Displaying reports 70841-70860 of 84530.Go to page Start 3539 3540 3541 3542 3543 3544 3545 3546 3547 End
Reports until 16:29, Monday 22 September 2014
H1 SEI
jim.warner@LIGO.ORG - posted 16:29, Monday 22 September 2014 - last comment - 16:40, Monday 22 September 2014(14074)
ETMY Senscor Issues

I spent today trying to sort out more sensor correction issues. Over the weekend I realized that sensor correction not working could be explained by bad blends. I looked at ETMY and sure enough, there are 2 T750 blends, one of which is an old version, not to be used. Additionally I found that the calibration filters for the STS weren't turned on. I also noticed that the 8hz peak that comes from HEPI looks a bit bigger than the same peak at ETMX, something Hugh had noticed in August(?) and we thought had been addressed. I tried a number of configurations, shown in the attached pictures. The first is how I found the chamber this morning, with bad blends, HEPI running and basically no sensor correction (there was a gain of .5, with the cal filter off that becomes  ~.05). The second is how I'm leaving it tonight, with the right blends, STS cal filters engaged and sensor correction on. I wouldn't really call it better. The sensor correction seems to be re-injecting noise at ~1 hz, but cutting some at .1-.2 hz. This is not what I've found on E/ITMX, where sensor correction seems to cut more noise between .1-1hz and reinjects some below. Not sure what is going on there.

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 16:40, Monday 22 September 2014 (14077)

Since I mentioned ETMX, I'll put up some plots. JeffK suggested I project the ISI's motion to the optic, so Friday tried to do that. These are the results for ETMX, showing that most of the optic's longitudinal motion (the black trace) is mostly explained by the ISI's X (blue) motion (makes sense) down to about .1hz, where the ISI RY becomes a significant contributor. Below that X takes over again, but that's probably actually still tilt.

Images attached to this comment
H1 SUS
betsy.weaver@LIGO.ORG - posted 16:18, Monday 22 September 2014 - last comment - 14:29, Tuesday 23 September 2014(14075)
3IFO QUAD6 status

B. Weaver, T. Sadecki, D. Sellers

Today we laced reaction chain cables and plugged in UIM and PenRe OSEMs.  We then began another round of R0 TFs.  TF results pending.  Offsets and gains for the lower stages are as follows:

L1UL 19173 1.565  -9586
L1LL 27412 1.094 -13706
L1UR 24370 1.231 -12185
L1LR 23612 1.271 -11806
L2UL 22949 1.307 -11474
L2LL 24546 1.222 -12273
L2UR 24855 1.207 -12428
L2LR 23608 1.271 -11804

These have been loaded to MEDM.

Comments related to this report
betsy.weaver@LIGO.ORG - 14:29, Tuesday 23 September 2014 (14096)

We noted that the Q6 UIM L1 UL BOSEM was very low voltage.  Today, Travis and Danny switched in a new one.  We've updated the open light voltage offset and gain and loaded to MEDM.

After this change, now with serial numbers:

LOC  SN  OLV   GAIN   OFFSET
L1UL 575 28826 1.041 -14413
L1LL 491 27412 1.094 -13706
L1UR 489 24370 1.231 -12185
L1LR 427 23612 1.271 -11806

L2UL 153 22949 1.307 -11474
L2LL 133 24546 1.222 -12273
L2UR 216 24855 1.207 -12428
L2LR 145 23608 1.271 -11804
H1 General
andres.ramirez@LIGO.ORG - posted 16:00, Monday 22 September 2014 (14073)
Ops Shift Summary
9:09 Heading to End to search for parts in TMS lab – Travis
9:13 Working on all Mid/End station fans – Bubba
9:55 Heading into the LVEA - Aidan
9:58 Opening GV-15 at Mid X – Kyle
10:09 Going to work in the LVEA (West Bay) – Betsy/Danny
11:08 HEPI work on HAM6 – Hugh
12:05 Taking measurements at End X – Gerardo
12:46 Back from End X – Gerardo
13:05 Returning to the LVEA to work on Test Stand – Danny/Travis
13:18 Back to work in the LVEA – Aidan
13:57 Searching for cables (LVEA) – Corey
14:05 Valve operation on IP1/2/3/4/6 and YBM/XBM turbos – Kyle
14:59 Finished working in the LVEA – Travis/ Danny
15:42 Removing RGA from PT-346 at MidX - Kyle


H1 SUS (SUS)
jeffrey.kissel@LIGO.ORG - posted 14:03, Monday 22 September 2014 (14071)
H1SUSSRM M3 Coil Balancing -- More Failure
J. Kissel

Remembering that we may need to re-balance the recycling-cavity HSTS drives after having replaced the lower stage drivers with increased-range, modified, coil drivers (PRM, PR2, SR2, SRM; see Integration Issue 936 for links log entries), I began to balance SRM M3 this morning. 

