(Alexa, Daniel, Sheila)
We measured the WFS sensing matrix for both PIT and YAW. The matrices were computed by exciting ITMX/ETMX PIT/YAW at 1.5Hz with 0.3 amplitude, and looking at the WFS transfer function. The results can be found in the following DTTs saved in /ligo/home/sheila.dwyer/ALS/HIFOX/WFS/:
Inverting the results gives...
YAW input sensing matrix:
-2.294 -3.155
0.529 -2.604
PIT input sensing matrix:
-0.494 -0.239
0.302 -0.489
These values are in microradians; however the factor of 1e-3 is taken care of in the gain of the filter module. We also adjusted the signs as follows:
Under this configuration, with all four DOFs engaged (DOF 1,2 P,Y) the WFS_A/B_I_YAW/PIT_OUTMON all converge to zero. The lock appears to be stable with a transmission betweeen 950 to 1050 counts. Futher investigation needs to be done...
Aidan. Dave H. Greg G. Thomas V.
We have, basically, finished installing all the electronics and cabling on the TCSX table. There are a few outstanding issues (summarized in the attached PDF). The ones that are preventing us from turning on the laser are listed below.
We hope to address all these early on Monday to get the laser running to check the output power.
Chiller settings:
We adjusted the bypass valve on the back of the TCSX chiller to achieve the following settings (to mimic what's at LLO)
Flow rate: 3.4GPM
Pressure: 60.5PSI
Laser housing leak.
The replacement laser installed on the TCSX table also has a slow leak from the laser housing (in the same place as the first one) - about 1 drip every 5 minutes. We will check the output power of this laser before we pull it. See photos 3 & 4
This trip happened around the time of a beckhoff restart. A beckhoff restart causes a crash of the IMC guardian, which causes it to stop. Its not clear to me why this should cause MC2 to get a large signal, but it seems to. This causes a cascade of trips. Even though I don't know why this happends, creating a safe state in the IMC guardian so that it handels missing channels better could help with the problem.
Another solution is to have the SUS WDs not trip HEPI. Can we get that fix soon?
HAM2 also tripped at the same time, Jeff and Hugh brought that back.
To be clear, IMC guardian did not "crash" in this particular situation. The guardian responded exactly as it's currently programmed to respond, which is to go into ERROR when it looses communication with any of the channels it's monitoring. I want to distinguish and ERROR condition, which is something that guardian handles, to a "crash", which means that the guardian process died unexpectedly.
Here's my guess for the sequence of events:
It's possible guardian could be made slightly more robust against loss of some of it's channels, but that only helps up to a point. Eventually guardian has to drop into some error condition if it can't talk to whatever it's trying to control. It could try to move everything to some sort of safe state, but that only works if it can talk to the front-ends to actually change their state.
A Beckhoff restart also causes the IMC servo board to be reset, as well as all whitening for photodiodes, wavefront sensors and QPDs. I assume that the resulting transient caused the MC to trip. It would be interesting to know, if this is due the length or alignment system. Is it the initial transient or a run-away integrator? In either case this should not result in a trip. A better action would be to simply turn off the ISC inputs.
After taking a look at the time of WD trips, it seems like HAM3 ISI trips before MC2, see green plot vs red plot (the X axis is the number of seconds after gps=1080083200)
Please mind the signs here at HAM4, there is an Optic suspended here. Commissioning to commence shortly.
