Displaying reports 66961-66980 of 85902.Go to page Start 3345 3346 3347 3348 3349 3350 3351 3352 3353 End
Reports until 16:52, Friday 03 July 2015
H1 ISC
jenne.driggers@LIGO.ORG - posted 16:52, Friday 03 July 2015 (19442)
MICH freeze attempt - inconclusive
Nic, Evan, Jenne

We tried looking at the efficacy of MICH freeze with DRMI today.   

First, we looked at the MICH fringe velocity in Michelson-only: With the MICH freeze engaged, the fringe velocity seems to slow down by a factor of about 2 versus without the freeze.  

Then we aligned the DRMI and tried to get some locking statistics (length of time waiting for lock) with the freeze on vs. off, but we aren't really getting any locks at all.  We waited more than 15 minutes without a lock with the freeze off, so we went to trying with the freeze engaged.  With MICH freeze engaged, we had 2 wait times of 2 or 3 minutes, but all other times have been more than 15 minutes.  (We tried changing trigger threshold settings a few times, which is what defined the ends of these 15 minute wait stretches).

So far, it's not clear to us whether the MICH freeze is having a significant effect at all.  We think we'll try again later.
H1 ISC
evan.hall@LIGO.ORG - posted 16:27, Friday 03 July 2015 (19441)
DRMI locking partially recovered

Nic, Jenne, Evan

In spite of the bad POPAIR situation, we were able to get DRMI to lock by increasing the whitening gain of POP18 (from 12 dB to 45 dB), and by lowering the trigger thresholds for MICH by a factor of 10, and SRCL by a factor of 5.

After DRMI locked, we were able to optimize the buildups of POP18 and AS90, mostly by moving PR3 positive in pitch, and then compensating by moving PR2 negative in yaw. In this way we increased the buildup of POP18 by a factor of 45. (We then undid the extra analog whitening gain.) So this seems to support the idea that our issues are caused (at least partially) by misalignment of the power recycling cavity.

We went onto the table and again tried to resteer onto POPAIR_B, but we got only 10 % more power on the PD. POP18 is now at about 6 ct (normalized), whereas we expect about 300 ct for DRMI without arms. So there is still a missing factor of 50 somewhere.

We measured the OLTFs of PRCL, MICH, and SRCL, and they seem fine. So the DRMI LSC seems healthy as far as we can tell; there's just some problem with the POP path.

H1 COC (ISC)
nicolas.smith@LIGO.ORG - posted 14:36, Friday 03 July 2015 - last comment - 15:28, Monday 27 July 2015(19440)
BS Butterfly Ringdown measurement

(evan jenne nic)

Evan said that the Q of the BS butterfly hadn’t yet been measured.

We let the system alone for 10 minutes in DRMI and analyzed the ringdown. The biggest SNR was in the PRCL error signal.

The resonance frequency is 2449Hz, the Q factor is (5.6 pm 0.2) 	imes 10^{6}. This means a time constant of 12 minutes.

Ringdown with fit is atached.

Non-image files attached to this report
Comments related to this report
nicolas.smith@LIGO.ORG - 15:28, Monday 27 July 2015 (19969)

(script attached)

Non-image files attached to this comment
H1 DAQ (DAQ)
stefan.countryman@LIGO.ORG - posted 12:32, Friday 03 July 2015 (19439)
Timing System Installation Diagram v. 1 on DCC
I've been working on a diagram of the timing system with specific locations and uses of every timing system component. I've put it up on DCC as https://dcc.ligo.org/LIGO-D1500201. I'd appreciate feedback regarding what else would be useful (while bearing in mind that I'd like this to stay clear and simple). I'm also happy to make extra pages if someone needs a detailed view of some particular element of the system that wouldn't fit into the overview shown.

Happy 4th!
Non-image files attached to this report
H1 ISC
evan.hall@LIGO.ORG - posted 23:57, Thursday 02 July 2015 (19436)
Initial alignment recovery progress

Jenne, Evan

Tonight we ran through initial alignment of the corner in order to get back to DRMI locking (without arms). This is a bit different than usual, since we cannot use the X arm as a reference for PR3 (via the COMM beat note) or for IM4 and PR2 (via the IR input pointing).