I used the same pringle-drive, demodulation-of-the-OSEM-sensors, balancing technique that we've used in the past (see e.g., LHO aLOGs 11392, or 9453). 

I was able to confirm that the balancing *is* now worse by a factor of ~6 (by not changing the previous COILOUTF gains and simply comparing the new results with the balanced old), but was unable to balance to many better than this newly bad value. My time was limited because of priority DRMI work, but I'll continue to investigate as time allows. For now, I've left the COILOUTF gains as they were balanced to the unmodified driver, 
 
H1:SUS-SRM_M3_COILOUTF_UL_GAIN  0.972
H1:SUS-SRM_M3_COILOUTF_LL_GAIN -1.014
H1:SUS-SRM_M3_COILOUTF_UR_GAIN -0.972
H1:SUS-SRM_M3_COILOUTF_LR_GAIN  0.964

See attached for plots.

Plots, Data, and Diary live in and have been committed to
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/SRM/Common/Data/
Non-image files attached to this report
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 11:41, Monday 22 September 2014 - last comment - 17:20, Wednesday 01 October 2014(14067)
DRMI online calibration running on projector1

I finished preparing the online calibration filters for the DRMI which I started in the last week (see alog 14030). The output spectra are now monitored on projector1 (see alog 14042) in the control room as shown in the above pitcture.

I checked the HSTS and BS suspension filters in the calibration paths in h1oaf. While the HSTS suspensions looked good to me, the BS suspension did not look similar to the suspension model. In particular, the location of resonances did not look correct in the first place. So I changed the BS model by running the BS design Matlab script in the SVN and, copying and pasting the result in foton. The DC gain was then re-adjusted by hand such that it is at 8.85 x 10-5 um/cnts (= 46.40 um / (4 x 217 counts)). Note that this 46.4 um came from DCC-T1100602-v2. And the factor of 4 in the denominator compensates to the Euler matrix in the suspension realtime controller.

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 11:52, Monday 22 September 2014 (14069)

The dtt xml file now lives in userapps/release/isc/h1/scripts and is named DRMI_spectra_whiten.xml.

Since all the signals are whitened in the oaf model, the spectra must be un-whitened when they are displayed. I put the anti-whitening calibrations in the dtt file and this calibration must be always applied to all three spectra in order to get the correct displacements. Note that the whitening filters are the ones copied from Livingston at some point in the past and they have the following poles and zeros: zpk([1;1], [100;100], 1, "n").

kiwamu.izumi@LIGO.ORG - 17:20, Wednesday 01 October 2014 (14261)

Calibration factors for the DRMI optica gains:

  • PRCL = 3.3e-7 um/cnts (REFL_A_RF9_I)
  • MICH = 1.16e-7 um/cnts (REFL_A_RF45_Q)
  • SRCL = 5.7e-6 um/cnts (REFL_A_RF45_I)
H1 SEI
hugh.radkins@LIGO.ORG - posted 10:51, Monday 22 September 2014 - last comment - 11:38, Monday 22 September 2014(14065)
WHAM6 HEPI V2 IPS sensor fix Saturday short lived

Saw the HAM6 HEPI was tripped this morning.  It tripped on a large Actuator spike Saturday about 4pm, just a few hours after I go it fixed.  Attached are the IPS spectra for HAM6 like I've plotted before--obviously V2 is back to its old ways.  Also attached are all the local IPS sensors when it tripped.  Again, V2 is not like the others and it is a bit hard to be sure but it looks like the big glitch on the V2 channels preceeds the other sensors moving off their previous values.

What to do?

Wiggle the cable some more and/or look closer at thecable going into the Pier Pod?

