Displaying reports 71301-71320 of 83104.Go to page Start 3562 3563 3564 3565 3566 3567 3568 3569 3570 End
Reports until 10:54, Saturday 31 May 2014
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 10:54, Saturday 31 May 2014 (12156)
CDS model and DAQ restart report, Friday 30th May 2014

model restarts logged for Fri 30/May/2014
2014_05_30 14:42 h1fw0

unexpected restart of h1fw0

H1 SYS
rainer.weiss@LIGO.ORG - posted 20:48, Friday 30 May 2014 - last comment - 13:20, Thursday 05 June 2014(12155)
Vibration Measurements of the beamtube
John Worden, R. Weiss
Vibration measurements of the beamtube with the new R10 insulation and aluminum jacket were made and described in the
appended document. The new insulation reduces the acoustic coupling to the beamtube but not as well as the initial insulation.
The data enables a more detailed estimate of the phase noise from the beamtube motion due to the faulty HR coatings of the new 
test mass mirrors. The DCC number of the document is T1400384.
Non-image files attached to this report
Comments related to this report
rainer.weiss@LIGO.ORG - 13:20, Thursday 05 June 2014 (12233)
Error in the document. Please see T1400384-v2 or LHO log entry 12232
H1 ISC (IOO, ISC, SEI, SUS)
daniel.hoak@LIGO.ORG - posted 19:02, Friday 30 May 2014 (12154)
Beams marked
Sheila & Dan

We marked the two IM4_TRANS beams (PRM_REFL and IO_TRANS) on the wall of the LVEA today.

To be clear: both of these beams are transmitted through IM4.  One is from the beam leaving the IMC, the other is the reflection coming back from the PRM.  The IO_TRANS beam is in the -X direction on the wall.

We noticed that there are also two green beams very close to the IR beam reflected off the PRM.  Only the Y-arm was locked at the time, so this must be the faint green reflection off of PR2 from the Y-arm.  We might want to use these as a reference for the BS alignment in the future.
H1 ISC
daniel.sigg@LIGO.ORG - posted 18:00, Friday 30 May 2014 - last comment - 13:23, Saturday 31 May 2014(12153)
EY WFS

(Keita, Daniel)

Today we measured and adjusted the gain/phase of the WFS RF signals. WFS_A looks fine, but WFS_B has segments 3 and 4 which are ~10x smaller than 1 and 2. Engaging DOF4 did result in a stable feedbakc but with a rather large offset. We added an offset to segment 4 of WFS_B to recover. Adding DOF3 did not result in a stable feedback.

Playing around with the individual pit/yaw offsets and the WFS DC centering one can find feedback topologies with all 4 DOFs engaged. However, they are susceptible to disturbances.

Comments related to this report
daniel.sigg@LIGO.ORG - 13:23, Saturday 31 May 2014 (12158)

For this measurement the WFS offsets are:

Bias WFS A WFS B
Pitch -2500 -1600
Yaw -5000 4700

 

 

 

First measurement shows an ETMY yaw misalignment by 1 µrad. The power drops by about 15%. The yaw offsets are of order 10,000 counts. We also see a significant pitch ofset. But, more importantly the DC offsets on the WFS heads are also very large. The WFS servo was off during the measurement.

The second plot shows the effect of moving the picomotor for WFS_B by about 0.2 in yaw and 0.3 in pitch. The resulting offsets are of order 7000. Since the WFS servo was off, there was no degradation of the cavity power.

Non-image files attached to this comment
LHO VE
kyle.ryan@LIGO.ORG - posted 17:58, Friday 30 May 2014 (12152)
~1715 hrs. local -> Combined BSC7's RGA with the XBM
This RGA is set up to scan every 60 minutes for this weekend in order to capture data during the Ring Heater "burn".
H1 SUS
arnaud.pele@LIGO.ORG - posted 17:13, Friday 30 May 2014 (12151)
ETMy L2L measurement with ESD

Yesterday night, after Richard re-rerouted the end-y ESD cabling, I took a L2L transfer function to check the length drive with the ESD. It seems to behave nicely, and similarly as EX. The attached plots shows the comparison of Ex and Ey TF in m/cts.