Initially we just restored the suspension slider values for PRM, PR2, PR3, SR2, SR3, SRM, and IM4 to their pre-vent values. Then we tried locking PRX, but found that we could not get decent fringing, even by moving the PRM a milliradian or so in pitch and yaw. So we instead restored the pitch and yaw values of PR3 as seen by the oplev, and then optimized the fringing using PR2 and PRM. This allowed us to lock PRX with a decent amount of light at the AS port (>1000 ct on ASAIR_LF with 10 W PSL power), and we then engaged the PRX WFS loops as usual.

We were able to lock the dark Michelson and optimize the BS angle without issue.

We were able to lock SRY using a similar philosophy as PRX: we restored the SR3 alignment to its pre-vent values as seen by the oplev. After that, we saw fringing and were able to optimize it by moving SRM. Then we were able to lock SRY and engage the SRC alignment loops without issue.

Then we tried moving on to DRMI locking. The flashes at the AS port look more or less normal, but there seems to be very little going on in POP18. We saw some flashing with peaks < 1 ct, but for normal DRMI acquisition we expect hundreds of counts. We went onto ISCT1, and found that the beams were low (by about 1 mm) on both the POPAIR diodes. We adjusted the spot on POPAIR_B so that it is more centered, but this only brough the flashing up to at most 15 ct (at 20 W PSL power).

Since it seemed that the POP path was somehow not aligned, we then relocked PRX (with WFS loops feeding back to PRM) and moved IM4 while watching the spots on the POP QPDs. We centered the spots as best we could, then ran through the other initial alignment steps again, and then tried locking DRMI. However, this did not improve the flashing in POP18. For the time being, we reverted our change to the POPAIR_B pointing on ISCT1.

JCD:  Since we were able to lock the individual pieces of the vertex, it seems like the vertex optics are all aligned to one another, however I am less confident in the overall alignment (e.g., the input pointing doesn't match what the Xarm will want, so we're not hitting all of our PDs simultaneously).  I am not quite sure how to resolve this, with the sensors that we have at our disposal right now.

H1 PSL (ISC, PSL)
sudarshan.karki@LIGO.ORG - posted 20:40, Thursday 02 July 2015 - last comment - 11:50, Monday 25 January 2016(19434)
ISS Loop Transfer Function Measurement

Inner Loop

ISS Inner Loop has  UGF of 22 KHz with a phase margin of about 50 degress. This was measured with variable gain set at 6 dB for the best phase margin. This is the normal operation settings for Inner Loop.

Outer Loop

Outer Loop has a UGF of 1 KHz ( designed for 4 KHz) with a phase margin of  about 30 degrees. The variable gain was set at 40 dB (max available) and an additional gain stage(?) was switched on as well.

Also tried moving the the Inner loop gain to see if it shows any improvement on the outer loop but no luck.

TF Plots are attached.

Images attached to this report
Comments related to this report
sudarshan.karki@LIGO.ORG - 18:18, Monday 06 July 2015 (19460)

These transfer function measurements were taken at ~10 W of PSL power.

sudarshan.karki@LIGO.ORG - 11:50, Monday 25 January 2016 (25141)

The data used for the plot above is attached.

Non-image files attached to this comment
H1 CAL (CAL, DetChar, INJ)
peter.shawhan@LIGO.ORG - posted 19:53, Thursday 02 July 2015 - last comment - 10:12, Friday 03 July 2015(19435)
Sign flip seen in ER7 hardware injections
I selected a loud sine-gaussian injection done during ER7 at each site and compared h(t) data from the HOFT frames against the intended strain waveform; this is documented at https://wiki.ligo.org/Main/HWInjER7#Injection_sign_check .  It looks like the injection came through with the wrong sign at LHO, but with the right sign at LLO.  This should be investigated.
Comments related to this report
peter.shawhan@LIGO.ORG - 10:12, Friday 03 July 2015 (19438)CAL, INJ
I double-checked the PCAL-based strain sign checks at both LHO and LLO and they seem sound.  So the sign flip seen for the ER7 hardware injections would have to be something related to the actuation path used for the hardware injections, e.g. the inverse actuation filter bank or perhaps the location of a minus sign in the main feedback loop (in the output matrix or elsewhere?).
H1 ISC
nicolas.smith@LIGO.ORG - posted 16:38, Thursday 02 July 2015 - last comment - 20:22, Friday 10 July 2015(19432)
Couldn't make fringe wrapping with bright michelson

(evan jenne nic)

We wanted to investigate the OMC alignment/backscatter problem. Driving the OMC SUS in Yaw has been known to cause backscatter noise due to the modulation of the optical path length when the OMC moves in Yaw.