Change the sensor?--Requires changing both V2 & H2 sensors as these are matched to the Kaman Box within the Pier Pod--And yes you guessed it, pull the Pier Pod box and mount the matching Kaman box within.

Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 11:38, Monday 22 September 2014 (14068)

Okay, just 'fixed' this again.  This time I unplugged and plugged in the cable repeatedly and twisted (I believe) the cable to maybe clean the contacts.  I don't think this is a 'back shell' shorting to the box issue.  If it were, we have this problem in many places given the design and how they look.

The HEPI control loops are back on, manual without boosts.

H1 SEI
jim.warner@LIGO.ORG - posted 10:33, Monday 22 September 2014 - last comment - 10:55, Monday 22 September 2014(14064)
Sensor correction is working on a few BSC's
Last week with Seb here, I started playing with sensor correction on the ETM ISI's. ETMX turned on and did good things around a half hz, as it should. ETMY so far has not done as well, so I'm still investigating there. This morning I looked at the ITM's, just seeing if filters were installed. ITMX seems to have had it's filters modified, so I used foton and gedit to fix the IIR and FIR filters. That did not happen as transparently as I would have hoped, so, my apologies to the ISC crew. After talking to Kiwamu and Sheila, they let me try to turn sensor correction on on ITMX this morning. Looks like it's working so far, only on Z ,see attached plots. The upper right plot on each picture shows most of the action, but there does seem to be a reduction in the low frequency motion on RX and RY as well, that is probably just an artifact of only having 20 minutes of data. We can't use sensor correction on other DOF's, as the blends are too low.
Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 10:55, Monday 22 September 2014 (14066)

Calibration filters on ETMY were not turned on after JeffK fixed the model, here. I just turned them on, hopefully that fixes that chamber. ITMY seems to have issues, as things started going crazy when I tried turning on sensor correction on at that chamber. More digging.

H1 General
andres.ramirez@LIGO.ORG - posted 10:28, Monday 22 September 2014 (14063)
PSL Weekly Check
PSL Check: 9/22/2014

Laser Status: 
SysStat: Warning “VP program online” is red
Output power is 29.6 W (should be around 30 W)
FRONTEND WATCH is active
HPO WATCH is red

PMC:
It has been locked 5 days,0 hr , and 59 minutes (should be days/weeks)
Reflected power is 2.2 Watts and PowerSum = 25.9 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 3 d, 20 h, and 38 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 2.03 V (should be at least 0.9V)

ISS:
The diffracted power is around 7.1 % (should be 5-15%) 

Last saturation event was 4 d, 22 h and 51 minutes ago (should be days/weeks)





H1 AOS
alexan.staley@LIGO.ORG - posted 10:14, Monday 22 September 2014 - last comment - 12:51, Monday 22 September 2014(14062)
BS L2P and L2Y Coupling preliminary results

Over the weekend I ran a transfer function on the BS with optical lever damping filters off, and FM2 of both L2P and L2Y filters on. These two filters were installed last week (alog 14046 and alog 14022). This transfer function was taken to determine the residual motion of the BS in yaw and pit given a drive in length on the M2 stage. We can compare this to the previous rediula motion with FM1 filters on for both L2Y and L2P as measured by Stefan alog 9394.  The attach screen shot shows a side by side comparion. The red traces are all YAW/L and the blue traces are all PIT/L. And the left most windows, or the traces with square symbols are the TF with FM2 (current version), while the right most windows, or the traces with the circle symbols are the TF with FM1 on (old version). The L1Y coupling improved by about 16dB at low frequency, and 24dB at high frequency. Meanwhile, in L2P the high frequency magnitude is about the same (it's actually slightly worse by about 2dB), but at low frequency we have imporoved by a factor of about 19dB. The new filter for L2P was intended to mostly correct the low frequency motion, so this result is about as expected. Also, in the process of creating the new filters, we did not concern ourselves with any of the resonances, which is why you don't see much of a change (and it may be a bit worse) compared to the older values. Jeff and I also checked that the BS coil balancing did not change between Stefan's measurement and mine, indicating that the improvement clearly came from the coupling filter change. I was unsure why the phase between the new and old version are no longer the same, given the that old and new filters have the same phase. Kiwamu explained to me that this came from whether we were pushing too much or too little to compensate for the L2Y/P coupling. The phase can change by 180 when we adjust for this.

