~duo.tao/ER9_single_lines/index.html
Here is a DCPD x-power spectrum (including a self-x-power for SUM, calibrated the same way). Note that the x-power has structure around few hunder Hz, including a humpo at 800Hz.
Jenne, Sheila, Stefan
We used two independent methods to verify that the beam spot positions do not move during the power increase:
Jenne and Sheila put dither lines on all optics. While they have not calibrated the output, the result is that during the power increase they did not observe any motion of the beam, but during the soft offset adjustment just afterwards they did see the spots move.
I used Kiwamu's cameras to make time lapse movies of the power-up (the first two gifs). No spot motion is observed. I also atteched two movies (gif 3 and 4) of the motion during the soft offset adjustment. The spots can be seen wiggling.
FInally, I attached a Striptool chart of the beam positions during power-up. The power increase starts at -30 minutes, and finishes at -22minutes. Then the soft loops start moving until about -16 minutes. At -12 minutes Jenne took the soft offsets out again.
How to make gif movies
Attached are two scripts:
snapper.py grabs a snap shot from each of the 4 test mass cameras every 10 seconds.
makemovie converts those images into a gif movie.
(prefix and run number are hard-coded.)
Nutsinee, Kiwamu,
WP5990
We have (re-) set up the polarization monitors on the HWS table by HAM4. We have confirmed that they are functional. For those who are interested in the polarization data, here are the channels to look at:
In theory, they should be in unit of watts as measured at the HWS table.
[Installation notes]
This time, we have newly installed a short pass optic (DMSP950L from Thorlabs) to pick off the main interferometer beam without getting too much contamination from either the SLED light (790 nm) or the ALS beam (532 nm). The short pass mirror was inserted between the bottom periscope mirror and the first iris (D1400252-v1). Looking at the green light at the table from the end stations, we learned that the beam size is already pretty small and (visually) small enough for the beams to fit into the PDA50Bs without a lens. So we decided to go without lenses as opposed to the previous setup (24046).
The short pass mirror reflects the interferometer beam toward the left on D1400252. We placed a PBS (CM1-PBS25-1064-HP) on the left side of the short pass and placed the PDA50Bs. The power reflectivity of the newly installedshort pass mirror was measured to be 5% +/-3% for 532 nm. The absolute power (assuming the Nova hand held power meter is accurate) of the reflected green light was measured to be 1 uW.
One thing we leaned today was that the green light is not so trustable to get the optimum alignment. We first aligned the optics with the green light and then noticed that the infrared beams were almost falling off of the PDA80Bs. So we then closed the shutters and aligned them with the actual infrared beam.
The manual gain settings are:
The digital gains were also changed accordingly so that the calibration of these channels should be accurate.
This is a first look at the polarization data with the new setup. Some analysis with the previous setting was reported by Aidan at 25442 back in this February with a focus on noise behaviors. This time, since we are looking for a cause of the degradation in the power recycling gain, we focused on the time series rather than the spectra.
We saw two behavior in the polarization data when PSL was ~ 40 W.
Based on the fact that the amount of S-pol decreases as a function of time (which should increase the power recycling gain at the same time, naively speaking), I am inclining to say that the variation in the polarization is not a cause for the smaller power recycling gain.
[An observation from last night, July 13th]
I have used a lock stretch from last evening starting at ~ 2016-07-13 1:00 UTC for 2-ish hours. The attached two plots show the measured polarization in time series.
At the beginning of the lock stretch, the input power was increased step by step up to 40-ish W. The power recycling gain hit 35 right after completing the power-up operation and then settled to a lower value of 29 or so. The power in P-pol was about a factor of 8 larger than that for the S-pol. Note that this is opposite to what Livingston observed (G1501374-v1) where the S-pol was bigger than the P-pol. Back-propagating the measured power to those at BS's AR surface (the ones propagating from ITMX to BS), we estimated the power ratio to be Pp/Ps ~ 2500. This separation ratio is better than what has been measured at Livingston (G1501374-v1) by a factor of roughly 13.
[Another observation from Jan 31st for comparison]
I also looked at a similar data set from Jan 31st of this year (25442) to see if the polarization in the past behaved in the same way or not. This data was with a 20 W PSL without the HPO activated. The behavior looked similar to what we have observed last night -- a slow decay in the S-pol and P-pol was larger than the S-pol by a factor of 6-ish. See the attached below.
