Displaying reports 62061-62080 of 77270.Go to page Start 3100 3101 3102 3103 3104 3105 3106 3107 3108 End
Reports until 19:50, Monday 15 December 2014
H1 SUS
betsy.weaver@LIGO.ORG - posted 19:50, Monday 15 December 2014 - last comment - 20:31, Monday 15 December 2014(15634)
BSC10 ETMy-HR surface inspection

Betsy, Travis, Rick

After Jim finished locking the ISI this afternoon, Travis hopped in the chamber.  Using the Green LED flashlight (part of the "green lantern" kit) he immediately identified the 3 spots on the HR surface of the ETMy that last week's pcal images highlighted - 2 fairly large chunks of FirstContact (FC) remnants and one quite small one.  I also went into the chamber and with green and white flashlights we then identified the ring observed at 3" diameter around the center of the optic also due to previous FC sheets.  The 3 macroscopic remnants near the center of the optics look very similar to remnants we have observed on other failed FC sheet pulls, and were likely left on there from the March cleaning FC sheet, and not from the original sheet.  Why Margot and I missed them in March is beyond me.

We then:

Installed the ACB locking brackets and dropped the ACB down such that it was partially swung back.  (We did not use the wedge, but instead just let it hang in it's free and open state.  This allowed pcal to still see the optic but also gave enough clearance to attach the green lantern and also to paint FC.)

Installed the Green Lantern on the ETMy structure and reinspected the optic.  Again the large FC remnants lit up well.  However, the ring feature did not show up under this lighting condition.  (The flashlights were better for this.  The ring feature looked more like a staining or clouding that started at the 3" ring boundary and spread ~uniformly outward towards the edge of the optic.  There did not appear to be large macroscopic contaminants or FC remnants at this 3" ring boundary.

Rick took a few pcal photos (see attached) of the remnants and the ring using the green lantern and the green flashlight.

We secured the ETM TM with TFE rails on the EQ stops around the barrel of the optic.

We proceeded to paint FC onto the HR surface, making a nice thick coating after a few layers were applied - this took ~45 mins.  Note, we only observed "spider webs" on the FC sheet/brush toward the end of the 45 min and they were more noticeable around the TFE rail that was nearest the HR face which was likely charged and attracting or even causing(?) the webs.

Set the PETG cap on the face of the optic.

 

Tomorrow we plan to come in mid morning and then catch up with Richard/Fil to make the ESD connector swap midday.  After that, we can pull the FC and hopefully see that we have removed the remnants and improved the ring feature. 

Comments related to this report
richard.savage@LIGO.ORG - 20:21, Monday 15 December 2014 (15635)COC
The first image shows the ETM surface illuminated with the "Green Lantern."

In the second image it is illuminated with a green LED flashlight.

The third, zoomed image is illuminated with both the green lantern and the green flashlight and shows the large first contact remnants (the two bright vertical objects with ragged edges and the smaller object to the left), as well as the edge of the "ring feature" identified with the ~3" window in the first contact that was applied previously.  Although the two vertical objects are roughly aligned with the edge of the ETM telescope baffle in the background, this photo and in-chamber visual inspection clearly showed that they are debris on the front surface of the ETM.
Images attached to this comment
richard.savage@LIGO.ORG - 20:31, Monday 15 December 2014 (15636)COC
A few images of the process of painting the First Contact on the ETM HR surface attached below.

The last image shows the PETG cover in place to slow the drying of the First Contact.
Images attached to this comment
H1 SEI (DetChar)
krishna.venkateswara@LIGO.ORG - posted 17:54, Monday 15 December 2014 (15632)
A case for closing RZ loops on BSC-ISIs

F. Matichard, K. Venkateswara

This is related to Sheila's alog (15607), where they had trouble locking MICH_DARK. This was being caused by excessive RZ motion as described in the comment to that post.

