Nic, Sheila
We put an excitation into the HAM6 ISI ISO Y filter bank, (30000 counts at 0.3 Hz) from about 3:17-3:19 UTC July 11. We then did a by eye fit (on a log log scale)for a fringe wrapping model. We expected the excitation to result in 30 um motion of the OMC, but we had to use 36 um to get the fringe speed right. We get an amplitude reflectivity of 1.6e-7 for the single pass shelf. (compare to 1e-5 measured in 17919) We see no evidence of a second shelf or a shelf in the null stream.
We plan to make measurements in exactly the same was as 17919, if we get a chance again tonight.
There is a typo in this alog, the reflectivity is r=160e-7, as the legend in the plot says, not 1.6e-7 as I wrote.
(stefan sheila evan nic)
We had to modify the mode damping settings to get the ETMX roll to decay. These might not be stationary, but here is the record of the settings that worked.
|
Stefan, Sheila, Nic, Evan
We remeasured the relative strength of the EX and EY ESDs in full lock. EX was driven with the high-range driver (40 V/V dc), and EY was driven with the low-noise driver (2 V/V dc). All other things being equal, we expect a relative strength of 20 V/V between the two actuation chains.
We found that the relative strength is instead 30 V/V (see attachment, which includes a digital gain of 30 ct/ct in the EY path). When we first did this measurement (back in May, pre-discharging), the relative strength was more like 50 V/V. So we are closer to the nominal value.
We successfully transitioned control of DARM from EX to EY with this new digital gain. We also took a quick DARM OLTF, both before an after the transition. The attachment shows an old, pre-vent OLTF (blue), today's OLTF with EX (green), and today's OLTF with EY (red).
A relatively small point, but the LV ESD Driver DC gain is actually closer to 1.9 V/V at low frequencies. There's a pole just above 150Hz.
Scott L. Ed P. Rodney H. 7/8/15 Cleaned 68 meters of tube, bellows, and crevices ending at HSW-2-093. Test results posted below. Removed lights and started relocating equipment to next section and hanging lights. 7/9/15 Cleaned 79.6 meters of tube and ending at HSW-2-089. Bellows and crevices still need to be cleaned. 7/10/15 The bellows and crevices on the previous section were cleaned and tested. Results posted below. The lights were removed and relocated to the next section. The crew started vacuuming the support tubes and capping them. 21 meters of tube cleaned. A total of 227.6 meters of Y-Arm tube has been cleaned.
Chris S. Joe D. 7/8/15 Installed aluminum strips on 39 joints beginning 604 meters from the corner station and continuing north and caulked 13 of those. 7/9/15 Finished caulking the joints from yesterday and cleaned 10 additional joints ending 756 meters north of the corner station.
plot 1: PSL FSS TPD (top), PMC TRANS (bottom) - 7 days of data - channels changed behavior about 4 days ago, and TPD fell below 0.9 sometime in the last day or so.
plot 2: PMC REFL (top), PMC TRANS (bottom) - 7 days of data - PMC REFL climbing faster in last 24 hours or so.
plot 3: Ch7: PSL OSC BOXHUM (upper right) seems to be correlated with other three channels in the first 1/2 of the plot.
- IFO in commissioner's hands most of the afternoon
- during lunch, IFO locked at 10W, and I raised to 15W about 30 minutes later, it held about 15 minutes, then LL
- minimal activities, no LVEA entry
Current:
- locked, commissioning going on
- PSL FSS TPD flashing yellow, since it's below 0.9V - see plots in alog
- RobertS and DickG and Nutsinee and Elli all have weekend work permits, #5343, #5344, and #5342, respectively
The temporary T240 as ground sensor for ETMX SEI sensor correction was swapped on Tuesday 7 July with an STS2. Attached are spectra comparing the signals. Obviously we can't compare signals at the same time so the T240 time is 3 July at 0200 local and the STS2 time is 9 July at 0200. Based on the wind trends, these are pretty quiet times for wind.
To assess the stationarity of the Spectra, I walked before and after this time a few hours on July 3 and the variability below a Hz can be many factors. With the 0200 July 3 signals as reference, I walked before and after 0200 on 9 July and really found no close match, again especially at lower frequencies. The Z DOF of the T240 below about 30mHz seems to be main outlier but above about 200 mHz the traces are maybe within a factor mostly. There is an exception to that centered around 0.6Hz with the T240 showing a bit higher amplitude especially on the Y DOF. Maybe we could argue that our calibrations for these two instruments are correct.