Next, I plan on inspecting the power spectrum of the BS oplev with the PRMI locked on the sideband. One question we have though, is what constitutes good enough residual motion?

Images attached to this report
Comments related to this report
alexan.staley@LIGO.ORG - 12:51, Monday 22 September 2014 (14070)

To confirm our results I also repeated the measurements taken by Rana in alog 14028. The first screenshot shows the BS oplev pit and yaw power spectrum under a few different configurations. The dashed green and brown traces are spectra taken by Rana with PRMI locked and the older L2Y/P coupling filters. Meanwhile, the pink and light blue traces are the spectra with PRMI locked and the new coupling filters. We see an imporovement in the power spectrum with the new filters at about 0.1-0.3Hz in Yaw, and and 0.001-0.8Hz in Pit.  The red and blue traces are the spectra with PRMI unlocked and the new filters. The second screenshot shows the coherence of the oplevs with the MICH loop. The trace color representation is as before. One can see a reduction in the coherenece for both Pit and Yaw with the new filters in place. 

 

Note: everything can be found under /ligo/home/alexan.staley/Public/BS_Coupling/

Images attached to this comment
H1 General
andres.ramirez@LIGO.ORG - posted 09:43, Monday 22 September 2014 (14061)
Installation Meeting Minutes
 Attendees: Fred, Jim, Vern, Greg, Peter, Ed, John, Kyle, Gerardo, Jeff B, Jodi, Jason, Bubba, Patrick, Alexa, Daniel, Richard, Corey, Aaron, Filiberto, Kiwamu, Betsy, Travis, Hugh, Sheila, etc …

Today's Tasks:


Hugh reported about the WHAM6 HEPI V2 IPS sensors Problem. New HEPI TFs will need to be run. 

Work on 3IFO QUAD will continue today in LVEA (Betsy/ Travis)

Getting ready to test 3IFO (Quad 9) in Staging Building – Jeff B

Circuit breaker failure (Replacement part needed) – Richard  

Moving TCS HWS table into position (HAM4) – Greg/Aidan

Illuminator control work in LVEA – Richard/Filiberto

Fans work on all Mid/End stations – Bubba



H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 07:19, Monday 22 September 2014 (14059)
CDS model and DAQ restart report, Sunday 21st September 2014

model restarts logged for Sun 21/Sep/2014
2014_09_21 03:27 h1fw1

unexpected restart of h1fw1

H1 ISC (ISC, SUS)
rana.adhikari@LIGO.ORG - posted 13:34, Sunday 21 September 2014 - last comment - 10:47, Monday 22 September 2014(14057)
PRC spot positions using Angle to Length coupling

Kiwamu, Rana, Sheila

On Friday, we measured the spot position on PR2 using the A2L technique:  vertical / horizontal   ~   5mm / 3 mm.


We wonder if the spots are far off center and might explain the low recycling gain that we see. The camera images are not detailed enough to tell us about positions better than a couple cm.

The concept of the A2L measurement is that we drive the optic angle and measure the response in the cavity length sensor (REFL 45, in this case). We assume that the LSC photodiode has a very small angle sensitivity (which is true as long as the cavity is non-degenerate and has a high finesse and the beam on the photodiode is a few times smaller than the diode active area and that the photodiode has a small non-uniformity). We also assume that the mechanical coupling from torque to longitudinal motion is smaller than the centering precision we want. I don't know if all of these assumptions are true, but we proceed as if they are.

To make the measurement robust, we want to avoid the GDS testpoint dropouts, and so we use the front end digital lockins. Unfortunately, the LSC does not have capability to drive mirror angles and the SUS/ASC do not have capability to demodulate the LSC signals. So we make a temporary workaround in the SUS DRIVEALIGN matrix by routing the L drive to P or Y**. Then the LSC drive can drive apply torque on PR2-M3 and demodulate REFL45_I. We then null the demodulated signal by also sending some of the torque to the L2L path. By taking into account the size of the optic (R = 8 cm from T0900435-v9), we can use the ratio of L2L/L2A gains to determine the beam position on the optic (we also take into account the non-unity gains in the EULER2OSEM matrix: L=0.25, P/Y=5.3).