  1. Fabrice and I then tried to understand what was producing this motion, because the ground is very quiet in RZ under normal circumstances. To do so , all the isolation loops were turned OFF (red) and then ON (blue) at ITMX. The result is shown in the first pdf. This led us to conclude that all the RZ/Yaw being produced is self-inflicted. RZ is increased by a factor of ~50 from normal.
  2. To see where this comes from, we looked at the coherence between RZ and X, Y  and Z channels. As seen in the second plot, it looks like Z dominates (through the known Z-RZ coupling) near 0.1 Hz, and Y and X cause the rest at lower frequencies.
  3. This excess RZ contributes to the problems locking MICH as seen in the third attachment. Note the high coherence between BS_CPS_RZ and the MICH_OUT at very low frequencies.
  4.  One possible solution is to close the RZ loop, as seen in the fourth pdf. First page shows the CPS_RZ, second shows the T240_RZ and the third shows the OpLev_Yaw. Green curve is with no sensor correction, blue is with Z sensor correction to HEPI and the red is with Z sensor correction to HEPI and RZ loop closed with a high blend (T750)  (along with Z to Rz subtraction at Stage 1).

Ideally, we should not even need Z to RZ subtraction and a simple position sensor loop ought to be sufficient to hold RZ constant. We will test this configuration soon and look at other chambers as well.

Currently, the RZ loop is engaged on the Beamsplitter.

Non-image files attached to this report
H1 SEI
jim.warner@LIGO.ORG - posted 17:04, Monday 15 December 2014 (15630)
ETMY ISI locked for de-schmutzing
LHO VE
kyle.ryan@LIGO.ORG - posted 16:52, Monday 15 December 2014 (15629)
Instrument-air compressor #1 at X-mid seized (several days ago) -> will replace at next opportunity


			
			
H1 ISC
alexan.staley@LIGO.ORG - posted 16:18, Monday 15 December 2014 (15628)
Snap files

I have made new snap files for h1iscey, h1iscex, h1asc, h1susetmx. Betsy had recently done h1susetmy, so I did not repeat it.

H1 AOS
kyle.ryan@LIGO.ORG - posted 16:17, Monday 15 December 2014 (15627)
Hard-closed GV18 -> Vented Y-end -> Removed North door of BSC10
Kyle, Gerardo-

Venting activities 


Kyle, Gerardo, Bubba, Betsy- 

BSC10 door removal activities
H1 IOO (IOO, PSL)
paul.fulda@LIGO.ORG - posted 16:09, Monday 15 December 2014 (15625)
PSL RFAM measurement setup

[Peter K., Alexa, Paul, Mackenzie]

(Written by Mackenzie, on Paul's login)

This morning we went into the PSL to set up the RFAM measurement and try to finalize the IO table layout.

We found that IO_AB_W1, which sends the beam to IO_AB_PD1 and IO_AB_PD3, was substituted previously with a temporary 1" HR mirror, so we located and installed the correct AR 2" window. We placed IO_AB_PD3 in transmission of IO_AB_W1 and placed IO_AB_PD1 (the RF 1811 PD) in reflection of the window. IO_AB_PD3 is saturating right now even at the lowest gain setting, so we'll have to attenuate that beam more somehow. 

Regarding the RFAM measurement, we were able to see the carrier with 45 MHz and 9 MHz sidebands on the OSA. The resolution was not great, and some of the sidebands were lost in the noise floor. Volker had previously performed this measurement, but needed an amplifier to boost the signal onto the EOM to see the sidebands well. We intend to go back in later today or tomorrow with an amplifier and try again. When we slowed the OSA scan down to try to resolve the peaks better, we found that the carrier peak was quite distorted (see attached pic). Maybe something to do with PZT non-linearity, we're not sure yet. Another attached pic shows the visible 45MHz sidebands along with two carrier FSRs. 

We looked at the RFPD spectrum, and were only able to see 9MHz sidebands. For some reason neither the 24MHz or 45MHz sidebands were visible. We had assumed RFAM/PM would be consistent between different sideband frequencies, but this doesn't seem to be the case. Again, we will know better once we get the amplifier in place. 