Our procedure was to lock the vertex optics in a bright michelson configuration (a state has been added to the IFO_ALIGN guardian to make this easy). Then we wanted to drive the OMC in yaw and choose the yaw->longitudinal matrix element such that the center of rotation would be about the input beam, rather than the omc center of mass. This would be determined by minimizing the scatter as measured by either OMC trans, or the MICH error point.

I was surprised that we were not able to induce any significant backscatter fringe wrapping noise in this configuration. We drove the OMC SUS in longitude up to the point that the beam was misaligning enough to noticibly affect the OMC DC trans.

We also drove the ISI table directly by putting a 1Hz 1mm injection into the Y isolation loop error point. 

Driving the path length, we both listened and had a live spectrum running. We saw no evidence of scatter in either OMC trans or MICH_IN1.

We will need to think if there is another configuration (available to use without arms) that will be more sensitive to backscatter.

Comments related to this report
nicolas.smith@LIGO.ORG - 16:52, Thursday 02 July 2015 (19433)

I forgot to mention that we turned on the AS fast shutter and OMC pzt high voltage supplies for HAM6.

nicolas.smith@LIGO.ORG - 20:22, Friday 10 July 2015 (19563)

This measurement didn't work because I was wrong about the calibration. The isolation loop error points are in nanometers, not micrometers. So we were moving the table 1000 times less than I thought.

H1 General
nutsinee.kijbunchoo@LIGO.ORG - posted 16:21, Thursday 02 July 2015 (19431)
Ops Summary

8:52 Sudarshan work on ISS

         Fil to EX (run cables)

9:16 Jordan to HAM6 checking on PEM sensors

9:44 Kyle to HAM6

10:01 Kyle grab He bottle at EY

10:15 Leo - EX ESD measurement

10:20 Kyle out of EY

11:25 Dave and Stefan to EY - Hook up GPS receiver

11:32 Fil & Andrea back from EY

11:50 Jordan to LVEA testing PEM accelerometer

12:04 Leo done with EX ESD measurement. Starting EY ESD measurement.

12:47 Jordan back

13:00 Kyle to EX

12:42 ISI HAM5 tripped

           -No correlation with PEM acc. sensors. Not sure what happened. Coincides with sharp peak on 1-3 Hz ground motion plot. Suspect electronics issue.

13:42 Fil and Stefan to EX by the rack

13:53 Jordan to EX

14:12 Leo finished EY ESD measurement

14:24 Rick and Sudarshan to LVEA - ISS work by PSL enc.

          Kyle back

14:37 Jordan to EY

15:00 Fil & Stefan back

15:10 Rick & Sudarshan back

15:19 Sudarshan & Jordan to IOT2R table

15:27 Kyle to HAM6

15:41 Kyle back. Shutdown HAM5, HAM6 annulus pump.

 

Other notes:

- Jordan discovered magnetometer at EY rack output straight line since June 22nd.

- Laser Hazard notifications added to Ops Overview.

 

HAPPY HOLIDAY!!!!

LHO VE
kyle.ryan@LIGO.ORG - posted 16:17, Thursday 02 July 2015 (19430)
~1530 hrs. local -> Isolated HAM5 and HAM6 annulus pump carts and shut them down


			
			
H1 SUS (SYS)
leonid.prokhorov@LIGO.ORG - posted 15:39, Thursday 02 July 2015 (19429)
OPLEV Charge measurements
More charge measurements. Plots are attached.
Images attached to this report
H1 DAQ (DAQ)
stefan.countryman@LIGO.ORG - posted 15:05, Thursday 02 July 2015 (19425)
Calibrated Cesium Standard CsIII Atomic Clock in MSR to Minimize Long-Term Linear Drift
Jim, Daniel

I calibrated the frequency of the CsIII Cesium Standard Atomic Clock in MSR by setting the Output Frequency on to 1 + (1010 * 10^-15). I did this at 12:26 PDT on July 2, 2015. I used the serial port on the back of the clock and the Monitor2 software installed on the IBM Thinkpad in MSR.

