H1 General (Lockloss, SUS)
anthony.sanchez@LIGO.ORG - posted 22:15, Saturday 16 November 2024 (81315)
Saturday Ops Eve Shift End

TITLE: 11/17 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: TJ
SHIFT SUMMARY:

3:01 UTC GRB-Short E 525935 We were not locked for the begining of this stand down event.


2024-11-17 3:27:53 NOMINAL_LOW_NOISE reached
3:30:27 OBSERVING reached

Violin ITMY Mode 5 is still rising on this lock.
I changed the Gain to 0 and waited 15 minutes to see if it will stop going up or even turn around.
I then tried about 3 minutes with the gain at -0.02 ( Nominal gain just negative) but it was obvious that the no gain is better than opposite gain.
I ended up just turnin goff the gain.

05:11 UTC Lockloss https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1415855535
This was a very fast lockloss , it doesn't look like the same IMC /FSS locklosses we have been seening. But the PSL-FSS_FAST_MON_OUT_DQ was still the first signal to change behavior.

Relocking notes:

Relocking has been made very difficult with the IMC constantly unlocking. Most of the Locklosses are in locking Green arms, but everyfew attempts it gets a little further like DRMI, then the IMC stops working again.
I tried turning the ISS off again, but it hasn't stopped it from unlocking the IMC.
I turned back on the ISS so  just in case H1 manager gets to about ISC_LOCK state 500 it will auto lock for him.

Images attached to this report
H1 General
anthony.sanchez@LIGO.ORG - posted 18:46, Saturday 16 November 2024 (81314)
Relocking notes & Mid shift .

I couldn't get Past DRMI 1f, and ended up in a PRMI & Check Mich loop.

I tried to do an Intial_Alignment, But IMC would not stay locked for long enough.
I eventually waited for the IMC to lock then turned off the ISS Auto Locker for the duration of the Initial alignment.

Initial alignment went well with the ISS turnt off.
Trying to relock again.

Images attached to this report
H1 General (SUS)
anthony.sanchez@LIGO.ORG - posted 16:37, Saturday 16 November 2024 - last comment - 17:39, Saturday 16 November 2024(81312)
Saturday Ops Eve Shift Start

TITLE: 11/17 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 159Mpc
OUTGOING OPERATOR: Ibrahim
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 15mph Gusts, 11mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.24 μm/s
QUICK SUMMARY:

H1 Has been locked for 1 Hour and 45 minutes.
DCPD's are diverging, Which I usually attribute to the Violins growing.
Turns out it's one of the usual suspects..... ITMY Mode 5 increasing.

I have turned the Gain off of ITMY m5 to see if it stays at that level or decreases.

The IFO stays locked until 1:25 UTC then I can do a Calibration sweep are the calibration time early today was missed by a lockloss.

Images attached to this report
Comments related to this report
anthony.sanchez@LIGO.ORG - 17:39, Saturday 16 November 2024 (81313)Lockloss, PSL

Lockloss https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1415840300
 

This looks like there is some IMC issue that cause this Lockloss.
The Lockloss plots show that at -906ms  there was blip on the PSL-FSS_FAST_MON _OUT_DQ channel.

Relocking the IMC was difficult as the IMC kept going into fault. I took ISC_LOCK to idel and took the IMC to idle for 10 minutes.
Relocking started fine, except I had to manual green arms.

Images attached to this comment
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:30, Saturday 16 November 2024 (81311)
OPS Day Shift Summary

TITLE: 11/16 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 157Mpc
INCOMING OPERATOR: Tony
SHIFT SUMMARY:

IFO is in NLN and OBSERVING as of 22:35 UTC

There was one IMC-caused Lockloss which ended our 13hr and 36 min lock acquisition. (alog 81310)

Stayed locked for the majority of this shift with one somewhat troublesome lock acquisiton for a number of reasons:

I was not able to take a calibration measurement due to the Lockloss, and we're not thermalized yet. I've informed Tony, the EVE operator, and he will take a calibration sweep once thermalized (01:24 UTC).

LOG:

None

H1 ISC (Lockloss)
ibrahim.abouelfettouh@LIGO.ORG - posted 11:17, Saturday 16 November 2024 (81310)
Lockloss 19:07 UTC

Lockloss seemingly caused due to the IMC with MC2 losing lock within 1ms of ASC (attached).

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:18, Saturday 16 November 2024 (81309)
Sat CP1 Fill

Sat Nov 16 10:10:51 2024 INFO: Fill completed in 10min 47secs

Images attached to this report
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 07:33, Saturday 16 November 2024 (81308)
OPS Day Shift Start

TITLE: 11/16 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 164Mpc
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 12mph Gusts, 9mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.20 μm/s
QUICK SUMMARY:

IFO is in NLN and OBSERVING as of 05:34 UTC (10 hr lock!)