Images attached to this report
H1 ISC
alexan.staley@LIGO.ORG - posted 16:05, Monday 15 December 2014 (15626)
Preparing for RFAM

Alexa, Peter K., Paul, Mackenize

We have begun the set up for the RFAM measurement in the PSL enclosure.

First we aligned the OSA and attached it to an oscilliscope. We could see the carrier and 45MHz sidebands, but only sort of see one of the 9 MHz sidebands. After we set this up we remembered an email from Volker indicating we must add an amplifier to the EOM in order to better see the sidebands from the OSA. We will do this when we return to the PSL enclosure. At the moment we have a beam dumb in front of the OSA until we continue.

Second we placed an 1811 PD on the table. The PSL layout indiciates that IO-AB_PD1 should be a 1811; however this was not installed on the table. Instead we had found IO-AB_PD3 in the postion of PD1 and the wedge IO_AB_W1 was actually a HR mirror. We moved PD3 to its specified location and put the proper wedge (2025-45P) on the table. We placed the 1811 as specified and connected it to a spectrum analyzer. We could only see the 9 MHz and not the 45 MHz sideband ?? Also PD3 is conntected to H1:PSL-EOM_A_DC_POWER, which reads 2.22 (not sure what this is calibrated in); we noticed we are actually saturating this PD even though we are at this lowest gain setting of the PD.

H1 General
edmond.merilh@LIGO.ORG - posted 16:01, Monday 15 December 2014 (15614)
Daily Ops Log

07:50 Karen in LVEA

08:00 EY VENT TODAY

08:50 I'm covering for Travis today, apparently :)

08:51 Kyle requested picomotor drives be disabled at EY

08:59 Bubba into LVEA

09:04 Switched ESD bias to ETMY L3 off

09:19 Ellie out to LVEA

09:23 Aaron out to LVEA

09:49 Fil out to LVEA

10:25 Compressed air alarm at Mid-X. John and Kyle notified

10:49 Corey out to LVEA - squeezer bay

12:24 Hugh to EY to lock HEPI

12:57 Betsy and Travis will be goin into the LVEA and then to EY - EY control will be turned over to Betsy after the vent for optic cleaning.

13:20 Fil and Essence out in the LVEA

13:22 Jeff at EX to check out contamination stuff13:25 Kyle getting ready to start Kobelco for at HAM1

13:40 Corey back out to squeezer bay

13:43 Cris to EX to stock garb

14:00 Jeff back from EX

14:29 posted PSL report

15:24 Suresh and Doug out of LVEA - replacing BSC3 op Lev LASER

15:35 Cyrus out to LVEA to check on a couple of vacuum computer

15:49 Cyrus out of LVEA

15:51 Dick out to ISC R1

16:00 Jeff maybe out to EY

H1 SEI
hugh.radkins@LIGO.ORG - posted 14:57, Monday 15 December 2014 (15624)
Tilt Decoupling Checked ITMY--Numbers from Feb Still Good

I did this back 7 February, the numbers still look like they work well.

See the attached where I measure the tilt with the previous values and bracketed values around the good number.  The blue traces are the original values and the green and red traces are with the vlues offset a bit.  Still looks like the old values are the best.  The first attachment is the inline measurements (drive Y) with the second attachment showing the croosline measurement (drive X.)

Images attached to this report
H1 PSL
edmond.merilh@LIGO.ORG - posted 14:29, Monday 15 December 2014 (15623)
PSL Status report
Laser Status: 
SysStat is good
Output power is 29.3 W (should be around 30 W)
FRONTEND WATCH is GREEN
HPO WATCH is RED

PMC:
It has been locked 4day, 22hr 0minutes (should be days/weeks)
Reflected power is 2.9 Watts  and PowerSum = 25.8 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 0h and 35 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 1.74V (should be 0.9V)
ISS:
The diffracted power is around 7% (should be 5-15%)
Last saturation event was  h and 16minutes ago (should be days/weeks)

There was a group of four working in the enclosure today taking RFAM measurements causing some grief to the normal stability of things.
H1 SEI
hugh.radkins@LIGO.ORG - posted 14:15, Monday 15 December 2014 (15622)
EndY BSC10 HEPI Pump Station Down Briefly

Since the platform was tripped (at the end of my locking up the HEPI) I turned the pump down to get zero pressure readings.  I use these to define the AOFF so the sensors don't have that zero bias offset.  Pump back on within a few minutes and servoing on 70psi differential pressure.