The measurement was taken using the same dtt template as for ETMx. The arm was locked without the slow control engaged. Sensor used was ALS-Y_REFL_CTRL (in um). Measurement lives in

/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMY/SAGL3/Data/2014-05-29_H1SUSETMY_L3_L2LPY_SweptSine.xml

2014-05-29_H1SUSETMY_L3_L2LPY_SweptSine.xml2014-05-29_H1SUSETMY_L3_L2LPY_Swep
Images attached to this report
H1 TCS (TCS)
thomas.vo@LIGO.ORG - posted 17:05, Friday 30 May 2014 (12150)
TCS Progress
CO2X
 - Zero-ed out the QPD segments to compensate for the differences in individual quadrants.
CO2Y:
 - Zero-ed out the QPD segments to compensate for the differences in individual quadrants.

HWS:
 - Built up HWS camera for both X&Y sides of the table, able to get serial connection and read out the status information
 - End station HWS table inventory progress: Cables and optics completely kitted.
 - Need a re-do of positioning of the HWS-Y camera because of interference with the table enclosure.

RH:
 - Prepared for a 24 hour ITM ring heater test. 30 minutes on full power (~30 Watts) and 30 minutes off. We'll do this work on Saturday and Sunday.
H1 ISC (ISC)
lisa.barsotti@LIGO.ORG - posted 16:59, Friday 30 May 2014 (12149)
Preparation for 90/10 BS installation in HAM3
I put together this  T1400371 procedure for aligning the 90/10 BS in HAM3. It describes the way they did the alignment at Livingston (Option 1, green beam from HAM1, IOTR table would need to move) and the way that Keita would like to do it here (IR beam from HAM3, no need to move IOTR).


All the parts needed are "mostly" ready. Here is a summary:

- BS 90/10, post, clamp (mirror is cleaned and installed in mirror mount)
- V-shaped dump, post, clamp (ready)
- IR laser diode source (in optics lab)
- spare clean sled (in optics lab, can be used to mount the laser source and collimator in HAM3, if needed)
- 2 spare 2" HR 1064 clean optics (on optics table in OSB optics lab - I left there also a clean 2" HR 532)
- Missing: IO style posts for mounting the steering optics (Corey and Cheryl working on it)
H1 General
jeffrey.bartlett@LIGO.ORG - posted 16:07, Friday 30 May 2014 (12148)
Ops Summary
Operators Log 05/30/2014
Day Shift Summary
LVEA Laser Hazard 

08:49 Filiberto – To End-Y working on cable repairs
09:00 Kyle - Moving RGA cart to BSC7 area
09:00 Hugh – Working on external cabling at HMA5/6
09:35 Dave & Thomas – Working on TSC-X and Hartman tables
09:45 Rite Choice on site to pick up pallets and scrap wood 
10:04 Keita & Sheila – Going to End-X
10:06 Kyle – Going to End-Y to recover test equipment
10:45 Betsy & Travis – Working in the West Bay cleanroom
12:49 Karen – Going to End-Y
13:08 Betsy & Travis – Working in the West Bay cleanroom
13:17 Dave – Working on TCS-X and Hartman tables
14:18 Karen – Back from End-Y
15:22 Betsy & Travis – Out of the LVEA
H1 SEI (INS, SUS)
hugh.radkins@LIGO.ORG - posted 15:49, Friday 30 May 2014 (12147)
WHAM5 ISI Transfer Functions Poised to begin tonight ~6pm--Run on OPSWS8
Arnaud may run some measurements on the SUS in HAM5 so I delayed these to start later today.  Running on OPSWS8
H1 SEI (INS, SUS)
hugh.radkins@LIGO.ORG - posted 15:46, Friday 30 May 2014 - last comment - 11:59, Tuesday 03 June 2014(12146)
WHAM5 & most of HAM6 In Air Cables Reattached & secured
When the septum was removed, the cabling between HAMs 5 & 6 were disconnected so the crew could deal with all the bolts.  I asked FClara to reconnect the cables to the WHAM6 NW feedthru, SUS & ISC I think.  I'm not familiar enough to distinguish these so I left those to him.  When they are sorted out I will complete the cable/feedthru protecting.
Meanwhile, HAM5 ISI sensors are live, don't disconnect any cabling, you'll have to cut zip ties to do so.
Comments related to this report
filiberto.clara@LIGO.ORG - 10:29, Monday 02 June 2014 (12171)
Installed SUS and ISC cabling to D3 flange of HAM6 (NW corner).