Leonid.Prokhorov, Jeffrey.Kissel We continue the charge measurements on ETMs. The data are beginning to be consistent with a small increase in charge, consistently negative. For the past few weeks since the discharging the charge has been small (<10 [V] Effective Bias Voltage), and roughly consistent with a parent Gaussian distribution with a mean about 0.0 [V] Effective Bias Voltage. However, as we get more data, we're beginning to see a case for a slow negative charging of the test masses (about 2-5 V/week) in at least half the quadrants (ETMX UR, UL, and LR; ETMY (the pitch estimates of) UR, UL, and LR). The uncertainty of these measurements is too large to state that the trend is statistically significant, but yellow flags have gone up. This argues that we should continue roughly daily (say, thrice a week) measurements of the charge, and we need to continue to do so for another week or three. There are three sets of plots, each which show the same data in a different way: (1) ETMX.png & ETMY.png show the "raw" data, where the result of each estimate of the effective bias voltage shown *since* the discharging. Each single data point is an estimate of the effective bias voltage, i.e. the charge, as determined by driving the test mass while varying the requested bias voltage and measuring the response with the optic's optical lever. (2) ETMX_Mean.png & ETMY_Mean.png shows both the mean and standard deviation, and weighted mean and sqrt(weighted variance) of the charge measurements for a given day (which can be from ~4 to ~15 of these 12 minute measurements per mean point). We believe this better shows the long-term trend of the charge. (3) ETMX2015.png & ETMY_2015.png shows the mean charging measurement results since January 2015, i.e. it includes data when the test masses were charged and shows the discharge which happened in Mid-June.
It is useful to note that if the bias is going in a negative direction, the charge on the test mass is going positive. This would be consistent with uv or xuv photoemission from the front surface of the test mass. Refer to DCC T1400647-v1. The slow growth of charge on the LLO etmy is also going in the same direction toward positive charge.
I modified some scripts to make easier processing of the charge measurements. 1. Couple of matlab scripts allow to analyze several sets of measured results (instead analyzing one point at time). You can input the set in ESD_Analyse_ManyMeasurements.m, it calls ESD_analysis.m which is a little bit adapted version of ESD_UL_LL_UR_LR_analysis_07_H1.m . As a result you have the set of .mat files in corresponding folders and one figure per set showing dependence of response with different bias voltage. You can processing ETMX and ETMY all together. You can process all data of the day in one script. 2. I modified the long trend matlab script to make two new kinds of plots (you can see it in alog https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=19458). Also, it create plots for several TMs, if you need. New version is Long_Trend_H1_v2.m . 3. I want to modify python script, but it seems that I can do only one improvement, when I want two and have else one trouble. - First, which seems to be done, is making several measurements in line, without pause or waiting for operator's reaction. ESD_UL_LL_UR_LR_charge_07-H1-ETMY_Long.py repeats the measurements as long as "ENTER" is not pressed. After this it finish round of the measurement and stops. For sure, it can be stopped by ctrl+c (and you'll have partly measured directory). - Second - we often have time on both ETMs together. I'd prefer to have one script which measure charge at both arms at the same time. Not in line, as it is realized now, but simultaneously. Now I use two scripts (one for each arm, but I'm not sure if it's the best choice). - Third, at first run scripts return an error. It happens not at start, but in several first minutes of work. Seems that it happens once for each terminal. These errors do not stop the script. Also, I saw that length of "waiting for response loop" (I mean, number of measured GPS seconds) may differs. It happens rare, but it happens. The measurement time usually becomes smaller, but once I saw more than 2 minutes of measurement for one loop (instead 30 seconds). For sure, any of these errors kills all the processing for the data set. So, if you want to use modified versions of scripts - you are welcome! If you have any suggestions or critic - you are welcome! If someone could improve python script - I would be happy! Links for all the committed files: https://redoubt.ligo-wa.caltech.edu/svn/sus/trunk/QUAD/Common/Scripts/ESD_analysis.m https://redoubt.ligo-wa.caltech.edu/svn/sus/trunk/QUAD/Common/Scripts/ESD_Analyse_ManyMeasurements.m https://redoubt.ligo-wa.caltech.edu/svn/sus/trunk/QUAD/Common/Scripts/Long_Trend_H1_v2.m https://redoubt.ligo-wa.caltech.edu/svn/sus/trunk/QUAD/Common/Scripts/ESD_UL_LL_UR_LR_charge_07-H1-ETMY_Long.py https://redoubt.ligo-wa.caltech.edu/svn/sus/trunk/QUAD/Common/Scripts/ESD_UL_LL_UR_LR_charge_07-H1-ETMX_Long.py
(times in PST)
0803 - Robert to the LVEA to get a few things
0825 - Robert out
0855 - Katie to LVEA to level magnetometer
0920 - Richard to roof
0927 - Katie out
I came in early to get some work done and decided to try to lock before I began. I very easily made it up to Bounce roll damping before it lost lock. I didnt see any modes badly rung up, but many suspensions were saturating according to VerbalAlarms.