For PR2, we used a drive frequency of 9.57 Hz and amplitude 555 counts. Since we do not have WFS feedback yet, there were large beam spot motions. The 9.57 Hz digital lockin we use to demodulate the 45 MHz demod outputs have I and Q outputs. The I-phase corrresponds to the the steady beam spot position and the Q-phase gives us information of how much the beam is moving (?).


** This hacky technique won't work for the optics being used for LSC feedback, so we need to come up with a better hack or make some model rewiring. Now that we have larger range coil drivers, we can swap the control from PRM to PR2 to allow us to measure the spots all over the DRMI. And, of course, we can use the front end dither system to automatically center the spots as well.

Comments related to this report
kiwamu.izumi@LIGO.ORG - 10:47, Monday 22 September 2014 (14060)
  • The beam was too-low vertically by 5.3 mm
  • The beam was off horizontally by 3.1 mm toward BS or West.

 

(some details)

As for the vertical measurement, we needed a coefficient of -2.5 in the L2L drive-align matrix element. Taking the Euler matrix into account, I computed the imbalance alpha (see the previous alog 13765 for more info about the definition of those numbers) to be 0.12. Note that the L2P path had a coefficient of -1 while the L2Y had 1 as usual. This indicates an offcentering of 5.3 mm downdward on the optic.

As for the horizontal measurement, we needed a coefficient of -1.5 in the L2L drive-align matrix element. This corresponds to an offcentering of 3.1 mm. Since the UL and LL coils needed to be pushed harder than that of Rs, the spot must be off toward the right side.

Note that we have not readjusted the coil balance on PR2 after the recent coil drive modification.

H1 SEI (DetChar, PEM, SYS)
jeffrey.kissel@LIGO.ORG - posted 00:39, Saturday 20 September 2014 - last comment - 16:41, Monday 22 September 2014(14047)
Tilt and Blend Science at H1 EX
J. Kissel

The Message: I've cross-checked the calibration of all the ground sensors at the X End-Station, and used that knowledge to gain further confidence in their assessment of ground motion and ground tilt (second attachment). With these confirmed sensors, I tried to figure out why no one can find coherence between the ISI T240 X and either the GND BRS RY or GND T240 X (first attachment). My conclusion is that the ISI ST1 X DOF is limited by re-injected noise from ISI ST1 RY DOF between 20 and 200 [mHz], because we've copied and pasted our X T240 blend filter to RY without being conscious of this tilt-horizontal-coupling path (fourth attachment). I *think* this noise, is T240 sensor noise in this 20 to 200 [mHz] frequency band (see fifth attachment). This can be solved by sacrificing unneeded higher-frequency performance in RY (say between 1-10 [Hz]), and moving the RY blend up a bit, and making the T240 high-pass roll-off more aggressive, or "faster," as a function of frequency (third attachment is current X / RY blends).

%%%%%%%%
% The Deets %
%%%%%%%%
Calibration:
------------
In the second attachment, 2014-09-18_H1EXGND.pdf, I've calibrated everything into translational acceleration units. I summarize here, then explain the details after.
Summary: 
(1) H1:ISI-GND_STS_ETMX_X_DQ                 1e-9 [(m/s) / ct] --> Let DTT differentiate once to acceleration units
(2) H1:ISI-ETMX_ST1_BLND_X_T240_CUR_IN1_DQ   1e-9 [(m/s) / ct] --> Let DTT differentiate once to acceleration units
(3) H1:ISI-GND_BRS_ETMX_RY_OUT_DQ            1.568e-8 [(m/s^{2}) / ct]
(4) H1:PEM-EX_SEIS_VEA_FLOOR_X_DQ            7.9e-9   [(m/s) / ct] --> Let DTT differentiate once to acceleration units
(5) H1:PEM-EX_TILT_VEA_FLOOR_X_DQ            5.5e-8   [(m/s^{2}) / ct]
(6) H1:PEM-EX_TILT_VEA_FLOOR_T_DQ            5.39e-7  [(m/s^{2}) / ct]