LHO VE
kyle.ryan@LIGO.ORG - posted 13:31, Monday 15 December 2014 (15619)
~1330 hrs. local -> Started Kobelco purge-air compressor and booster pump at the Corner Station
In preparation for HAM1 door removal
H1 SEI
hugh.radkins@LIGO.ORG - posted 13:11, Monday 15 December 2014 (15618)
ETMY EndY BSC10 HEPI Locked onto Stops

Secure for pushing & pulling.

H1 TCS
eleanor.king@LIGO.ORG - posted 12:49, Monday 15 December 2014 (15617)
ITMy and ITMx Hartmann sensor realignment

ITmy and ITMx HWS have both been aligned so that the green beam from the X-arm returns through the two irises and onto the respective HWS's.  To do this I had to move both of the upper and both of the lower periscope mirrors about an inch to the right. No other optics were touched.  I took some pictures of the original beam path.  The beam going to the ITMy HWS looked clean as came onto the table (image 1), but clipped the periscope (images 2 & 3).  There were a few beams coming from ITMx, but the brightest one missed the periscope entirely.  I aligned the brightest beam to the HWS.

 

Once the green beams were returning to both HWS cameras, I tried looking at the HWS camera images to fine tune the alignement.  I could open a camera image using the command

 /opt/Hartmann_Sensor_SVN/release/bin/stream_image_X/distrib/stream_image_X/opt/Hartmann_Sensor_SVN/release/bin/stream_image_X/distrib/stream_image_X

but I couldn't see a beam in this image.  I tried shining a flashlight on the cameras, including without the Hartmann plate for ITMy camera, but the image didn't change.  Either the camera exposure needs to be increased (seems doubtful), or the camera image isn't updating properly.   Once I've figured out how to get the camera image I'll go back and fine tune the alignment.
 

Images attached to this report
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 09:27, Monday 15 December 2014 (15581)
Long Term Frontend Laser Performance
In summary, the health of the laser is okay.



The first plot is of the output power of the frontend laser pump diodes, D1, D2, D3, and D4.  The output power decline for D1 and D2
is fairly small, declining about 2.5% over the last 6 months.  For diodes D3 and D4 the picture is less clear as the signal is a little
noisier, but it looks like about 5% over the last 6 months.

The second plot is of the diode currents.  DCUR1 corresponds to the currents for D1 and D2.  DCUR2 for D3 and D4.  Considering that
none of us have explicitly changed the diode current this should appear constant (and it does).

The third plot is of the output power of the various stages in the power amplifier.  The fluctuations in the power monitors for stages
1 through 3 might be due to relative humidity fluctuations in the LAE.  The output of the frontend laser decreased by ~2 W about 120
days ago.  This might be a real power decrease or the alignment of the beam onto the monitoring photodiode.

The NPRO output power has steadily declined.  Attached are 3 plots of the NPRO output power.  6 months, 2 years and going back to
May 2012.  To recover some out the output power, we should consider increasing the diode current.