Day Shift Summary LVEA Laser Safe Apollo - Bolting dome on BSC10, working on view ports at End-Y Mitch – Check ACB at End-Y Hugh – Working on HAM5 HEPI Richard – Y-Arm TCS cabling 08:35 Mitch – Going to End-Y 08:53 Dave – LVEA working on TCS cabling 09:00 Justin – Moving laser barriers around the HAM4 TCS table area 09:18 Andres – Working on dog clamp shelves next to HAM2 09:20 Hugh & Jim – Working on HAM5 HEPI 09:29 Betsy & Margo – Going to End-Y for BSC10 closeout 11:26 Filiberto – Going to Mid-Y to pick up TCS panels 13:00 Justin – Transition LVEA to laser hazard 13:20 Dave & Thomas – Turning on TCS chillers 14:00 Filiberto – Going to End-Y to check ACB feed through 14:05 Andres – Working on dog clamps shelves next to HAM2 15:30 Jax – Briefly transitioning End-Y to laser hazard to make a PZT check 15:36 Michael – Going to End-Y to check chamber close status
At Richard's request I disabled the End Y picomotors, so that they will not be used durring the pumpdown.
Tested the forward voltage for the ACB diodes. BSC10 Flange F1-3C1 (air side). Pin 1&2 PD forward voltage: 0.431V Pin 4&5 PD forward voltage: 0.423V Pin 7&8 PD forward voltage: 0.431V Pin 10&11 PD forward voltage: 0.423V All results within spec.
Margot and I performed all of BSC10 closeout tasks listed in my previous alog regarding this closeout. We pulled the FC from the ETMy optic a little after 11am, and the door was going on at ~1:30. It took "so long" (ha) because we ended up having to spend some extra time calling in the troops about one of the ISC viewports having what appeared to be a scratch on it's inside surface. More details to come as needed.
Here are some further pictures of the ACB swing back and the FirstContact spray cone attachment to the HR side of the ETMy QUAD structure from yesterday. Yes, the cone protrudes into the baffle. Yes, Margot still was able to fit her head and an arm with the spray bottle in the cone.
The last picture shows the placement of the horizontal wafer and verticle 1" witness optic in the center of the chamber, placed today just before the door went on the chamber.
Notes and dcc numbers for Contamination Control Samples (ones that came out and ones that went in) to WBSC10 before doors went on: 1. Vertical 1’’ optics on QUAD: T1400246 (SN1195) out and T1400247 (SN 262) in. 2. Vertical 1’’ optic under quad: T1400248 (SN261) placed at end of closeout. 3. Vertical wafer on quad: T1400249 attached to quad under HR of ETM. 4. Horizontal wafer on floor under quad:T1400250 placed at end of closeout. 5. Horizontal 1’’ optic: T1400251 was left on beam tube floor between BSC10 and BSC6. Looking up SN/history. Didn’t take it out, because we had no clean optic containers handy. 6. 24’’ PET swipe sample taken in tube near purge today close to end of closeout. Labeled, and taken to PET microscope area.
Jim locked the HAM5 ISI and then we floated the platform on HEPI. A couple of Spring adjustments and we are level on the Optical Table to +-0.1mm. Not sure about the elevation yet but it is likely close to Spec. This will be good enough for continuing the ISI commissioning which we may get started before days end.
Please mind the signs--There are no optics here but HEPI is unlocked & floating; also please mind the Dial Indicators--Thanks--H
Elevation of Optical Table is currently ~1mm low. Target elevation is -325mmGz(D0901131-v1); the table is now at -326.2mm. Easy to correct during Initial Alignment.
ITMY HEPI tripped. Reset.
For further record, it's actually HEPI-HAM3
I've created a new folder on the desktop for WD trip plots, if operators could help us to plot and post these each time there is a trip it would be greatly appreciated.
You can create the plot by opening the WD screen, before untripping idenify the type of trip by looking for a red box in the First Trig column. Use the blue buttons labaled Plot ACT trip to make the plots (it will pop up after a while). Use the magnifying glass in the plot window to zoom the plot to the relevant time scale, then take a screen shot. You can store this in the WDTrip folder on the desktop, and then post it in the alog.
Jamie, Fabrice,
I am posting for the records and for discussion the WD trips we observe while tuning the turn on procedure for the guardian.
no restarts reported.
Here is an update on the instruction for untripping seismic, with some new tips from Fabrice, especially for ETMX.