The previous setting was +1078 (parts per 10^15); I calculated the proper value to be 1010 (details below). I used the time difference between the atomic clock's 1PPS signals and our timing system's 1PPS signals. I took a linear regression to determine the rate of linear drift of the CsIII, since an improperly calibrated Cesium clock shows very little short-term jitter but does show long-term linear drift. Our timing system is long-term accurate since it's GPS backed.

I interpreted the meaning of the time difference according to an explanation from Daniel: a negative time difference means that the external (i.e. Cesium) 1PPS signal is ahead, while a positive time difference means it's behind. I calculated the drift rate to be approximately -6.8e-14 corresponding to -175 ns per 30-days.; the CsIII 1PPS is getting earlier and earlier relative to the timing system 1PPS, suggesting that the period is a bit short. I calculated the proper setting for the CsIII, as explained below, and talked to Microsemi's customer support to confirm that the CsIII's output frequency setting is what one would expect (the instructions were vague). He agreed with my interpretation but didn't have deep knowledge of the system. So we'll have to keep an eye out and check the drift rate in a few weeks to see if it is indeed close to zero (I expect it
will be fine).

The offset to be entered was calculated as follows. The current 1PPS period is given by tau_{measured} - 1 = m_{drift} where m_{drift} is the drift rate. If the clock's "natural" frequency is f_{natural}, then the current offset (Delta{f})_{measured} (which I found to be 1078 * 10^{-15} according to the Monitor2 interface) relates to the current measured period as (1 + (Delta{f})_{measured})f_{natural} = f_{measured} = frac{1}{tau_{measured}}. We want f_{calibrated} = 1, so we need to set (1 + (Delta{f})_{calibrated})f_{natural} = 1 = (1 + (Delta{f})_{measured})f_{natural}tau_{measured}, which simplifies to 1 + (Delta{f})_{calibrated} = (1 + (Delta{f})_{measured})tau_{measured}, giving (Delta{f})_{calibrated} = (1 + (Delta{f})_{measured})tau_{measured} - 1 = (tau_{measured} - 1) + (Delta{f})_{measured}tau_{measured}, or, in terms of the drift rate, (Delta{f})_{calibrated} = m_{drift} + (Delta{f})_{measured}(1 + m_{drift}), which we then divide by 10^{-15} in order to get the value to be entered into the Monitor2 software. See the attached code (written in Julia) for the precise method used. The output (besides the graphs, also attached) includes the above calculations:

Drift rate in Time Code Generator in MSR: -6.786047272268983e-14
Drift rate in Time Code Translator in EX: -6.788794482601771e-14
Drift rate in Time Code Translator in EY: -6.780539996010063e-14
mean_drift_rate = mean($(Expr(:comprehension, :(channel.drift_rate), :(channel = channels)))) => -6.785127250293606e-14
delta_f_measured = 1.078e-12 => 1.078e-12
delta_f_calibrated = mean_drift_rate + delta_f_measured * (1 + mean_drift_rate) => 1.0101487274969909e-12
what_to_enter_into_monitor2 = delta_f_calibrated / 1.0e-15 => 1010.1487274969908

Note that the drift rates calculated for each channel are in very close agreement. I used this to pick the new setting of 1010 for the Cesium clock. The channels I looked at were:

H1:SYS-TIMING_C_MA_A_PORT_2_SLAVE_CFC_TIMEDIFF_1 (Time Code Generator in MSR)
H1:SYS-TIMING_X_FO_A_PORT_9_SLAVE_CFC_TIMEDIFF_1 (Time Code Translator in EX)
H1:SYS-TIMING_Y_FO_A_PORT_9_SLAVE_CFC_TIMEDIFF_1 (Time Code Translator in EY)

The timeseries (with linear regressions) and histograms are included. The period analyzed is from the beginning of ER7 until last week. I had to remove a chunk of spurious data from the EY channel since Dave and I were using the BNC cable for a different test and polluted the EY channel during that time.