Flange    Cable           Description
D3-1C1    H1:SUS_HAM6-10  OMC TOP
D3-1C2    H1:SUS_HAM6-11  OMC Right/Side

D3-2C1    H1:ISC_HAM6-265 ASC_AS_A (WFS)
D3-2C2    H1:ISC_HAM6-266 ASC_AS_B (WFS)

Document D1300122, internal in-vacuum cabling calls out for cable ISC-265 and ISC-266 to be connected to flange D5-3C1 and D5-3C2. Document should be updated.


Filiberto Clara
hugh.radkins@LIGO.ORG - 11:59, Tuesday 03 June 2014 (12187)
HAM6 NW Cables stress relieved and E-FeedThrus protected.
H1 ISC
keita.kawabe@LIGO.ORG - posted 15:34, Friday 30 May 2014 (12144)
POP QPD sled in HAM3 centered

In preparation for the vent, I centered POP QPDs so that we can use it as an alignment reference when we will install HAM3 beam splitter.

Right now both POPA and POPB PIT and YAW are all within +-0.015.

QPDA sum is 56.3, QPDB 51.7 using straight shot.

POPA picomotor counts are: X -331, Y -28257.

POPB picomotor counts are: X 4151, Y 20661.

Not unlike TMS picomotors, these are sort of degenerate. You can use pico B to center QPDA, then pico A to center QPDB, and repeat thousands of times to eventually have a good centering. You can also overshoot repeatedly to make it quicker. For example, when QPDB is too low while QPDA is centered, you use pico A to bring QPDB up, then pico B to bring QPDA down, and repeat.

The sled is sensitive to even 1 step of the picos. Medium step (100 steps) is safe to use without worrying too much about losing the beam, but if you want to use Large (1000 steps), be prepared to lose the beam on both of the QPDs.

Images attached to this report
H1 AOS
betsy.weaver@LIGO.ORG - posted 15:17, Friday 30 May 2014 (12145)
3IFO Q6 upper structure under way (1 of 2 ITMs)

Travis, Betsy

Over the last few days, we assembled an upper structure which will eventually attach to one of the ITM lower structure that is currently trapped in chamber.  We finished assembling the 4 top stage blade cartridges and 2 top masses for this unit.  We then mounted all of them plus top wire segments and parts of the table cloth to the upper structure.  We will mount the whole cube to the test stand on Monday and then it will wait for it's other 2/3.

H1 AOS (CDS)
david.barker@LIGO.ORG - posted 14:34, Friday 30 May 2014 (12143)
Hartmann wavefront sensor camera on X-arm table communicating with MSR

Thomas, Jim, David and Dave

We were able to establish serial communications between the h1hwsmsr computer in the MSR with the HWS  Teledyne Dalsa Pantera 1M60 camera on the X-arm TCS table in the LVEA. We suspect communications were not working earlier due to serial synchronization problems, which were cleared when the remote transceiver (RCX CLink) and camera were power cycled. We discovered several MEDM/Beckhoff/Wiring issues with the power control of these devices which Thomas is going to resolve over the next few days.

H1 General
jeffrey.bartlett@LIGO.ORG - posted 09:04, Friday 30 May 2014 (12142)
Morning Meeting Minutes
Hugh - Working on external cabling at HAM5 and possibly HAM6
Dave - Working on the TCSX table and prepping for ring heater test
Richard - Working on end stations ESD
Kyle - Moving RGA cart to BSC7 area
Lisa & Sheila - Setting fiducials at HAM2  
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:17, Friday 30 May 2014 (12141)
CDS model and DAQ restart report, Thursday 29th May 2014

no restarts reported.

H1 TCS
david.hosken@LIGO.ORG - posted 19:02, Thursday 29 May 2014 (12140)
TCS
David / Greg / Thomas

TCSX:
- QPDs tested and appear to be working correctly. Need to locate different translation stage to calibrate.
- laser will be running overnight. (TCSY laser, power supplies and chiller are all OFF).
- FLIR camera mount drawing added to DCC ( https://dcc.ligo.org/LIGO-D1400168 )

HWS End-X:
- mounts and optics have been staged

Corner HWS clamping plate: modifications to be made by Apollo

Polarisers:
- guys at Caltech looking at a way to bond these so that nylon screw not required to secure to the mount.
H1 SYS
daniel.sigg@LIGO.ORG - posted 18:38, Thursday 29 May 2014 (12139)
Wrap up of the current integration period

The idea is to wrap up commissioning by Saturday night. The ring heater stress test can proceed on Sunday. The vent can start Monday.