/* Note the frontend diode box was switched out on June 6th, 2014. */
Images attached to this report
H1 SEI (ISC, SEI)
sheila.dwyer@LIGO.ORG - posted 19:05, Sunday 14 December 2014 - last comment - 13:32, Monday 15 December 2014(15607)
HEPI Z sensor correction is increasing RMS motion of michelson by about a factor of 10

Stefan, Sheila

Today we had difficulty locking DRMI, we tried PRMI and found we had difficulty locking that as well.  Then I found that I couldn't even keep MICH DARK locked.  Stefan was able to keep it locked by turning the gain up to -1000, removing the limiter from the BS M3 ISCINF once it locks, and adding a boost (FM4 in LSC_MICH).  I compared the control signal to what we saw on decmber 3rd at a time when DRMI was locked, and see that we have about a facotr of ten higher RMS, all due to motion below 0.2 Hz.  I then tried turning off different parts of sensor correction, first just the ISI XY, then also the HEPI Z. turning off HEPI Z brought the RMS back down to the old value.  For now I'm leaving it all off.  

Images attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 21:08, Sunday 14 December 2014 (15608)
BS M2 DRIVEALIGN L2Y gain changed from -1 to -0.8.

This significantly improved the BS DC balancing.
krishna.venkateswara@LIGO.ORG - 12:34, Monday 15 December 2014 (15616)

K. Venkateswara

Looks like Z sensor correction is producing excessive Rz (Yaw) motion of Stage 1 at all three CS BSCs, but particularly so at the Beamsplitter. Attached plot shows the ASD of the CPS_RZ from the same period. The REFs were taken when Z sensor correction (to HEPI) was ON for all three, and the normal traces were with it OFF. While the sensor correction reduces motion at ~0.15 Hz, it is unexpectedly inducing excess RZ in all three, but much more so in BS. The coherence plot on Page 2 shows that RZ of the BS was the reason the MICH was difficult to control.

I don't have an immediate answer but will investigate some more.

Non-image files attached to this comment
krishna.venkateswara@LIGO.ORG - 13:32, Monday 15 December 2014 (15620)

I think one solution to the above problem is to turn on Z to Rz subtraction and enable the Rz loop. Jim and I tried this last week briefly and the result appeared agreeable. So I tried it again at ITMX and the results are shown in the attached file. First page shows the CPS_RZ, second shows the T240_RZ and the third shows the OpLev_Yaw. Green curve is with no sensor correction, blue is with Z sensor correction to HEPI and the red is with Z sensor correction to HEPI and Z to Rz subtraction at Stage 1.

I will also try this at BS and ITMY chambers.

Non-image files attached to this comment
H1 ISC
stefan.ballmer@LIGO.ORG - posted 16:39, Saturday 13 December 2014 - last comment - 14:13, Monday 15 December 2014(15599)
Full Green WFS on both arms: a little more bandwith
Evan, Stefan

Did some more WFS/camera loop work on both arms:
 - Added a 0.3Hz LP with a notch at the main pendulum modes into the WFS output filter modules. This allows for a ~3-5 times higher gain.
 - Carefully measured the output matrix for the camera loops (X and Y arm, yaw and pitch) by dragging the ETM around and letting the WFS follow.
   The new output matrix values are in the snaphots #3 below. With them the camera loop no longer relies on gain hierarchy - it's gain was increased ~x10.
 - Some gains were redistributed - see the attached snaps.
 - Cleaned up the arm Guardian. Removed all old slow feed-back stages, and instead added a state commanding Daniel's new setup.
 - Added a WFS relief step to the Guardian.
Images attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 17:34, Saturday 13 December 2014 (15600)
We noticed that the higher Camera gain can cause a too big kick during the WF engaging process.

Thus we used the existing FM triggering to set up a 10 second delay. During those 10 seconds the DoF3 (P&Y) , i.e. the camera loops, have 20 times lower gain. This seems to take care of the engaging problem.

The attached snapshot contains all elements that changed.
Images attached to this comment
stefan.ballmer@LIGO.ORG - 14:13, Monday 15 December 2014 (15621)
The experimentally measured output matrix to move TMS, ETM and ITM simultaneously to center the ITMX green camera (without misaligning the beam) is
      X PIT:           X YAW: 
ITMX: 0.68            -0.43
ETMX: 1.00             1.00
TMSX: 1.43             1.00

Y PIT:                 Y YAW:
ITMY: 1.43            -1.23
ETMY: 1.00             1.00
TMSY: 1.54             1.15

The expected matrix would be
PIT:
ITM:  1.06 = -g1 
ETM:  1.00
TMS:  1.00

YAW:
ITM: -1.06 = g1
ETM:  1.00
TMS:  1.00

This means that our angular actuation calibration is somewhat fishy...