The linear regression shows a very clear and steady trend, as expected. Standard deviation of the atomic clock offset is reduced by a factor of five once linear drift is accounted for, and the histogram more closely follows a Gaussian distribution:

std(z) => 1.1845358220722136e-8    # Jitter about linear trend
std(y) => 5.853090568967089e-8      # Absolute time difference
Images attached to this report
Non-image files attached to this report
H1 DAQ (DAQ)
stefan.countryman@LIGO.ORG - posted 15:00, Thursday 02 July 2015 (19426)
Tested and Installed New CNS II GPS Clocks in EX and EY
Filiberto, Jim, Dave

Installed one new CNS Clock II in each end station in receiving, directly above the Timing IRIG-B Module in each rack. We tested each unit to make sure it successfully locks to a GPS signal.

Filiberto ran an extension cable for the GPS antenna from VEA into receiving in both end stations in order to provide signal for the new CNS II as well as a BNC cable from receiving to CER to deliver the CNS II 1PPS signal to the third time difference port of the Timing Comparator located there. The cable lengths for the 1PPS signals are the same to within an inch.
LHO VE
kyle.ryan@LIGO.ORG - posted 14:58, Thursday 02 July 2015 (19428)
One more joint leak tested at X-end -> climbed on tube near BSC9
Upon completing the leak testing of the Y-end yesterday, Gerardo and I realized that we had failed to test the 4.5" blank which had replaced the 1st generation ESD gauge on BSC9 at the X-end.  So... 

~1345 - 1415 hrs. local 

Replaced X-end turbo scroll baking pump with LD and sprayed 20 seconds of audible helium flow at untested CFF -> OK -> Restored pump configuration to as found -> Shut down LD 

LHO VE
bubba.gateley@LIGO.ORG - posted 13:37, Thursday 02 July 2015 (19427)
Beam Tube Washing
Scott L. Ed P.

6/30/15 
Cleaned 62 meters ending 1.5 meters north of HNW-4-097.
 
7/1/15 
Cleaned 47.4 meters ending 13 meters north of HNW-4-099.

7/2/15 
Finished crevice cleaning and testing. Results posted here.


X-ARM IS COMPLETE.
Non-image files attached to this report
H1 DetChar (DetChar)
thomas.dent@LIGO.ORG - posted 11:31, Thursday 02 July 2015 (19423)
LHO DQ Shift: 12-15 Jun

This is the (delayed due to conference travel) summary of my DQ shift for 12-15 June mentored by Andy L, the detailed report is at https://wiki.ligo.org/viewauth/DetChar/DataQuality/DQShiftLHO20150612

Friday 12th: No science data

 

Saturday 13th: Few minutes of science data

Several blip glitches occurred during that time, eg this one

DAC glitches were noticed in LSC-MCL_IN1, these seem to come from SUS-MC2_M3 as shown by the lineup plot :

MC2_glitching002.png
 

 

Sunday 14th: Several hours lock, several hours of science data

Much of the data was clean with good sensitivity - but there were significant bad periods of excess broadband and/or low frequency noise during hour 8 UTC (spectrogram link) and hour 10 UTC (link) which killed inspiral range.  The excess noise appears also in a large number of ASC and LSC channels but no clear cause.  Does anyone have an idea what could have been going on at these times?

One curious feature was a band of excess noise over 60-75 Hz appearing at almost exactly 12:00 and gradually moving up and down in frequency until hour 17 UTC when it suddenly disappeared.  See spectrogram link for an overview.  Again this is mysterious and any clues from on side would be welcome.

It is clear from the spectrograms and omicron triggers that there were a few (about 1 per hour) really loud glitches - SNR > 1000.  It has been suggested these are related to beam tube cleaning and its aftermath (e.g. particles loosened & falling through the beam).  See omega scan of the loudest

Another feature seen previously is a set of near-periodic glitches at 60Hz, approximately every 72 minutes.  These are associated with activity in the SUS-ETMY_L2 auxiliary channel, 'something happening' at the Y arm end station with ~4300s period.

 

Monday 15th: Tiny amount of data

Images attached to this report
LHO General
richard.oram@LIGO.ORG - posted 15:34, Wednesday 01 July 2015 - last comment - 10:51, Thursday 02 July 2015(19415)
LLO PSL HPO diode power supplies turned off and locked out

Danny Sellers, LLO Laser Safety Officer, unplugged and locked out the PSL HPO diode box power supplies to prevent the accidental activation of the High Power Oscillator. 