H1 ISC
keita.kawabe@LIGO.ORG - posted 18:29, Thursday 29 May 2014 (12138)
WFS whitening chain today

I asked EE to make/fix the whitening cables and/or install temporary new cables that are shorter than nominal, whichever is quicker. The shorter ones are to be replaced in the future.

Two cables were fixed, and the two new short ones were put in place. After this, I still saw that one of the two fixed cables (H1:ISC-Y-65, this is the first cable for WFSB) still had a short circuit for one pin. Since they didn't have enough short new ones, and since the broken bit was only for one of the whitening stages which I can leave off, I decided to keep the defective cable for now and live with it.

I also asked them to swap both of the whitening chassis, as the whitening gain was all over the place even though the BIO readback looked good, which didn't make sense, and we were getting suspicious about the whitening. The new ones were put in place.

I measured the dark noise, and both of WFSs had some problem.

For WFSA, one of the 8 channels (WFSA-Q4) would suddenly rail when the whitening gain is 24dB or larger. Up to 21dB it looks completely normal, DC output increases as I increase the gain, but when the gain reaches 24dB it suddenly jumps from a few hundred counts to 32k. Disconnected the analog, repeated it, the same result. Swapped the BIO cable, the same result. Since this seemed like a problem of the whitening chassis itself, I put one of the old whitening chassis (S1101600, which is one of the two units pulled out this morning) back and it behaved good. The problematic one is S1101636.

For WFSB, one of the channels (WFSB-I2) had much larger low frequency noise (f<1Hz) than everything else. You can kind of see that in the plot I posted yesterday, but today the noise looked larger. Disconnecting the demod didn't do anything as far as the low frequency noise is concerned, disconnecting the analog out DB9 on the whitening chassis didn't do anything, but when I went to the remote rack I immediately noiticed that the DB9 cable connector on the AA chassis was loose. I disconnected it and the noise was gone. I reseated everything, and things look good.

To summarize,

  1. Whitening chain is in a workable state after fixing some nonsense problems.
  2. Two short cables should be replaced, one bad cable (H1:LSC-Y-65) should be fixed/replaced, after this commissioning window.
  3. Whitening chassis S1101636 seems to have a problem for Q4 channel (that's the last channel of 8-ch chassis).
  4. It's not clear if S1101601 (which was pulled out this morning) has a problem or not, but we have some suspicion.
  5. S1101600 for WFSA and S1101602 for WFSB are in use now.
Images attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 16:16, Thursday 29 May 2014 (12135)
Ops summary
doors being craned for HAM5 and rigged such that closing the gate valve is not required

08:40 Kyle to start pump for RGA on BSC8 (WP 4647)
08:46 Aaron and Filiberto to end Y to fix short in WFS cables
08:47 Jeff B. to HAM5 to take PET swipe
09:07 Hugh to HAM5 to move equipment out of the way for door installation
09:08 Arnaud to HAM5 to take hammer measurements
09:22 Richard to end Y to drop components off for Aaron and Filiberto
09:30 Travis to LVEA test stand area to retrieve documents
10:05 Praxair driver approximately 15 minutes away
10:31 Luis Z. to end Y to install grounding wrist strap connectors on the electronics racks
10:50 Jeff B. and Andres done at HAM5
11:27 Corey to LVEA to put away equipment
12:13 Praxair truck leaving
13:11 Jodi, Travis and Betsy to the LVEA west bay
13:22 Kingsoft water to take water samples
13:27 Hugh to LVEA test stand to work on sub-assemblies for 3rd IFO ISIs
15:21 Keita to end Y to work on WFS
15:48 The septum has been installed between HAM5 and HAM6. The doors are tacked on HAM5. Sheila is transitioning the LVEA to laser hazard.
Displaying reports 71301-71320 of 83104.Go to page Start 3562 3563 3564 3565 3566 3567 3568 3569 3570 End