Displaying reports 65621-65640 of 85578.Go to page Start 3278 3279 3280 3281 3282 3283 3284 3285 3286 End
Reports until 02:59, Thursday 13 August 2015
H1 ISC
sheila.dwyer@LIGO.ORG - posted 02:59, Thursday 13 August 2015 - last comment - 10:26, Thursday 13 August 2015(20497)
commissioning this afternoon and evening

The main story of the day is related to the non stationary low frequency noise first seen early in the morning. It seems to come from CHARD noise, which could be coupling to DARM more now because of a change in alignment.   

A few ideas for next steps:

We also worked on a few other things today...

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 03:04, Thursday 13 August 2015 (20499)

Here is a measurement of CHARD Yaw at high power, overlaid with yesterday's measurements at 23W.  The 23W measurement includes the MsBoost, but not any 23W boost or the lead-plus-cutoff filters that Sheila designed.

Images attached to this comment
gabriele.vajente@LIGO.ORG - 10:26, Thursday 13 August 2015 (20507)

Here is how I retuned the A2L. I injected some band limited noise (ellip band pass 1-100 Hz, amplitude 20000 cts) on ETMX_L2 L2L, P2P and Y2Y paths, with the P2L and Y2L gains set to zero. The measurements were good between 20 and 100 Hz. The ratios -P2P/L2L and -Y2Y/L2L are what we need to implement in the correction paths. Those trasnfer functions are quite constant above 30 Hz, but not so much below 30 Hz. We would need a better (sweep sine) measurement if we want to improve the decoupling below 20 Hz.

I changed the gains of the P2L and Y2L of ETMX as follows:

P2L from 1.18 to 1.03

Y2L from 1.33 to 1.23

Coherence of DARM with CHARD reduced at low frequency. However, we reverted to the old numbers to investigate the low frequency non stationary noise.

Images attached to this comment
H1 ISC
jenne.driggers@LIGO.ORG - posted 02:42, Thursday 13 August 2015 - last comment - 10:43, Thursday 13 August 2015(20498)
A2L run

I ran Hang's latest A2L script (see aLog 20013), after the realignment work that was done tonight (Sheila is writing up her entry as I type). 

We stil have excess noise at low frequency, but maybe there's a bit less than when I started the script?  The noise we're seeing is totally non-stationary, so it's hard to say.  Certainly the A2L didn't eliminate it.

Comments related to this report
hang.yu@LIGO.ORG - 10:43, Thursday 13 August 2015 (20508)

I checked the results of the latest a2l run. It seemed that the decoupling worked only for ETMX pitch and ITMX pitch, and the optimal gains changed only by 7% and 2%. On the other hand, it failed for ETMY pitch and ITMX yaw. I attached a worked result and a failed one for comparison.

By examining the results, the bad ones had a flatter slope and were more likely to have "outliers". We thus might be able to get a better result by increasing the steps between two measurements, or increase the number of gains to be measured. Nonetheless, it seemed to also indicate that we were already near the optimal spot that such an linear, single frequecy decoupling could achieve...

Images attached to this comment
H1 ISC
nutsinee.kijbunchoo@LIGO.ORG - posted 00:37, Thursday 13 August 2015 (20496)
Violin First Harmonics Damped!

Dan, Nutsinee

Tonight we identified and damped more ITM first harmonics.

Frequency Test Mass Damp Setting
991.7478 ITMY 260dB, -60deg
991.9345 ITMY 260dB, 120deg
992.4256 ITMX 260dB, +/-180deg
997.8868 ITMY 260dB, +/-180deg

They will be added to the violin wikipage.

Images attached to this report
LHO General
corey.gray@LIGO.ORG - posted 00:09, Thursday 13 August 2015 (20491)
Ops EVE Summary

All Times in UTC

H1 SUS (SUS)
corey.gray@LIGO.ORG - posted 23:02, Wednesday 12 August 2015 (20495)
ITMy Roll Mode Addressed

(Dan, Corey)

Around 5:30UTC (22:50PT), Dan noticed a rung up roll mode on DARM spectrum, and from the Bounce/Roll monitor, the culprit was ITMy. 

Brought up the bounceroll.stp StripTool from the Ops Template & could see ITMroll channel high.  From the ITMy SUS screen, opened the DARM BR DAMP filter bank screen.

Initial Values:  -60degrees & a gain of +40.

New Values:  +60degrees & a gain of -80.

