Displaying reports 60041-60060 of 87874.Go to page Start 2999 3000 3001 3002 3003 3004 3005 3006 3007 End
Reports until 09:31, Monday 25 July 2016
H1 CDS
james.batch@LIGO.ORG - posted 09:31, Monday 25 July 2016 (28613)
Reboot h1digivideo1, power cycled h1cam18
The IOC for h1cam18 was not running properly, an old image was being displayed and centroid values were frozen.  Tried restarting the IOC, it didn't clear the problem.  Tried rebooting h1digivideo1, it came back OK, but the IOC for h1cam18 wouldn't run.  Tried deleting all of camera 18's log files and restarting, still wouldn't run.  Logged on to network switch to power cycle the h1cam18, then restarted the IOC and it appears to be behaving OK now.
H1 PSL
edmond.merilh@LIGO.ORG - posted 09:29, Monday 25 July 2016 (28614)
PSL Weekly Trends - FAMIS# 6106

Multiple LASER trips are evident in the trends. Also, apparently, H1:PSL-OSC_DB3_CUR is reporting 100A. This is inconsistent and virtually impossible. Something wrong with the channel. FRS# 5955 filed.

Images attached to this report
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 05:55, Monday 25 July 2016 (28612)
laser tripped
Came in to find that the laser had tripped.  Error messages were as before.  Found that although I could
reset the laser I could not turn it on without rebooting the Beckhoff computer.  All the TwinCAT channels
were live and so on.  When the command was issued to turn on the pump diodes, the pump diode current(s)
did not increase.

    The crystal chiller displays "Error Flow - sensor 1", as before, even though the laser is up and
running.

    One other observation is that the mouse on the Beckhoff computer appears to be dead.  Whilst its LED
is on, moving the mouse does not move the cursor.  Not sure that the two problems are related but it
certainly doesn't help.
H1 ISC (ISC)
carl.blair@LIGO.ORG - posted 02:01, Monday 25 July 2016 - last comment - 15:44, Friday 19 August 2016(28611)
PI dark region

I have been taking spectra from the SR785 (WP6005) whenever I get a chance over the last week to see if there is any evidence of three mode interactions in the 60kHz to 70kHz region that we will not be sensitive to with the aliased OMC DCPD HF channels that we normally analyze.
There is a consistent peak at 62935Hz, this peak is present with no optical power in the arm cavities.
There are several other more transient peaks, one of the times several had large amplitudes is shown in the first figure.
The largest peak is at 63776Hz  The maximum amplitude seen was 5E-6  This is about the same as the 18040Hz mode when it is 2 orders of magnitude above thermal noise and 2 orders of magnitude under unlocking the cavity.
The second largest is at 70160Hz and the third largest at 62336.

There is no evidence of peaks in DARM at this time at the expected aliased frequencies 1760Hz,  3200Hz or 4624Hz and the peaks that appear in the HF channels that do not appear in the normal DCPD channels do not coincide in frequency, see the second image.

Images attached to this report
Comments related to this report
terra.hardwick@LIGO.ORG - 21:50, Monday 25 July 2016 (28631)

We disconnected this SR785 around 11am local time today. This closed work permit #6005.

jeffrey.kissel@LIGO.ORG - 15:44, Friday 19 August 2016 (29204)
J. Kissel, T. Hardwick

We've taken the liberty of rifling through Carl's home directory in hopes to find the raw data from this entry to re-plot for clarity. We found it!

The newly attached plot now highlights the PI modes that Carl mentions in his aLOG, and also shows the anticipated ADC noise. Thus, any modes below 6.3e-6 [V/rtHz] should not be resolved by the ADC, and therefore will show up aliased into digitized signals in the detection band. 

Terra notes that the mode Carl mentions at 70160 Hz is the largest of several peaks at 69.84, 69.95, and 70.03 kHz (not highlighted), which is likely a mode cluster.

Other details:
The raw data lives here (determined by Terra knowing that Carl keep his GPIB data in his home folder, then lining up the data on the figure with the filename): 
/ligo/home/carl.blair/gpib/netgpibdata/dataSPSR785_24-07-2016_212424.txt

This data (as described in the referenced work permit 6005), is the raw analog output of the TMSY's red QPD's whitening chassis.