Matt later pointed out that there is a possibility that my measurement set up could be unintentionally rotated with respect to interferometer's polarization plane. In this case, depending on the rotation angle, the S-pol can appear to decrease even though the actual S-pol in the interferometer increases. I did a back of envelop calculation and confirmed that the measurement setup needs a rotation of about 20 deg to get such confusion [ angle = atan(sqrt(1/8) )]. I don't think we have such a big rotation in our setup. So it seems that the S-pol really decreases at the beginning of the lock stretch.
Here are some photos of our set up.
J. Kissel, S. Dwyer, S. Ballmer We continue to have trouble with the FSS oscillating after a lock loss, in that it'll often either take several minutes to relax, or it requires manual intervention such as briefly reducing the common gain of the FSS loop. As such, Sheila took a look at the IMC PDH loop to look for problems and instabilities there. I looked over her shoulder at her results, and saw some areas for improvement in the loop design. The current loop design has an UGF at 66 [kHz], with a phase margin of 68 [deg]. However the gain margin around ~200 [kHz] is pretty dismal because of what looks to be some icky features in the physical plant. These features have been shown to be directly influenced by the FSS common gain (see second attachment in LHO aLOG 28183). I figure, given that we've got oodles of phase margin, what harm could be done by just adding a simple 200 [kHz] pole in loop, and reducing the gain by ~2 [dB]? As such I took Sheila's data, which lives here /opt/rtcds/userapps/release/isc/common/scripts/netgpib/netgpibdata/TFAG4395A_12-07-2016_163422.txt (also attached) and added these modifications offline as a design study. In the attached plots, I compare the as-measured IMC PDH Open Loop Gain, G, Loop Suppression, (1/1+G), and the Closed Loop Gain, (G/1+G), against one modified as described above (blue is as measured, and green is the modified design study). The results are encouraging: a still-substantial UGF of 47 [kHz], and a very-healthy phase margin of 58 [deg]. However, as can bee seen in the loop suppression and the closed loop gain, there is far less gain-peaking and/or a much great gain margin and we would no longer have to worry about the icky features in the plant that are so sensitive to the FSS common gain. Where to stick such an analog filter? It's of course dubious to claim that the MEDM screen for such a system is representative of the analog electronics, but assuming it is, one can see that there is the possibility of a switchable daughter board in the FAST path that gets shipped off to the PSL AOM for the FSS. Because it's switchable, we can toss whatever simple filter in there that we like, and then compare and contrast the performance for ~1 week to see if it improves the stability problems we've been having. What impact would this have on the full IFO's CARM loop? I'll remind you of Evan's loop analysis of the whole frequency stabilization spaghetti monster in LHO aLOG 22188. There he suggests that the CARM UGF is around 17 kHz, so as long as the Closed Loop Gain around there is the same, then this change in the IMC PDH loop should have little impact [[I just made this sentence up based on just a few words from Sheila who asked me to look at the CLG. I'm not confident of its truth. Experts should chime in here]]. Indeed the third .pdf attachment shows that G/(1+G) of the IMC PDH loop, regardless of modification remains unity out to 100 [kHz].
How does this compare with the Pomona box from anno domini?
Description of the notch in pamona box 5141 (this was in the loop for a few years, but was removed serveral months ago, I think before O1)
Thanks for finding the aLOG entry Shiela! @Daniel -- though she doesn't say it explicitly, the aLOG shows that the Pomona Box notch was centered about ~700 kHz. As shown by my OLGTF model, if we add this ~200 kHz pole, then not only will any features at 200 kHz be suppress significantly, but whatever might happen at 700 kHz is even further suppressed. In otherwords this pole just shapes the high-frequency, super-UGF portion of the OLG to better handle *any* non-sense, instead of the focused bandaid fixes that any notch would provide.
J. Kissel In the rush and panic of ER9 last week, we neglected to aLOG that we reverted the ETM ESD bias flip that I'd done last Tuesday because -- for some unknown reason -- the sign flip *this* time caused problems for ALS DIFF and subsequently switching to EY later in the acquisition sequence. As such, we didn't gain as much in this week's charge mitigation, as shown by the plots attached below, which includes a new measurement from today. The effective bias voltage is still within the ~10 [V] range, so the reversion did little harm. Since ER9, we haven't had the time to explore why the flipping failed us. I'll continue to try to push for figuring this out, but we'll see what kind of priority it gets with respect to all other commissioning tasks.