For (1) and (2), myself and the SEI group have graciously calibrated these channels into 1 [(nm/s) / ct] in the front end, following the electronics chain as described in D1001575. So I merely have to convert to (m/s), and let DTT handle the differentiation by requesting m/s^{2} / Hz^{1/2} on the units menu.
For (3), Krishna and I have installed a similarly dead-reckoned calibration that we believe is in 1 [nrad/ct]. However, converting to translational acceleration by multiplying by g = 9.8 [m/s^{2}/rad] and by 1e-9 [m/nm], leaves a discrepant factor of 1.6 between the GND T240 and the GND BRS (see pg 1 of 
2014-09-18_H1EXGND.pdf), where there is great coherence, between 10 and 100 [mHz] and we expect the signals to be the same. Also notice the how the harmonics is the 8 [mHz] resonance pollute the spectrum (the BRS has been rung up to +/- 200 [ct] during this measurement period). That's when I invoked the PEM sensors, hoping they would be coherent enough between the sensors to cross-check, but alas, in the 10 to 100 [mHz] region, they're too noisy to really tell if the GND BRS or GND T240 are "right," so I added in the extra factor of 1.6 assuming the T240s were correct, hence 1.6 * 9.8 * 1e-9 = 1.568e-8 [(m/s^{2}) / ct]
For (4), I used the pem.ligo.org prescribed 7.6e-9 [(m/s) / ct], it matched the GND T240 very well (within the 22% quoted precision) in the frequency region where we expect them both to be sensitive to translation, i.e. above 100 [mHz].
For (5) and (6), since the instrument has not yet been successfully calibrated (see LHO aLOG 13623) I assumed the that GND T240 and GND PEM Guralp were correct, and simply scaled the PEM TILT X channel to match them above 100 [mHz], ending up with 5.5e-8 [(m/s) / ct] (and let DTT do the differentiation). I then blindly assumed that the Tilt channel uses the same calibration value, but for rotational displacement, i.e. 5.5e-8 [rad/ct]. Scaling by g = 9.8 [m/s^{2} / rad] that yeilds the above 5.39e-7  [(m/s^{2}) / ct]. It seems to match up reasonably well, and it's not hard to imagine that the electronics chain is the same for both channels, but the sensor appears to be limited by some noise incoherent with the GND BRS in the 10 to 100 [mHz] region.

The Tilt-Horizontal Coupling Model:
-----------------------------------
On the final page of 2014-09-18_H1EXGND.pdf, I plotted the ISI performance against all of the ground sensors, and noted the hump between 10 and 100 [Hz] that looked suspiciously like a blend filter bump. Going on a hunch I've had for a while Similar to what I did in my thesis, knowing that we've thus far only copied and pasted our X blend filters to the RY DOF, I used the same 1-stage, MISO model (e.q. 5.3 on pg 80) to predict how much the residual platform tilt (RY) motion is coupling into the X DOF,
                     G_x         g        x
x              =  -------   *  ----- *  F          * res RY
  from res RY      1 + G_x      w^2       T240 HP
Thankfully, at these low frequencies (f < 1 [Hz]), the ISIs have loop gain, G_x, much much greater than 1, the closed loop gain (the first term) is well-approximated by unity, and I only have to know the blend filter F^{X}_{T240 HP}.

This model shows varying degrees of success.
(1) Between 50 and 300 [mHz], this predicts the X ST1 motion exactly. ISI T240 RY doesn't show coherence, however, but I'm confident that's because it's incoherent sensor noise of the RY loop in this band -- at least up until 100 [mHz]. I'm still confused why the double-peaked microseism (100-200 [mHz]) in ISI X is very coherent with GND X, but (a) doesn't show up in nor is it coherent with the GND RY spectrum, and (b) perfectly matches the shape of the ISI RY motion projected into ISI X. 
(2) The model WAY over predicts the X displacement between 10 and 60 [mHz]. I've triple-checked my blend-filter-multiplication-via-DTT-calibration, and I'm confident I'm doing it right -- see third attachment. Steps:
    - Grab a matlab version of the blend filter from ${SeiSVN}/seismic/BSC-ISI/Common/Complementary_Filters_BSC-ISI/aLIGO/TSheila.mat
    - Ask matlab for its poles and zeros via [Z,P,K]=zpkdata(High_Pass_Filters(1)), where X is the first DOF
    - Turn them from [rad/s] into [Hz], by dividing by -2*pi
    - Copy them into foton, bode plot, and find the correct normalization factor such that the filter asymptotes to 1 at high-frequency (-160.202 [dB])
    - Copy the normalization gain, poles and zeros into DTT and multiply the correct displacement calibration, 
         gain: 1/(2*pi) [rad / (rad/s)] * 1e-9 [rad/nrad] * 9.8 [(m/s^2) / rad] * 1 / (2*pi)^2 [m / (m/s^2)]
         poles: 0, 0, 0
         zeros: [none]
(3) Independent of the model's confusion, at least it shows lots of room for improvement with GND X to ISI X in this region (100 - 700 [mHz]).
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 16:41, Monday 22 September 2014 (14076)
J. Kissel, S. Karki

