Mon Jun 09 10:09:58 2025 INFO: Fill completed in 9min 55secs
Travis confirmed a good fill curbside.
FAMIS 31089
PMC REFL and RefCav TPD both had a bit of a drop-off over the past 3-4 days, but PMC TRANS is largely unchanged. We've seen correlation before with changing outdoor temperatures and RefCav alignment, so I suppose it's possible the recent heat wave is having an effect on the PMC as well.
The PID controller for EY CP7 cryopump has opened its liquid level control valve (LLCV) to 100% but this is proving to be insufficient to maintain a 92% fill level in the pump. The pump level started deviating from nominal around 4pm Friday 06jun2025 PDT.
Vacuum team is investigating.
Note: no alarms issued at this point, the cyropump alarm level is < 80%, CP7 currently is at 85%.
I had noted the LLCV issue for CP7 here. Tomorrow we'll attempt to re-zero the LLCV actuator.
TJ, Camilla WP 12605, WP 12593
As HAM1 is now low enough in pressure, we removed the HAM1 +Y yellow VP covers, re-attached bellows and removed guillotines. ISCT1 was moved back into place on Friday 84850.
After TJ transitioned to Laser Hazard, we opened the ALS and PSL light pipes and turned back on the SQZ, CO2s and HWSs lasers.
After okay from VAC, IMC locked without issue, PRC and MICH alignments look good enough on AS AIR camera and we will begin realigning ISCT1 soon.
TITLE: 06/09 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 4mph Gusts, 2mph 3min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.12 μm/s
QUICK SUMMARY:
We relocked the IMC at 15:34, and Gate Valves 1 & 2 are now open.
Sun Jun 08 10:08:17 2025 INFO: Fill completed in 8min 13secs
Sat Jun 07 10:05:59 2025 INFO: Fill completed in 5min 55secs
Erik, Dave:
Web ndscope trends stopped updating around 09:45 this morning. Some nomachine sessions are having issues.
We are investigating.
I'm running the web ndscope trends on cdsws33 while opslogin0 is having issues.
Opslogin0 is reporting an ldap failure.
I rebooted opslogin0 about 12:10 PST and it's seems to be working normally.
Gerardo, Travis, Randy, Janos Today we executed WP 12595, reventing, troubleshooting, closing, and re-pumping HAM1. The times of the operations are approximate. 08:00 - 08:30: GV1 and GV2 large gate valves have been closed. HAM1 turbo and Ion pump (IP13) have been valved out. The annulus IPs have been switched off, the annulus volume has been vented. Also, IP3 was valved in to the corner, after the successful troubleshooting yesterday (see aLog 84826). 08:40 - 09:53: The chamber was vented with bottled Nitrogen. Approximately 2 full bottles were needed. Only metallic parts were used between the venting port and the bottle (no plastic tubes), and the tube was flushed with Nitrogen beforehand. 10:00 - 10:40: The Y+ door has been taken off. The O-rings have been inspected, then removed. A human hair was found at 6:30 o'clock, and a little piece of white tape was found in the O-ring's groove at 8:00 o'clock, and overall, the O-rings and grooves were substantially dirty. Pictures about these in the comments. Also, the cleanroom was needed to be moved Y+ direction, to give enough room for the forklift. 10:40 - 11:40: Chamber work. During this time, Randy stayed with the forklift, so the door was hanging from it all the time, and inside the cleanroom. 11:45 - 13:30: O-ring installation: it took a few tries, but both O-rings were installed nicely. Then, installing back the door: smooth process, with shoving it in as the last step, to avoid the O-rings falling out. 13:20 - 13:50: Pumping of the annulus systems, and roughing the chamber. The HAM1 side annulus needed ~10 mins to go below the pressure it reached in the last 2 weeks, so the troubleshooting was very successful! The roughdown goes well, too. 16:15 - 16:45: Transitioning to turbo
Because of the safety concerns of the high amount of Nitrogen in the chamber after venting, and then opening the door, Oxygen sensors have been used during the door opening process. Moreover, because of these same reasons, and because of the lack of purge air, a crossflow from a HEPA-filtered fan was also used (additionally to the cleanroom's HEPA fans), when the chamber was open.
Left HAM1 with SS500 cart pumping, it was able to reach 5.0X10-05 Torr, at such pressure the cart system will allow the interlock to be engaged, this interlock is there to protect the chamber in case of an anomaly (power interruption, component failure, etc.) and its function is to close two valves on the cart. HAM1 pressure when I left the site was a bit higher than the cart, 6.1X10-05 Torr. Pumpdown appears nominal.
The annulus for HAM1 and HAM2 are doing good. HAM1 was at 2.66x10-05 Torr, and the pressure on HAM2 annulus was at 7.57x10-06 Torr.