For reference, the H1 installed optics radius of curvature are (from https://galaxy.ligo.caltech.edu/optics/,  after coating measurements):
ETMX: ETM-08: RoC=2242m ± 2m, (LIGO-C1103233), g2= -0.7817
ITMX: ITM-03: RoC=1940m,      (LIGO-C1103237), g1= -1.0590
X Arm length: L=3994.4704 m ± .3mm (alog 9626) 

ETMY: ETM-12: RoC=2240 ± 2mm, (LIGO-C1103257), g2= -0.7832
ITMY: ITM-11: RoC=1939.4m,    (LIGO-C1103255), g1= -1.0596
Y Arm length: L=3994.4691m ± .7mm (alog 11611)


H1 COC (COC)
richard.savage@LIGO.ORG - posted 08:01, Saturday 06 December 2014 - last comment - 17:38, Monday 15 December 2014(15479)
ETM high-resolution images with Green and IR light resonating
SudarshanK, TravisS, EvanH, AlexaS, RickS

Using the Pcal beam localization cameras at both end stations, we took images of the ETM surfaces under three conditions: IR and Green resonating; IR only resonating, and Green only resonating.

Attached below are two composite images composed of four separate images taken with the same camera settings:
Upper Left: Xarm Green
Lower Left: Xarm IR
Upper Right: Yarm Green
Lower Right: Yarm IR

The images in the first composite were taken with the following camera settings: F8, ISO 200, 30 second exposure, WB-cloudy.

For the second composite image the aperture was F29 (~13 times less light)

The Yend camera was re-focused for the IR-only images, but the Xend camera was not re-focused.
Images attached to this report
Comments related to this report
richard.savage@LIGO.ORG - 21:51, Monday 08 December 2014 (15502)
Thomas Abbott at LLO applied the Pcal beam localization analysis the the LHO ETMY image to calculate the position of the center of the optic in the image from last Friday.

The image below contains lines that indicate the center of the optic using the Pcal image analysis.
Images attached to this comment
betsy.weaver@LIGO.ORG - 12:19, Wednesday 10 December 2014 (15536)

Attached is a picture of the original FirstContact (FC) sheet, circa ~Jan 2014, showing the "IAS window" which is a thinner film of FC in the central 3" of the larger sheet.  To me, the shape of the FC window looks similar to the 3" ring showing up in green on the recent optic photo above.  SYS is working with us to get our cleaning game plan together in order to remove the ring.  As well, they are investigating other possible scenarios of where the ring came from if not the window.  Note, there was a full FC sheet re-cleaning in March that apparently did not remove all of the ring that was left behind apon the removal of the first sheet.  To be continued...

Images attached to this comment
dennis.coyne@LIGO.ORG - 17:38, Monday 15 December 2014 (15631)COC, SYS
I've attached an overlay of (a) the SolidWorks CAD view of ETMy along the PCal camera path and (b) the PCal camera image of H1 ETMy (scaled and rotated). Since SolidWorks does not diffract the image viewed through the ETM optic, I indicate the shift in the ETM Telescope Baffle aperture as well. Three of the 4 bright areas are along the ETM Telescope Input aperture/baffle edge (a coincidence?). (The upper one is red.) One of the 4 bright spots does not correspond to any feature in the CAD image and is likely a spot of residual First Contact.
As subsequently shown by the zoomed in PCal image using the Green Lantern flashlight (green LED) after venting (see entry #15635), it is simply a coincidence that the two prominent bright areas appeared to be along the ETM Telescope baffle aperture edge.
Non-image files attached to this comment
Displaying reports 62061-62080 of 77270.Go to page Start 3100 3101 3102 3103 3104 3105 3106 3107 3108 End