LHO General
corey.gray@LIGO.ORG - posted 22:00, Friday 15 November 2024 (81305)
Fri Ops Eve Summary

TITLE: 11/16 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Aligning
INCOMING OPERATOR: TJ
SHIFT SUMMARY:

Most of the meat of the shift has already been posted earlier.

Had a return of the FSS/IMC glitches this afternoon/evening.  Have made it back to Observing at the end of the shift though.  

For the night we are set for locking for the OWL.  But if locking becomes an issue that lasts over 2hrs, plan is to take ISC LOCK to IDLE for the night.
LOG:

H1 PSL (IOO, PSL)
corey.gray@LIGO.ORG - posted 20:47, Friday 15 November 2024 (81306)
02:08utc NLN Lockloss due to FSS/IMC Glitch After 1hr Lock

Had about 20hrs with no IMC/FSS glitches from 9pm Thurs night until about 340pmPT Fri afternoon.  But have now had several for the last 5hrs.

Also for one ~60min stretch, H1 almost made it to NLN, but had rung up violins which kept us at OMC Whitening for 20min until the next IMC glitch.

Here are some running notes for the rough 5hrs filled with glitches (currently have H1 in IDLE).

Notes During Return of IMC/FSS Glitch Locklosses

2340utc (340pmPT) LOCKLOSS with IMC tag was the beginning of IMC/FSS glitch locklosses (where we went almost 20hrs since the last glitch around 5utc (9pmPT).

0111-0208:  OBSERVING

0208utc (608pmPT):  Attached screenshot shows a coincident glitch lockloss seen on FSS & IMC.  Here is the LOCKLOSS page also tagged with IMC glitch.

Starting an FRS Ticket to mark the DOWNTIME (FRS-32644) for 3.5hrs of DOWNTIME so far.  Where there was notable part of downtime due to IMC not locking easily.

VIOLINS Rung Up A Bit: (haven't had this since my shift on Tues night!)   So were stuck at OMC WHITENING for over 20min before H1 had....

So for the last 3hrs:

Not sure what is more useful.  Keep having ISC LOCK continue to locking, or to leave it in IDLE for IMC/PMC/FSS monitoring....

Images attached to this report
H1 PEM (PEM)
corey.gray@LIGO.ORG - posted 19:12, Friday 15 November 2024 (81307)
HVAC Fan Vibrometers FAMIS Check (FAMIS 26340)

For FAMIS #26340:  All looks well for the last week for all site HVAC fans (see attached trends).

Images attached to this report
H1 AOS
ibrahim.abouelfettouh@LIGO.ORG - posted 16:43, Friday 15 November 2024 (81304)
OPS Day Shift Summary

TITLE: 11/16 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Aligning
INCOMING OPERATOR: Corey
SHIFT SUMMARY:

IFO is LOCKING in ENGAGE_ASC_FOR_FULL_IFO

Just had a lockloss and trying to get back to observing. IFO has been quite well behaved today. Both lock acquisitions were fully automatic with H1 needing no intervention.

That being said, we did have two locklosses.

Lockloss 1: ETM glitch caused.

Lockloss 2: IMC caused.

LOG:

None

LHO General
corey.gray@LIGO.ORG - posted 16:24, Friday 15 November 2024 (81303)
Fri Ops Eve Transition

TITLE: 11/16 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Aligning
OUTGOING OPERATOR: Ibrahim
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 14mph Gusts, 11mph 3min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.28 μm/s
QUICK SUMMARY:

Walked in to a locking h1, which then has moved on to INITIAL ALIGNMENT.  Ibrahim shared how H1 has had a better time--so am cautiously optimistic for the night---with the goal of OBSERVING!

Environmentally we are becoming quieter with µseism dropping below the 95th percentile and winds not too breezy.

H1 ISC (Lockloss)
ibrahim.abouelfettouh@LIGO.ORG - posted 16:04, Friday 15 November 2024 - last comment - 16:23, Friday 15 November 2024(81301)
Lockloss 23:40 UTC

Ibrahim, Ryan S

Seems like an IMC lockloss. Interestingly, there was no obviously glitchy behavior (as previously observed) before or after the lockloss. Attached screenshot of ASC and IMC losing lock.

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 16:23, Friday 15 November 2024 (81302)

Adding more channels for the sake of investigation. As Ibrahim mentions, no glitchy behavior was seen leading up to this IMC lockloss, so this was either a large fast glitch out of nowhere or something else giving out. Seemingly all at once (or at least very close to each other), the FSS_FASTMON, RefCav trans, NPRO power, PMC high voltage, PMC mixer, and all IMC signals quickly change, so it's hard to tell which is truly happening first as there is lots of feedback between these signals.