Plot comparing the current vs the previous pumpdown. Blue trace is the current pumpdown.
WP12596 ADF VCXO IPC between IOP-LSC0 and OMCPI
Daniel, Vicky, Kevin, EJ, Erik, Dave
New h1ioplsc0 and h1omcpi models were installed to add two IPCs at 64kHz rate between lsc (sender) and omcpi (receiver)
H1:FEC-7_IPC_IOP_PCI_ADF_VCXO_COS
H1:FEC-7_IPC_IOP_PCI_ADF_VCXO_SIN
After the new models were restarted we noticed that the new receivers were flashing with a single RX error (1 out of 65536 in that second) every few seconds, randomized between the two new channels.
At this point we upgraded h1omc0 from RCG-5.3.0 to RCG-5.5.0, the new system has code to correct for such receive errors. Erik had the new 5.5.0 boot server ready to go, it is h1vmboot5-5 (a new naming scheme).
This required the rebuild and restart of all the models on this frontend (h1iopomc0, h1omc, h1omcpi). The new RCG added slow channels, so a DAQ restart was required for all these models.
The builds on RCG-5.5.0 use EJ's new rev-lock system, whereby all the source code (mdl and C) must be committed to subversion, and only the locked revision of these files are used to build the model, not what happens to be in the userapps models directory at the time. This worked like a charm.
Later Vicky added some filtermodules to h1omcpi, which required a second restart of h1omcpi and the DAQ.
WP12597 Quarterly reboot h1guardian0
TJ
TJ rebooted the guardian machine, it came back with no problems.
WP12603 HEPI HAM1 model cleanup and removal of TT-L4C IPC senders, switch SEI-PROC to ISI HAM1 GS13 IPC
Jim, Erik, Dave:
New h1hpiham1 and h1seiproc models were installed with no problems. DAQ was restarted, both models had new INI files.
Move HWS data from h1hwsmsr local RAID to /ligo NFS
Erik, Dave:
Erik reconfigured the HWS machines to mount /ligo/data/hws as /data locally. Some files needed to be copied to the new location before the HWS cameras could operate and write current files to the new location. Archived data is being transferred to preserve lookbacks.
DAQ Restarts (plural)
Erik, Dave
Restart 1
11:02 DAQ was restarted for two changes, first the expected changes to h1ioplsc0 and h1omcpi. Second for the unexpected upgrade of h1omc0 to RCG-5.5.0 which added channels to h1iopomc0 and h1omc, and additional channels to h1omcpi.
Clean restart except GDS1 needed a second restart
Restart 2
13:52 DAQ was restarted for the h1hpiham1 and h1seiproc model changes.
Clean restart except GDS1 needed a second restart
Restart 3
14:33 DAQ was restarted for additional change to h1omcpi.
Clean restart except, you guessed it, GDS1 needed a second restart
Note that the Vacuum upgrade which would have required a DAQ restart did not go ahead today.
CDS Overview is showing h1omc0's new RCG-5.5.0 as a dark green tag. Reminder that h1susex is at RCG-5.3.0 (LIGO-DAC) and everything else is RCG-5.1.4
h1omc0 models' rev_lock revisions summary (recent changes in bold):
h1iopomc0:
22838 cds/common/src/MAX_MIN_CALC.c
24547 cds/common/src/STAT_CALC.c
30625 cds/h1/models/h1iopomc0.mdl
27113 isc/common/models/LSC_TRIGGER.mdl
h1omc:
26004 cal/common/models/CAL_OSC_MASTER.mdl
6334 cds/common/models/FILTBANK_MASK.mdl
6334 cds/common/models/SCHMITTTRIGGER.mdl
18154 cds/common/models/TRIG_RAMP.mdl
18648 cds/common/src/RAMP_VALUE.c
4225 cds/common/src/wait.c
22840 isc/common/models/FILTBANK_TRIGGER.mdl
27113 isc/common/models/LSC_TRIGGER.mdl
18148 lsc/common/models/lsc.mdl
30547 omc/common/models/omc.mdl
30548 omc/h1/models/h1omc.mdl
h1omcpi:
24137 cds/common/src/BAND_SELECTION.c
24174 cds/common/src/FIXED_PHASE_OSC_WITH_CONTROL.c
24176 cds/common/src/LOW_FREQ_DEMUX.c
24176 cds/common/src/LOW_FREQ_MUX.c
31648 omc/h1/models/h1omcpi.mdl
31645 sqz/common/models/FC_LIB.mdl
26600 sus/common/models/PI_MASTER_V2.mdl
Randy, TJ, Fil, Richard, Camilla, WPs: 12593, 12594
After Fil and Richard un-cabled ISCT1, TJ and I added guillotines to HAM1 +Y VPs, removed bellows and attached yellow VP covers. Randy then fork-lifted ISCT1 away from HAM1 and placed it next to IOT2L.