This data also happens to cover the frequency region surrounding the 65536 [Hz] native sampling frequency of the General Standards ADC, and the corresponding notch in all Anit-Aliasing (AA) chassis. One can see, delightfully, that there is very little noise or lines in this frequency band on this channel that might also otherwise be aliased down to low frequency. 

We should perform a similar spectral analysis of the OMC DCPD whitening chassis output voltage to check if their AA chassis is also sufficiently notched so as to not contribute noise in to the DARM sensitivity.
Images attached to this comment
Non-image files attached to this comment
H1 ISC (ISC)
carl.blair@LIGO.ORG - posted 23:29, Sunday 24 July 2016 - last comment - 04:42, Tuesday 26 July 2016(28609)
Phase locked loop tracking PI modes

[Matt, Carl]

The phase lock loop is now installed as an optional tool in the PI armory. I have used the settings and method Matt demonstrated to me on the test stand to lock onto OMC PI signals in the last 50W lock. I tried locking onto QPD signals but was unable to at quiescent amplitudes. The phase locked loop in the locked state is shown in the first image. It is tracking a 15521Hz ITMX mode (sorry for the poor labeling in the figure, there's still a few bugs in medm screens).

The settings used were:
Filter I - 100Hz LP Gain 1
Filter Q - 100Hz LP Gain 1
Freq Filter 1 - gain 1
Freq Filter 2 – gain 0.02, 20mHz integrator + low pass see below
FC Count - 10mHz low pass (this was not low enough as the frequency estimate was still fluctuating by 20Hz
Ampl Filt er - 1Hz low pass gain 1
Lock Filter - 1Hz low pass gain 1

Lock was acquired by setting a set frequency close to the mode frequency observed in a spectrum of the OMC HF channels. The loop was engaged with a 100Hz low-pass in Freq Filter 2 then put in a narrow band mode by engaging a 1Hz low-pass in Freq Filter 2, the loop lost lock when I tried to further reduce its bandwidth by either reducing the gain of decreasing the frequency of the low-pass. The figure is in the high bandwidth mode.  The PLL and iwave can be accessed from the mode block for any mode and a matrix is used to select a control signal, see the second image.

Images attached to this report
Comments related to this report
matthew.evans@LIGO.ORG - 01:12, Monday 25 July 2016 (28610)

I think the best way to transition from PLL "aquisition mode" (high-gain and wide-band) is:

  1. start with gain of 1 in FREQ_FILT1 and 2 (FF1 and FF2), no filtering in FF1, and LP10 in FF1
  2. lower the gain in FF1 to ~0.02
  3. engage a low-pass filter in FF1 (e.g., LP0.1)

This should keep the output of FF2 (which contains an integrator) fairly constant, and thus keep the PLL locked during the transition.

matthew.evans@LIGO.ORG - 04:42, Tuesday 26 July 2016 (28634)

Well, step 1 was supposed to read:

start with gain of 1 in FREQ_FILT1 and FREQ_FILT2 (FF1 and FF2), no filtering in FF1, and Int20mHz + LP10 in FF2

H1 ISC (ISC)
jenne.driggers@LIGO.ORG - posted 17:42, Sunday 24 July 2016 (28608)
ISS 3rd loop measured at 50W

We had a lockloss at 50W due to what looked like the ISS 3rd loop, so I re-measured with higher recycling gain.  It seems that when we recover recycling gain by moving the PRM, we get some loop interactions that reduce the gain of the 3rd loop right where we need it.  So far, just increasing the gain of the loop has been enough to keep things under control.

Attached is a set of measurements of the 3rd loop, with the state of the IFO noted in the magnitude legend on the upper left.  "g" refers to the gain H1:PSL-ISS_TR_GAIN. 

Once, while the 3rd loop gain was -1.5, it started to oscillate.  I put the gain to -2, then down to -1.  I think it liked -1 better at that time, and that's how I left it for the rest of that lock, but -2 could also have been okay perhaps.  I'm not totally sure.  Perhaps we need to give this thing a UGF servo?

Images attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 14:42, Sunday 24 July 2016 - last comment - 16:27, Sunday 24 July 2016(28604)
1430 hrs. local -> Kyle woking in Vac Prep lab
Will make log entry when leaving.  