This took care of ITMy, and should be fine for this current lock.  Whether we want to keep ITMy like this remains to be seen.  Should see how we look after a few locks.

H1 PSL
sheila.dwyer@LIGO.ORG - posted 17:20, Wednesday 12 August 2015 - last comment - 08:20, Monday 17 August 2015(20490)
PMC transmitted power dropped after PSL temperature excursion

After the PSL temperature adventure on monday afternoon, it seems as though the PMC alignment shifted.  We now get less transmitted power, although the sum of refl and trans has not changed much. 

Images attached to this report
Comments related to this report
peter.king@LIGO.ORG - 08:47, Thursday 13 August 2015 (20502)
In looking at the data around the period of the temperature excursion, all the power monitoring photodiodes indicate a
change in power monitored.  The ones located before the pre-modecleaner return to their previous values.  The ISS
photodiodes after the pre-modecleaner return to their previous values however the transmission and reflection signals
are lower than before.  The pre-modecleaner reflected spot doesn't look substantially different, however the pre-modecleaner
heater output did not return to its previous level.  It has remained at an elevated level since the temperature excursion.

The coefficient of thermal expansion for aluminium is approximately 22 microns per degK.  The attached plot suggests the body
of the pre-modecleaner changed by 0.5 degK and stayed at the higher temperature.  It should be noted that the pre-modecleaner
heater did its job and relieved the PZT high voltage.  In doing so the length of the spacer is different and this may be the
cause of any mis-alignment.  The plot also shows the Laser Room temperature.

It might be worth trying reducing (presumably reducing, might also be increasing) the heater offset to bring the pre-modecleaner
temperature back down to 304.5 degK from its current value of 305.0 degK to see if this peaks the power transmitted by the
pre-modecleaner.
Images attached to this comment
peter.king@LIGO.ORG - 08:20, Monday 17 August 2015 (20591)
Attached is the output of the quadrant photodiode in the ISS photodiode box with the corresponding temperature
measurement.  Clearly a change in the beam position is indicated, with the vertical not returning to its
previous value.  The horizontal seems to track the room temperature.
Images attached to this comment
H1 CDS
james.batch@LIGO.ORG - posted 16:35, Wednesday 12 August 2015 (20489)
The h1hwinj0 computer has been renamed to h1hwinj2
FRS 3247

The h1hwinj0 computer has been renamed to h1hwinj2. 
H1 General
jim.warner@LIGO.ORG - posted 16:03, Wednesday 12 August 2015 (20488)
Shift Summary
9:00 Richard, Carlos, Andres,9:00 Richard, Carlos, Andres, Ken working on CR monitor install, lots of banging on wall between LVEA and CR
9:00 Richard and crew in the CR install new monitors all day on wall next to LVEA
9:30 Kyle to Ymid, back 10:30
10:15 Robert to LVEA to do PEM injections
11:30 Robert to LVEA to do PEM injections
~19:00 Earthquake breaks lock
19:10 Ken fork lifting TV pallets around the corner station ~30 mins
19:30 Elli to ISCT6, out 20:50
19:30 Elli to ISCT6, out 20:50 Ken working on CR monitor install, lots of banging on wall between LVEA and CR
9:30 Kyle to Ymid, back 10:30
10:15 Robert to LVEA to do PEM injections
11:30 Robert to LVEA to do PEM injections
~19:00 Earthquake breaks lock
19:10 Ken fork lifting TV pallets around the corner station ~30 mins
19:30 Elli to ISCT6, out 20:50
LHO VE
bubba.gateley@LIGO.ORG - posted 15:57, Wednesday 12 August 2015 (20487)
Beam Tube Washing
Scott L. Ed P. Rodney H.

8/11/15
After filling the water tank, the crew moved the fans and re-strung the lights. 45 meters of tube and bellows were cleaned ending at station HSW-2-012. Cleaning results posted here.

8/12/15
60.6 meters of tube and bellows cleaned ending at station HSW-2-009. 

Non-image files attached to this report
H1 ISC
eleanor.king@LIGO.ORG - posted 15:15, Wednesday 12 August 2015 (20479)
Beam parameter measurement at AS_Air camera

Sheila, Elli

Summary:

Yesterday morning durng maintenance we went out quickly to measure the beam parameter ar AS-Air.   IThe beam waist is 3.7e-4 (1e-4) m at the AS_Air camera.  The fit of the beam divergence is pretty bad, we were hurrying to get done quickly and I didn't take enough data points to get a good fit for the beam divergence.  Using measured waist size and the (very poor) fit of beam divergence, the beam parameter at the AS_AIR camera is calculated to be 0.14+0.22i in the vertical axis and 0.16+0.29i in the horizontal axis.  UPDATE:  Today during the Earthquake at lunchtime I went out and took some more measurement points.  I am fitting these new points at the moment, and hopefully they will improve this measurement.