At the end of the day on Friday, Randy fork-lifted ISCT1 back to HAM1. Location and height was adjusted to match markings. We left the yellow VP covers on and bellows off as HAM1 VAC had started pumping.
Today at ~11:10 am, IP3 (LVEA 2x 1250 l/s IP) failed leading to a slight pressure rise in the corner. Seen on all LVEA pressure gauges (BSC2, HAM6, etc.). Attached snapshot shows the two pumps behavior at the time of failure, pump ion current on left column, output voltage on right.
Gerardo and I went to the mechanical room to diagnose the controller, upon arrival both channels on the Dual Vac controller had tripped off. We reset the high voltage and powered on the IP. Channel 1 (corresponding to Pump A in CDS) would only reach 300 V and then would trip the controller. Channel 2 (Pump B would reach 7kV no issue but would trip when Channel 1 tripped). We then swapped to a Gamma MPC controller to test if the controller was the problem. We saw the same exact behavior where Pump A would only reach 300V and Pump B was ok at 7kV.
We then swapped back to the DualVac controller after confirming it was not the issue, but with only Pump B connected and powered on. We are continuing to monitor to make sure Pump B remains steady. In parallel, Travis and Mark are testing the HV cable to the pump to see if there are any shorts. Results to follow.
If the cable is not the culprit we will HiPot the pump to see if there are any shorts, pump will be isolated from main volume during this test. To be continued.
Using the Agilent FieldFox N9912A we measured the cable length with the following parameters:
Mode = DTF (VSWR), F_Start = 2.0MHz, F_Stop = 450.0MHz, Points 801, Resolution 261mm
Start Distance 5.0m, Stop Distance 170.0m
The cable end was detected at 80.79m from the start of the launch cable, launch cable was about 1m.
The cable measurement was similar to the other cables measured.
When measured with the DMM, this cable measured open, but when the connector was moved, it displayed ~6M ohms, this would normally not be an issue, but due to the high voltage in this location it may be an issue, recommend hipot.
M. Pirello, F. Clara, G. Morenao, T. Sadecki, J. Vanosky
Problem solved, see in aLog 84826
Daniel, Kevin, (Erik, Dave), Vicky
Kevin is interested in taking ADF sweeps around the second higher order mode arm resonance near 10.4 kHz.
We (Daniel) made model changes in h1ioplsc0 and h1omcpi that should enable higher freq ADF demods. Daniel built the models successfully. Then Dave also built these models successfully for the custom RCG running on h1omcpi for the variable duotone. We put in a WP to boot in these changes to h1ioplsc0 and h1omcpi next Tuesday 6/10.
From h1ioplsc0, we added 2 new PCIe channels to write the digital ADF LO oscillator out to PCIe. These are H1:IOP-PCI_ADF_VCXO_{COS,SIN} in screenshot 1. We also changed a rogue limiter such that the ADF should be able to scan +/-1 MHz according to the PLL.
In h1omcpi, we use the fast 64 kHz OMC PI user model to do a fast digital demodulation of the OMC_DCPD_SUM signal using the above ADF-LO PCIe channels. This is beating the fast dcpd output signal (H1:OMC-PI_DCPD_64KHZ_AHF), against the ADF LO's cosine and sine components from PCIe (H1:IOP-PCI_ADF_VCXO_{COS,SIN}), to get the demodulated ADF I/Q signals. The demod signals will have names like H1:OMC-PI_ADF_DEMOD_{I,Q}. Screenshots 2-4.
Operationally nothing should change for the ADF or for OMC-based fast PI damping. We are just using PCI to send 2 IPC channels from an IOP model to a user model on a different computer at ~65kHz, and doing the demod in a 64kHz user model (but the demod does not go anywhere, it is just used for squeezer diagnostics).
Model change summary:
h1ioplsc0
+2 IPC senders to H1.ipc
+2 slow channels to DAQ
h1omcpi
(+2 IPC receivers)
+33 slow channels to DAQ
Model changes today seem relatively successful. No way to verify in hardware (yet), but the ADF PLL claims it successfully locks at -11kHz from carrier, so at least the limiter fix in h1ioplsc0 seems to work; in principle the ADF sweep limiter is now set at +/- 1 MHz.
I added a quick ADF demod of the 64kHz OMC DCPD SUM channel into the normal ADF screen, screenshot attached, see the very bottom. New filter banks initialized with values and filters as in the other adf demod chains. Also tried to clean it up in SDF: channels are initialized, gains and phase are monitored, and saved into sdf (h1omcpi model).