The mic in the EY electronics bay had been reported by Vinny Roma as having unusually low signal, requiring a calibration factor 10 times larger than the other microphones. I determined that the fault was with the signal conditioning box - when plugged in to a different channel, the microphone performed normally. Additionally, I verified that the power supply was delivering the proper voltage to the signal conditioning box. Using the second channel in the signal conditioning box also did not improve the signal. We should plan to replace the signal conditioning box next week.
asc filters
The h1asc filters which have been pending install since before ER9 were loaded.
framewriter testing
Jim, Dave
we continued testing of h1fw0 to investigate its instability. The daqd code was rebuilt against a later version of framecpp (2.4.2 instead of 1.19.32). When fw0 was instructed to write all frame types, it was as unstable with the older framecpp. The size of the science frames between the two writers at this point was different, with fw0's frame 2 bytes smaller than fw1's frame. We suspect the framecpp version string is encoded in the header, "2.4.2" is two chars shorter than "1.19.32". To get the frames the same size, we reverted fw0 back to framecpp 1.19.32.
The next test was to install a local 1TB 7200rpm SATA hard disk drive on h1fw0 and build an ext4 file system on it. The daqd process was directed to write to the local disk instead of the NFS mounted SATABOY. The disk access immediately sped up by many seconds. When we got the writer to write all four types of files, very surprisingly the frame writer still crashed, but took longer to do so. Writing all frames to the NFS-QFS crashed in less than 5 minutes, with local ext4 it crashing in 9 to 10 minutes. The crash time is not related to writing second trends, we saw several successful trend file writes over the several tests we ran. Monitoring the threads showed that with local disk the thread did not go into the "D" diskIO wait state. Another surprise was with local disk writing we did see a retransmission request about 40 seconds prior to the slew of requests at the time of crash.
We reconfigured fw0 backj to writing to ldas-h1-frames so they could be used to back-fill fw1's gaps.
As a test we had fw0 write commissioning frames only rather than science frames only. The data loading goes up from 0.9GB to 1.7GB per 64 second time period. fw0 is stable in this configuration. It can handle either 64 second frame, just not both.
We are leaving fw0 writing science frames overnight.
Y sled has been replaced recently and now that X sled is getting very dim, I decided to replace X sled as well. This way it's easier to keep track of when the sleds got replaced (this time both were replaced within one week). The old sled number is 12.05.21 replaced with 07.14.256.
Kiwamu, Stefan
Yesterday Kiwamu realigned the red IR cameras. I reset the centroid setting today. The configuration files are:
ITMX:
[Camera Settings]
Camera Name = H1 ITMX (h1cam21)
maxX = 659
maxY = 494
Exposure = 100000
Analog Gain = 1023
Auto Exposure Minimum = 150
Name Overlay = True
Time Overlay = True
Calculation Overlay = True
Do Calculations = True
Calculation Mask = Circle
Circle Mask X = 373
Circle Mask Y = 150
Circle Mask Radius = 250
Calculation Subtraction File = None
Auto Exposure = False
Calculation Noise Floor = 25
Snapshot Directory Path = /ligo/data/camera
Frame Type = Mono12
Number of Snapshots = 1
Archive Image Minute Interval = 0
Archive Image Directory = /ligo/data/camera/archive/
[No Reload Camera Settings]
Base Channel Name = H1:VID-CAM21
Camera IP = 10.106.0.41
Multicast Group = 239.192.106.41
Multicast Port = 5004
Height = 480
Width = 640
X = 0
Y = 0
ITMY:
[Camera Settings]
Camera Name = H1 ITMY (h1cam23)
maxX = 659
maxY = 494
Exposure = 100000
Analog Gain = 1023
Auto Exposure Minimum = 150
Name Overlay = True
Time Overlay = True
Calculation Overlay = True
Do Calculations = True
Calculation Mask = Circle
Circle Mask X = 295
Circle Mask Y = 220
Circle Mask Radius = 250
Calculation Subtraction File = None
Auto Exposure = False
Calculation Noise Floor = 25
Snapshot Directory Path = /ligo/data/camera
Frame Type = Mono12
Number of Snapshots = 1
Archive Image Minute Interval = 0
Archive Image Directory = /ligo/data/camera/archive/
[No Reload Camera Settings]
Base Channel Name = H1:VID-CAM23
Camera IP = 10.106.0.43
Multicast Group = 239.192.106.43
Multicast Port = 5004
Height = 480
Width = 640
X = 0
Y = 0