Also, BSC8 annulus ion pump railed again, same as back in May.  See also, https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=27325 and https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=27385 

Will likely replace pump during Tuesday Maintenance period
Comments related to this report
kyle.ryan@LIGO.ORG - 16:27, Sunday 24 July 2016 (28607)
Started pumps in room 169 @ 1615 hrs. local -> will be running until further notice 

1700 hrs. local -> Leaving site now
H1 ISC (ISC)
jenne.driggers@LIGO.ORG - posted 14:08, Sunday 24 July 2016 - last comment - 15:41, Sunday 24 July 2016(28603)
Locked at 50W on ETMY

[Jenne, Carl, Terra]

To facilitate the damping of the 18kHz ETMX mode, we are currently locked at 50W using ETMY so that ETMX can be swapped to its low noise driver. SRM dither loop is open.

To do this, it's okay to (once IncreasePower is finished) go to Manual mode and select LowNoise_ESD_ETMY. Last week we lost lock trying to do this because this state also (used to) switch the L2 coil drivers to their low noise state, which we aren't ready for.  So, the noise is acceptable for the lownoise ESDs, but not the low noise L2 drivers.  On the to-do list is to decide where in the sequence to put the L2 switching.

PI and other work is ongoing.

Comments related to this report
peter.fritschel@LIGO.ORG - 15:05, Sunday 24 July 2016 (28605)

For the L2 coil drivers, if I understand correctly you are set up to transition to the lowest noise (and lowest range) state, state 3. As an alternative you might transition to state 1, where the ACQ mode is switched OFF, but the low-pass remains OFF. I believe the low-pass is a 1 Hz pole, 10 Hz zero, so this would afford more range above 10 Hz than state 3, and still puts the L2 noise low enough that it's not a significant limit, at least for our O2 goal. See Chris W's entry 28264.

jenne.driggers@LIGO.ORG - 15:41, Sunday 24 July 2016 (28606)

Ah, yes.  I think you had mentioned this, but I forgot.  I'll give that a try.

H1 AOS
terra.hardwick@LIGO.ORG - posted 20:58, Saturday 23 July 2016 - last comment - 21:25, Saturday 23 July 2016(28600)
PI at 50, 60 W today

Kiwamu, Carl, Terra

18041 Hz (ETMY) successfully damped through several locks. 15542 Hz (ETMX) unstable after about 1 hour at 50 W but we cannot damp without going to LOWNOISE_ESD_ETMY.

We had a few ~1 hour locks at 50 W today, with a bit of time at 60 W. 

18041: damped first with +30 phase, +30K gain but rang up in the second lock despite. Changing to zero phase damped it in the second lock and prevented any ring up during the third lock.

15542: rings up almost immediately at 60 W. Rings up after an hourish at 50 W. This mode belongs to ETMX but we've been sitting at INCREASE_POWER during locks so we cannot use the LNLV ETMX driver to damp. 

Comments related to this report
chris.whittle@LIGO.ORG - 21:01, Saturday 23 July 2016 (28601)
Because ETMX PI couldn't be damped, I tested my ITM DARM charge measurement script, which broke lock. I know the cause and will be able to fix it for another attempt tomorrow.
carl.blair@LIGO.ORG - 21:25, Saturday 23 July 2016 (28602)

The SUS_PI guardian is running again.  It had got stuck with spm differences after the model restarts.  
During these lock stretches settings for damping several modes were found and put in the gaurdian.  
15522Hz ITMX (mode2) was ringing up at the end of the last lock aand was damped with gain -300 and phase -60 deg.
15541Hz ETMY (mode25) was stable but the driven up to find a damping gain 1000 and phase +60deg and the 
18041Hz ETMY phase has been adjusted to 0 deg (gain stays 1000)


 

H1 ISC
kiwamu.izumi@LIGO.ORG - posted 19:12, Saturday 23 July 2016 (28599)
SRM dither alignment 50 W

This is a quick note on the SRM dither loop.