See photo. The appropriate laser safety documentation concerning the operation of the PSL is being reviewed and updated to reflect the implementation of this new administrative/ engineering control

Note: Having the HPO diode box power supplies off does not interfere in the operation of the 35W Front End laser. 

Also note that LHO made the same change today.

Images attached to this report
Comments related to this report
vernon.sandberg@LIGO.ORG - 10:51, Thursday 02 July 2015 (19424)PSL

The adding the capability to lock out the pump diodes for the HPO is an important change in operational configuration and how we operate the sites' PSLs.  For the record, this is the path followed at LHO to investigate the possibility of an engineering control (the physical locking out of power to the pump diodes), the installation of the control , and the communication with our sister site.

1.  Investigation of the way to disable power to the pump diodes and to determine the effect on auxilliary systems, e.g., FE (front-end monitors and coputer controlled interlocks).

LHO WORK PERMIT #5320

Date: Wed Jul 1 11:19:11 PDT 2015

Serial number: 5320

Task leader: Jason Oberling

People performing work: Jason, Rick

Facility: LHO

A lock and tag will not be used.

Period of Activity: 7/1/2015 - 7/1/2015

Area of Activity: H1 PSL, Laser Diode Room

Hazards:

Activity: In light of the event last week at LLO, to avoid accidental turning on of the PSL High Power Oscillator (HPO) we are going to test turning off the HPO diode box power supplies to see if the Front End (FE) still runs. The laser may trip during this test. In discussion with AEI Hanover, who has been talking with NeoLase, we have been told that this should be possible.

2.  Announcement to the community of a change in configuration, email:

Subject: PSL HPO Diode Box Power Supplies Locked Out
Date: Wed, 01 Jul 2015 12:58:38 -0700
From: Jason Oberling <oberling_j@ligo-wa.caltech.edu>
To: lho all <lho-all@ligo-wa.caltech.edu>



FYI for all:

I have turned off, unplugged, and locked out the H1 PSL HPO diode box power supplies.  This is to prevent the accidental turning on of the HPO, and does not interfere with the operation of the 35W Front End laser.  Lockout IDs are 3986 and 3770.  If for some reason access to these power supplies is needed/desired, come see me.

Jason

3.  Communication of change and status to a broader group; in particular, our LLO colleagues:

Subject: H1 PSL HPO diode power supplies turned off and locked out
Date: Wed, 01 Jul 2015 13:04:59 -0700
From: Jason Oberling <oberling_j@ligo-wa.caltech.edu>
To: lsc-psl@ligo.org, Janeen Romie <janeen@ligo-la.caltech.edu>, Richard Oram <roram@ligo-la.caltech.edu>
CC: fred.raab@ligo.org



As the subject says, I have turned off, unplugged and locked out the PSL HPO diode box power supplies to prevent the accidental turning on of the HPO.  Benno reported in this morning's PSL meeting that this should be possible and I have confirmed: having the HPO diode box power supplies off does not interfere in the operation of the 35W Front End laser.  I recommend LLO implement this as soon as they are able.

Jason

 

 

H1 CAL (AOS, CAL)
sudarshan.karki@LIGO.ORG - posted 10:01, Wednesday 17 June 2015 - last comment - 14:41, Sunday 12 July 2015(19186)
Gravitational Wave Strain h(t) sign convention determination using Pcal

Calibration Team

Sign of h(t):

The gravitational wave strain h(t) is given by  h(t) = Delta L/L where Delta L  is is computed using

                         Delta L = ± (Lx - Ly)

The sign of Delta L can be determined using Pcal actuation on the test mass. Pcal only introduces a push force  so pcal readout signal (truly pcal excitation) is minimum when the testmass is away from the corner station (closer to pcal laser). From the first plot the phase between DARM/PCAL is ~ -180 degrees (DARM lags PCAL) which suggests that DARM signal from ETMX will be maximum when pcal is minimum (ETMX further away from corner station). Similarly, from second plot, since DARM and PCAL have a phase difference of ~-360 degrees (essentially 0 degrees), the  DARM signal from ETMY is minimum when the pcal is minimum. This shows that the sign convention for the Delta L is '+'

Time Delay between Pcal and DARM:

Also the slope of the curve gives the time delay between Pcal and DARM signal chain. The time delay is about 125±20 us. This time delay can be accounted for, within the uncertainity, from the difference in signal readout chain outlined in Figure 3 attached.