Typo in the above entry -- I used 7.6e-9 [(m/s) / ct] for the H1:PEM-EX_SEIS_VEA_FLOOR_X_DQ (PEM guralp), which is much closer to the new pem.ligo.org value of 7.39e-9 [(m/s) / ct], which Sudarshan has recently updated. My value of 7.6e-9 [(m/s) / ct] was from the previous value reported on pem.ligo.org, which I naively assumed hadn't changed. It's a 3% discrepancy; well within the quoted 22% uncertainty reported for the PEM Guralp.

One more example of use wanting to be able to edit aLOGs more the 24-hours in the past...
H1 ISC (ISC)
daniel.hoak@LIGO.ORG - posted 17:39, Friday 19 September 2014 - last comment - 21:21, Sunday 21 September 2014(14045)
modulation depth measurement using the OMC

Dan, Koji (from a distance), Rana

We used an OMC mode scan to measure the modulation depth of the 9 and 45MHz sidebands, to close the loop on the recent changes to the amplification path before the EOM.  Koji had done this previously, and measured Gamma1 (9MHz) = 0.198, Gamma2 (45MHz) = 0.305.

Today, we measured Gamma1 = 0.208, Gamma2 = 0.240 +/- 0.01.  The error in Gamma2 is due to an asymmetry between the upper and lower sidebands; using the ratio of the 45MHz USB to the carrier returns Gamma2 = 0.251, while the LSB returns 0.233.  This 10% discrepancy between the 45MHz LSB and USB is consistent across several sweeps of the cavity.  The analysis code used today was thrown together a little quickly and doesn't do a sophisticated job of integrating the curve around the peaks (in fact, doesn't do any integration, just compares the peak heights to the carrier), but when applied to the same data that Koji used earlier this month it finds values similar to his to within a few percent. 

The attached plot has an overlap of five cavity sweeps, with the peaks of the 9 and 45MHz sidebands used in the calculation marked with a cross.  The data are here.  The mode scan was performed off a single bounce from ITMX.  This weekend we'll do a more careful scan for both ITMX and ITMY to calculate (more?) accurate numbers of the mode-matching for each path.

Images attached to this report
Comments related to this report
rana.adhikari@LIGO.ORG - 22:18, Friday 19 September 2014 (14049)

Daniel, Kiwamu, Rana

The modulation depth mystery for 45 MHz is still unsolved. Educated guesses are welcome.

In December of 2013, the people have measured that the input power to the EOM was 11 dBm and the modulation depth was 0.07 radians.

Today, we checked at the CDS electronics room, the field rack, and all around the PSL table and near the EOM, for the presence of any more un-documented amplifiers. We found none.

We measured the power at the field rack to be +12 dBm. So we think that based on the EOM calibration from December that we should be getting < 1 radian.

However, the OMC scan seems to show a modulation depth of ~0.24 radians for the 45 MHz sideband. Unless the carrier peak is saturated in the DC readout PDs with a single bounce Michelson beam, this seemsto be impossible to square with the EOM calibrations.

In December of 2012, Volker measured the modulation coefficient to be (0.33 rad / 10 Vpp) = 0.066 rad/V.    10 Vpp ~ 24 dBm (for 50 Ohms). This agrees well with Kiwamu's entry.

 

We have the further evidence of high modulation that the REFL and AS demod signals didn't change much (or at all) when the ZHL-1A was pulled out of the modulation chain.

What's going on here?

rana.adhikari@LIGO.ORG - 21:21, Sunday 21 September 2014 (14058)ISC

Measured the gain of ZHL-1A amplifier that we removed at 45.5 MHz through a 20 dB attenuator into a 50 Ohm loaded scope; gain is nominal over a wide range of input power levels.

Images attached to this comment
Displaying reports 70841-70860 of 84530.Go to page Start 3539 3540 3541 3542 3543 3544 3545 3546 3547 End