Details:

We used a straight shot through corner station with PRM, ITMX, SRM misaligned.  We placed a flipper mirror 0.434m upstream of the AS_Air camera, which sends the AS-Air beam across ISCT6 where we used a Thorlabs BP-104 slit scanning beam profiler to measure the beam profile at 5 locations (see optical layout diagram).  This beam scanner is mounted on its side, so that Y is the horizontal axis and X is the vertical axis.  The laser power was turned up to 24W to get enough light on Thorlabs beamscan to take a measurment (and the backgorund noise was still high at 24W).

The background noise was to high for the beam profiler to automatically fit beam waist, so we saved the beam shape in the x and y axes and fitted a waist size to these beam profiles after the measurement was over.  Puzzlingly, the maximum intensaity in the X and the Y directions is not the same.The fit of the gaussian beam in x and y is good.  Then plotted beam radius vsd distance to fit the waist location and size with a hyperbolic function.  There aren't enough points/too much error in each point to get a satisfying fit.  I have attached a fit which looks kind of reasonable, but really there is too much variation in the points to call this a good fit .(See picture of alternative fit which is almost as good, and very different.)  So the Beam waist location/size can't be pinned down using this meaurement.  This measurement could be inproved by taking more data points.

Images attached to this report
H1 ISC (DetChar, GRD)
sheila.dwyer@LIGO.ORG - posted 14:16, Wednesday 12 August 2015 (20484)
ALS shutters temporarily disabled

The ALS end station shutters have temporarily been commented out of the guardian. (They will no longer close, even in the shuttered state).  The goal is to have green light locked in the arms while we are at a good sensitivty, with the idea that this could possibily be helpfull for determining if some of the glitches we see are due to dust.  ( alogs 20395 20355 20328 )

We had originally added the closing of the shutters since we heard that it improved the noise at LLO, even though we saw no evidence that it improved our noise here.  After a few weeks of engineering run we should revert this change.

H1 General
jeffrey.bartlett@LIGO.ORG - posted 13:54, Wednesday 12 August 2015 - last comment - 14:30, Wednesday 12 August 2015(20483)
Dust Monitor Alarm Levels

   Reset the dust monitor alarm levels for the LVEA, VEAs, and PSL from the tighter levels associated with the June vents to the levels for the O1 environment. The current values are:

Room Air Size Minor Alarm Major Alarm
LVEA & VEAs Clean-10000
  0.3-Microns 14,000 20,000
  0.5-Microns 7,000 10,000

 

Anti & Laser Room Size Minor Alarm Major Alarm
PSL Enclosure Clean-1000
  0.3-Microns 1,400 2,000
  0.5-Microns 700 1,000

The Diode Room and Optics Lab alarm levels will be adjusted when I can get confirmation of their particle level requirements

Comments related to this report
jeffrey.bartlett@LIGO.ORG - 14:30, Wednesday 12 August 2015 (20485)

Set the Diode Room alarms to the same Clean-1000 levels as the PSL.

H1 ISC
kiwamu.izumi@LIGO.ORG - posted 12:20, Wednesday 12 August 2015 - last comment - 11:50, Monday 19 September 2016(20481)
dtt correction filter updated for the control room DARM spectra

I have updated the correction filter for the dtt DARM spectra that are projected in the wall and screens in the control room. I hope Evan will like it.

This does not affect the CAL-CS or GDS calibrations. This is only for display purpose in diaggui.

 


[The DARM spectra]

Here is a comparison of the newly corrected and previous DARM spectra.

As shown in the attached screenshot, the high frequency excess wing that has been shown above 3 kHz is now gone. The noise level at around 7 kHz now seems a bit too low compared with the f-shaped GWINC curve (shown in green). This maybe a real calibration error because I have applied the best and latest knowledge about the sensing chain for correcting the spectrum.