The SRM dither loop does not seem to give us the optimum SRC alignment when the PSL power is above 50 W in this evening. I don't remember seeing this problem two days ago (28577). Today at one point, we had a lockloss at 50 W when slowly introducing an offset in the PRM pointing from 0 to 0.5 QPD count over 5 min. During this engagement process, AS90 kept decreasing while the carrier recycling gain kept increasing. In the next lock stretch, we opened up the SRM dither loop and started manually aligning SRM. At the optiumum point where the signal recycling gain (i.e., AS90 / POP90) was maximized, the dither-based error signals had a non-negligible offset in both pitch and yaw (on the order of 0.005 at the input of the control filters for both).

H1 PSL (PSL)
peter.king@LIGO.ORG - posted 16:18, Saturday 23 July 2016 (28598)
Laser trip
Kiwamu called, told me the laser was off.

    In what is becoming an all too often occurrence, the laser tripped again around 9 pm last night.

    The Beckhoff status screen is attached.  I checked the chillers: crystal chiller was off, diode
chiller was still running.

    Reset the laser, tried starting it.  It failed.

    Previously the suspect was perhaps one of the flow sensors in the power meter cooling circuit.
I decided to change one of the crystal chiller tripping points.  I changed the exception time from
000 seconds to 001 seconds.  Thinking that if a fast transient caused the exception then perhaps a
1 second delay might "filter" it out.  If there was a real problem with the flow rate then 1 second
might not make much difference before things shut down.  I need to check the chiller manual.

    Tried restarting the laser again but it failed.  I rebooted the Beckhoff PC and was then able
to start the laser.  No sooner as I had left the diode room, I noticed that the crystal chiller
was complaining about a "flow sensor problem".  I re-entered the diode room to check the laser status -
it was still on.  So it appears that the change in exception time did the trick (for now).

    My conclusion from this is that the vortex flow sensor in the crystal chiller is the prime suspect
at this stage and not a flow sensor in one of the laser's cooling circuits, since there is no external
signal feed back to the crystal chiller.  If it were an external sensor, the interlock box would have
kicked in.
Images attached to this report
H1 General
jeffrey.bartlett@LIGO.ORG - posted 21:02, Friday 22 July 2016 (28597)
Ops Evening Shift Summary
Title:  07/22/2016, Day Shift 23:00 – 07:00 (16:00 – 00:00) All times in UTC (PT)
State of H1: IFO locked at INCREASE_POWER, 19.9W. Wind is a Moderate Breeze to Near Gale (18 - 38mph), and gusty. Seismic is elevated, as expected, at the End Stations. Microseism is low but has been rising for the past hour. Locking during the day shift has not been difficult.          
Commissioning: Commissioners are commissioning.
Outgoing Operator:  Travis
 
Activity Log: All Times in UTC (PT)

23:00 (16:00) Take over from Travis


Title: 07/22/2016, Day Shift 23:00 – 07:00 (16:00 – 00:00) All times in UTC (PT)
Support:  Jenne, 
Incoming Operator: N/A

Shift Detail Summary: IFO is locked at DC_READOUT. No commissioners on site. No planned commissioning work tonight. I have locked the doors and turned off the light. 
H1 ISC (ISC)
carl.blair@LIGO.ORG - posted 17:02, Friday 22 July 2016 (28596)
First damping of PI with new architecture

[Terra, Kiwamu, Carl]

We have demonstrated damping PI with the new architecture.  the 15.009kHz ETMY mode (MODE26) was driven up with noise several times before getting teh settings correct for damping.

Images attached to this report
H1 ISC
jenne.driggers@LIGO.ORG - posted 16:28, Friday 22 July 2016 (28595)
Lockloss due to ISS 2nd loop

Since we're "stuck" at ~20W while Team PI finishes re-commissioining the new system, I asked JeffB to take us to NomLowNoise @20W so that we could make sure we can set the Intent Bit when we're ready to actually do so. 

Unfortunately, the ISS second loop error signal wandered away and the 2nd loop unlocked.  The IFO was still locked for almost 1 second, but the HARD ASC loops were very unhappy and caused some saturations that I believe caused the lockloss.  I'm not sure why the 2nd loop ran away though.

Images attached to this report
Displaying reports 60041-60060 of 87874.Go to page Start 2999 3000 3001 3002 3003 3004 3005 3006 3007 End