Refer to LLO alog #18406 for the detailed explanation behind this  conclusion.

Images attached to this report
Comments related to this report
peter.shawhan@LIGO.ORG - 10:04, Friday 03 July 2015 (19437)CAL
I believe this sign check and the sign check at LLO are correct.  For the record, below is how I reached that conclusion:

The photon calibrator laser can only push, but there is a nonzero baseline intensity and you modulate the intensity around that.  The question is, if you apply a positive voltage to the PCAL system input, do you get more force or less force on the test mass?  Figure 21 of the PCAL final design document seems to show that the undiffracted beam through the AOM is what is sent to the test mass, so increasing the amplitude of the 80 MHz drive to the AOM REDUCES the force on the test mass.  However, the AOM driver electronics could introduce a sign flip when it conditions the input voltage.  To check that, I pulled up PCAL excitation and receiver photodiode data (e.g. H1:CAL-PCALX_EXC_SUM_DQ and H1:CAL-PCALX_RX_PD_OUT_DQ) and plotted a short time interval at GPS 1117933216.  I saw that the PCAL photodiode signal variations are basically in phase with the PCAL input excitation, with just a ~30-40 degree phase lag at ~500 Hz, presumably from filter delay.  So, applying a positive voltage to the PCAL system input causes more force on the test mass, and anyway the PCAL receiver photodiode measures intensity directly.  I confirmed this for all four PCALs (H1 and L1, X and Y) and also confirmed that the transmitter and receiver photodiodes vary together.

The PCAL pushes on the front of the ETM, i.e. on the face that the primary interferometer beam reflects off of.  This being a pendulum, the ETM is closest to the laser (i.e., the arm is shortest) when the force is at its MAXIMUM.  LLO alog 18406 has a comment consistent with that: "Theory of pendulums suggests that Pcal signal will be minimum when ETM swings further away from corner station".  LHO alog 19186, above, has a statement, "pcal readout signal (truly pcal excitation) is minimum when the testmass is away from the corner station (closer to pcal laser)", which is more ambiguous because the ETM being away from the corner station would put it FARTHER from the PCAL laser.  But both draw the correct conclusion from the data: with the intended sign convention, DARM should be at its positive maximum when the X arm is longest (ETMX is farthest from the corner station; PCALX intensity is at its minimum) or when the Y arm is shortest (ETMY is closest to the corner station; PCALY intensity is at its maximum), and that is what was reported at both sites.
darkhan.tuyenbayev@LIGO.ORG - 10:33, Tuesday 07 July 2015 (19466)

Peter,

I disagree with one assumption in your argument, but it does not disprove (or support) the rest of your conclusions.

"The question is, if you apply a positive voltage to the PCAL system input, do you get more force or less force on the test mass? Figure 21 of the PCAL final design document seems to show that the undiffracted beam through the AOM is what is sent to the test mass, so increasing the amplitude of the 80 MHz drive to the AOM REDUCES the force on the test mass. However, the AOM driver electronics could introduce a sign flip when it conditions the input voltage."

As far as I know there's no sign flip in AOM electronics. Undiffracted beam gets dumped in BD2, while diffracted beam is sent to the ETM.

Unfortunately I couldn't find an explicit noting of it in our recent DCC documents.

peter.shawhan@LIGO.ORG - 14:41, Sunday 12 July 2015 (19580)CAL, INJ
Oh, the diffracted beam gets sent to the test mass?  Then I agree, there isn't a sign flip in the electronics.  (In figure 21 in the document, it looks like the undiffracted beam went to the test mass.)

BTW, I've posted a multi-frequency look at the hardware injection actuation sign (and amplitudes and time delays) at https://wiki.ligo.org/Main/HWInjER7CheckSGs.
Displaying reports 66961-66980 of 85902.Go to page Start 3345 3346 3347 3348 3349 3350 3351 3352 3353 End