The correction filter is made by a matlab script:

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER8/H1/Scripts/ControlRoomCalib/H1DARM_FOM_correction.m

This script creates a text file which contains the correction filter in [freq, dBmag, phasedeg] format from 1 Hz to 7444 Hz. The text file can be found at

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER8/H1/Scripts/ControlRoomCalib/DARM_FOM_calibration.dat

Both script and text files are checked in to the svn. Also I updated the latest FOM dtt template, H1_DARM_FOM_20150722.xml so that it now uses the new correction filter.

 

[Corrections newly included]

Here is a list of what I newly included.

Note that the previous correction was made only from the IOP down sampling filter and the numerical whitening zpk([1,1,1,1,1], [100,100,100,100,100]) which are still included in the new correction.

Images attached to this report
Comments related to this report
evan.hall@LIGO.ORG - 03:51, Friday 14 August 2015 (20525)

For those of us who like to look at the DCPD streams, I modified the script to also produce the appropriate correction TF (using the uncompensated preamp poles, the AA filter, and the downsampling filters only).

There is still a little bit of sloping at high frequencies, even in the null stream (which we expect to be flat above 100 Hz). Maybe the AA model needs to be tuned?

Non-image files attached to this comment
kiwamu.izumi@LIGO.ORG - 11:50, Monday 19 September 2016 (29795)

I happened to come across this code again to reproduce the calibration correction filter for O1. I then found some bugs in the code which are now fixed.

  • The code was looking at the latest foton filter
    • Now it looks at an archived filter from Sep.10.2015
  • The code was picking up a wrong filter module (FM2 instead of FM1) for whitening filter of DELTAL_EXTERNAL.
    • Now it looks for FM1.
  • Also, the ascii version of the filter (DARM_FOM_calibration.dat) was also dur to the above errors.
    • I reproduced and saved it in svn.
  • The code is saved to SVN.
H1 ISC (DetChar, ISC)
gabriele.vajente@LIGO.ORG - posted 11:52, Wednesday 12 August 2015 - last comment - 10:36, Thursday 13 August 2015(20480)
Scattering noise

I took a look at the scattering noise we see in all lock since last night .

I computed the band-limited RMS between 20 ans 120 Hz, and this is a good indicator of the scattering noise level. Then I looked at correlations with all suspension motions (using M0 and M1 signals, as Keita did for the OMC).

So I'm able to reconstruct the noise variation over time, using a linear combination of all the suspension signals and their squared values. However, I'm not able to pick point one single mirror which is moving more, as shown in the ranking of the most important channels for the BLRMS reconstruction.

I compared the suspension motion spectra from tonight (GPS 1123422219) and few days ago (GPS 1123077617). The most relevant difference is that all test masses YAW motion have now a large bump at 3 Hz. ITMY also has large lines at 0.45 and 0.63 Hz. Finally ETMY pitch shows a large line at 6 Hz and some excess noise above that frequency.

Not sure if all of this is really relevant...

Images attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 10:36, Thursday 13 August 2015 (20509)


		
		
H1 General
jeffrey.bartlett@LIGO.ORG - posted 11:11, Wednesday 12 August 2015 (20478)
End-Y Dust

   Checked the End-Y building and found no apparent problems with doors, doors seals, or penetrations into the building. Given general air quality (high winds, smoke, dust, and no rain) over the past few of weeks, spikes or upward trends in counts would not be surprising.

   At End-Y, there were 17 spikes over 1000 0.5-micron particles over the last 60-days. The 60-day trends do not show any upward slope. Over the last 24 hours, there were seven spikes over 1000 0.5-micron particle. There were seven recorded entries into End-Y during Tuesday’s maintenance window, at least one where the roll-up door open between receiving and the VEA was open. The 0.5-micron plot shows a sharp spike during the maintenance window, which trends upward, then drops. There is a second sharp spike last night/this morning. There is some coloration with the wind speeds during this period. The air filtration system takes several hours to clean the air within the VEA.

   The alarm level were set much tighter during the vent activities when chambers were open. They will be relaxed to the class-10000 levels for O1.       

Several things to keep in mind:

1). The 227b monitor at End-Y is archaic and not entirely dependable. It is slated for replacement.

2). The 227b samples for 20 seconds then multiples the sample by 30 to correct it to the standard 0.1CFM sample rate. Sample times of less than 60 seconds are known to be less reliable.

3). The VEAs are required to maintain a clean-100000 level. The dust monitors are set to alarm at less than a clean-10000 level. The counts recorded are well below these levels.

4). The reference particle size for a class designation is 0.5-microns not 0.3-microns. The 0.3 particle size counts tend to be 10x those of the 0.5 particle.