Images attached to this comment
H1 SQZ (SQZ)
victoriaa.xu@LIGO.ORG - posted 15:53, Friday 15 November 2024 (81300)
More low pass filtering for ADF SQZ readbacks

Sheila, Vicky. 

We changed the low pass filtering on the ADF SQZ angle servo to be lower, going from 1 Hz low pass to 0.1 Hz low pass on the ADF I/Q demod filter banks. I think this reduced the sqz angle readback noise a bit (trends before/after). It didn't make a huge impact on the control signal that goes to the CLF_RF6 demod phase (bottom purple trace), and it might be worth considering low pass filtering that control signal too. I accepted the SDFs for this filter bank LPF change.

Images attached to this report
H1 GRD
sheila.dwyer@LIGO.ORG - posted 12:10, Friday 15 November 2024 (81298)
automation got stuck in a PRMI loop overnight

Ibrahim, Sheila, Ryan S

Ibrahim noticed that the guardian was in a loop of locking PRMI something like 20 times overnight.

What happened was that PRMI was poorly aligned (POP18 and 90 I ERR  both below 20 counts), but PRMI was able to grab lock for a couple of seconds, but would not hold lock as the guardian engaged the top mass offloading, adjsuted gains and filters.  Eventually this hit the timer in H1 manager to do an initial alignmemt, after just over an hour. 

There is also a timer in ISC that checks if PRMI hasn't locked in 10 minutes it should move onto MICH fringes.  This didn't happen in this case because of the very short locks.  Ryan and I read through the ISC_LOCK AQUIRE_PRMI state, and realized that it had a check (line 1371) for DRMI arrived (in the state PRMI locked), which didn't include a check for the state being done, which would return true in ISC LOCK before ISC_DRMI finished the final steps in PRMI locked  .  In the second, zoomed screenshot, you can see that as soon as the DRMI guardian enters state 35 (PRMI locked), ISC lock moves on immediately from state 50 (ACQUIRE PRMI) to 51 and 52 (PRMI ASC), which resets the timer so that we never went to MICH fringes. 

We added a check for the ISC_DRMI state to be both arrived and done in PRMI locked, so that PRMI will have to survive the offloading and boost engagement for the timer to be reset.  Ibrahim has now reloaded this.  We think that if this situation came up again, now we would only spend 10 minutes relocking PRMI, before going to MICH fringes.  If H1 is under the manager control, it would run initial alignment if running MICH fringes didn't help. 

 

Images attached to this report
H1 General
ryan.crouch@LIGO.ORG - posted 11:27, Friday 15 November 2024 - last comment - 12:08, Friday 15 November 2024(81296)
Range_compare for a range drop midlock

I ran the range_compare script using the 10-12 minute low range stretch from the most recent lock and some time before it of the same lock where the range was better, the construction crew by staging was also seen moving large equipment around this time on the cameras.

To produce this figure I ran "python3 /ligo/gitcommon/ops_tools/rangeComparison/range_compare.py --span 600 1415725641 1415728983", I only did 600 seconds of data. The range drop looks to be from low frequency noise, mostly below ~300Hz, the most obvious peak difference looks to be at 46/47 (mostly on sensmon), 36 (mostly on sensmon), and 60Hz (both sensmon and DARM) for the low range stretch. The lines are hard to see, but if you open the pdf in LibreOffice (linux default pdf viewer) and right-click -> Arrange -> Behind Object, then you can hover your cursor over the curves and it will be much easier to see the differences.

Images attached to this report
Non-image files attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 12:08, Friday 15 November 2024 (81299)

The SQZ blrms didn't really change during this time either.

Images attached to this comment
H1 ISC (Lockloss)
ibrahim.abouelfettouh@LIGO.ORG - posted 10:57, Friday 15 November 2024 - last comment - 11:21, Friday 15 November 2024(81295)
Lockloss 18:36 UTC

Ryan S, Ibrahim

Unexpected lockloss being investigated but confirmed not the IMC nor the environment. Ryan S is seeing that it is looking like the ETM Glitch.

Comments related to this report
ryan.short@LIGO.ORG - 11:21, Friday 15 November 2024 (81297)

Does not look like an IMC/PSL lockloss; trends attached. The NPRO temp has been glitching this morning, not drastically, which has been causing the EOM drive to be higher in general, but these glitches were not happening at the time of this lockloss. The shape of of the signal on AS_A is indicative of an arms lockloss and happens before the IMC loses lock according to MC2_TRANS. Additionally, the lockloss tool shows glitches on ETMX starting up to second before the lockloss, although the ETM_GLITCH tag was not applied in this case as the glitches did not meet the required threshold.

(The ndscope template I'm using lives in my home directory ~/templates/psl_glitch_hunting.yaml)

Images attached to this comment