Sheila, Stefan, Elli, Nic
This afternoon we got a little more time with the IFO. We redid inital alingment and were able to engage the ASC which increased our recycling gain. We were able to power up to 15 Watts without a problem, when we increased the power to 23 Watts a 0.4 Hz oscillation that showed up mostly in SRC1 PIT and SRC2 PIT knocked us out of lock. We've had difficulty in the early part of the evening with ALS, we have the glitches that show up in the arms even when only the arms are locked, reminding us of the glitches seen in alogs 17576 and 15242. As before, these glitches went away a couple of hours and we are now locking OK again.
We have reduced the time for the REFL trans ramp from 5 seconds to 1 second, this seems to be fine and we could think about making it shorter or reducing other ramp times in the final stages of CARM offset reduction. We haven't had the REFL trans lockloss today, but this problem has always been intermittent so we will have to continue locking for several days to see if speeding it up actually helped.
Screen shots of several of these ALS locklosses that seemed to go away after a few hours.
We've had a similar problem this morning. We saw one glitch at around 18:11:24 UTC, at this time we were sitting with only the arms locked and tidal from the green PDH to the ETMs.
Now the problem has gone away on its own again.
The current of the REFLAIR diode while in lock is about a factor of 10 below where shot noise is equivalent to dark noise.
In order to better understand the out-of loop behavior of the Common Mode loop, I did a lightbulb test on the REFLAIR_A diode. The thermal radiation from the lightbulb acts like a quantum limited light source at 9MHz.
The light was powered by a benchtop DC power supply. I mounted the bulb right in front of the diode and varied the DC light level (as seen by the REFLAIR_LF channel) and recorded the noise floor in the REFLAIR 9MHz demodulated channels. The whitening gain was temporarily changed from 12dB to 42dB for this test to overcome the ADC noise level.
The attached pdf file shows the noise vs current curve, and a fit. EDIT: apparently the LF channel is already calibrated in milliWatts. The 9MHz channel is not calibrated to physical units (though the calibration could be determined from the shot noise). As one can see, the light level we use in lock is about 10 times smaller than where shot noise starts to overcome dark noise.
REFLAIR also has an ND filter stack screwed directly on the diode box. Since we have so little light, maybe we should remove it.
data files
The amount of rf coming out of REFL9 in full lock (at 20+ W) is about −10 dBm, or 70 mV. It should be fine to switch over.
END X has been transitioned to the NEG pump. The main turbo is isolated but still spinning.
Pressure has remained at ~2.2e-8 torr.
Kyle, Gerardo, John
The NEG pump has been valved into the END Y volume and all appears well.
The plot shows 2 hours of the BSC chamber pressure. Initially the main turbo was pumping the station along with the LN2 pump and open to the tube. The LN2 pump had been opened to the BSC earlier in the day and we let the pressure settle over lunch.
With the turbo still connected you can see the pressure fall as we open the NEG to the volume. Next the turbo was valved out and the pressure returns to near the starting value.
So far so good!
(evan jenne nic)
We wanted to investigate the OMC alignment/backscatter problem. Driving the OMC SUS in Yaw has been known to cause backscatter noise due to the modulation of the optical path length when the OMC moves in Yaw.
Our procedure was to lock the vertex optics in a bright michelson configuration (a state has been added to the IFO_ALIGN guardian to make this easy). Then we wanted to drive the OMC in yaw and choose the yaw->longitudinal matrix element such that the center of rotation would be about the input beam, rather than the omc center of mass. This would be determined by minimizing the scatter as measured by either OMC trans, or the MICH error point.
I was surprised that we were not able to induce any significant backscatter fringe wrapping noise in this configuration. We drove the OMC SUS in longitude up to the point that the beam was misaligning enough to noticibly affect the OMC DC trans.
We also drove the ISI table directly by putting a 1Hz 1mm injection into the Y isolation loop error point.
Driving the path length, we both listened and had a live spectrum running. We saw no evidence of scatter in either OMC trans or MICH_IN1.
We will need to think if there is another configuration (available to use without arms) that will be more sensitive to backscatter.
I forgot to mention that we turned on the AS fast shutter and OMC pzt high voltage supplies for HAM6.
This measurement didn't work because I was wrong about the calibration. The isolation loop error points are in nanometers, not micrometers. So we were moving the table 1000 times less than I thought.
Attached is a noise budget for the DARM spectrum from 2015-06-07 00:00:00Z–02:00:00Z. Notes:
Conclusions:
Attached are the MICH and SRCL coupling TFs into OMC-DCPD_SUM.