5). Based upon particle and air temperature trends the internal air filtration system is working properly when the doors are closed.    

End-Y Measured Particle Counts 08:09 through 08:23 08/12/015

Location Particle Size Counts
Particle Counts End-Y
Outside    
  0.3-micron 2,529,000
  0.5-Microns 209,600
  1.0-Microns 38,690
Receiving    
  0.3-Microns 124,880
  0.5-Micorn 9,870
  1.0-Microns 2,330
Change Room    
  0.3-Microns 2,980
  0.5-Microns 1,750
  1.0-Microns 1,940
VEA next to 227b    
  0.3-Microns 1,080
  0.5-Microns 480
  1,0-Microns 200
Non-image files attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 10:13, Wednesday 12 August 2015 (20477)
ADC lists generated after yesterday's model changes

DetChar has requested we generate the front end model ADC channel lists. This is a new RCG feature in which all the ADC channels a model uses are listed along with the name and type of the first connected part in the simulink model. This feature is in the trunk release of RCG and not available in RCG-2.9.6. To generate the lists I performed a "make -i World" in the trunk build area (taking care to save the H1.ipc file before and ensuring it was not modified). I then copied the files from their build area into the location these files will eventually occupy (/opt/rtcds/lho/h1/chans/adc/). I then copied these files over to the exports area of the web server for offsite access:

https://lhocds.ligo-wa.caltech.edu/exports/adc/adclist_11aug2015/

Note that the ADC numbering always starts from zero and is not necessarily the physical card number. For example, h1tcscs.mdl has two adc parts (called ADC0 and ADC1), which are card_num 2 and 3 (the third and fourth ADCs in the chassis). In the h1tcscs_adclist.txt file the cards are referenced as 0 and 1.

H1 ISC (GRD)
evan.hall@LIGO.ORG - posted 07:27, Wednesday 12 August 2015 - last comment - 19:08, Wednesday 12 August 2015(20466)
Some LSC loop checks

Some random commissioning tasks from tonight:

LSC rephasing

In full lock, the phases of POP9 and POP45 were adjusted to minimize the appearance of PRCL in POP9Q and the appearance of SRCL in POP45Q. Then the input matrix element for POP9I→SRCL was tuned to minimize the appearance of a PRCL excitation in the SRCL error signal. New settings attached.

We should take a sensing matrix measurement sometime soon.

LSC OLTFs

I took OLTFs of PRCL, MICH, and SRCL. The data are attached.

[I also did some noise injections into CARM for frequency budgeting purposes. Those are attached too.]

Front-end LSC triggering

Jamie and I started this a few weeks ago, and now it is completed.

There are occasions when the DOWN state of the ISC_LOCK guardian (which, among other things, turns off feedback to the suspensions) is not run immediately after a lockloss (e.g., because the guardian is paused or in manual mode). Therefore, Jamie and I set up the LSC trigger matrix so that PRCL, MICH, SRCL, DARM, and MCL are turned off if POP DC goes below 100 normalized counts. This is set in the DRMI_ON_POP state.

CARM gain redistribution in the Guardian

The state REFL_IN_VACUO now redistributes the CARM gain slightly in order to improve the noise performance of the CARM loop. This state has not been tested and has been left commented out.

Images attached to this report
Non-image files attached to this report
Comments related to this report
evan.hall@LIGO.ORG - 19:08, Wednesday 12 August 2015 (20492)

The CARM gain code was uncommented and seems to work fine.

H1 SUS
vernon.sandberg@LIGO.ORG - posted 17:53, Tuesday 11 August 2015 - last comment - 10:13, Wednesday 12 August 2015(20448)
Replacement of Oscillating SUS-TMSX OSEM Satellite Box

V. Sandberg & R. McCarthy

The SUS-TMSX OSEM Satellite Box was driving a ~2kHz oscillation on its DC power line again. (See https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=20428) This time we replaced the satellite box.  Verified that H1:SUS-TMSX_M1_OSEMINF_RT_IN1 and H1:SUS-TMSX_M1_OSEMINF_SD_IN1 were behaving themselves.  Work was done during the interval 18:00 - 18:45 UTC (11:00am - 11:45am PDT).

Apologie: I failed to follow my own rules and neglected to notify the operator on duty before we left and changed out the satellite box at End X.  Shame and humiliation duly noted.  :-(  Fortunately, no damage was done.