Untrip HEPI WD, use Commands2 window to isolate level 1 on all chambers except BS which is level2
Untrip ISI WD (you may need to raise T240 Threshold)
reset CPS offsets,
Check that everything is in high gain: (this is also good to check if you are already isolated but have a lot of motion)
If you have been struggling with switching the blend filters, you may want to zero the T240s. You can do this in the three T240 screens by tuning on the AutoZ filter module in the X filter module, then turning it off.
Use commands, isolate level 3 for most chambers. For ETMX start with Commands, Isolate Stage 1 Level 2. Once this script is finished, isolate stage 2 level 3.
Switch Blends:
good luck...
We just tried again. The following sequence seems robust for ETMX:
a) Isolate Stage 1 Level 2
b) Isolate Stage 2 Level 3
c) Swith Stage 2 to TCrappy (Using Switch All)
d) Swith Stage 1 to TCrappy (Using Switch All)
We are working on the Guardian to simplify all of this. It will handle all the parameters and sequences specific to each chamber (controls level, blend filters, alignement dofs to be restored...)
If you have been struggling with switching the blend filters, you may want to zero the T240s. You can do this in the three T240 screens by tuning on the AutoZ filter module in the X filter module, then turning it off.
It is my understanding that you should NOT zero the T240, certainly not just to try it. In a new installatioin or the T240 is suspected of being out of level, then one should look at the U V W signals and AutoZ if they are not close to zero once the machine has stabilized. Before I was told this, it was my experience that pressing the AutoZ did NOT speed up the Trillium settleing process. Rich Mittleman--please weigh in here.
I agree with Hugh that re-zeroing the T240 should not be done while the T240 is in the midst of settling from a large dynamic shift. The re-zeroing is used to reset the mass-center of the instrument. This can shift with temperature or DC level changes. One should be sure the instrument has fully stabilized, otherwise you run the risk of 'recentering' to some off-center location. the Manufacture's (Nanometrics) instuctions can be found in section 2.4.2 of the T240 OBS manual (1). https://dcc.ligo.org/LIGO-T0900323 (2) -Brian (1) OBS - Ocean Bottom Seismometer. we use this because it is smaller, and we provide our own pressure vessel. (2) yes, we have received permission to put a copy in the DCC.
Thank you for clarifying...
Fabrice asks that we also add a step at the begining of the procedure where we plot the WD trip, zoom in and post it in the alog. Hopefully this we result in seismic people analyzing them and figuring out why we have so many trips.
This is important because our current rate of trips (and the time it takes to recover from them) will be a serious impediment to commsioning once we need to use 7 chambers all at the same time.
Aidan, Dave H, Thomas More cabling and plumbing. The leaky laser from ALOG 11019 was replaced with a new one and a matching RF driver. We're in a position to be able to turn on the laser mid-morning tomorrow.
The new ITMx monolithic welding has been successfully completed and was left suspended overnight last night. The majority of the modal measurements have also been completed today, with the exception of the violin modes, which are proving to be trickier to measure than in previous suspensions. We will continue with these measurements tomorrow with some different equipment. Stay tuned for subsequent posts with IAS and modal measurement numbers.
Modal frequencies of the test mass (ITMX) with the PUM nominally locked are: ITMX value resolution Pendulum 0.649 0.002 Pitch 1.1153 0.002 Yaw 1.0928 0.002 Bounce 6.74 0.016 Roll 9.195 0.016 Violin modes fibre serial number BR 499.9 0.06 S1400155 BL 501.2 0.06 S1400156 FR 500.8 0.06 S1400138 FL 502.2 0.06 S1400153 [just as a note to self as much as anything - remember to use the Uniform window on the spectrum analyser for the violin modes when using the chirped sweep output]
Forgot to post this last week. Final alignment numbers from the ITMx welding (all directions/rotations reported from the view of the IAS equipment, i.e. looking at the HR face of the ITMx):