Comments related to this report
andrew.lundgren@LIGO.ORG - 02:01, Wednesday 12 August 2015 (20463)DetChar
After the replacement there was a DC readout lock (Aug 12 4 UTC). The TMSX QPDs look cleaner during this lock, although not quite perfect. The first plot is the spectrum of the TMSX QPD sum at three times - ER7 (blue), Aug 1 (red), and today (green). The second plot is the same for TMSY. It looks like there might still be some excess noise with the same shape but a much smaller amplitude. We'll check this again when there's a nominal lock.

Contrary to what I reported in my previous alog, it is possible to see the 1821 Hz noise in the OSEMs even though they are recorded at 256 Hz. The third plot shows the OSEM RT signal (and SD is the same) at four different times. Blue is the time when Matt and Evan alogged the problem. Red is the lock after the amplifier had been smacked and the audible oscillation was gone. Green is Aug 8 (a week later), and yellow is today. The noise is gone today after the amplifier was replaced.

The confusing thing is that the red trace is the same time as the plot in my alog. It seems that the high frequency noise in the amplifier was not there, but there was still just as much excess noise on the TMSX QPD. That might mean that the decrease in the excess noise today is just coincidental. We'll have to keep monitoring this.
Images attached to this comment
andrew.lundgren@LIGO.ORG - 07:11, Wednesday 12 August 2015 (20468)DetChar
There's a later DC lock today that seems to have a more nominal configuration. The noise in the TMSX QPD is like it was before the amplifier swap, so it hasn't been fixed and must be unrelated to the satellite amplifier problem. The previous lock, where the noise looked smaller, had a factor of ten lower DC level than the current value. The second plot is just a quick check showing that the high frequency line from the satellite amplifier is still gone since the swap.
Images attached to this comment
vernon.sandberg@LIGO.ORG - 10:13, Wednesday 12 August 2015 (20476)

The S number and S/N for the "old" satellite box for SUS-TMSX_M1_OSEM SD and  RT channels are:

S-number bar code  S1100176

S/N  3202-022

The identifiers for the new box will be obtained at a future oportunistic entry to the EX electronics racks area.

H1 PSL (PSL)
gabriele.vajente@LIGO.ORG - posted 20:15, Monday 10 August 2015 - last comment - 11:58, Wednesday 12 August 2015(20394)
ISS is re-injecting the jitter peaks

Sudarshan, Gabriele

Today when the IFO was locked in low noise, we could see the 300-400 Hz peaks in the sensitivity again. Those are know to be due beam jitter caused by the PSL periscope.

We looked into the performance of the ISS second loop, which was closed at the time. The first strange thing that we can't understand is that both the in-loop and out-of-loop signals are dominated by a flat  noise background at a level of 1e-8 /rHz. We would expect the out-of-loop to be at this level, but the in-loop should be squeezed down by the larg loop gain.

We changed the ISS gain, anmd discovered that if we turned down the gain at the minimun (-26-20 dB with respect to nominal), and disengage both boost and integrator, the in and out of loop ISS signals got worse, but the sensitivity got better at the 300-400 Hz peaks. So out guess is that beam jitter on the ISS array is actually causing an excess sensing noise on the ISS diodes, which is then re-injected into the laser beam as real intensity noise when the ISS second loop is closed with high gain.

We tried briefly to optimize the centering of the beam on the ISS array. We couldn't find any spot better than the initial one in terms of power on the diodes. One possible strategy would be to move the beam into the ISS array with the IFO in full lock, and use the coupling of a PZT jitter line to the sensitivity as a figure of merit.

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 20:58, Monday 10 August 2015 (20397)

I saw the same strange noise floor that were at a level of 1x10-8 RIN/sqrtHz a week ago. See the spectra from this alog:

https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=20148

sudarshan.karki@LIGO.ORG - 11:58, Wednesday 12 August 2015 (20482)

In light of this strange noise floor, I looked at some of the RIN spectrum I had taken peridocally. This change in In-loop PD noise seem to have happened after we started using the actual DC signal from the PD to obtain RIN ( we were using a hardcoded offset that was an estimate of DC signal). So we implemented a low pass filter with two poles at 10 mHz  in the DC signal this morning and the problem is gone. Attached is a spectrum before and after the low pass filter was implemented.  SDF table is appropriately updated.

Images attached to this comment
Displaying reports 65621-65640 of 85578.Go to page Start 3278 3279 3280 3281 3282 3283 3284 3285 3286 End